CADD 21-2120-0102-02Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of CADD (123)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This ICU Medical CADD infusion pump delivers fluids or medication into the body. Its 380 reports mostly describe failure to start or deliver fluid, false alarms and degraded parts. The reports include no injuries and no deaths. Reports fell from 74 in the previous 12 months to 38 in the last 12 months. There are 3 open recalls: two concern false upstream occlusion alarms that can interrupt an infusion, and one concerns the pump's indicated uses.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 381
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 381 adverse event reports that list this exact device (DI 10610586042591), received between Jun 27, 2023 and Sep 16, 2026 — classified as 381 malfunctions.
The most frequently reported specific device problem is “Degraded” (64 reports).
3 FDA recall records name this device's identifier, 3 still open.
4 other recalls cover devices cleared under the same FDA submission (K111275); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 10, 2025Class IOpen, Classified
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Z-1667-2025 · Root cause: Labeling design · Recall details · FDA record ↗
- Apr 10, 2025Class IOpen, Classified
Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Z-1668-2025 · Root cause: Labeling design · Recall details · FDA record ↗
- Feb 27, 2024Class IOpen, Classified
The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Z-2421-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗
Most reported problems
Device problems
- Degraded64
- Failure to Infuse51
- Device Sensing Problem40
- Device Alarm System39
- Excess Flow or Over-Infusion27
- Defective Component21
- Insufficient Device Problem Information20
- Failure to Sense19
Patient problems
- No Clinical Signs, Symptoms or Conditions239
- Insufficient Information145
- Pain2
- Discomfort1
Recent adverse event reports
The 25 most recent of 381 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Inaccurate Delivery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50795957 ↗MalfunctionSep 16, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50793338 ↗MalfunctionAug 31, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50527894 ↗MalfunctionAug 26, 2026Failure to Infuse
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50461952 ↗MalfunctionAug 13, 2026Device Sensing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50241113 ↗MalfunctionJul 27, 2026Failure to Sense
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50024512 ↗MalfunctionJun 25, 2026Failure to Infuse
THE CUSTOMER RETURNED THE DEVICE STATING, "DEVICE IS FOR REPAIR". DURING DEVICE EVALUATION IT WAS FOUND LEC 44509. THIS IS A HIGH-PRIORITY ERROR INDICATING A CRITICAL CONDITION WHICH MAY RESULT IN AN INTERRUPTION OF THERAPY. THERE WAS NO PATIENT INVOLVED AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25627932 ↗MalfunctionJun 22, 2026Failure to Infuse; Obstruction of Flow
IT WAS STATED THAT PUMP TRIGGERED REPEATED DOWNSTREAM OCCLUSION ALARM. THE HIGH-PRIORITY ALARM OCCURRED BEFORE PATIENT USE. HOWEVER, IF THE ISSUE REOCCURS DURING INFUSION, IT WILL INTERRUPT THERAPY AND POTENTIALLY AFFECT THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25585695 ↗MalfunctionMay 21, 2026Crack
IT WAS REPORTED THAT AN INFUSION PUMP HAD A CRACKED DOWNSTREAM OCCLUSION SENSOR (DSO) SEAL. THERE WERE NO PATIENT INVOLVEMENT AND NO PATIENT HARM REPORTED AS THE EVENT WAS NOTED DURING PREVENTIVE MAINTENANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25246713 ↗MalfunctionMay 12, 2026Device Sensing Problem; Material Integrity Problem
AN EVENT INVOLVING A CADD PUMP REPORTED THAT THE DOWNSTREAM OCCLUSION (DSO) SEAL WAS FOUND OPEN DURING PRE-TESTING, WHICH COULD RESULT IN IMPROPER OCCLUSION DETECTION AND ALLOW FLUID TO ENTER THE PUMP, POTENTIALLY LEADING TO THERAPY INTERRUPTION. ADDITIONALLY, THE PUMP HAD A DIM LCD SCREEN AND A CRACKED BATTERY COMPARTMENT. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM/ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25149049 ↗MalfunctionMay 12, 2026Detachment of Device or Device Component; Device Sensing Problem
AN EVENT INVOLVING A CADD PUMP REPORTED THAT THE DOWNSTREAM OCCLUSION (DSO) SEAL WAS FOUND OPEN DURING PRE-TESTING, WHICH COULD RESULT IN IMPROPER OCCLUSION DETECTION AND ALLOW FLUID TO ENTER THE PUMP, POTENTIALLY LEADING TO THERAPY INTERRUPTION. ADDITIONALLY, PUMP DISPLAYED AN ERROR CODE. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25149009 ↗MalfunctionMay 12, 2026Degraded; Improper Flow or Infusion
IT WAS REPORTED THAT THE PUMP HAD A MOTOR SERVICE DUE AND A DOWNSTREAM OCCLUSION SEAL OPEN. REPORTER ALSO ALLEGED ERROR CODE 46440 WAS BEEPING. THE EVENT OCCURRED DURING PRE-TESTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25148891 ↗MalfunctionMay 12, 2026Degraded
