Alaris 8110 (UDI 10885403811043)Recalls & reported problems

Carefusion 303, Inc. · Alaris Syringe Module · Class 2

All models of BD Alaris Infusion System (29) · Alaris (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Alaris 8110 syringe module is an infusion pump component. Its 4,298 adverse event reports mostly describe broken, cracked, worn or corroded parts, and some describe a set of pumps shutting down during an infusion. The reports include no deaths and 9 injuries. The last 12 months had 111 reports, down from 893 in the 12 months before. 2 open recalls name this device, covering bent connectors that can delay infusions and syringe labeling that lists unvalidated syringes.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,299
Deaths reported
0
Injuries reported
10
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4,299 adverse event reports that list this exact device (DI 10885403811043), received between Jan 25, 2022 and Sep 9, 2026 — classified as 10 injuries, 4,289 malfunctions.

The most frequently reported specific device problem is “Break” (1,803 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 322 reportsDec 2023: 246 reportsJan 2024: 192 reportsFeb 2024: 171 reportsMar 2024: 157 reportsApr 2024: 133 reportsMay 2024: 57 reportsJun 2024: 58 reportsJul 2024: 68 reportsAug 2024: 53 reportsSep 2024: 92 reportsOct 2024: 74 reportsNov 2024: 124 reportsDec 2024: 48 reportsJan 2025: 110 reportsFeb 2025: 87 reportsMar 2025: 79 reportsApr 2025: 41 reportsMay 2025: 55 reportsJun 2025: 22 reportsJul 2025: 33 reportsAug 2025: 252 reportsSep 2025: 24 reportsOct 2025: 18 reportsNov 2025: 24 reportsDec 2025: 19 reportsJan 2026: 13 reportsFeb 2026: 10 reportsMar 2026: 16 reportsApr 2026: 11 reportsMay 2026: 7 reportsJun 2026: 4 reportsJul 2026: 7 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports202420252026322
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3141-2024 · Root cause: Process control · Recall details · FDA record ↗

  2. Sep 15, 2023Class IOpen, Classified

    Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

    Z-0185-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1,803
  • Deformation Due to Compressive Stress1,635
  • Naturally Worn861
  • Crack848
  • Corroded829
  • Contamination619
  • Physical Resistance/Sticking522
  • Failure to Calibrate251

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,270
  • Insufficient Information28
  • Low Blood Pressure/ Hypotension4
  • Bradycardia2
  • Decreased Level of Consciousness1
  • High Blood Pressure/ Hypertension1
  • Low Oxygen Saturation1

Recent adverse event reports

The 25 most recent of 4,299 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2026Excess Flow or Over-Infusion

    It was reported that a pediatric ICU patient was noted at the start of shift to be significantly less responsive than on prior days despite decreasing sedation in preparation for extubation. The nurse stopped the morphine infusion and later dexmedetomidine and subsequently observed that the morphine syringe contained far less medication than expected. Clinical findings, including decreased responsiveness, urinary retention, rising CO2 levels, ventilator support requirements, constricted pupils, and reduced cough strength, were consistent with opioid toxicity. Although naloxone was initially deferred, administration of 0.05 mg naloxone later that day resulted in immediate improvement in alertness, eye tracking, responsiveness, and cough/gag reflexes, confirming opioid effect as the likely cause of the altered mental status.Biomedical Engineering investigated the Syringe Infusion Pump Module and found no pump programming errors, alarms, or equipment faults. A cracked morphine syringe was identified, with the defect located beneath the syringe label and difficult to detect visually. Customer testing demonstrated that air could enter through the crack, allegedly allowing medication to

    Patient problems: Decreased Level of Consciousness

    MAUDE report 50685190 ↗
  • MalfunctionJul 20, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25872774 ↗
  • MalfunctionJul 15, 2026Loose or Intermittent Connection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25823548 ↗
  • MalfunctionJul 13, 2026Calibration Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25793746 ↗
  • MalfunctionJul 10, 2026Break; Deformation Due to Compressive Stress; Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25778805 ↗
  • MalfunctionJul 9, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25761223 ↗
  • MalfunctionJul 9, 2026Display or Visual Feedback Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25761220 ↗
  • MalfunctionJul 3, 2026Naturally Worn

    IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703647 ↗
  • MalfunctionJun 17, 2026Pumping Stopped; No Apparent Adverse Event

