ARROW IPN033776Recalls & reported problems
Teleflex Incorporated · OnControl(R) Bone Marrow Biopsy System Needle Set, 152mm · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10816000012236), received between Jul 19, 2021 and Jul 22, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K142377); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 22, 2021Break; Break
AS REPORTED: PHYSICIAN USED ONCONTROL BONE MARROW SYSTEM NEEDLE (3.0MMX152MM) WITH ONCONTROL POWER DRIVER. WHILE COMING OUT FROM THE LESION, THE DISTAL LOCKING ASSEMBLY OF THE AFOREMENTIONED NEEDLE GAVE AWAY FROM THE ATTACHED POWER DRILL, RESULTING IN THE BREAKAGE OF THE CANNULA AT SUB TROCHANTERIC ENTRY POINT. THE PROXIMAL PART OF THE CANNULA IS STUCK BETWEEN THE SUBTROCHANTERIC REGION AND THE HEAD OF THE FEMUR. TROCAR AND EXTRA-OSSEOUS PART OF THE BROKEN CANNULA WERE REMOVED. THE PATIENT WAS TREATED IN MUMBAI AND IS BACK IN DELHI WITH A PART OF THE NEEDLE IN THE BONE.
Patient problems: Foreign Body In Patient
MAUDE report 12209890 ↗MalfunctionJul 19, 2021Break; Break
CUSTOMER REPORTED THAT THE DOCTOR, REGEN ORTHOSPORT USED ARROW ON CONTROL BONE MARROW SYSTEM NEEDLE (3.0MMX152MM) WITH ONCONTROL POWER DRIVER. PROCEDURE PLANNED: APPROACHING THE INTRAOSSEOUS LESION IN THE HEAD OF THE FEMUR UNDER ALL ASEPTIC PRECAUTIONS. WHILE COMING OUT FROM THE LESION, THE DISTAL LOCKING ASSEMBLY OF THE AFOREMENTIONED NEEDLE GAVE AWAY FROM THE ATTACHED POWER DRILL, RESULTING IN THE BREAKAGE OF THE CANNULA AT SUB TROCHANTERIC ENTRY POINT. CLINICAL CONSEQUENCES: THE PROXIMAL PART OF THE CANNULA IS STUCK BETWEEN THE SUBTROCHANTERIC REGION AND THE HEAD OF THE FEMUR. TROCAR AND EXTRA-OSSEOUS PART OF THE BROKEN CANNULA WERE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12188441 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142377. One submission often covers many models, so check the product and lot details in each record.
- Z-0679-2017 · Nov 8, 2016 · TerminatedARROW OnControl Ported Aspiration System Tray, Sterile, Rx only, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marrow aspiration of the iliac crest of adul
- Z-0676-2017 · Nov 8, 2016 · TerminatedARROW OnControl Bone Marrow Biopsy System Comprehensive Tray, Sterile, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marrow aspiration of the iliac crest o
- Z-0678-2017 · Nov 8, 2016 · TerminatedOnControl Biopsy System Ported Needle Tray, Sterile, Rx only, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marrow aspiration of the iliac crest of adult a
- Z-0675-2017 · Nov 8, 2016 · TerminatedARROW OnControl Bone Marrow Biopsy System Tray, Sterile, Rx only, Manufactured for: Teleflex Medical, The Arrow OnControl Bone Marrow Aspiration System is intended for bone marro
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall118 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall29 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.