ECHELON FLEX PSEE45ARecalls & reported problems

Ethicon Endo-Surgery, LLC · The ECHELON FLEX Powered Plus Articulating Endoscopic Linear Cutters are sterile, single patient use instruments that simultaneouslycut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON FLEX 45 Powered Plusinstruments have a staple line that is approximately 45 mm long and a cut line that is approximately 42 mm long. The shaft can rotate freelyin both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operativesite.The instruments are packaged with a primary lithium battery pack that must be installed prior to use. There are specific requirements fordisposing of the battery pack. Refer to the Battery Pack Disposal section.The instruments are packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple legpoints during shipping and transportation. The instruments’ lock-out feature is designed to prevent a used or improperly installed reloadfrom being refired or an instrument from being fired without a reload. · Class 2

All models of ECHELON FLEX (7)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon ECHELON FLEX endoscopic stapler is a surgical instrument that cuts and staples tissue during surgery. Many reports describe the stapler failing to fire, being difficult to open or close, or jamming, and some describe staples not forming properly. The FDA database lists 1,255 reports, with 113 injuries and 5 deaths, and reports fell in the last 12 months, to 102 from 198. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,253
Deaths reported
5
Injuries reported
114
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,253 adverse event reports that list this exact device (DI 10705036015352), received between Dec 28, 2017 and Sep 18, 2026 — classified as 5 deaths, 114 injuries, 1,134 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (435 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 16 reportsJan 2024: 18 reportsFeb 2024: 11 reportsMar 2024: 15 reportsApr 2024: 17 reportsMay 2024: 11 reportsJun 2024: 19 reportsJul 2024: 17 reportsAug 2024: 20 reportsSep 2024: 27 reportsOct 2024: 26 reportsNov 2024: 23 reportsDec 2024: 16 reportsJan 2025: 16 reportsFeb 2025: 15 reportsMar 2025: 8 reportsApr 2025: 13 reportsMay 2025: 31 reportsJun 2025: 13 reportsJul 2025: 24 reportsAug 2025: 12 reportsSep 2025: 17 reportsOct 2025: 10 reportsNov 2025: 9 reportsDec 2025: 13 reportsJan 2026: 14 reportsFeb 2026: 16 reportsMar 2026: 8 reportsApr 2026: 10 reportsMay 2026: 4 reportsJun 2026: 12 reportsJul 2026: 3 reportsAug 2026: 10 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202631
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire435
  • Difficult to Open or Close431
  • Failure to Form Staple183
  • Mechanical Jam156
  • Activation Failure105
  • Adverse Event Without Identified Device or Use Problem63
  • Activation Problem31
  • No Apparent Adverse Event31

Patient problems

  • No Clinical Signs, Symptoms or Conditions917
  • No Consequences Or Impact To Patient209
  • Hemorrhage/Blood Loss/Bleeding57
  • No Code Available43
  • Blood Loss28
  • Failure to Anastomose22
  • Insufficient Information12
  • Pain9

Recent adverse event reports

The 25 most recent of 1,253 reports, in the reporter's own words as published by the FDA.

  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that on (b)(6) 2026, patient underwent Gastric Bypass Gastroplasty procedure. Patient has hypertension as comorbidity. During the initial procedure, the surgery was uneventful. The patient maintained blood pressure within normal parameters and no unexpected bleeding was observed during the procedure. A methylene blue test was performed and nothing was detected in the staple line and anastomosis. \\On (b)(6) 2026, a surgical reapproach was necessary after endoscopy to identify the presence of bleeding, associated with a drop in hemoglobin to 8.0. \During the reapproach, a hematoma was identified in the region of the gastroenteric anastomosis, with the presence of small blood clots, but no active bleeding in the cavity. Thus, the surgeon performed a new suture on the staple line and anastomosis. A new endoscopy was performed in the room, which indicated internal bleeding with a large amount of clots. The medical suspicion is that the bleeding occurred in the anastomosis (gastro-entero). So far, it has been reported that the patient received a transfusion of two bags of blood. Patient is still hospitalized and under observation. Additional information requested.

