Skip to content

PROXIMATE TX60BRecalls & reported problems

Ethicon Endo-Surgery, LLC · Proximate Reloadable Linear Stapler · Class 2

All models of PROXIMATE (41)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Proximate TX60B reloadable linear stapler is a surgical stapler used to join tissue during surgery. Of its 170 reports, 44 are filed as injuries and 3 as deaths; many describe staples that do not form or fire, and some describe staple lines that opened, requiring repeat surgery. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
171
Deaths reported
3
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 171 adverse event reports that list this exact device (DI 10705036003342), received between Dec 15, 2017 and Aug 12, 2026 — classified as 3 deaths, 45 injuries, 123 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (65 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple65
  • Failure to Fire54
  • Adverse Event Without Identified Device or Use Problem33
  • Insufficient Device Problem Information8
  • Difficult to Open or Close5
  • Delivered as Unsterile Product4
  • Misfire4
  • Device Operates Differently Than Expected3

Patient problems

  • No Clinical Signs, Symptoms or Conditions71
  • No Consequences Or Impact To Patient53
  • Failure to Anastomose33
  • No Code Available12
  • Hemorrhage/Blood Loss/Bleeding4
  • Unspecified Tissue Injury4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Blood Loss2

Recent adverse event reports

The 25 most recent of 171 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 12, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that during a right hemi colectomy on (b)(6)2026. He used GLC and TX60B. The staple lines looked good during the case, but the patient came back with an anastomotic leak and had a revision on (b)(6)2026. During the revision, the GLC staple line looked good but the whole TX staple line was open. He is not sure if it was due to the stapler and wasn't necessarily fully blaming the stapler as it could've just been patient factors. There was patient consequence reported. The patient needed a revision surgery to fix the leak. As stated, we are not 100% sure if the stapler is at fault for the leak.

    Patient problems: Failure to Anastomose

    MAUDE report 50227561 ↗
  • InjuryAug 5, 2026Failure to Form Staple; Adverse Event Without Identified Device or Use Problem

    It was reported that during an Unknown procedure that the staple line unraveled during a leak test. They had to over-sew the staple line that did not pass the test, then had to create a stoma. The patient will have to have re-operate in 3 months.

    Patient problems: Failure to Anastomose

    MAUDE report 50142867 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING AN ILEOCECECTOMY PROCEDURE, THE SURGEON CLOSED THE COMMON ENTEROTOMY USING A TX STAPLER. ON POSTOPERATIVE DAY 1, THE PATIENT DEVELOPED A LEAK AND WAS TAKEN BACK TO SURGERY. UPON RE-EXPLORATION, IT WAS OBSERVED THAT THE STAPLED SITE APPEARED COMPLETELY OPEN, MEASURING APPROXIMATELY 3¿4 CM IN LENGTH. THE SURGEON SUSPECTED A POTENTIAL FAILURE OF THE TX STAPLER; HOWEVER, NO ISSUES WERE NOTED WITH THE DEVICE DURING THE INITIAL PROCEDURE.

    Patient problems: Failure to Anastomose

    MAUDE report 25331327 ↗
  • MalfunctionMay 25, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE STAPLERS DID NOT DEPLOY STAPLES WHEN FIRED. NO PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25270385 ↗
  • MalfunctionMay 4, 2026Failure to Fire

    IT WAS REPORTED THAT DURING A COLECTOMY, THE STAPLER FIRED ON COLON TISSUE AS USUAL, BUT NO STAPLES DEPLOYED FROM STAPLER. NO STAPLES COULD BE VISUALIZED IN STAPLER. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073134 ↗
  • InjuryApr 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A LEFT OPEN SIGMOID COLECTOMY ON MARCH 20TH. PATIENT WAS TAKEN BACK INTO SURGERY ON MARCH 27TH. AFTER REOPENING PATIENT, THE SURGEON SAID THAT THE STAPLE LINE WAS GAPING OPEN.

