TEMNO EVOLUTION CTT1811Recalls & reported problems
Carefusion 2200, Inc · TEMNO EVOLUTION COAXIAL W/18G X 11CM NDL · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885403089343), received between Jan 15, 2020 and Jul 1, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Fracture1
- Material Separation1
Patient problems
- Foreign Body In Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2020Fracture; Material Separation; Detachment of Device or Device Component; Fracture; Material Separation; Detachment of Device or Device Component
THE ACCOUNT ALLEGES THAT DURING LUNG BIOPSY PROCEDURE, THE BIOPSY NEEDLE TIP DETACHED WITHIN THE PATIENT. THE DETACHED TIP WAS LEFT WITHIN THE PATIENTS [LUNG] TUMOR. THE ACCOUNT STATES THAT THE TIP WILL BE EXCISED FROM THE TUMOR AT A LATER DATE. PATIENT WAS DISCHARGED TO HOME AS EXPECTED, PAIN FREE.
Patient problems: Foreign Body In Patient
MAUDE report 10221200 ↗InjuryJan 15, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE ACCOUNT ALLEGES THAT DURING LUNG BIOPSY PROCEDURE, THE BIOPSY NEEDLE TIP DETACHED WITHIN THE PATIENT. THE PATIENT WAS PREVIOUSLY DIAGNOSED WITH COPD AND SUFFERED FROM AICHMOPHOBIA. THE HOSPITAL ALLEGED THAT THE PATIENT WOULD NOT LIE STILL DURING THE BIOPSY PROCEDURE. EVENTUALLY THE PATIENT'S OXYGEN SATURATIONS STARTED TO DECREASE, AND A CODE WAS CALLED. HOSPITAL BLS PROTOCOLS WERE ADMINISTERED REQUIRING INTUBATION AND NECESSARY RESUSCITATION ATTEMPTS. THE NEEDLE TIP WAS SUCCESSFULLY REMOVED VIA SURGICAL EXCISION.
Patient problems: Foreign Body In Patient
MAUDE report 9594321 ↗
Other Biopsy needle devices
- MAGNUM MN1410Bard Peripheral Vascular, Inc. · Biopsy needle1 recall118 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4021Olympus Medical Systems Corp. · Biopsy needle1 recall94 reports
- EchoTip G31520Cook Ireland Ltd · Biopsy needle1 recall84 reports
- Single Use Aspiration Needle NA-201SX-4021Olympus Medical Systems Corp. · Biopsy needle1 death126 reports
- EchoTip ProCore G34785Cook Ireland Ltd · Biopsy needle97 reports
- MAGNUM MN1820Bard Peripheral Vascular, Inc. · Biopsy needle1 recall37 reports
- Single Use Aspiration Needle NA-U401SX NA-U401SX-4022Olympus Medical Systems Corp. · Biopsy needle1 recall29 reports
- Interject M00518250Boston Scientific Corporation · Biopsy needle70 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.