Skip to content

D.A.S.H. G34780Recalls & reported problems

Cook Incorporated · D.A.S.H. Preloaded with Acrobat Wire Guide · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10827002347806), received between Feb 21, 2017 and Nov 15, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2023Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK D.A.S.H.® PRE-LOADED WITH ACROBAT WIRE GUIDE. IT WAS INITIALLY REPORTED DURING THE FIRST PASS WITH THE EXTRACTOR BALLOON, THE GUIDE WIRE BENT AND SUBSEQUENTLY WERE UNABLE TO USE IT. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2023 STATING THAT CHANGE OF GUIDE WIRE FOR ANOTHER RE-CANNULATION WITH THE BALLOON EXTRACTOR (LOSS OF WIRE GUIDE ACCESS AND COATING DAMAGE). [SUBJECT OF REPORT] A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18142266 ↗
  • MalfunctionNov 17, 2021Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK D.A.S.H.® PRE-LOADED WITH ACROBAT WIRE GUIDE. IT WAS REPORTED THAT DURING THE PROCEDURE, THE COATING OF THE WIRE CAME OFF. A NEW PRODUCT WAS OPENED AND THE PROCEDURE WAS CARRIED OUT AS PLANNED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12830978 ↗
  • MalfunctionFeb 21, 2017Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK DASH PRE-LOADED ACROBAT WITH WIRE GUIDE. DURING THE PROCEDURE, THE WIRE GUIDE COATING STRIPPED OFF OF THE WIRE [COATING DAMAGE]. THE FOLLOWING INFORMATION WAS PROVIDED ON 2/2/17: "ALL THE DAMAGE OCCURRED AT THE PATIENT END OF THE WIRE WITHIN 20-25 CM FROM THE TIP."

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6346334 ↗

Other Unit, electrosurgical, endoscopic (with or without accessories) devices

All Unit, electrosurgical, endoscopic (with or without accessories) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page