PROXIMATE TLC75Recalls & reported problems
Ethicon Endo-Surgery, LLC · Proximate Reloadable Linear Cutter With Safety Lock-Out, 75mm · Class 2
Quick summary
This Ethicon Proximate surgical stapler is used to cut and close tissue during surgery, and its 510 reports often describe the stapler failing to fire or form staples, with staple lines coming apart. Several say the blade stopped partway or the staples came out of the cartridge, and some surgeons switched to another stapler to finish. Reports include 14 deaths and 176 injuries, though many report no patient harm. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 507
- Deaths reported
- 14
- Injuries reported
- 175
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 507 adverse event reports that list this exact device (DI 10705036013211), received between Jul 21, 2015 and Jul 7, 2026 — classified as 14 deaths, 175 injuries, 318 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (150 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire150
- Adverse Event Without Identified Device or Use Problem139
- Failure to Form Staple135
- Difficult to Open or Close27
- Mechanical Jam17
- Firing Problem16
- Insufficient Device Problem Information15
- Patient Device Interaction Problem13
Patient problems
- No Clinical Signs, Symptoms or Conditions217
- Failure to Anastomose131
- No Consequences Or Impact To Patient100
- No Code Available45
- Hemorrhage/Blood Loss/Bleeding29
- Blood Loss16
- Insufficient Information12
- Unspecified Infection10
Recent adverse event reports
The 25 most recent of 507 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2026Failure to Form Staple; Failure to Fire; Difficult to Open or Close; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED BY THE SALES REP THAT DURING A LEFT UPPER LOBECTOMY PROCEDURE, THE PATIENT¿S LOBE WAS DESCRIBED AS THICK, PREVENTING THE STAPLER FROM CLOSING INITIALLY. THE SURGEON SUBSEQUENTLY FORCED THE DEVICE TO CLOSE AND HELD COMPRESSION FOR TWO MINUTES. UPON FIRING, A LOUD POP WAS HEARD, AND THE KNIFE DID NOT ADVANCE FURTHER. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25738995 ↗MalfunctionMay 5, 2026Structural Problem
IT WAS REPORTED THAT DURING A SMALL-BOWEL RESECTION, WHEN REMOVING THE INSTRUMENT¿S ORANGE PROTECTIVE COVER, THE STAPLES CAME OUT OF THE CARTRIDGE SPONTANEOUSLY. OPENING AN INDIVIDUAL CARTRIDGE PRODUCED THE SAME PROBLEM: THE STAPLES CAME OUT OF IT. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25083370 ↗InjuryMay 4, 2026Use of Device Problem
IT WAS REPORTED THAT AFTER-HOURS ACUTE CASE (UROLOGY RE-LOOK PLUS BOWEL RESECTION), THE COLORECTAL COMPONENT WAS PERFORMED BY DR. DURING THE BOWEL RESECTION, A SIGNIFICANT ISSUE OCCURRED INVOLVING THE ETHICON TLC75 DEVICE. THE DEVICE WAS FIRED ONCE WITHOUT INCIDENT AND THE PROCEDURE CONTINUED. THE DEVICE WAS THEN CALLED UPON FOR A SECOND FIRING. THE ASSISTING NURSE (ORTHOPEDIC NURSE, UNFAMILIAR WITH THE DEVICE) HAD NOT, OR DID NOT RECALL TO, UNLOAD, CLEAN, AND RELOAD THE DEVICE. AS A RESULT, THE DEVICE WAS FIRED A SECOND TIME WITH THE FIRST SPENT RELOAD STILL IN PLACE. CONCERNINGLY, THE DEVICE DID NOT LOCK OUT OR OTHERWISE PREVENT FIRING WITH A PREVIOUSLY USED RELOAD. THE KNIFE TRANSECTED THE BOWEL, HOWEVER NO STAPLES WERE DEPLOYED. UPON FIRING, THE SURGEON NOTED AN ABSENCE OF THE NORMAL RESISTANCE HE WOULD EXPECT, BEING VERY FAMILIAR WITH THE DEVICE. THE STANDARD PRESSURE HE APPLIED RESULTED IN THE KNIFE RAPIDLY ADVANCING TO THE END OF THE KNIFE TRACK. ON REFLECTION, THE SURGEON ALSO NOTED THAT THE CLOSING HANDLE PRIOR TO THE SECOND FIRING OPENED WIDER THAN WOULD NORMALLY BE EXPECTED, OPENING TO APPROXIMATELY 90 DEGREES RELATIVE TO THE INSTRUMENT BODY. DR RECTIFIED THE SITUATION B
Patient problems: Insufficient Information
MAUDE report 25073267 ↗MalfunctionApr 24, 2026Failure to Fire
IT WAS REPORTED THAT, DURING GASTRECTOMY, THE DEVICE WAS ATTEMPTED TO FIRE, BUT THE DEVICE COULD NOT BE FIRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24982148 ↗MalfunctionApr 24, 2026Failure to Fire
IT WAS REPORTED THAT DURING AND UNK PROCEDURE THE STAPLER FIRING, THE BLADE LOCKED EARLY, WHERE IT WAS NOT POSSIBLE TO COMPLETE THE SHOT OR THE FORMATION OF A STAPLE LINE. THE STAPLER WAS REPLACED AND THE SURGERY WAS COMPLETED. IN THE FIRING THERE WERE NO STAPLES THAT CAME LOOSE OR DAMAGE TO THE SURGERY. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24982107 ↗MalfunctionApr 9, 2026Manufacturing, Packaging or Shipping Problem
IT WAS REPORTED THAT BEFORE AN UNK PROCEDURE, WHEN THEY REMOVED THE PRODUCT FROM PACKAGING BOX, IT WAS BROKEN. THE METAL COMPONENT SEPARATED FROM THE PLASTIC COMPONENT. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24831232 ↗MalfunctionApr 9, 2026Tear, Rip or Hole in Device Packaging
