Skip to content

Olympus A22255CRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14042761036839), received between May 30, 2019 and Mar 19, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 8, 2023Class IIOpen, Classified

    Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip

    Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component1
  • Sparking1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THE TIP OF THE HIGH FREQUENCY RESECTION ELECTRODE BROKE WHILE INSIDE THE PATIENT DURING A HYSTEROSCOPY/DILATION AND CURETTAGE (D&C) PROCEDURE. THE BROKEN PORTION WAS RETRIEVED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THE DEVICE WAS DISCARDED. THERE WAS NO REPORTED PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18935494 ↗
  • MalfunctionMay 6, 2022Break; Sparking

    SHORTLY AFTER THE PROCEDURE IS STARTED, A LOUD POP IS HEARD AND IMMEDIATELY REALIZED THAT MONOPOLAR CORD SPARKED AND BROKEN OFF WHERE IT ATTACHES TO ESU (ELECTRO-SURGICAL UNIT) MACHINE.

    Patient problems: Insufficient Information

    MAUDE report 14308730 ↗
  • MalfunctionMay 30, 2019Break; Break

    THE SERVICE CENTER WAS INFORMED THAT DURING AN UNSPECIFIED PROCEDURE, THE DOCTOR OBSERVED THAT THE STERILE DISPOSABLE ELECTRODE NEEDLE BROKE OFF AND FELL INTO THE PATIENT¿S BLADDER. THE DOCTOR WAS ABLE RETRIEVE THE METAL DEVICE FRAGMENT. THE INTENDED PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8654958 ↗

FDA clearance and classification

The FDA regulates product code KNS (unit, electrosurgical, endoscopic (with or without accessories)) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Olympus A22255C

Has the Olympus A22255C been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2413-2023 (Class II), began on Jun 8, 2023: Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Olympus A22255C?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between May 30, 2019 and Mar 19, 2024. In those reports, the most common specific device problems are “Break” (3), “Detachment of Device or Device Component” (1) and “Sparking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Olympus A22255C?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Olympus A22255C, and is it still sold?

The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model A22255C). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, Electrosurgical, Endoscopic (With Or Without Accessories) devices

All Unit, Electrosurgical, Endoscopic (With Or Without Accessories) devices →

Sources and updates