    IT WAS REPORTED THAT THREE PUMPS INVOLVED IN AN INCIDENT AND THE ENTIRE SET SHUT DOWN ON ITS OWN. THE PATIENT WAS RECEIVING AN INFUSION OF HEPARIN (1UNIT/ML IN 0.45% NACL) AT A RATE OF 1ML / HOUR THE EVENT OCCURRED BETWEEN 2300-0100 HOURS. THE DEVICES WERE PLUGGED INTO AC POWER AT THE TIME OF THE EVENT. PICC LINE WAS FLUSHED AND SUBSEQUENTLY HAD TO BE REPLACED BUT NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25530731 ↗
  • MalfunctionJun 16, 2026Inaccurate Delivery; No Apparent Adverse Event

    IT WAS REPORTED THAT THE ALARIS SYRINGE PUMP MODULE ADMINISTERING CONTINUOUS ALPROSTADIL INFUSION REPEATEDLY ALARMED FOR ¿PATIENT OCCLUSION,¿ WITH INCONSISTENT DOCUMENTED INFUSION VOLUMES (0.06ML AT 8PM, 0.09ML AT 9PM, AND A DECREASED VOLUME OF 0.08ML AT 10PM), LIKELY RELATED TO FREQUENT INTERRUPTIONS FROM OCCLUSION ALARMS ON (B)(6) 2026. AFTER SWITCHING TO A DIFFERENT SYRINGE PUMP, THE INFUSION VOLUMES WERE ACCURATE. BIOMED EVALUATED THE RETURNED SYRINGE ONLY (P119377), AS THE PCU AND TUBING WERE UNAVAILABLE, AND WAS ABLE TO REPLICATE DECREASING VOLUME READINGS WHEN OCCLUSION ALARMS WERE CLEARED WITHOUT RESOLVING THE OCCLUSION. CUSTOMER REPORTED "MINOR HARM, AS THE PATIENT MAY HAVE RECEIVED LESS ALPROSTADIL THAN WHAT WAS TRACKED BY THE PUMP". NO FURTHER INFORMATION WAS MADE AVAILABLE. CUSTOMER BIOMED HAS PERFORMED INVESTIGATION ON THE DEVICE. PCU AND LINES WERE NOT ABLE TO BE SET ASIDE. CUSTOMER ABLE TO REPLICATE THE DECREASING VOLUME WHEN THE PUMP OCCLUSION ALARM IS CLEARED WITHOUT CLEARING THE OCCLUSION. DEVICES WILL NOT BE SENT TO BD FOR INVESTIGATION.

    Patient problems: Insufficient Information

    MAUDE report 25521552 ↗
  • MalfunctionJun 16, 2026Display or Visual Feedback Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT THE PATIENT WAS RECEIVING PROSTIN VIA A PICC LINE USING A SYRINGE PUMP MODULE (CHANNEL C) WHEN A DISCREPANCY IN INFUSION VOLUME WAS NOTED. AFTER RESETTING THE ¿VOLUME INFUSED¿ TO TRACK HOURLY DELIVERY, THE RN OBSERVED 0.05ML INFUSED AT ~0920 BEFORE GOING ON A BREAK. UPON RETURN AT 1000, THE VOLUME INFUSED DISPLAY WAS BLANK DESPITE THE PUMP INDICATING THE INFUSION WAS ONGOING, PREVENTING CONFIRMATION OF DELIVERED MEDICATION. AT ~1030, THE DISPLAY REAPPEARED BUT SHOWED 0ML INFUSED, LEAVING THE AMOUNT ADMINISTERED DURING THAT PERIOD UNKNOWN. THE PUMP FUNCTIONED NORMALLY AGAIN AFTER 1100 BUT WAS REPLACED DUE TO SAFETY CONCERNS. THERE WAS PATIENT INVOLVEMENT, BUT NO HARM. BIOMED RECEIVED THE PCU AND LATER IDENTIFIED THE ASSOCIATED SYRINGE PUMP MODULE, THOUGH NO SYRINGE OR TUBING WAS AVAILABLE FOR EVALUATION. THE ISSUE COULD NOT BE REPLICATED, AND INVESTIGATION IS ONGOING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25521547 ↗
  • MalfunctionJun 15, 2026Naturally Worn

