ECHELON FLEX PCEE45ARecalls & reported problems
Ethicon Endo-Surgery, LLC · The ECHELON FLEX Powered Plus Articulating Endoscopic Linear Cutters are sterile, single patient use instruments that simultaneouslycut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON FLEX 45 Powered Plusinstruments have a staple line that is approximately 45 mm long and a cut line that is approximately 42 mm long. The shaft can rotate freelyin both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operativesite.The instruments are packaged with a primary lithium battery pack that must be installed prior to use. There are specific requirements fordisposing of the battery pack. Refer to the Battery Pack Disposal section.The instruments are packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple legpoints during shipping and transportation. The instruments’ lock-out feature is designed to prevent a used or improperly installed reloadfrom being refired or an instrument from being fired without a reload. · Class 2
All models of ECHELON FLEX (7)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Echelon Flex powered stapler has 390 reports in the FDA database, with 64 injuries and 1 death. The reports often describe the stapler failing to fire, jaws that are hard to open or close, staples that do not form properly, and jams. Some describe a collapsed lung (pneumothorax) or bleeding during surgery. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 389
- Deaths reported
- 1
- Injuries reported
- 64
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 389 adverse event reports that list this exact device (DI 10705036015338), received between Dec 7, 2017 and Sep 17, 2026 — classified as 1 death, 64 injuries, 324 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (126 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire126
- Difficult to Open or Close103
- Failure to Form Staple51
- Activation Failure34
- Mechanical Jam33
- Adverse Event Without Identified Device or Use Problem32
- No Apparent Adverse Event20
- Activation Problem19
Patient problems
- No Clinical Signs, Symptoms or Conditions242
- No Consequences Or Impact To Patient83
- No Code Available29
- Pneumothorax18
- Failure to Anastomose14
- Hemorrhage/Blood Loss/Bleeding13
- Unspecified Tissue Injury7
- Blood Loss5
Recent adverse event reports
The 25 most recent of 389 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Manufacturing, Packaging or Shipping Problem
It was reported that during an unknown procedure, it was observed that the device was damaged after opening the package. Another device was used to complete the surgery. There were no adverse consequences to the patient. No additional information could be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50806196 ↗MalfunctionAug 17, 2026Failure to Fire; Difficult to Open or Close; Noise, Audible
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50286942 ↗MalfunctionJun 24, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT LOCK-OUT OCCURRED IN TWO DEVICES IN A SINGLE CASE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25612376 ↗MalfunctionMay 13, 2026Failure to Fire
IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT FIRE AT ALL WHEN ABOUT TO FIRE ON THE PULMONARY TISSUE. THE BLADE DID NOT CUT AND NO CLIPS LIGATING. THEY OBSERVED THAT THERE WAS NO ENERGY IN THE BATTERY. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25163817 ↗MalfunctionMay 13, 2026Failure to Cut; Failure to Fire; No Apparent Adverse Event; Activation Failure
IT WAS REPORTED THAT DURING A VATS PROCEDURE THE PATIENT HAD INTERSTITIAL LUNG DISEASE. POWERED STAPLERS ARE NOT ABLE TO DIVIDE TISSUES. USED 2 BLACK RELOADS WITH THE MANUAL STAPLER BEFORE IT JAMMED. REVERSE BUTTON NOT WORKING. REQUIRED DIVISION OF LUNG TISSUE WITH HOOK DIATHERMY ¿ AIR LEAK.