    Patient problems: Hematoma; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50829248 ↗
  • MalfunctionAug 31, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50526491 ↗
  • MalfunctionAug 26, 2026Failure to Fire; Difficult to Open or Close; Mechanical Jam; Activation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459368 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    It was reported that during an unspecified surgical procedure, it was observed that the device would not fire. Changed to a new one to complete the procedure. There was no patient consequence nor surgical delay reported. No additional information provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50432540 ↗
  • MalfunctionAug 25, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50431322 ↗
  • MalfunctionAug 24, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50410045 ↗
  • InjuryAug 20, 2026Use of Device Problem; Failure to Fire; Difficult to Open or Close

    It was reported that during a Laparoscopic Lobectomy procedure, after several fires, a CECR45W was used to treat the main pulmonary artery. After closing the jaw, noted could not fire. It cannot be clarified whether the vessel has been cut, so changed to open operation to open the jaw. The jaw was opened without any difficulty and the vessel was not cut. Changed other devices to complete the surgery. Additional information requested.

    Patient problems: Insufficient Information

    MAUDE report 50351211 ↗
  • MalfunctionAug 19, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323832 ↗
  • MalfunctionAug 14, 2026Insufficient Device Problem Information

    It was reported that during an unknown procedure, it was observed that the device did not respond after firing. A new PSEE45A was subsequently used, and the procedure was completed successfully. There was no adverse patient consequence nor surgical delay reported. No additional information provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50253022 ↗
  • MalfunctionAug 4, 2026Failure to Form Staple

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50129789 ↗
  • MalfunctionAug 3, 2026Positioning Failure; Failure to Cut; Difficult to Open or Close; Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50114918 ↗
  • MalfunctionJul 27, 2026Failure to Fire; Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50046128 ↗
  • MalfunctionJul 27, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50029512 ↗
  • MalfunctionJul 1, 2026Failure to Cut

    IT WAS REPORTED THAT DURING A GASTROPLASTY PROCEDURE, WHEN ACTIVATING THE ECHELON FLEX PLUS REF PSEE45A LAPAROSCOPIC LINEAR STAPLER, THE SAME STAPLED, BUT DID NOT CUT THROUGH THE TISSUE SEPARATION. THE RELOAD WAS CHANGED TWICE TO MAKE SURE THAT IT WOULD NOT BE A DEFECT IN THE RELOAD BUT IT WAS NECESSARY TO OPEN ANOTHER STAPLER. WAS THERE ANY REPORTED PATIENT OR USER HARM? UNKNOWN WAS THERE A SIGNIFICANT DELAY? UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25677566 ↗
  • MalfunctionJun 24, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE MANUAL OVERRIDE PANEL ON THE MAIN BODY WAS DETACHED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25611373 ↗
  • MalfunctionJun 24, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25611042 ↗
  • MalfunctionJun 17, 2026Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE FAILED TO DEPLOY. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25538084 ↗
  • MalfunctionJun 17, 2026Difficult to Open or Close; Mechanical Jam; Activation Failure

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE KNIFE MOVED TO THE DISTAL END BUT DID NOT RETURN KNIFE AUTOMATICALLY. KNIFE REVERSE BUTTON DID NOT WORK. USED MANUAL OVERRIDE LEVER TO RETURN KNIFE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25537413 ↗
  • MalfunctionJun 17, 2026Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT AFTER THEY INSTALLED THE RELOAD AND CLOSED THE JAW, THEY COULD NOT OPEN THE JAW ANYMORE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25537228 ↗
  • MalfunctionJun 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448503 ↗
  • MalfunctionJun 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25448196 ↗
  • MalfunctionJun 9, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446076 ↗
  • MalfunctionJun 5, 2026Failure to Form Staple

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE STAPLES WERE MALFORMED AFTER FIRING. CHANGED TO A NEW ONE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414646 ↗
  • MalfunctionJun 3, 2026Difficult to Open or Close

    IT WAS REPORTED TO THAT DURING A NEPHRECTOMY PROCEDURE, THE PSEE45A WAS STUCK CLOSED AFTER FIRING AND WOULD NOT OPEN. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25374838 ↗
  • MalfunctionJun 2, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT FIRE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25362356 ↗

FDA clearance and classification

The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the ECHELON FLEX PSEE45A

Has the ECHELON FLEX PSEE45A been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036015352), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ECHELON FLEX PSEE45A?

The FDA's MAUDE database lists 1,253 adverse event reports for this device, received between Dec 28, 2017 and Sep 18, 2026. In those reports, the most common specific device problems are “Failure to Fire” (435), “Difficult to Open or Close” (431) and “Failure to Form Staple” (183); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (57), “Blood Loss” (28) and “Failure to Anastomose” (22). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ECHELON FLEX PSEE45A?

5 reports are classified as deaths, 114 as injuries and 1,134 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ECHELON FLEX PSEE45A, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON FLEX, model PSEE45A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other STAPLE, IMPLANTABLE devices

All STAPLE, IMPLANTABLE devices →