IT WAS REPORTED THAT THE PUMP HAD A MOTOR SERVICE DUE AND A DOWNSTREAM OCCLUSION (DSO) SEAL OPEN. THE EVENT OCCURRED DURING PRE-TESTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25148790 ↗MalfunctionMay 12, 2026Degraded
IT WAS REPORTED THAT THE PUMP HAD A MOTOR SERVICE DUE AND A DOWNSTREAM OCCLUSION (DSO) SEAL OPEN. THE EVENT OCCURRED DURING PRE-TESTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25148788 ↗MalfunctionMay 5, 2026Failure to Infuse
IT WAS REPORTED THAT THE PUMP HAD DISPLAYED RED SCREEN WITH AN ERROR CODE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25081454 ↗MalfunctionApr 27, 2026Device Sensing Problem
AN EVENT INVOLVED A CADD-SOLIS VIP AMBULATORY INFUSION PUMP REPORTED THAT THE PUMP DOES NOT DETECT THE CASSETTE ANYMORE. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24998768 ↗MalfunctionApr 15, 2026Device Sensing Problem
IT WAS REPORTED THAT THE PUMP HAD AN EVENT OF ALARM: CASSETTE IMPROPERLY INSTALLED. REINSTALL THE CASSETTE. WHICH HAPPENED REGARDLESS OF THE CASSETTE INSTALLED PROPERLY. THE ALARM SOMETIMES REMAINS ACTIVE EVEN AFTER REMOVING THE CASSETTE. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24885058 ↗MalfunctionApr 8, 2026Failure to Sense
IT WAS REPORTED THAT THE PUMP REPEATEDLY INDICATED THE CASSETTE WAS NOT ATTACHED PROPERLY AND WAS REATTACHED. THERE WAS UNKNOWN PATIENT INVOLVEMENT, NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818775 ↗MalfunctionMar 18, 2026Excessive Heating
IT WAS REPORTED THAT UNIT OVERHEATED DURING PATIENT ADMINISTRATION. THERE WAS PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24631809 ↗MalfunctionMar 18, 2026Failure to Infuse
THE DEVICE WAS RETURNED AS IT WAS REPORTED THAT THERE WAS AN ERROR CODE FAILURE 42224 SYSTEM DURING POWER UP. THE RESULTS OF THE INVESTIGATION CONFIRMED THE REPORTED ISSUE IN THE DEVICE'S HISTORY LOG. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24629824 ↗MalfunctionMar 13, 2026Calibration Problem
IT WAS REPORTED THAT THE FLOW RATE IS OUT OF TOLERANCE. THERE WAS NO PATIENT INVOLVEMENT AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597391 ↗MalfunctionMar 12, 2026Failure to Infuse
IT WAS REPORTED THAT THE DEVICE EXHIBITED ERROR CODE 46507. THE TIME DOES NOT INCREMENT WHEN THE DEVICE IS CLOSED (BATTERY FOR THE INTERNAL CLOCK NEEDS REPLACING). THERE WAS UNKNOWN REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24581067 ↗MalfunctionMar 11, 2026Failure to Infuse
THE CUSTOMER RETURNED THE PRODUCT FOR ERROR CODE 45169 AND ALARMING. WATER/LIQUID (DISPLAY LCD) DEFECTIVE, MISSING INSULATORS POGO PIN. DURING DEVICE EVALUATION, ERROR CODE 44000 AND A RED SCREEN WAS IDENTIFIED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569077 ↗MalfunctionMar 9, 2026Failure to Infuse; Device Sensing Problem
IT WAS FOUND DURING THE INVESTIGATION OF A RETURNED PUMP THAT ALARM FALL OUT LATCH WON'T TOGGLE ERROR WAS REPLICATED UPON POWERING THE DEVICE. ADDITIONALLY, WATER SPILLS OBSERVED ON LCD DISPLAY DURING PRE-TEST. THERE WAS NO PATIENT INVOLVEMENT, AND NO HARM/ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24550657 ↗MalfunctionFeb 18, 2026Defective Alarm; Complete Blockage; Defective Component
IT WAS STATED THAT THE DEVICE WAS BEEPING IN OCCLUSION WITH NO REASON AND THE CUSTOMER HAD TO REMOVE THE BATTERY TO TURN IT OFF. DURING INVESTIGATION FOUND THE FOUND DAMAGED DSO SEAL. THIS MALFUNCTION MAKES THE EVENT REPORTABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24383192 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111275. One submission often covers many models, so check the product and lot details in each record.
- Z-1682-2025 · Apr 10, 2025 · Open, ClassifiedCADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP
- Z-0222-2025 · Jul 16, 2024 · Open, ClassifiedCADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-
- Z-0466-2025 · Jun 26, 2024 · Open, Classifiedsmiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
- Z-0107-2022 · Mar 24, 2020 · TerminatedCADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 2
Other Pump, infusion devices
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
- Alaris 8100Carefusion 303, Inc. · Pump, infusion4 recalls78 deaths344,514 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion2 recalls45 deaths148,093 reports
- BD® Alaris™ Pump Module, Model 8100Carefusion 303, Inc. · Pump, infusion5 recalls30 deaths51,256 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls4 deaths21,956 reports
- Alaris 8110Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths15,547 reports
- Alaris 8015Carefusion 303, Inc. · Pump, infusion3 recalls6 deaths13,022 reports
- CADD 21-6500-51Icu Medical, Inc. · Pump, infusion2 deaths8,231 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page