Skip to content

Autotome™ RX 39 M00545190Recalls & reported problems

Boston Scientific Corporation · Cannulating Sphincterotome · Class 2

All models of Autotome RX 39 (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Autotome RX39 cannulating sphincterotome is used to cut tissue during ERCP procedures on the bile duct. Most reports describe the cutting wire breaking or bending during the procedure, and devices being hard to position or advance. There are 70 reports, with3 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
68
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 68 adverse event reports that list this exact device (DI 08714729444763), received between Apr 11, 2017 and May 5, 2026 — classified as 3 injuries, 65 malfunctions.

The most frequently reported specific device problem is “Break” (50 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 3 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break50
  • Adverse Event Without Identified Device or Use Problem12
  • Material Deformation7
  • Difficult to Advance6
  • Positioning Problem5
  • Device Dislodged or Dislocated4
  • Deformation Due to Compressive Stress3
  • Failure to Align3

Patient problems

  • No Clinical Signs, Symptoms or Conditions53
  • No Consequences Or Impact To Patient13
  • Electric Shock1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 68 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILIARY DISEASE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 39 BROKE WHEN SPHINCTEROTOMY WAS PERFORMED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25085394 ↗
  • MalfunctionMay 5, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILIARY DISEASE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 39 BROKE WHEN SPHINCTEROTOMY WAS PERFORMED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT CONDITION FOLLOWING THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25085388 ↗
  • MalfunctionMay 4, 2026Use of Device Problem; Deformation Due to Compressive Stress; Difficult to Advance; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CUT WIRE WAS BENT AND DIFFICULT TO CANNULATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE TIP LOOKED DEFORMED DURING EVALUATION UPON RECEIPT HOWEVER THE DEVICE WAS STILL USED. PER THE INSTRUCTION FOR USE (IFU), DO NOT USE THE DEVICE IF ANY DEFECT, MALFUNCTION, OR ABNORMALITY IS OBSERVED DURING INSPECTION, AS THIS MAY RESULT IN PATIENT INJURY OR DEVICE FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25071923 ↗
  • MalfunctionApr 29, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT CHOLANGITIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUT WIRE BROKE OFF FROM THE TIP OF THE TOME. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25038307 ↗
  • MalfunctionApr 25, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUT WIRE BROKE. THE PROCEDURE WAS COMPLETED WITH THE SAME ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24991740 ↗
  • MalfunctionApr 10, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT PRIMARY SCLEROSING CHOLANGITIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, A SPHINCTEROTOMY WAS PERFORMED. AT THE COMPLETION OF THE CUT, THE PROXIMAL END OF THE EXPOSED CUTTING WIRE SNAPPED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. A PHOTO OF AN UNKNOWN DEVICE WAS PROVIDED AND SHOWED A BROKEN WIRE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24841022 ↗
  • MalfunctionMar 30, 2026Break; Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE WHEN MAKING THE SPHINCTEROTOMY. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THE CUTTING WIRE WAS BROKEN AND KINKED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24716187 ↗
  • MalfunctionMar 23, 2026Deformation Due to Compressive Stress; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING THE PREPARATION, IT WAS REPORTED THAT THE DEVICE CAME OUT OF THE PACKAGE TWISTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS: THE CUTTING WIRE KINKED. PLEASE REFER TO BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24665079 ↗
  • MalfunctionMar 2, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE CUTTING WIRE BROKE WHEN ENERGIZED WITH CAUTERY. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED DUE TO THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24479057 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2026. PRIOR TO THE PROCEDURE, PHYSICIAN WENT TO BEGIN SPHINCTEROTOMY AND ENERGY WAS NOT BEING DELIVERED. UPON BOWING AND UNBOWING, THE PHYSICIAN REALIZED THAT THE CUT WIRE WAS BROKEN. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401790 ↗
  • MalfunctionJan 9, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE CUTTING WIRE OF THE AUTOTOME RX 39 BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT, THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24023233 ↗
  • MalfunctionOct 31, 2025Peeled/Delaminated; Material Deformation; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Material Deformation; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE AMPULLA OF VATER DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE TIP OF THE TOME WAS TORN, AND THE PAINT WAS PEELED OFF THE EXTRUSION. THE PROCEDURE WAS COMPLETED WITH THE SAME ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT AFTER THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF CUTTING WIRE KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23441612 ↗
  • MalfunctionOct 21, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE DUODENUM OR AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILE DUCT STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHEN BOWING THE SPHINCTEROTOME THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23352208 ↗
  • MalfunctionOct 17, 2025Break; Failure to Align; Break; Failure to Align