    IT WAS REPORTED THAT THE DEVICE PLUNGER HAS RETURNED TO BEING STUCK/STICKY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25508103 ↗
  • MalfunctionMay 22, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE DEVICE HAD PLUNGER FORCE ACCURACY FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25252381 ↗
  • MalfunctionMay 18, 2026Accessory Incompatible; Use of Device Problem; Insufficient Flow or Under Infusion; Naturally Worn

    IT WAS REPORTED THAT THE SYRINGE PUMP ALARMED MULTIPLE TIMES AND EVENTUALLY DISPLAYED A GLITCH SCREEN. AS A RESULT, THE NURSING STAFF WERE UNABLE TO DETERMINE HOW MUCH VOLUME HAD BEEN INFUSED TO THE PATIENT. THE CUSTOMER PROVIDED THE FOLLOWING INFORMATION: PC UNIT USED DURING THE INCIDENT WAS IDENTIFIED FROM NICU EVENT AND ERROR LOGS DOWNLOADED FOR BOTH PCU AND SYRINGE MODULE. NO RELEVANT ERRORS IDENTIFIED IN THE ERROR LOGS FOR EITHER DEVICE. PCU EVENT LOGS REVIEWED BY HOSPITAL BIOMED: DRUG: ALPROSTADIL , RATE: 1.39 ML/H, DRUG AMOUNT: 5 MCG, DILUENT VOLUME: 1 ML , CONCENTRATION: 5 MCG/ML , DOSE: 0.03 MCG/KG/MIN , PATIENT WEIGHT: 3.86 KG , TIME UNITS: MINUTES , SYRINGE TYPE: BD 50 ML. EVENT LOG REVIEW SHOWED: O BETWEEN 01:00 04:00, THE INFUSOR STOPPED AND ALARMED APPROXIMATELY 8 TIMES PRIOR TO TUBING REPLACEMENT. TUBING REPLACEMENT RECORDED AT 03:13. CHANNEL TURNED OFF AT 03:54. EACH ALARM WAS FOLLOWED BY PUMP RESTART AND REPROGRAMMING. RECORDED VOLUME INFUSED VALUES WERE INCONSISTENT AND DECREASED ACROSS EVENTS, INDICATING UNRELIABLE VOLUME TRACKING, LIKELY DUE TO REPEATED RESTARTS/REPROGRAMMING. EVENT SEQUENCES SHOW THE PUMP WAS REPEATEDLY RESTARTED IMMEDIATELY AFTER OCCLUSION ALA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25211754 ↗
  • MalfunctionMay 13, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT THE DEVICE HAD FAILED BARREL CLAMP OPEN TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25158711 ↗
  • MalfunctionMay 13, 2026Break; Deformation Due to Compressive Stress; Naturally Worn

    IT WAS REPORTED THAT THE DEVICE HAD BAD MOTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25158700 ↗
  • MalfunctionMay 13, 2026Break; Contamination; Output Problem

    IT WAS REPORTED THAT THE DEVICE HAD BAD MOTOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25158257 ↗
  • MalfunctionMay 13, 2026Break; Circuit Failure; Deformation Due to Compressive Stress; Naturally Worn

    IT WAS REPORTED THAT THE DEVICE HAD DRIVE ARM MOVING FREELY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25157836 ↗
  • MalfunctionMay 6, 2026Calibration Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE DEVICE HAD DEFECTIVE BARREL CLAMP THAT IS NOT WORKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25092579 ↗
  • MalfunctionApr 27, 2026Break; Loose or Intermittent Connection; Deformation Due to Compressive Stress; Output Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE DEVICE HAD RATTLING INTERNALLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24998502 ↗
  • MalfunctionApr 24, 2026Break; Deformation Due to Compressive Stress; Naturally Worn; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE DEVICE HAD FAILED PLUNGER FORCE CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24979566 ↗
  • MalfunctionApr 21, 2026Break; Deformation Due to Compressive Stress

    IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24938551 ↗
  • MalfunctionApr 17, 2026Device Sensing Problem

    IT WAS REPORTED THAT A FAILURE WAS OBSERVED DURING A PLANNED PREVENTIVE MAINTENANCE OR RECALL REMEDIATION SERVICE EVENT. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24906744 ↗
  • MalfunctionApr 6, 2026Device Alarm System

    IT WAS REPORTED THAT THE DEVICE HAD CHANNEL ERROR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788617 ↗
  • MalfunctionApr 6, 2026Break; Contamination

    IT WAS REPORTED THAT THE DEVICE HAD CRACKED FRONT CASE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24788556 ↗

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page