Endo GIA 030454Recalls & reported problems
Covidien LP · Universal Articulating Loading Unit · Class 2 · Implant
All models of Endo GIA (48)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien Endo GIA is a stapler reload used to cut and seal tissue during surgery. Its 252 reports often describe the stapler failing to fire, staples not forming, and reloads breaking. The reports include 62 injuries and 5 deaths. Reports dropped from 9 in the previous 12 months to none in the most recent 12 months. There is 1 closed recall, begun April 2018, over a possibly missing part that can stop staples from forming.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 252
- Deaths reported
- 5
- Injuries reported
- 62
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 252 adverse event reports that list this exact device (DI 10884523002850), received between Aug 3, 2017 and Jul 1, 2025 — classified as 5 deaths, 62 injuries, 185 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (102 reports).
1 FDA recall record name this device's identifier.
72 other recalls cover devices cleared under the same FDA submission (K111825); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 26, 2018Class IITerminated
Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.
Z-1817-2018 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Fire102
- Failure to Form Staple61
- Adverse Event Without Identified Device or Use Problem39
- Misfire39
- Difficult to Open or Close19
- Break16
- Detachment of Device or Device Component11
- Human-Device Interface Problem11
Patient problems
- No Clinical Signs, Symptoms or Conditions71
- No Known Impact Or Consequence To Patient64
- No Code Available40
- No Consequences Or Impact To Patient36
- Unspecified Tissue Injury19
- Hemorrhage/Blood Loss/Bleeding15
- Blood Loss9
- Insufficient Information7
Recent adverse event reports
The 25 most recent of 252 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE ON TISSUE RESECTION, THE DOCTOR FOUND THAT TWO CONSECUTIVE STAPLER RELOADS FELL OFF AND BROKE, WHICH SERIOUSLY AFFECTED THE PROGRESS OF THE OPERATION. THE SUBSEQUENT DOCTOR COULD ONLY SUTURE MANUALLY WITH SUTURES, AND THE OPERATION WENT SMOOTHLY.
Patient problems: Insufficient Information
MAUDE report 22371523 ↗MalfunctionJul 1, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE ON TISSUE RESECTION, THE DOCTOR FOUND THAT TWO CONSECUTIVE STAPLER RELOADS FELL OFF AND BROKE, WHICH SERIOUSLY AFFECTED THE PROGRESS OF THE OPERATION. THE SUBSEQUENT DOCTOR COULD ONLY SUTURE MANUALLY WITH SUTURES, AND THE OPERATION WENT SMOOTHLY.
Patient problems: Insufficient Information
MAUDE report 22371519 ↗MalfunctionJun 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A LAPAROSCOPIC NEPHRECTOMY INVOLVING ONE RELOAD, THERE WAS BLEEDING IN THE RENAL ARTERY AND STAPLE LINE BLEEDING AFTER FIRING THE RELOAD. THE STAPLE LINE REQUIRED REPLACEMENT WITH SUTURING TO CONTROL THE BLEEDING. THESE ISSUES RESULTEDIN EXTENDED SURGICAL TIME FOR 30 MINUTES. SUTURING WAS PERFORMED BOTH TO CONTROL THE BLEEDING AND AS A REPLACEMENT FOR THE STAPLE LINE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22349843 ↗MalfunctionJun 13, 2025Failure to Fire; Failure to Fire
IT WAS REPORTED THAT THE STAPLER EXPERIENCED FIRING ISSUES WITH THE RELOAD COMPONENT. THE STAPLER CARTRIDGE WAS INSERTED INTO THE SHOOTER BUT DID NOT PUNCH OR CUT WHILE IN USE. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE WITH A NEW CARTRIDGE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22201482 ↗MalfunctionApr 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A ROUX-EN-Y GASTRIC BYPASS, THERE WAS A STAPLE LINE LEAKAGE AT THE OPENING AT THE END OF A SIDE-TO-SIDE ANASTOMOSIS AFTER A LEAK TEST WAS PERFORMED. THERE WAS NO TENSION AND NO COMPLICATION. THE SURGEON PERFORMED A MANUAL SUTURING AS A STAPLE LINE REINFORCEMENT.