IT WAS REPORTED THAT BEFORE AN UNK PROCEDURE, THE OUTER PLASTIC BOX IS DEFECTIVE. THE HANDLE ON THE UPPER PART IS REMOVED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24831206 ↗InjuryApr 6, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE DOCTOR REPORTS DELAYED ACTIVE BLEEDING 48 HOURS AFTER SURGERY AT THE STAPLE LINE. THE ANTICOAGULANT WAS DISCONTINUED IN ORDER TO STABILIZE HIM. STABILIZATION WAS ACHIEVED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24792679 ↗MalfunctionMar 26, 2026Mechanical Jam
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, CAN'T ADVANCE FIRING KNOB. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24703478 ↗MalfunctionMar 23, 2026Failure to Fire
IT WAS REPORTED THAT DURING COLON RESECTION PROCEDURE, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24665694 ↗MalfunctionMar 23, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24665460 ↗MalfunctionMar 16, 2026Mechanical Jam
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE STAPLE WHICH INSERTS INTO INTRALUMINAL, STAPLER JAMMED/THE BLADE OF STAPLE WOULD NOT RETRACT AND WAS PROTRUDING OUT ON STAPLE AFTER USE AND WAS NOT ABLE TO BE REMOVED FROM THE STAPLER ITSELF SO ANOTHER HAD TO BE OPENED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24610755 ↗InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, A PATIENT CAME BACK WITH POSTOP GI BLEED AFTER USING THE TLC75. CAME BACK 2 TO 3 WEEKS POST OP.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24475030 ↗InjuryFeb 27, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, A PATIENT CAME BACK WITH POSTOP GI BLEED AFTER USING THE TLC75. CAME BACK 2 TO 3 WEEKS POST OP.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24475024 ↗InjuryFeb 24, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A COLECTOMY WITH COLOSTOMY AND ANASTOMOSIS PROCEDURE. A SMALL ANASTOMOTIC LEAK WITH ASSOCIATED ABSCESS WAS SUBSEQUENTLY IDENTIFIED. THE DEVICE USED DURING THE PROCEDURE WAS DISCARDED AND IS NOT AVAILABLE FOR RETURN OR EVALUATION. NO OTHER INFORMATION PROVIDED.
Patient problems: Abscess; Failure to Anastomose; Post Operative Wound Infection
MAUDE report 24433279 ↗MalfunctionFeb 24, 2026Failure to Form Staple; Mechanical Jam
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY THE THIN FIRING KNOBS WERE BUMPED FORWARD, THE TECHNICIAN DID NOT RELOAD THE STAPLER; THEREFORE THE FIRST COUPLE OF STAPLES PROXIMAL TO THE STAPLE LINE DID NOT FORM. THERE WERE NO PATIENT ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24432807 ↗MalfunctionFeb 20, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING A ILEOSTOMY REVERSAL THE STAPLES AT THE END OF THE STAPLE LINE WERE NOT FULLY CLOSED. THEREFORE, THE TISSUE LAYERS WERE NOT FULLY CLOSED. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24405414 ↗MalfunctionFeb 20, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE INSTRUMENT WAS LOCKED DURING FIRING AND SECOND ONE HAD THE SAME PROBLEM. THERE WAS NO DELAY IN THE PROCEDURE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401060 ↗MalfunctionFeb 20, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, THE INSTRUMENT WAS LOCKED DURING FIRING AND SECOND ONE HAD THE SAME PROBLEM. THERE WAS NO DELAY IN THE PROCEDURE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401029 ↗MalfunctionFeb 18, 2026Failure to Form Staple; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THAT THE STAPLER FAILED. STATED THAT THE STAPLER WAS NOT WORKING. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24377614 ↗MalfunctionFeb 6, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24280764 ↗MalfunctionFeb 5, 2026Failure to Fire
IT WAS REPORTED THAT, DURING AN UNKNOWN SURGERY, THE FIRST ONE WAS FIRED WITHOUT ANY PROBLEMS. IT COULD NOT BE FIRED WHEN LOADING THE SECOND TCR75. THE CARTRIDGE COLOR WAS BLUE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24270864 ↗MalfunctionJan 20, 2026Failure to Fire
IT WAS REPORTED THAT DURING THE COLON RESECTION SURGERY, COULD NOT FIRE. CHANGED ANOTHER RELOAD TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24122206 ↗MalfunctionJan 5, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THEY CANNOT LOCKED THE HANDLE AND THE STAPLES WERE MALFORMED. NO FURTHER INFORMATION PROVIDED. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23960878 ↗MalfunctionDec 31, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING THE UNK SURGERY, COULD NOT FIRE FARTHER AFTER FIRED A HALF. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23933828 ↗
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.