Olympus WA22039CRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 14042761051603), received between Oct 13, 2021 and Oct 13, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 8, 2023Class IIOpen, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 13, 2021Material Separation; Material Separation
IT IS REPORTED BY THE CUSTOMER (AND IN MW5103123), DURING A TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) USING A SINGLE USE HF-RESECTION ELECTRODE, THE ELECTRODE BROKE INSIDE THE PATIENT'S URETHRA. MOST OF THE DEVICE FRAGMENTS WERE REMOVED USING OTHER UNSPECIFIED INSTRUMENTS. POST-OPERATIVE X-RAYS DID NOT DEMONSTRATE ANY RETAINED FOREIGN BODY. THE CUSTOMER COULD NOT CONFIRM IF THERE WERE ANY PROCEDURAL COMPLICATION OR ANATOMICAL CHALLENGES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT AS A RESULT OF THIS OCCURRENCE. THE PATIENT'S CURRENT CONDITION IS REPORTED AS: PATIENT WAS DISCHARGED FROM THIS HOSPITAL ADMISSION. THE SETTINGS USED FOR THE PROCEDURE WERE 120/200 CUT. THE CUSTOMER WAS UNABLE TO PROVIDE ANY FURTHER INFORMATION REGARDING THE PATIENT, DEVICES USED, OR EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12624268 ↗
Other Unit, Electrosurgical, Endoscopic (With Or Without Accessories) devices
- THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip TB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 recall2 deaths404 reports
- SONICBEAT 5 mm, 35 cm, Front-actuated Grip SB-0535FCOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)379 reports
- Single Use Electrosurgical Knife KD-611LOlympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)1 death357 reports
- Autotome™ RX 44 M00545170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall262 reports
- TRUEtome 44 M00584170Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)237 reports
- Dreamtome™ RX 44 M00584040Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall165 reports
- Dreamtome™ RX 44 M00584050Boston Scientific Corporation · Unit, electrosurgical, endoscopic (with or without accessories)1 recall127 reports
- Single Use 3-Lumen Sphincterotome V KD-V411M-0725Olympus Medical Systems Corp. · Unit, electrosurgical, endoscopic (with or without accessories)160 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page