    Patient problems: Failure to Anastomose

    MAUDE report 24840957 ↗
  • MalfunctionMar 30, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE SURGEON CLOSED THE STAPLER ON NORMAL THICKNESS SMALL INTESTINE AND FIRED IT. THE STAPLER APPEARED TO FIRE FINE. UPON REMOVAL, THE SURGEON NOTED THAT THE DISTAL END OF THE STAPLE LINE HAD COMPLETELY UNFORMED STAPLES, BUT THE PROXIMAL SIDE WAS CORRECTLY FORMED. THE SURGEON WAS ABLE TO EXTRACT SEVERAL STAPLES THAT HAD NOT FORMED. THE SURGEON STATE THAT THE TISSUE WAS NOT ABNORMALLY THICK, AND THE STAPLER APPEARED TO FIRE NORMALLY, AND THAT HE HELD FOR TISSUE COMPRESSION. THEY OPENED A NEW STAPLER TO COMPLETE THE PROCEDURE. NO HARM OR DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24735815 ↗
  • InjuryFeb 20, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE CLAMP WAS PLACED AND THE STAPLER WAS FIRED. THE OPERATOR THEN PERFORMED A TRANSECTION USING THE COLD BLADE AT THE HIGH RECTAL LEVEL. WHEN THE CLAMP WAS OPENED, IT WAS FOUND THAT THERE WERE NO STAPLES IN THE CARTRIDGE. THE RECTUM THEN ENDED UP OPEN INTO THE ABDOMINAL CAVITY. THE SURGEON HAD TO PERFORM A LOWER TRANSECTION AT THE MID-RECTAL LEVEL. AS A CONSEQUENCE, THE PATIENT REQUIRED THE CREATION OF A PROTECTIVE ILEOSTOMY, WHICH HAD NOT BEEN PLANNED IN THE INITIAL OPERATIVE STRATEGY. A DIFFERENT TX60G CLAMP WAS THEN USED TO TRANSECT THE MID-RECTUM.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 24401741 ↗
  • MalfunctionFeb 4, 2026Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING AN UNK PROCEDURE THE STAPLER PAPER FELT TOO THIN AND RIPPED EASILY - SEEMED TO NOT HAVE INTEGRITY. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260778 ↗
  • MalfunctionOct 20, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING A COLON RESECTION THAT THE TX WAS APPLIED TO THE BOWEL AND FIRED. WHEN THE STAPLED AREA WAS EXAMINED, NONE OF THE STAPLES WERE FORMED PROPERLY. THE MAJORITY OF THE STAPLE LEGS WERE STRAIGHT WITH ONLY A FEW THAT WERE FORMED IN A C SHAPE. STAPLE LINE WAS OVERSEWN. PROCEDURE WAS DELAYED BY 25 MINUTES. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23336385 ↗
  • MalfunctionSep 9, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLES DID NOT FULLY CLOSE. THE DOCTOR WANTED TO STAPLE THE COLON AND NOT ALL THE STAPLES WENT THROUGH AND CLOSED PROPERLY. THEY PROCEEDED TO USE REGULAR SUTURES TO MAKE SOME POINTS. THEY INCLUDED SOME OF THE STAPLES AS PROOF THAT THEY WERE MALFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23003233 ↗
  • InjuryJul 8, 2025Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT WITH HISTORY OF COPD, CORONARY ARTERY DISEASE, OBESITY AND SURGICALLY TREATED PROSTATE CANCER WHO UNDERWENT RIGHT HEMICOLECTOMY, INCISIONAL HERNIA REPAIR, AND PARTIAL OMENTECTOMY TO TREAT CECAL AND ASCENDING COLON CANCER. OPERATIVE NOTE INDICATES MASS WAS JUST PROXIMAL TO THE CECUM AND THE ASCENDING COLON AND THE RIGHT COLON WAS MOBILIZED TO THE LEVEL OF THE HEPATIC FLEXURE. THE ILEUM WAS ALSO MOBILIZED GIVEN EXTENSIVE ADHESIONS. ¿LIGASURE¿ WAS USED TO TRANSECT TO DIVIDE THE RIGHT COLON AND A ¿75MM GIA STAPLER¿ WAS USED TO TRANSECT THE ILEUM AND MID TRANSVERSE COLON AND TO CREATE THE ANASTOMOSIS. A ¿TX60¿ WAS USED TO CLOSE THE ENTEROTOMY SITES. NO DIFFICULTIES WERE NOTED WITH ANY STAPLERS. 4 DAYS POST-OPERATIVELY, THE PATIENT SHOWED SIGNS OF LEAK AND UNDERWENT AN EXPLORATORY LAPAROTOMY WITH CREATION OF ILEOSTOMY AND DRAINAGE OF INTRA-ABDOMINAL ABSCESS. OPERATIVE NOTE INDICATES A ¿4 MM AREA OF DEHISCENCE OF BOWEL CONSISTENT WITH THE SITE OF HIS ANASTOMOTIC LEAK.¿ AN ILEOSTOMY WAS CREATED AND PATIENT WAS DISCHARGED ONE WEEK LATER. 10 DAYS FOLLOWING DISCHARGE, HE PRESENTED WITH A LEAKING STOMA. POST-OPERATIVELY, THE PATIENT¿S ILEOSTOMY HAD BECOME