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTME RX 39 WAS ATTEMPTED TO BE USED IN THE BILIARY DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CUTTING WAS BROKEN, AND THE TIP WOULD NOT RELAX WHEN BOW WAS DOWN. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTME RX 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23320212 ↗
  • MalfunctionSep 29, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES. THE EXACT PROCEDURE DATE WAS UNKNOWN. DURING THE PROCEDURE, THE CUTTING WIRE OF THE AUTOTOME RX 39 BROKE PRIOR TO SPHINCTEROTOMY. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23174759 ↗
  • MalfunctionAug 20, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILIARY STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CUTTING WIRE BROKE WHILE PERFORMING SPHINCTEROTOMY. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22853465 ↗
  • MalfunctionAug 18, 2025Positioning Problem; Positioning Problem

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO AUTOTOME RX 39 USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE PAPILLA DURING A PAPILLOTOMY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, AFTER ROTATING THE AUTOTOME, THE WIRE DID NOT FACE THE INTENDED POSITION. THE SAME PROBLEM OCCURRED WITH A SECOND AUTOTOME. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22836252 ↗
  • MalfunctionAug 18, 2025Positioning Problem; Positioning Problem

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO AUTOTOME RX 39 USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE PAPILLA DURING A PAPILLOTOMY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, AFTER ROTATING THE AUTOTOME, THE WIRE DID NOT FACE THE INTENDED POSITION. THE SAME PROBLEM OCCURRED WITH A SECOND AUTOTOME. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22836231 ↗
  • MalfunctionAug 4, 2025Break; Material Deformation; Adverse Event Without Identified Device or Use Problem; Break; Material Deformation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE BILIARY AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE WIRE BROKE DURING SPHINCTEROTOMY. A PHOTO OF THE COMPLAINT DEVICE WAS PROVIDED AND SHOWED THAT THE CUTTING WIRE WAS BROKEN AND KINKED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22709261 ↗
  • InjuryJul 3, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE DUODENUM AND COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES PERFORMED ON (B)(6) 2025. DURING SPHINCTEROTOMY, THE TECH BOWED THE TOME SLIGHTLY AND THE CUT WIRE POPPED OFF THE CATHETER. THE DETACHED CUT WIRE WAS SUCCESSFULLY RETRIEVED USING A NET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22399143 ↗
  • MalfunctionMar 5, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE PANCREATIC DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. THE EXACT PROCEDURE DATE WAS UNKNOWN. DURING THE PROCEDURE, A DOUBLE WIRE CANNULATION WAS PERFORMED. WHEN THE CUTTING WIRE CAME IN CONTACT WITH THE GUIDEWIRE IN THE PANCREATIC DUCT, THE CUTTING WIRE OF THE AUTOTOME RX 39 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21529982 ↗
  • MalfunctionSep 25, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS USED IN THE PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, IT WAS NOTICED THAT THE CANNULATION WAS DIFFICULT. THE PATIENT HAD A NARROW DUCT, AND A PANCREATIC DUCT STENT WAS IN PLACE WITHIN THE DUCT. THE WIRE ANCHOR DISLODGED FROM THE CATHETER BUT THE CUTTING WIRE WAS INTACT. NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER AUTOTOME RX 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20298736 ↗
  • MalfunctionAug 28, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS BEING PREPARED FOR USE IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING PREPARATION OUTSIDE THE PATIENT, UPON OPENING THE PACKAGE, IT WAS NOTICED THAT THE CUTTING WIRE WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20085560 ↗
  • MalfunctionAug 23, 2024Difficult to Advance; Device Contamination with Chemical or Other Material; Difficult to Advance; Device Contamination with Chemical or Other Material

    IT WAS REPORTED, TO BOSTON SCIENTIFIC CORPORATION. THAT AN AUTOTOME RX 39 WAS ATTEMPTED TO BE USED IN THE COMMON BILE DUCT (CBD), DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE GUIDEWIRE COULD NOT ADVANCE INSIDE THE DEVICE. IT WAS REPORTED, THAT A FOREIGN OBJECT OR SUBSTANCE WAS INSIDE THE GUIDEWIRE PORT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER AUTOTOME RX 39 DEVICE. THERE WERE NO PATENT COMPLICATIONS REPORTED, AS A RESULT OF THIS EVENT. NOTE: A PHOTO OF THE DEVICE, OUTSIDE THE PATIENT WAS PROVIDED BY THE CUSTOMER. AND A BROWN COLORED FOREIGN SUBSTANCE CAN BE SEEN INSIDE THE WORKING LENGTH OF THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20049867 ↗
  • MalfunctionAug 21, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AUTOTOME RX 39 WAS BEING PREPARED FOR USE IN THE DUODENUM AND BILIARY TREE DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2023. DURING PREPARATION OUTSIDE THE PATIENT, IT WAS NOTICED THAT THE DEVICE WAS BROKEN AS THE WIRE APPEARED TO BE CUT AND LOOSE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20020515 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page