Endo GIA EGIA30AVMRecalls & reported problems
Covidien LP · Articulation Reload with Tri-Staple Technology · Class 2 · Implant
All models of Endo GIA (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien Endo GIA reload stapler cartridge used in surgery has 94 reports, including 6 injuries and 2 deaths. Common reports describe staples not firing or not forming, cuts made without staples, and bleeding along the staple line. Reports fell from 4 in the prior 12 months to 1 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 93
- Deaths reported
- 2
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 93 adverse event reports that list this exact device (DI 10884523003123), received between Aug 10, 2017 and Sep 2, 2025 — classified as 2 deaths, 6 injuries, 85 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (58 reports).
73 other recalls cover devices cleared under the same FDA submission (K111825); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire58
- Failure to Form Staple12
- Adverse Event Without Identified Device or Use Problem10
- Misfire9
- Difficult to Open or Close7
- Failure to Cut4
- Firing Problem3
- Human-Device Interface Problem3
Patient problems
- No Known Impact Or Consequence To Patient33
- No Consequences Or Impact To Patient20
- No Clinical Signs, Symptoms or Conditions19
- Tissue Breakdown9
- Hemorrhage/Blood Loss/Bleeding6
- Insufficient Information6
- No Code Available3
- No Information3
Recent adverse event reports
The 25 most recent of 93 reports, in the reporter's own words as published by the FDA.
InjurySep 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT FOLLOWING A GASTRIC BYPASS PROCEDURE USING 5 RELOADS WITH A POWERED STAPLER, THE PATIENT EXPERIENCED EXTENDED HOSPITALIZATION FOR TWO ADDITIONAL DAYS AND BLEEDING IN STOOL BETWEEN 24 AND 48 HOURS POSTOPERATIVELY. STAPLE LINE BLEEDING WAS REPORTED. NO INTERVENTION WAS REQUIRED FOR THE STAPLE LINE BLEEDING.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown
MAUDE report 22930443 ↗InjuryAug 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT INTRAOPERATIVELY WHEN FIRING THE RELOADS, THE SURGEON OBSERVED EXCESSIVE BLEEDING THAN NORMAL. NO INTERVENTION WAS DONE OR REQUIRED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown
MAUDE report 22903376 ↗MalfunctionMar 7, 2025Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT DURING THE USE OF THE DEVICE, THERE WERE FIRING ISSUES WITH THE RELOAD, AS THE RELOAD DID NOT FIRE WITH THE STAPLE, AND THE STAPLES DID NOT DEPLOY, RESULTING IN A CUT BUT NO STAPLES. THE DEVICE WAS REPLACED WITH ANOTHER RELOAD OF THE SAME SIZE AND MODEL. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21546594 ↗MalfunctionJan 6, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE INVOLVING THE STAPLER, THERE WERE FIRING ISSUES WITH THE HANDLE. THE SURGEON WAS UNABLE TO SQUEEZE THE MANUAL HANDLE, ALTHOUGH THE GREEN SAFETY BUTTON WAS PRESSED SUCCESSFULLY. THE INSTRUMENT DID NOT FIRE. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE WITH A NEW HANDLE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT FOUND THAT THE RELOAD WAS PARTIALLY FIRED WITH THE INTERLOCK ENGAGED. STAPLE PUSHERS WERE VISIBLE AT THE 4CM CUT LINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21085664 ↗MalfunctionOct 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, AFTER FIRING, THERE WAS STAPLE LINE BLEEDING ON THE FOUR RELOADS. IT WAS NOTED THAT THE SURGICAL TIME WAS EXTENDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20390368 ↗MalfunctionOct 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER DURING A PROCEDURE, AFTER FIRING, THERE WAS STAPLE LINE BLEEDING ON THE FOUR RELOADS. IT WAS NOTED THAT THE SURGICAL TIME WAS EXTENDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20380169 ↗MalfunctionSep 5, 2024Misfire; Firing Problem; Misfire; Firing Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE RELOAD WAS DIFFICULT TO FIRE. A NEW RELOAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT FOUND THAT THE RELOAD WAS PARTIALLY FIRED WITH THE INTERLOCK ENGAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20152109 ↗MalfunctionAug 9, 2024Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE STAPLE LINE WAS INCOMPLETE. A NEW RELOAD WAS USED TO RESOLVE THE ISSUE AND THE SURGEON RESECTED AND RE-STAPLED TO COMPLETE THE STAPLE LINE.
