Alaris 8110 (UDI 10885403811036)Recalls & reported problems

Carefusion 303, Inc. · Alaris Syringe Module · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Alaris 8110 syringe module is a pump module that delivers medication from a syringe. Its 741 reports mostly describe malfunctions, including broken or cracked parts, software freezing or failing to start, and corroded or worn components. The reports include no deaths and 6 injuries. One recall, open since August 2024, concerns a bent connector that could delay the start of infusion or monitoring.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
748
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 748 adverse event reports that list this exact device (DI 10885403811036), received between Apr 13, 2020 and Sep 22, 2026 — classified as 6 injuries, 742 malfunctions.

The most frequently reported specific device problem is “Break” (230 reports).

1 FDA recall record name this device's identifier, 1 still open.

5 other recalls cover devices cleared under the same FDA submission (K072105); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 15 reportsDec 2023: 4 reportsJan 2024: 11 reportsFeb 2024: 25 reportsMar 2024: 21 reportsApr 2024: 5 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 8 reportsAug 2024: 2 reportsSep 2024: 6 reportsOct 2024: 2 reportsNov 2024: 19 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 4 reportsApr 2025: 1 reportsMay 2025: 11 reportsJun 2025: 3 reportsJul 2025: 8 reportsAug 2025: 2 reportsSep 2025: 16 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 8 reportsJan 2026: 1 reportsFeb 2026: 5 reportsMar 2026: 3 reportsApr 2026: 3 reportsMay 2026: 16 reportsJun 2026: 2 reportsJul 2026: 5 reportsAug 2026: 1 reportsSep 2026: 6 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2024Open, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3141-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Break230
  • Application Program Freezes or Fails to Launch169
  • Deformation Due to Compressive Stress113
  • Corroded104
  • Crack93
  • Naturally Worn80
  • Physical Resistance/Sticking64
  • Device Alarm System55

Patient problems

  • No Clinical Signs, Symptoms or Conditions706
  • No Patient Involvement30
  • Insufficient Information8
  • No Consequences Or Impact To Patient7
  • Extravasation1
  • Hypoglycemia1
  • Infiltration into Tissue1
  • No Information1

Recent adverse event reports

The 25 most recent of 748 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50853266 ↗
  • MalfunctionSep 16, 2026Device Alarm System

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785705 ↗
  • MalfunctionSep 15, 2026Calibration Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50768188 ↗
  • MalfunctionSep 13, 2026Break; Corroded; Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50747821 ↗
  • MalfunctionSep 9, 2026Break; Corroded; Component Misassembled

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50683620 ↗
  • MalfunctionSep 3, 2026Break; Corroded; Physical Resistance/Sticking

    It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]"; "[Failed calibration]". There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50579670 ↗
  • MalfunctionAug 26, 2026Device Alarm System

    It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[8110 sn: (b)(6) gives error 351.6740.0]". There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456655 ↗
  • MalfunctionJul 31, 2026Naturally Worn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50098290 ↗
  • MalfunctionJul 27, 2026Break; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50034013 ↗
  • MalfunctionJul 27, 2026Break; Crack; Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50033491 ↗
  • MalfunctionJul 27, 2026Break; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50033490 ↗
  • MalfunctionJul 1, 2026Calibration Problem

    It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[failed force syringe calibration]". There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25676630 ↗
  • MalfunctionJun 5, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTATIVE MAINTENANCE, RECALL REMEDIATION, OR REPAIR ORDER SERVICE EVENT. FIELD SERVICE; RELEASE LATCH ERROR - KEEPS STICKING AND JAMMING. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25413719 ↗
  • MalfunctionJun 5, 2026Device Alarm System; Circuit Failure; Component Misassembled

    It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Depot Repair]";"[Error message ¿channel error¿ and unable to be used afterwards.]". There was no reported patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25407209 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT,  THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25172964 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25170030 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168583 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168472 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT,  THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168155 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168150 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168084 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168065 ↗
  • MalfunctionMay 14, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS.  THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25168055 ↗
  • MalfunctionMay 13, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25160292 ↗
  • MalfunctionMay 13, 2026Device Alarm System

    IT WAS REPORTED THAT THE CUSTOMER IS ENCOUNTERING AND INCREASED NUMBER CHANNEL ERRORS WHEN CONNECTING THE DEVICE TO THE PCU. COMMON ERROR CODES SEEN ARE 211.2 AND 221.201. CUSTOMER HAS CONDUCTED PRELIMINARY TESTING AND STATES THAT AFTER FLASHING THE PUMP, THE DEVICE WILL OPERATE NORMALLY FOR A FEW MINUTES, AFTER WHICH THE SAME CHANNEL ERROR WILL REAPPEAR. THE CUSTOMER HAS REPLACED THE LOGIC BOARD IN MANY OF THE DEVICES. FOLLOWING LOGIC BOARD REPLACEMENT, THE DEVICE WILL OPERATE AS NORMAL, WITHOUT CHANNEL ERRORS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25160284 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K072105. One submission often covers many models, so check the product and lot details in each record.

  • Z-1645-2026 · Feb 11, 2026 · Open, Classified
    BD Alaris" System with Guardrails" Suite MX with Point of Care Unit
  • Z-0184-2024 · Sep 15, 2023 · Open, Classified
    BD Alaris PCU REF 8015
  • Z-0186-2024 · Sep 15, 2023 · Open, Classified
    Alaris PCA Module 8120
  • Z-0185-2024 · Sep 15, 2023 · Open, Classified
    BD Alaris Syringe Module, REF 8110
  • Z-1192-2023 · Jan 24, 2023 · Open, Classified
    Alaris PC Unit 8015 utilizing 802.11 b/g or 802.11 a/b/g wireless network cards, with IP addresses from x.x.x.224 to x.x.x.255

Other Pump, infusion devices