Patient problems: Failure to Anastomose
MAUDE report 21774092 ↗MalfunctionMar 25, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A VIDEO THORACOSCOPIC LOBECTOMY, THE STAPLER SUTURED ON THE FIRST RELOAD, BUT DID NOT STAPLE DURING THE NEXT RELOAD. THE USER REPLACED THE RELOAD; THE SAME THING HAPPENED. A STAPLER AND RELOAD FROM ANOTHER COMPANY WERE USED, AND THE OPERATION WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21686088 ↗MalfunctionMar 25, 2025Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A VIDEO THORACOSCOPIC LOBECTOMY, THE STAPLER SUTURED ON THE FIRST RELOAD, BUT DID NOT STAPLE DURING THE NEXT RELOAD. THE USER REPLACED THE RELOAD; THE SAME THING HAPPENED. A STAPLER AND RELOAD FROM ANOTHER COMPANY WERE USED, AND THE OPERATION WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21686051 ↗MalfunctionFeb 6, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A PROCEDURE, THERE WERE ISSUES AFTER FIRING; THERE WAS A STAPLE FORMATION; IT WAS NOT HOLDING AND FORMED WELL. THE END OF THE STAPLE LINE WAS NOT COMPLETED. THE RELOAD COMPONENT WAS USED WITH A MANUAL HANDLE. TO RESOLVE THE ISSUE, A NEW LOADING UNIT WAS USED.
Patient problems: Tissue Breakdown
MAUDE report 21324224 ↗MalfunctionJan 6, 2025Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC LEFT HEPATECTOMY, WHEN CUTTING OFF THE LEFT HEPATIC PEDICLE, THERE WAS POOR STAPLE FORMATION. THE STAPLES PARTLY FORMED. REINFORCEMENT WITH SUTURE WAS DONE TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 21081898 ↗MalfunctionSep 13, 2024Misfire; Failure to Form Staple; Firing Problem; Misfire; Failure to Form Staple; Firing Problem
ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE FIRST RELOAD WAS FIRED FOR A HALF AND NEAR THE LAST 15MM. WITH NO PROPER STAPLE LINE FORMATION. IT WAS MENTIONED THAT THE JAW OF THE SECOND RELOAD GOT STUCK AND WAS UNABLE TO OPEN IN THE MIDDLE OF THE FIRING. THE JAW OF THE SECOND RELOAD HAD TO BE OPEN USING ANOTHER HAND INSTRUMENTS. A NEW RELOAD WAS USED TO RESOLVE THE ISSUE AND COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Tissue Breakdown
MAUDE report 20213355 ↗MalfunctionAug 27, 2024Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE PHYSICIAN TRIED TO FIRE THE RELOADS, BUT IT COULD NOT BE FIRED. THE HANDLE AND RELOAD WERE REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20076403 ↗MalfunctionAug 27, 2024Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE PHYSICIAN TRIED TO FIRE THE RELOADS, BUT IT COULD NOT BE FIRED. THE HANDLE AND RELOAD WERE REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20076087 ↗MalfunctionApr 22, 2024Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC LOBECTOMY PROCEDURE, IN CLOSING THE TISSUE, THERE WAS POOR STAPLE FORMATION. TO RESOLVE THE ISSUE, ANOTHER RELOAD WAS USED.