Patient problems: Tissue Breakdown
MAUDE report 19951036 ↗DeathJul 12, 2024Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A THORACIC PROCEDURE, THE STAPLES DID NOT FORM A B-SHAPE, ANOTHER RELOAD WAS USED TO CLOSE THE WOUND. AFTER USING SEVERAL LOADS IN A LUNG RESECTION THE PATIENT WAS CLOSED AND A DRAINAGE WAS USED, AND IMMEDIATELY THE RESERVOIR WAS FILLED WITH 1 LITER OF BLOOD SO THE PATIENT HAD TO BE OPENED AGAIN TO LOOK FOR THE LEAKAGE, IT WAS FOUND THAT THE LEAKAGE WAS FROM A VASCULAR STUMP. THE SURGICAL TEAM THEN APPLIED MORE CLIPS ON THE PATIENT AND A BLOOD TRANSFUSION WAS PERFORMED. AFTER THAT THE PATIENT WAS SENT TO THE INTENSIVE CARE UNIT (ICU). DURING THE PROCEDURE THE USER HAD TO USE ANOTHER COMPANY'S CLIP APPLIERS. THE PATIENT WENT INTO CARDIAC ARREST AND DIED AT AROUND 2 AM.
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding; Tissue Breakdown
MAUDE report 19735881 ↗MalfunctionJan 2, 2024Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, DURING AN OPEN HEPATECTOMY, DURING CUTTING OF THE ANATOMICAL PIECE, THE DEVICE FIRED UNTIL THE END BUT STAPLE LINE WAS INCOMPLETE DISTALLY. A SECOND RELOAD WAS FIRED TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 18432599 ↗MalfunctionDec 13, 2023Misfire; Failure to Form Staple; Failure to Cut; Misfire; Failure to Form Staple; Failure to Cut
ACCORDING TO THE REPORTER, DURING THE ROBOTIC LAPAROSCOPIC PROCEDURE, THE STAPLE LINE WAS INCOMPLETE CENTRALLY AND THE THREE RELOADS FAILED TO CUT THE TISSUE. A NEW PRODUCT WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information; Tissue Breakdown
MAUDE report 18319958 ↗MalfunctionDec 13, 2023Misfire; Failure to Form Staple; Failure to Cut; Misfire; Failure to Form Staple; Failure to Cut
ACCORDING TO THE REPORTER, DURING THE ROBOTIC LAPAROSCOPIC PROCEDURE, THE STAPLE LINE WAS INCOMPLETE CENTRALLY AND THE THREE RELOADS FAILED TO CUT THE TISSUE. A NEW PRODUCT WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information; Tissue Breakdown
MAUDE report 18319955 ↗MalfunctionDec 13, 2023Misfire; Failure to Form Staple; Failure to Cut; Misfire; Failure to Form Staple; Failure to Cut
ACCORDING TO THE REPORTER, DURING THE ROBOTIC LAPAROSCOPIC PROCEDURE, THE STAPLER DID NOT CUT THE TISSUE COMPLETELY, AND FAILED TO FIRE TO THE END. THERE WAS NO PROBLEM AT THE STAPLE LINE TO THE PART WHERE THE TISSUE WAS CUT. A NEW PRODUCT WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information; Tissue Breakdown
MAUDE report 18319935 ↗MalfunctionNov 28, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, IN AN OPEN NEPHRECTOMY, THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON AND SQUEEZE THE HANDLE, BUT THE DEVICE DID NOT FIRE. THE HANDLE DID NOT ADVANCE, AND NO STAPLES WERE DEPLOYED. ANOTHER RELOAD WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18221050 ↗MalfunctionNov 14, 2023Failure to Form Staple; Failure to Form Staple
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC SEGMENTECTOMY, WHEN DETACHING BLOOD VESSELS AND DETACHING LUNG TISSUE, STAPLER FIRED, BUT STAPLES BURST OUT FOR BOTH RELOADS. THE STAPLE LINE OF THE FIRST RELOAD WAS INCOMPLETE DISTALLY, AND SECOND RELOAD WITH INCOMPLETE STAPLE LINE CENTRALLY. THE RELOADS WERE REPLACED TO RESOLVE THE ISSUE.