Patient problems: Unspecified Tissue Injury
MAUDE report 25160434 ↗MalfunctionMay 12, 2026Failure to Fire
IT WAS REPORTED THAT DURING A LAPAROSCOPIC VIDEO-ASSISTED THORACIC PROCEDURE FOR LUNG CANCER, IT WAS OBSERVED THAT THE FIRING STOPPED MIDWAY. THE TARGETED TISSUE WAS UNKNOWN. THE CARTRIDGE COLOR WAS GREEN. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25149999 ↗MalfunctionMay 6, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25097622 ↗MalfunctionMay 6, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE, THE ENDOSTAPLER PRESENTED A DEFECT / LOCKING IN THE STAPLING, AND IT WAS NECESSARY TO REPLACE IT WITH A NEW ENDOSTAPLER. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25097226 ↗MalfunctionMay 6, 2026Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN THORACIC PROCEDURE, AFTER THE STAPLE LINE REINFORCEMENT WAS ADHERED TO THE JAWS, THERE WAS A DEFECT. THE DEVICE BECAME UNUSABLE. THE CARTRIDGE COLOR WAS GOLD. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25094048 ↗MalfunctionApr 21, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT STAPLE AS ONE OF THE CARTRIDGES DID NOT STAPLE HALFWAY THROUGH. THE SECOND CARTRIDGE WORKED NORMALLY. THE CARTRIDGE COLOR WAS BLUE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24944071 ↗MalfunctionApr 13, 2026Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event; Activation Failure
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE SURGEON USED PCEE45A WITH A GREEN LOAD, FIRED THE STAPLER, IT SOUNDED LIKE IT WAS GOING AND THEN IT SOUNDED LIKE THE BATTERY DIED. SURGEON SAID THAT IT WOULD NOT OPEN WITH THE REVERSE KNIFE BUTTON SO SURGEON HAD TO USE THE MANUAL OVERRIDE. SURGEON OPENED A NEW STAPLER WITH GREEN RELOAD AND IT DID THE SAME THING. SO SURGEON THEN OPENED A PCEE60A AND IT DID IT AGAIN. SURGEON THEN OPENED A ECH60S AND IT WORKED. THE SURGEON SAID THAT THE TISSUE TORE WHILE USING THE MANUAL RELEASE FUNCTION AND THAT SINCE THE STAPLER DID END UP FIRING A FEW STAPLES AND CUTTING THAT SURGEON HAD TO RESECT MORE TISSUE. IT ADDED AN ADDITIONAL 30 MINUTES TO THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Unspecified Tissue Injury
MAUDE report 24864319 ↗MalfunctionApr 13, 2026Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event; Activation Failure
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE SURGEON USED PCEE45A WITH A GREEN LOAD, FIRED THE STAPLER, IT SOUNDED LIKE IT WAS GOING AND THEN IT SOUNDED LIKE THE BATTERY DIED. SURGEON SAID THAT IT WOULD NOT OPEN WITH THE REVERSE KNIFE BUTTON SO SURGEON HAD TO USE THE MANUAL OVERRIDE. SURGEON OPENED A NEW STAPLER WITH GREEN RELOAD AND IT DID THE SAME THING. SO SURGEON THEN OPENED A PCEE60A AND IT DID IT AGAIN. SURGEON THEN OPENED A ECH60S AND IT WORKED. THE SURGEON SAID THAT THE TISSUE TORE WHILE USING THE MANUAL RELEASE FUNCTION AND THAT SINCE THE STAPLER DID END UP FIRING A FEW STAPLES AND CUTTING THAT SURGEON HAD TO RESECT MORE TISSUE. IT ADDED AN ADDITIONAL 30 MINUTES TO THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: Unspecified Tissue Injury
MAUDE report 24864143 ↗MalfunctionApr 7, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A VATS PROCEDURE, IT WAS OBSERVED THAT THE DEVICE HAD AN UNSPECIFIED MALFUNCTION. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24800782 ↗MalfunctionMar 31, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE DEVICE HAD AN UNSPECIFIED MALFUNCTION. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24746734 ↗MalfunctionMar 26, 2026Unintended Ejection
