Skip to content

TRUEtome 39 M00584190Recalls & reported problems

Boston Scientific Corporation · Cannulating Sphincterotome · Class 2

All models of TRUEtome 39 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

The TRUEtome 39 is a Boston Scientific cannulating sphincterotome, a thin electrosurgical cutting wire used during ERCP procedures to open the sphincter. Most of the 45 reports describe the cutting wire snapping or breaking while the tip was bowed against tissue during the procedure, sometimes bending or deforming before use, and in several cases the piece stayed attached rather than detaching in the patient. Reports mention no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 08714729827658), received between Apr 3, 2018 and Sep 10, 2026 — classified as 46 malfunctions.

The most frequently reported specific device problem is “Break” (39 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break39
  • Adverse Event Without Identified Device or Use Problem6
  • Material Deformation6
  • Use of Device Problem5
  • Positioning Problem2
  • Deformation Due to Compressive Stress1
  • Device Dislodged or Dislocated1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions37
  • No Consequences Or Impact To Patient8

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2026Break; Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50723725 ↗
  • MalfunctionSep 3, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50599746 ↗
  • MalfunctionAug 31, 2026Break; Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50533767 ↗
  • MalfunctionApr 17, 2026Failure to Conduct; Material Integrity Problem

    NOTE: THIS REPORT PERTAINS TO A JAGWIRE REVOLUTION, TRUETOME 39, TWO TRAPEZOID RX LITHOTRIPTER BASKETS, AND CRE PRO WIREGUIDED DILATATION BALLOON USED DURING THE SAME PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE REVOLUTION, TRUETOME 39, TWO TRAPEZOID RX LITHOTRIPTER BASKETS, AND CRE PRO WIREGUIDED DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) AND CBD STONE REMOVAL PROCEDURE TO JAUNDICE AND CBD STONE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, AFTER THE STONE WAS CAPTURED, THE BASKET COULD NOT BE PULLED THROUGH THE PAPILLA. THE TRAPEZOID RX BASKET WAS THEN CONNECTED TO AN ALLIANCE HANDLE TO PERFORM LITHOTRIPSY. DURING LITHOTRIPSY, THE BASKET FRACTURED AT THE HANDLE END, AND THE SAFETY RELEASE MECHANISM DID NOT ACTIVATE TO ALLOW RELEASE OF THE STONE FROM THE BASKET. THE PHYSICIAN THEN PERFORMED ANOTHER CRE DILATION, BUT THE CRE BALLOON DEVELOPED A SMALL RUPTURE DURING INFLATION. FOLLOWING THE RUPTURE OF THE CRE BALLOON, THE PHYSICIAN ATTEMPTED A PARTIAL PAPILLOTOMY USING A TRUETOME. HOWEVER, THE CUTTING EFFICIENCY WAS POOR, POSSIBLY DUE TO INTERFERENCE FROM A CBD STONE OR THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24918055 ↗
  • MalfunctionFeb 2, 2026Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT OBSTRUCTION PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, WHEN THE TOME WAS BOWED AND CAUTERY APPLIED FOR SPHINCTEROTOMY, THE WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE DEVICE WAS DETACHED AND FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24234859 ↗
  • MalfunctionJan 21, 2026Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE AMPULLA/PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING PROCEDURE, UPON BOWING OR STEPPING ON THE PEDAL, THE CUT WIRE BROKE OFF FROM THE TIP OF THE TOME. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS REPORTED TO BE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24140803 ↗
  • MalfunctionJan 9, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS BEING PREPARED FOR USED IN THE PAPILLA DURING A PROCEDURE TO TREAT COMMON BILE DUCT STONE TO BE PERFORMED ON (B)(6) 2025. DURING PREPARATION, THE CUTTING WIRE WAS BENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24025547 ↗
  • MalfunctionDec 24, 2025Break; Use of Device Problem; Break; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT BILIARY STONES PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHEN BOWING THE TOME THE CUTTING WIRE SNAPPED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT, THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED. NOTE: IT WAS REPORTED THAT THE DEVICE WAS ENERGIZED WHEN NOT IN CONTACT WITH TISSUE; HOWEVER, PER THE INSTRUCTIONS FOR USE (IFU), PRECAUTION: IT IS RECOMMENDED TO MAINTAIN DIRECT AND CONSTANT CONTACT WITH TISSUE WHEN APPLYING ELECTROCAUTERY CURRENT. FAILURE TO DO SO MAY RESULT IN BROKEN CUT WIRE, DAMAGE TO THE ENDOSCOPE AND/OR PATIENT INJURY. ADDITIONALLY, IT WAS ALSO INDICATED THAT THE DEVICE WAS ENERGIZED PRIOR TO BOWING; HOWEVER, ACCORDING TO THE IFU, THE DEVICE DOES NOT NEED TO BE ENERGIZED PRIOR TO PERFORMING SPHINCTEROTOMY AS ENERGIZING THE CUTTING WIRE PRIOR TO USE WILL CAUSE PREMATURE CUTTING WIRE FATI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888521 ↗
  • MalfunctionSep 23, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT JAUNDICE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE TRUETOME WAS INSERTED INTO ENDOSCOPE, AND THE TIP WAS ALSO PARTIALLY INSERT INTO PAPILLA. WHEN THE TECHNICIAN TRIED TO USE THE HANDLE TO BEND THE TIP AND CUT PAPILLA, THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23124759 ↗
  • MalfunctionJul 18, 2025Entrapment of Device; Use of Device Problem; Entrapment of Device; Use of Device Problem