Patient problems: Tissue Breakdown
MAUDE report 18134901 ↗MalfunctionSep 6, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE STAPLER WAS NOT ABLE TO FIRE. THE RELOAD WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17695718 ↗MalfunctionJan 31, 2023Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING TESTING, THE STAPLER DID NOT FIRE. THE RELOAD WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16273175 ↗MalfunctionJan 11, 2023Failure to Cut; Failure to Cut
ACCORDING TO THE REPORTER, DURING AN OPEN THORACOTOMY PLUS LOBECTOMY DUE TO PNEUMONIA, THE DEVICE STAPLED BUT DID NOT CUT. TO RESOLVE THE ISSUE, A COLD CUT USING SCISSORS WAS MADE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16137983 ↗MalfunctionDec 8, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, THE RELOAD DID NOT FIRE. TO RESOLVE THE ISSUE, A NEW RELOAD WAS USED. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15945957 ↗MalfunctionOct 24, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, WHILE STAPLING, BEFORE USING THE FOUR RELOADS ON THE PATIENT, THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON OF THE TWO HANDLES BUT UNABLE TO SQUEEZE AND THE DEVICES DID NOT FIRE. A COMPETITOR INSTRUMENT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15661273 ↗MalfunctionOct 24, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, WHILE STAPLING, BEFORE USING THE FOUR RELOADS ON THE PATIENT, THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON OF THE TWO HANDLES BUT UNABLE TO SQUEEZE AND THE DEVICES DID NOT FIRE. A COMPETITOR INSTRUMENT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15658394 ↗MalfunctionOct 24, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, WHILE STAPLING, THE SURGEON USED FOUR RELOADS AND WAS ABLE TO PRESS THE GREEN BUTTON OF THE TWO HANDLES BUT UNABLE TO SQUEEZE AND THE DEVICES DID NOT FIRE. A COMPETITOR INSTRUMENT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15658356 ↗MalfunctionOct 24, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, WHILE STAPLING, BEFORE USING THE FOUR RELOADS ON THE PATIENT, THE SURGEON WAS ABLE TO PRESS THE GREEN BUTTON OF THE TWO HANDLES BUT UNABLE TO SQUEEZE AND THE DEVICES DID NOT FIRE. A COMPETITOR INSTRUMENT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15658331 ↗MalfunctionMay 14, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HERNIA PROCEDURE, THE CARTRIDGE WAS NOT COMING OUT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14402906 ↗MalfunctionMar 10, 2022Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PROCEDURE, THE STAPLER DID NOT FIRE. THE SURGEON USER WAS NOT ABLE TO PRESS THE GREEN BUTTON AND WAS NOT ABLE TO SQUEEZED THE HANDLE. TO RESOLVE THE ISSUE, NEW PRODUCT WAS EXCHANGED.
Patient problems: Insufficient Information
MAUDE report 13726119 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111825. One submission often covers many models, so check the product and lot details in each record.
- Z-0747-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Stapler with DST Series Technology 30mm-V 3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection or tra
- Z-0748-2021 · Nov 23, 2020 · CompletedCovidien TA Auto Suture Vascular Loading Unit with DST Series Technology 30 mm- V3. Intended for abdominal, gynecological, pediatric and thoracic surgical procedures for resection
- Z-0003-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 3.5 mm, Item Code 030452 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0007-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit, Item Code 030457 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric and thoracic
- Z-0001-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Gray Universal Articulating Loading Unit 30 mm - 2.0 mm, Item Code 030450 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gyn
- Z-0002-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 30 mm - 2.5 mm, Item Code 030451 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0006-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 3.5 mm, Item Code 030455 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0005-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Loading Unit 45 mm - 2.5mm, Item Code 030454 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecologic, pediatric
- Z-0009-2021 · Aug 12, 2020 · Open, ClassifiedCovidien Best Practices Procedure Kits containing Auto Suture Endo GIA Universal Loading Units, Parent Codes 00Z2760 - Product Usage: The Endo GIA universal staplers have applicat
- Z-0008-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 60 mm - 3.5 mm, Item Code 030458 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-0004-2021 · Aug 12, 2020 · Open, ClassifiedEndo GIA Auto Suture Universal Articulating Loading Unit 45 mm - 2.0 mm, Item Code 030453 - Product Usage: The Endo GIA universal staplers have applications in abdominal, gynecolo
- Z-1264-2020 · Jan 17, 2020 · TerminatedCovidien Endo GIA Ultra Universal Stapler 12 mm-intended for resection, transection in general surgery Item Code: EGIAUSTND
- Z-1826-2019 · May 17, 2019 · Open, ClassifiedEGIA 45 ARTICULATING MED THICK SULU, Item Code EGIA45AMT
- Z-1832-2019 · May 17, 2019 · Open, ClassifiedSIGNIA TRI-STAPLE 2.0 30MM MEDIUM/ THICK RELOAD , SIG30AMT
- Z-1827-2019 · May 17, 2019 · Open, ClassifiedEGIA 45 ARTICULATING XTRA THICK SULU, Item Code EGIA45AXT
…and 58 more.
Other STAPLE, IMPLANTABLE devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page