Dexcom, Inc. · G4 PLATINUM (Pediatric) Receiver with Share Kit US Blue · Class 3

Quick summary

This Dexcom G4 Platinum receiver has 261 adverse event reports, all of them malfunctions, with no injuries and no deaths. Reports mostly describe error messages on the screen, lost wireless connections, and alarms that did not sound, and all note no patient impact. There are no recalls that name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
261
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 261 adverse event reports that list this exact device (DI 10386270000207), received between Aug 17, 2016 and Nov 18, 2019 — classified as 261 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (112 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message112
  • Wireless Communication Problem35
  • No Audible Alarm30
  • Defective Alarm28
  • Device Inoperable23
  • Inappropriate or Unexpected Reset11
  • Unable to Obtain Readings11
  • Display or Visual Feedback Problem7

Patient problems

  • No Consequences Or Impact To Patient261

Recent adverse event reports

The 25 most recent of 261 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2019Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9337423 ↗
  • MalfunctionSep 20, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9099809 ↗
  • MalfunctionAug 16, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE TEST WAS PERFORMED AND PASSED. A RECEIVER BOOT TEST WAS PERFORMED AND FAILED DUE TO CALL TECH SUPPORT. THE LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8902586 ↗
  • MalfunctionMay 6, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8581943 ↗
  • MalfunctionApr 26, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8554868 ↗
  • MalfunctionMar 22, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN 'ERR 121' ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8444317 ↗
  • MalfunctionJan 30, 2019Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, A LOSS OF CONNECTION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE REPORTED ALLEGATION WAS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8291381 ↗
  • MalfunctionDec 11, 2018Unexpected Shutdown; Unexpected Shutdown

    DEXCOM WAS MADE AWARE ON 11/20/2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8151025 ↗
  • MalfunctionDec 11, 2018Unexpected Shutdown; Unexpected Shutdown

    DEXCOM WAS MADE AWARE ON 11/20/2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8150832 ↗
  • MalfunctionNov 20, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8091763 ↗
  • MalfunctionOct 17, 2018Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7974643 ↗
  • MalfunctionOct 9, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR CALL TECH SUPPORT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7947676 ↗
  • MalfunctionAug 20, 2018Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7797329 ↗
  • MalfunctionAug 20, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7797289 ↗
  • MalfunctionAug 7, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. DATA WAS PROVIDED FOR EVALUATION. LOSS OF CONNECTION WAS CONFIRMED. THE PROBABLE CAUSE WAS THE MOBILE DEVICE UPDATED IT¿S OPERATING SYSTEM WHICH CLOSED THE APP. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7758380 ↗
  • MalfunctionJul 30, 2018Unexpected Shutdown; Unexpected Shutdown

    DEXCOM WAS MADE AWARE ON 07/10/2018 THAT ON (B)(6) 2018 THE RECEIVER CEASED TO FUNCTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE CONFIRMATION OF THE COMPLAINT IS UNDETERMINED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7731701 ↗
  • MalfunctionJul 25, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE RECEIVER HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7720776 ↗
  • MalfunctionJul 25, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7720536 ↗
  • MalfunctionJul 19, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR 121". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7702971 ↗
  • MalfunctionJul 15, 2018Wireless Communication Problem; Wireless Communication Problem

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, A LOSS OF CONNECTION OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF A LOSS OF CONNECTION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7688773 ↗
  • MalfunctionJul 13, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBBT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7686871 ↗
  • MalfunctionJul 12, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBLE". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7684346 ↗
  • MalfunctionJul 6, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR CALL TECH SUPPORT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7666357 ↗
  • MalfunctionJun 29, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER DISPLAYED "ERR HWBAT". NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE ERROR ICON COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7653283 ↗
  • MalfunctionJun 26, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7639761 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a