    NOTE: THIS REPORT PERTAINS TO ONE TRUETOME 39 AND ONE JAGWIRE DEVICE USED IN THE SAME PATIENT AND PROCEDURE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STRICTURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE DEVICE WAS LOADED OVER A 0.035 JAGWIRE, AND CANNULATED SUCCESSFULLY, BUT COULD NOT BE EXCHANGED AS THE WIRE WAS STUCK IN THE DEVICE. IT WAS THEN REALIZED THAT THIS WIRE WAS NOT COMPATIBLE WITH THE TOME. BOTH TOME AND WIRE WERE REMOVED FROM PATIENT, AND A NON-BOSTON SCIENTIFIC DEVICE WERE ATTEMPTED TO BE USED, BUT UNABLE TO RECANNULATE THE BILE DUCT. THE PROCEDURE WAS NOT COMPLETED AND WAS CANCELED AFTER THE PATIENT HAD BEEN SEDATED WITH GENERAL ANESTHESIA. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THE COMPLAINANT REPORTED THAT THE DEVICE WAS LOADED OVER A 0.035 JAGWIRE. HOWEVER, ACCORDING TO THE DIRECTIONS FOR USE (DFU), THE TRUETOME 39 SPHINCTEROTOME IS CAPABLE OF ACCEPTING A 0.025 IN (0.64 MM) GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22548345 ↗
  • MalfunctionJun 30, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE DUODENUM DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, WHEN THE PHYSICIAN BOWED THE TRUETOME FOR THE FIRST TIME, THE CUTTING WIRE AT THE DISTAL TIP BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22355828 ↗
  • MalfunctionJun 25, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE AMPULLA OF VATER (AOV) DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2025 DURING THE PROCEDURE, WHEN THE PHYSICIAN TRIES TO CUT THE AOV, THE CUTTING WIRE OF THE TRUETOME 39 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22331167 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CUTTING WIRE DISCONNECTED. IT WAS REPORTED THAT THE CUTTING WIRE BROKE AND NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21962016 ↗
  • MalfunctionApr 30, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CUTTING WIRE WAS DISCONNECTED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936556 ↗
  • MalfunctionMar 20, 2025Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE DEVICE WAS ROTATED TWICE DURING CANNULATION AND THE CUTTING WIRE OF THE TRUETOME 39 BROKE WHILE PERFORMING SPHINCTEROTOMY. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21658312 ↗
  • MalfunctionFeb 12, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE SPHINCTER OF ODDI DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, CANNULATION WAS PERFORMED, AND THE DEVICE WAS CONNECTED TO THE GENERATOR. APPROPRIATE SETTINGS WERE MADE. THE PHYSICIAN PRESSES THE PEDAL TO START THE CUTTING AND THE CUT WIRE OF THE TRUETOME 39 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: PHOTO OF THE COMPLAINT DEVICE OUTSIDE THE PATIENT WAS PROVIDED BY THE CUSTOMER AND SHOWED THE CUTTING WIRE WAS BROKEN AND KINKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21369280 ↗
  • MalfunctionJan 10, 2025Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21123683 ↗
  • MalfunctionDec 31, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS ATTEMPTED TO BE USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE DISCONNECTED. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21050398 ↗
  • MalfunctionOct 3, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS BEING PREPARED FOR USED IN THE COMMON BILE DUCT (CBD) DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING PREPARATION, THE CUTTING WIRE BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20360052 ↗
  • MalfunctionOct 1, 2024Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS INTENDED TO BE USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PREPARATION OUTSIDE THE PATIENT, IT WAS NOTED THAT THERE WAS A HOLE IN PRODUCT PACKAGING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE REPORTED EVENT MAY SUGGEST THAT THE STERILE BARRIER WAS COMPROMISED DUE TO THE DAMAGED PACKAGING. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20346227 ↗
  • MalfunctionOct 1, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PREPARATION OUTSIDE THE PATIENT, THE CUTTING WIRE OF THE TRUETOME 39 BROKE. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20345794 ↗
  • MalfunctionSep 25, 2024Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE PAPILLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE BROKE OFF WHILE CUTTING. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER TRUETOME 39. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20301916 ↗
  • MalfunctionSep 6, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE PAPILLA DURING A PAPILLOTOMY PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE OF THE TRUETOME 39 BROKE. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20158520 ↗
  • MalfunctionSep 3, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE PAPILLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) AND ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE WAS BROKEN WHEN THE PHYSICIAN ATTEMPTED TO CUT THE PAPILLA. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20130952 ↗
  • MalfunctionAug 13, 2024Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A TRUETOME 39 WAS USED IN THE AMPULLA DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE CUTTING WIRE WAS DISCONNECTED AT THE MIDDLE PORTION. IT WAS REPORTED THAT NO PART OF THE CUTTING WIRE DETACHED AND FELL INTO THE PATIENT THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19970214 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page