    Patient problems: Abscess; Adhesion(s); Failure to Anastomose; Hernia; Necrosis

    MAUDE report 22439856 ↗
  • MalfunctionApr 2, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT DURING A SMALL BOWEL RESECTION THE PACKAGING TORE IN 2 LOCATIONS WHEN THE RN OPENED THE DEVICE. PROCEDURE WAS COMPLETED WITHOUT ANY SURGICAL DELAY. NO PATIENT CONSEQUENCES WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21751345 ↗
  • MalfunctionMar 6, 2025Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE CHARGER PRESENT ON THE CLAMP DID NOT WORK DURING STAPLING. OPENING A NEW CHARGER FROM AN INDIVIDUAL PACKAGE. NO PATIENT CONSEQUENCE. THE TRIGGER ADVANCEDBUT NO STAPLES WERE DEPLOYED ON THE DIGESTIVE TUBE. AND ONCE THE LOADER WAS REMOVED FROM THE CLAMP THERE WERE B-SHAPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21535492 ↗
  • InjuryNov 26, 2024Failure to Cut; Failure to Fire; Difficult to Open or Close; Adverse Event Without Identified Device or Use Problem; Noise, Audible; Firing Problem; Failure to Cut; Failure to Fire; Difficult to Open or Close; Adverse Event Without Identified Device or Use Problem; Noise, Audible; Firing Problem

    IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE THAT PRIOR TO CLOSURE THE LEAK TEST WAS PREFORMED AND A LEAK WAS FOUND. ON THE ATTEMPT TO FIX THE LEAK THE SURGEON WENT DISTAL TO THE FIRST STAPLE LINE WITH THE TX60B BUT THE STAPLER WAS DIFFICULT TO CLOSE AND IT MAKES A FUNNY NOISE. THE SURGEON "FORCED" IT TO FIRE AND WHEN EXAMINED, NO STAPLES HAD BEEN DEPLOYED AND THE KNIFE DID NOT CUT. A SECOND TX60B WAS OPENED. A SECOND LEAK TEST WAS DONE AND NO LEAK WAS DETECTED. THE SURGEON WAS NOT COMFORTABLE WITH THE OUTCOME SO ERRING ON THE SIDE OF CAUTION AND DID A DIVERSION AND CREATED AN ILEOSTOMY. THE PATIENT WILL RETURN IN THREE TO SIX MONTHS TO HAVE THE STOMA CLOSED.

