PROXIMATERecalls & reported problems across all models
Ethicon Endo-Surgery, LLC · 41 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2,027
- Deaths reported
- 22
- Injuries reported
- 585
- Recalls
- 0
We found 41 models of PROXIMATE by Ethicon Endo-Surgery, LLC in the FDA's device registry (GUDID). Together they are named in 2,027 adverse event reports in the FDA's MAUDE database, received between Apr 1, 2013 and Sep 30, 2026 — classified as 22 deaths, 585 injuries, 1,420 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (572 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported PROXIMATE problems
Device problems
- Failure to Form Staple572
- Failure to Fire514
- Adverse Event Without Identified Device or Use Problem403
- Mechanics Altered133
- Firing Problem108
- Difficult to Open or Close74
- Insufficient Device Problem Information62
- Failure to Cut57
Patient problems
- No Clinical Signs, Symptoms or Conditions891
- No Consequences Or Impact To Patient537
- Failure to Anastomose223
- No Code Available217
- Blood Loss146
- Hemorrhage/Blood Loss/Bleeding136
- Pain116
- Wound Dehiscence52
PROXIMATE models
41 models, most reported first. Each model has its own page with its reports, recalls and device details.
- PROXIMATE TLC75STAPLE, IMPLANTABLE14 deaths507 reports
- Proximate PPH03STAPLE, IMPLANTABLE389 reports
- PROXIMATE TX60BSTAPLE, IMPLANTABLE3 deaths171 reports
- PROXIMATE TCR75STAPLE, IMPLANTABLE142 reports
- Proximate PMW35STAPLE, REMOVABLE (SKIN)131 reports
- PROXIMATE TLC10STAPLE, IMPLANTABLE1 death98 reports
- Proximate PPH01STAPLE, IMPLANTABLE96 reports
- PROXIMATE TX30VSTAPLE, IMPLANTABLE63 reports
- PROXIMATE TX60GSTAPLE, IMPLANTABLE45 reports
- Proximate TCR10STAPLE, IMPLANTABLE43 reports
- PROXIMATE TLC55STAPLE, IMPLANTABLE41 reports
- PROXIMATE TX30GSTAPLE, IMPLANTABLE39 reports
- Proximate PXW35STAPLE, REMOVABLE (SKIN)37 reports
- PROXIMATE TX30BSTAPLE, IMPLANTABLE27 reports
- Proximate PRW35STAPLE, REMOVABLE (SKIN)20 reports
- PROXIMATE TCT75STAPLE, IMPLANTABLE20 reports
- PROXIMATE XR30VSTAPLE, IMPLANTABLE3 deaths20 reports
- PROXIMATE TCR55STAPLE, IMPLANTABLE18 reports
- PROXIMATE TLH30STAPLE, IMPLANTABLE1 death17 reports
- PROXIMATE TL90STAPLE, IMPLANTABLE15 reports
- PROXIMATE TRT10STAPLE, IMPLANTABLE14 reports
- PROXIMATE TL60STAPLE, IMPLANTABLE10 reports
- Proximate TCT10STAPLE, IMPLANTABLE9 reports
- PROXIMATE TLH90STAPLE, IMPLANTABLE8 reports
- PROXIMATE TLH60STAPLE, IMPLANTABLE7 reports
- PROXIMATE TVC55STAPLE, IMPLANTABLE6 reports
- PROXIMATE XR60BSTAPLE, IMPLANTABLE6 reports
- PROXIMATE TL30STAPLE, IMPLANTABLE4 reports
- PROXIMATE TRT75STAPLE, IMPLANTABLE4 reports
- PROXIMATE TRD75STAPLE, IMPLANTABLE3 reports
- PROXIMATE XR30GSTAPLE, IMPLANTABLE3 reports
- Proximate TCD75STAPLE, IMPLANTABLE2 reports
- PROXIMATE TVR55STAPLE, IMPLANTABLE2 reports
- PROXIMATE XR30BSTAPLE, IMPLANTABLE2 reports
- PROXIMATE XR60GSTAPLE, IMPLANTABLE2 reports
- Proximate PXR35STAPLE, REMOVABLE (SKIN)1 report
- Proximate TCT55STAPLE, IMPLANTABLE1 report
- PROXIMATE TLV30STAPLE, IMPLANTABLE1 report
- PROXIMATE TR30STAPLE, IMPLANTABLE1 report
- PROXIMATE TR60STAPLE, IMPLANTABLE1 report
- PROXIMATE TR90STAPLE, IMPLANTABLE1 report
Recent PROXIMATE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Failure to Form Staple; Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946430 ↗MalfunctionSep 21, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50840738 ↗MalfunctionSep 15, 2026Failure to Form Staple
It was reported that during an unknown surgery, the TX60B stapler did not form the staples correctly. Was there any harm to the reported patient or user? No. Was there a significant delay? Yes. How long was it delayed (minutes)? 30-50 minutes, they had to do the anastomosis manually, as the staples were not formed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50772686 ↗MalfunctionSep 8, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50668577 ↗MalfunctionAug 12, 2026Failure to Form Staple; No Apparent Adverse Event
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 50222145 ↗InjuryJul 31, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
It was reported that Postoperatively, a staple line leak was observed. The patient required reoperation, involving the re-transection of the bowel. The anastomosis was re-stapled. The patients stay is extended due to the reoperation. The patient is currently stable.