Patient problems: Tissue Breakdown
MAUDE report 19153175 ↗InjuryMar 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A SUB-COSTAL HEPATECTOMY, THERE WAS NO ISSUE WHEN 30 AND 45 MILLIMETER WHITE RELOADS WERE USED. HOWEVER, FIVE HOURS LATER, THE PATIENT WAS TAKEN BACK TO THE EMERGENCY ROOM DUE TO BLEEDING. THE STAPLE LINE OF BOTH RELOADS WERE OPEN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 18988190 ↗MalfunctionMar 11, 2024Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING THE LAPAROSCOPIC HEPATECTOMY, WHILE STAPLING ON TISSUE, THE STAPLER WAS NOT ABLE TO FIRE. THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON BUT UNABLE TO SQUEEZE THE HANDLE. THE ISSUE OCCURRED ON TWO MORE RELOADS. A NEW RELOAD WAS THEN USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18882661 ↗MalfunctionDec 29, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC LEFT UPPER LOBECTOMY, THE CUTTING STAPLER WAS BLOCKED, RESULTING IN A FAILURE TO FIRE NORMALLY. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18422562 ↗MalfunctionNov 7, 2023Difficult to Open or Close; Difficult to Open or Close
ACCORDING TO THE REPORTER, DURING A HEMIHEPATECTOMY, IN CLOSING THE TISSUE INTO THE ABDOMINAL CAVITY, THE FRONT END OF THE RELOAD WAS INCOMPLETELY CLOSED. TO RESOLVE THE ISSUE, A NEW RELOAD WAS USED WITH THE SAME HANDLE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18089361 ↗MalfunctionNov 6, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW RELOAD AND HANDLE WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18077742 ↗MalfunctionNov 6, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW RELOAD AND HANDLE WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18077616 ↗MalfunctionNov 6, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW RELOAD AND HANDLE WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18077422 ↗MalfunctionNov 6, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW RELOAD AND HANDLE WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18077289 ↗MalfunctionNov 2, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18059283 ↗MalfunctionNov 2, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18059198 ↗MalfunctionNov 2, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18059072 ↗MalfunctionNov 2, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC HEPATECTOMY PROCEDURE, AT THE TIME OF LIVER HEPATIC VEIN TRANSECTION, THE SURGEON WAS UNABLE TO SQUEEZE THE HANDLE AFTER PRESSING THE GREEN BUTTON. THEREFORE, THE DEVICE DID NOT FIRE. THE ISSUE OCCURRED IN THREE HANDLES AND FIVE RELOADS. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18058981 ↗
FDA clearance and classification
Cleared through 510(k) K111825 on Oct 25, 2011 (decision: Substantially Equivalent). Submitted by Covidien, Formerly US Surgical A Divison of Tyco H as “Endo Gia(Tm) Staplers, Dst Series(Tm) Gia(Tm) Staplers, Dst Series(Tm) Ta(Tm) Staplers”. The FDA review took 119 days. FDA record ↗
The FDA regulates product code GDW (staple, implantable) under 21 CFR 878.4750 as a Class II device. 21 CFR 878.4750 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111825. One submission often covers many models, so check the product and lot details in each record.
- Z-0747-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or tra
- Z-0748-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection
- Z-0005-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit 45 mm - 2.5mm, Item Code 030454 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric
- Z-0006-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 3.5 mm, Item Code 030455 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0008-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0001-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gyn
- Z-0004-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0009-2021 · Aug 12, 2020 · Open, ClassifiedCovidien Best Practices Procedure Kits containing Auto Suture Endo GIA Universal Loading Units, Parent Codes 00Z2760 - Product Usage: The Endo GIA universal staplers have applicat
- Z-0007-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit, Item Code 030457 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic
- Z-0003-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0002-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 2.5 mm, Item Code 030451 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-1264-2020 · Jan 17, 2020 · TerminatedCovidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
- Z-1827-2019 · May 17, 2019 · Open, ClassifiedEGIA 45 ARTICULATING XTRA THICK SULU, Item Code EGIA45AXT
- Z-1826-2019 · May 17, 2019 · Open, ClassifiedEGIA 45 ARTICULATING MED THICK SULU, Item Code EGIA45AMT
- Z-1825-2019 · May 17, 2019 · Open, ClassifiedEGIA 30 ARTICULATING MED THICK SULU, Item Code EGIA30AMT
…and 57 more.
Common questions about the Endo GIA 030454
Has the Endo GIA 030454 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1817-2018 (Class II), began on Apr 26, 2018: Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Endo GIA 030454?
The FDA's MAUDE database lists 252 adverse event reports for this device, received between Aug 3, 2017 and Jul 1, 2025. In those reports, the most common specific device problems are “Failure to Fire” (102), “Failure to Form Staple” (61) and “Misfire” (39); the most common patient problems are “Unspecified Tissue Injury” (19), “Hemorrhage/Blood Loss/Bleeding” (15) and “Blood Loss” (9). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Endo GIA 030454?
5 reports are classified as deaths, 62 as injuries and 185 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Endo GIA 030454 FDA approved?
It was cleared through 510(k) K111825 on Oct 25, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Endo GIA 030454 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Endo GIA 030454, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Endo GIA, model 030454). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,660 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,727 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,350 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,279 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,006 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,967 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,910 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,044 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page