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE RELOAD DID NOT "CLICK" INTO PLACE WHEN LOADED AND KEPT FALLING OFF. ANOTHER DIFFERENT ( BATCH NUMBER DIFFERENT) STAPLE WAS LOADED TO IDENTIFY WHAT WAS CAUSING THE ISSUE BUT THE SAME THING HAPPENED AGAIN. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24697336 ↗MalfunctionFeb 17, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT DURING AN OPEN HPB PROCEDURE, THE SURGEON WAS USING A COMPACT ECHELON FLEX STAPLER AND REPORTED THAT THE STAPLER "MISFIRED". CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24368150 ↗MalfunctionFeb 11, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, COULD NOT FIRE. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24316085 ↗MalfunctionFeb 10, 2026Failure to Fire; Difficult to Open or Close; Activation Problem
VATS LOBECTOMY DEVICE LOCKED ON LUNG TISSUE AND AFTER ATTEMPTS TO REMOVE USING BOTH THE REVERSE BUTTON AND THE MANUAL OVERRIDE, JAWS WOULD STILL NOT OPEN. DEVICE HAD TO BE MANUALLY OPENED USING ANOTHER INSTRUMENT. AN ALTERNATIVE DEVICE WAS OPENED AND USED TO COMPLETE THE PROCEDURE. NO ADVERSE EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24304703 ↗MalfunctionFeb 9, 2026Excessive Heating
IT WAS PREOPERATIVELY TO A VATS PROCEDURE; IT WAS OBSERVED BY THE STAFF THAT THE BATTERY AND HANDLE APPEARED TO BE OVERHEATING AND FELT HOT TO TOUCH. THIS WAS OUTSIDE OF THE PATIENT. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24293083 ↗MalfunctionJan 29, 2026Failure to Cut
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLER WOULD DEPLOY STAPLES BUT WOULD NOT CUT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. UNKNOWN AS TO HOW THE PROCEDURE WAS COMPLETED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210813 ↗MalfunctionJan 29, 2026Failure to Cut
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE STAPLER WOULD DEPLOY STAPLES BUT WOULD NOT CUT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED. UNKNOWN AS TO HOW THE PROCEDURE WAS COMPLETED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210811 ↗MalfunctionJan 6, 2026Failure to Fire
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23976895 ↗MalfunctionDec 19, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE OF THE LIVER, IT WAS OBSERVED THAT THE HANDLE ON THE DEVICE WOULD NOT RELEASE AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE WITH A DELAY OF 5-10 MINUTES. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23853655 ↗MalfunctionDec 18, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT DURING A LOBECTOMY PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23840692 ↗MalfunctionDec 15, 2025Failure to Fire; Difficult to Open or Close; Mechanical Jam; Failure to Fire; Difficult to Open or Close; Mechanical Jam
IT WAS REPORTED THAT DURING A VIDEO-ASSISTED THORACOSCOPIC SURGERY (VATS) PROCEDURE, IT WAS OBSERVED THAT THE STAPLER DID THE INITIAL FIRE WHEN THEY CLOSED IT ON THE LUNG TISSUE BUT WOULD NOT RETRACT SO THEY COULD NOT OPEN THE JAWS. THEY USED THE REVERSE BUTTON, TOOK THE BATTERY OUT THEN BACK IN BUT THE DEVICE STILL WOULD NOT FIRE. THEY USED MANUAL OVERRIDE TO RETRACT THE BLADE AND OPEN THE JAWS. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23810102 ↗
FDA clearance and classification
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ECHELON FLEX PCEE45A
Has the ECHELON FLEX PCEE45A been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036015338), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECHELON FLEX PCEE45A?
The FDA's MAUDE database lists 389 adverse event reports for this device, received between Dec 7, 2017 and Sep 17, 2026. In those reports, the most common specific device problems are “Failure to Fire” (126), “Difficult to Open or Close” (103) and “Failure to Form Staple” (51); the most common patient problems are “Pneumothorax” (18), “Failure to Anastomose” (14) and “Hemorrhage/Blood Loss/Bleeding” (13). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON FLEX PCEE45A?
1 report is classified as a death, 64 as injuries and 324 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ECHELON FLEX PCEE45A, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON FLEX, model PCEE45A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page