Skip to content

Olympus WA22607DRecalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 14042761075487), received between Mar 8, 2019 and Mar 8, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jun 8, 2023Class IIOpen, Classified

    Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip

    Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2019Break; Break

    OLYMPUS RECEIVED A MEDWATCH (REPORT# (B)(4)), WHICH STATED THAT WHILE THE DOCTOR WAS PERFORMING A TURP AND USING THE BUTTON TURP LOOP, THE LOOP BROKE WHILE CAUTERIZING THE PROSTATE. THE ENDS OF THE LOOP CURLED UP. THE ENTIRE LOOP IS STILL THERE, IT JUST CURLED UP AT THE ENDS. A NEW LOOP WAS OBTAINED AND THE PROCEDURE CONTINUED WITH NO HARM TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8405938 ↗

FDA clearance and classification

The FDA regulates product code FAS (electrode, electrosurgical, active, urological) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Olympus WA22607D

Has the Olympus WA22607D been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2413-2023 (Class II), began on Jun 8, 2023: Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Olympus WA22607D?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 8, 2019. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Olympus WA22607D?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Olympus WA22607D, and is it still sold?

The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model WA22607D). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrode, Electrosurgical, Active, Urological devices

All Electrode, Electrosurgical, Active, Urological devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page