EVOTECH™ 50004Recalls & reported problems
Advanced Sterilization Products Services Inc. · EVOTECH™ Endoscope Cleaner and Reprocessor (ECR) · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705037030958), received between Apr 6, 2021 and Apr 6, 2021 — classified as 1 malfunction.
1 FDA recall record name this device's identifier.
4 other recalls cover devices cleared under the same FDA submission (K082392, K140977, K152189); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 27, 2016Terminated
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.
Z-0965-2017 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
A CUSTOMER REPORTED A CYCLE CANCELLATION OF AN A.10.3 OVERFLOW OR SYSTEM LEAK WITH THEIR EVOTECH¿ ECR AND THE CANCELLED CYCLE WAS RELEASED FOR USE ON PATIENTS PRIOR TO REPROCESSING. THERE IS NO REPORT OF INFECTION, INJURY OR HARM TO PATIENT(S) ASSOCIATED WITH THIS ISSUE. ALTHOUGH THERE IS NO REPORT OF PATIENT INJURY OR HARM, ADVANCED STERILIZATION PRODUCTS (ASP) HAS DETERMINED IN THIS SITUATION HIGH LEVEL DISINFECTION CANNOT BE ASSURED. THEREFORE, ASP HAS DECIDED TO REPORT ALL INCIDENTS OF LOADS THAT ARE RELEASED FROM CANCELLED CYCLES PRIOR TO REPROCESSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11620164 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082392, K140977, K152189. One submission often covers many models, so check the product and lot details in each record.
- Z-1715-2015 · May 11, 2015 · TerminatedEVOTECH Endoscope Cleaner and Reprocessing System 220-230V, P/N 50004-002. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for us
- Z-1714-2015 · May 11, 2015 · TerminatedEVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with h
- Z-1150-2013 · Mar 28, 2013 · TerminatedEVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.
- Z-1123-2013 · Mar 22, 2013 · TerminatedEVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes.
Other ACCESSORIES, CLEANING, FOR ENDOSCOPE devices
- ENDOSCOPE REPROCESSOR OER-ProOlympus Medical Systems Corp. · ACCESSORIES, CLEANING, FOR ENDOSCOPE3 recalls1,163 reports
- ENDOSCOPE REPROCESSOR OER-EliteOlympus Medical Systems Corp. · Accessories, cleaning, for endoscope4 recalls323 reports
- SINGLE USE DISTAL COVER MAJ-2315Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls294 reports
- CONNECTING TUBE MAJ-2110Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall213 reports
- CONNECTING TUBE MAJ-2111Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall164 reports
- CONNECTING TUBE MAJ-2113Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall159 reports
- CONNECTING TUBE MAJ-2112Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall143 reports
- ENDOSCOPE REPROCESSOR OER-MINIOlympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope2 recalls11 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.