Patient problems: Failure to Anastomose
MAUDE report 50101922 ↗InjuryJul 31, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
It was reported that Postoperatively, a staple line leak was observed. The patient required reoperation, involving the re-transection of the bowel. The anastomosis was re-stapled. The patients stay is extended due to the reoperation. The patient is currently stable.
Patient problems: Failure to Anastomose
MAUDE report 50101825 ↗InjuryJul 7, 2026Failure to Form Staple; Failure to Fire; Difficult to Open or Close; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED BY THE SALES REP THAT DURING A LEFT UPPER LOBECTOMY PROCEDURE, THE PATIENT¿S LOBE WAS DESCRIBED AS THICK, PREVENTING THE STAPLER FROM CLOSING INITIALLY. THE SURGEON SUBSEQUENTLY FORCED THE DEVICE TO CLOSE AND HELD COMPRESSION FOR TWO MINUTES. UPON FIRING, A LOUD POP WAS HEARD, AND THE KNIFE DID NOT ADVANCE FURTHER. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25738995 ↗MalfunctionJun 18, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING GASTRECTOMY PROCEDURE AFTER FIRED, NOTED THE STAPLES MALFORMED. USED SUTURE TO OVERSEW. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25553290 ↗MalfunctionJun 15, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503848 ↗MalfunctionJun 15, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING THE UNK SURGERY, AFTER FIRED, NOTED THE STAPLES MALFORMED. CHANGED THE NEW ONE TO COMPLETE SURGERY. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25503798 ↗InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING AN ILEOCECECTOMY PROCEDURE, THE SURGEON CLOSED THE COMMON ENTEROTOMY USING A TX STAPLER. ON POSTOPERATIVE DAY 1, THE PATIENT DEVELOPED A LEAK AND WAS TAKEN BACK TO SURGERY. UPON RE-EXPLORATION, IT WAS OBSERVED THAT THE STAPLED SITE APPEARED COMPLETELY OPEN, MEASURING APPROXIMATELY 3¿4 CM IN LENGTH. THE SURGEON SUSPECTED A POTENTIAL FAILURE OF THE TX STAPLER; HOWEVER, NO ISSUES WERE NOTED WITH THE DEVICE DURING THE INITIAL PROCEDURE.
Patient problems: Failure to Anastomose
MAUDE report 25331327 ↗MalfunctionMay 25, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE STAPLERS DID NOT DEPLOY STAPLES WHEN FIRED. NO PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25270385 ↗MalfunctionMay 14, 2026Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, A PIN FELL OUT FROM THE DEVICE AND DEVICE WAS UNABLE TO BE USED AFTERWARDS. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25176112 ↗MalfunctionMay 5, 2026Structural Problem
IT WAS REPORTED THAT DURING A SMALL-BOWEL RESECTION, WHEN REMOVING THE INSTRUMENT¿S ORANGE PROTECTIVE COVER, THE STAPLES CAME OUT OF THE CARTRIDGE SPONTANEOUSLY. OPENING AN INDIVIDUAL CARTRIDGE PRODUCED THE SAME PROBLEM: THE STAPLES CAME OUT OF IT. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25083370 ↗
Other Ethicon Endo-Surgery, LLC products
- Harmonic ACE9 models · 12,862 reports
- ECHELON FLEX7 models · 9,393 reports · 1 recall
- Harmonic13 models · 8,752 reports · 2 recalls
- ECHELON ENDOPATH31 models · 6,543 reports · 2 recalls
- LIGACLIP26 models · 3,254 reports
- ENDOPATH37 models · 2,785 reports
- ECHELON CIRCULAR4 models · 2,532 reports
- ECHELON Flex ENDOPATH7 models · 2,334 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page