    Patient problems: Failure to Anastomose

    MAUDE report 20777976 ↗
  • InjuryOct 31, 2024Failure to Form Staple; Failure to Fire; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Failure to Fire; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT POST OP OF A ROBOTIC SIGMOID COLECTOMY PROCEDURE THAT THE PATIENT WAS BROUGHT BACK TO THE OR DUE TO A HOLE IN THE STAPLE LINE. THE HOLE WAS REPAIRED WITH SUTURE AND A DRAIN WAS PLACED. PATIENT CURRENT STATUS UNKNOWN BUT PATIENT WAS STABLE AFTER THE PROCEDURE.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 20577276 ↗
  • MalfunctionSep 24, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A PART RESECTION FOR THE SMALL INTESTINE, THEY USED THE TX60B, BUT THE STAPLE LINE WAS NOT COMPLETE. THEY TOOK ANOTHER XR60B RELOAD WITH THE SAME RESULT. THEY OPEN ANOTHER TX60B AND THEY FINISHED THE OPERATION WITHOUT ANY PROBLEMS. THE PATIENT SUFFERED NO HARM!

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20291150 ↗
  • MalfunctionSep 4, 2024Delivered as Unsterile Product; Difficult to Open or Remove Packaging Material; Delivered as Unsterile Product; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE PACKAGING/TYVEK TORE UNEVENLY COMPROMISING STERILITY ON TX60B. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20135210 ↗
  • MalfunctionSep 3, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A SIGMOIDOSCOPY PROCEDURE, DOCTOR FIRED STAPLER TO CLOSE AND NO STAPLES CAME OUT. UPON OPENING THE STAPLER THEY SAW THAT NO STAPLES HAD FIRED. OPENED ANOTHER STAPLER TO COMPLETE CASE. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20127656 ↗
  • MalfunctionAug 22, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLE WAS MALFORMED AFTER FIRING.THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20035128 ↗
  • MalfunctionAug 8, 2024Failure to Form Staple; Insufficient Device Problem Information; Failure to Form Staple; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY THE TX60G STAPLES DIDN'T DEPLOY PROPERLY. SURGEON HAD TO TAKE ANOTHER TX TO COMPLETE THE CASE. BOWEL INJURY WAS REPORTED, BUT NOT ELABORATED. OPENED ANOTHER TX60 DEVICE TO COMPLETE THE CASE, THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 19941594 ↗
  • MalfunctionJul 19, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE AND AFTER FIRING, STAPLES WAS NOT FORMED WELL ON TISSUE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19782051 ↗
  • InjuryJun 25, 2024Failure to Fire; Failure to Fire

    SEE ATTACHED MW (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19605649 ↗
  • MalfunctionMar 29, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A LOOP ILEOSTOMY CLOSURE, SURGEON FIRED THE DEVICE AND NO STAPLES PRESENTED FROM THE PROXIMAL/BOTTOM END OF THE JAW. THEREFORE THE OSTOMY REMAINED PARTIALLY OPEN. SURGEON RECOGNIZED IMMEDIATELY UPON EXAMINING THE STAPLE LINE AND OVERSEWED THE TISSUE TO COMPLETE THE CLOSURE. THE SURGERY WAS DELAYED 2 MINUTES. SEWED THE REMAINING TISSUE CLOSED WITH SUTURE, THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19002810 ↗
  • InjuryJan 19, 2024Failure to Fire; Failure to Fire

    IT WAS REPORTED THAT DURING A COLECTOMY THE TX 60B DID NOT FIRE OUT STAPLES. PATIENT REQUIRED A COLOSTOMY BAG. THERE WAS A 20 MINUTE DELAY FROM THE RESULT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18546166 ↗

FDA clearance and classification

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the PROXIMATE TX60B

Has the PROXIMATE TX60B been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036003342), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PROXIMATE TX60B?

The FDA's MAUDE database lists 171 adverse event reports for this device, received between Dec 15, 2017 and Aug 12, 2026. In those reports, the most common specific device problems are “Failure to Form Staple” (65), “Failure to Fire” (54) and “Difficult to Open or Close” (5); the most common patient problems are “Failure to Anastomose” (33), “Hemorrhage/Blood Loss/Bleeding” (4) and “Unspecified Tissue Injury” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PROXIMATE TX60B?

3 reports are classified as deaths, 45 as injuries and 123 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PROXIMATE TX60B MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the PROXIMATE TX60B, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand PROXIMATE, model TX60B). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other STAPLE, IMPLANTABLE devices

All STAPLE, IMPLANTABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page