ORISE™ ProKnife M00519371Recalls & reported problems
Boston Scientific Corporation · ORISE PROKNIFE 3.0 mm Electrode · Class 2
All models of ORISE ProKnife (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729999904), received between May 13, 2021 and Nov 27, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORISE PROKNIFE WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, AFTER A FEW APPLICATIONS, THE TIP CAME LOOSE FROM THE CATHETER WHEN CLEANING IT OUTSIDE THE PATIENT. THE TIP WAS FIXED WITH MICROPORE TO THE HANDLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20797458 ↗MalfunctionJun 18, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORISE PROKNIFE WAS USED IN THE URETER DURING AN ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) PROCEDURE PERFORMED ON (B)(4) 2021 DURING THE PROCEDURE, AFTER ABOUT 4 HOURS, THE ELECTRODE TIP OF THE PROKNIFE BEGAN TO DEFORM. THE ELECTRODE TIP EVENTUALLY BROKE OFF AND FELL INTO THE PATIENT. THE TIP WAS NOT REMOVED FROM THE PATIENT AND WAS LEFT TO PASS NATURALLY. THE PROCEDURE WAS COMPLETED WITH ANOTHER ORISE PROKNIFE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12024637 ↗InjuryMay 13, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ORISE PROKNIFE WAS USED IN THE GASTRIC CORPUS ON A 2CM LESION DURING AN ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE TIP OF THE KNIFE WAS BENT AND IT WAS DIFFICULT TO EXTEND AND RETRACT OUT OF THE SHEATH. HOWEVER, THE USAGE WAS CONTINUED AND AFTER ABOUT 90 MINUTES IN THE PROCEDURE, THE TIP OF THE KNIFE DETACHED INSIDE THE PATIENT. THE DEVICE WAS REMOVED FROM THE ENDOSCOPE AND THE DISLODGED ELECTRODE TIP WAS RETRIEVED USING GRASPING FORCEPS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821098 ↗
Other Unit, electrosurgical, endoscopic (with or without accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page