ORISE™ ProKnife M00519390Recalls & reported problems
Boston Scientific Corporation · ORISE 2.0 mm Electrode - Kitted Device · Class 2
All models of ORISE ProKnife (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729995593), received between Mar 22, 2021 and Mar 22, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Oct 10, 2022Open, Classified
***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention
Z-0226-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Twisted/Bent1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 22, 2021Use of Device Problem; Detachment of Device or Device Component; Material Twisted/Bent; Use of Device Problem; Detachment of Device or Device Component; Material Twisted/Bent
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORISE PROKNIFE WAS USED IN THE ESOPHAGUS DURING AN ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, AFTER SOME USAGE, THE ELECTRODE WAS NOTICEABLY BENT INSIDE THE PATIENT. THE PHYSICIAN REMOVED THE ORISE PROKNIFE FROM THE ENDOSCOPE AND PATIENT. THE USER TRIED TO MANUALLY BEND THE ELECTRODE TO STRAIGHTEN IT, HOWEVER THE ELECTRODE BROKE AND DETACHED. THE PROCEDURE WAS COMPLETED WITH A NON-BSC RESECTION KNIFE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT AN ERBE VIO 3 GENERATOR WAS USED FOR THIS PROCEDURE. HOWEVER, ACCORDING TO THE INSTRUCTIONS FOR USE (IFU), "THE ORISE PROKNIFE IS ONLY FOR USE WITH ERBE VIO 200D/300D GENERATORS". ADDITIONALLY, THE INFORMATION PROVIDED INDICATES THAT THE CLEANING TOOL WAS USED INCORRECTLY; THE ELECTRODE WAS "BEING JAMMED INTO THE CLEANING TOOL VERY AGGRESSIVELY WITH IT EXTENDED." ACCORDING TO THE IFU DOCUMENT: "CLEANING TOOL TO BE USED WITH THE ORISE PROKNIFE REMOVED FROM THE SCOPE AND ELECTRODE RETRACTED". HOWEVER, IT WAS NOTED THAT THE ELECTRODE WAS EXTENDED WHEN THE CLEANING TOOL WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11547483 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page