Olympus WA22544SRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jun 8, 2023Open, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · FDA record ↗
Other Electrode, Electrosurgical, Active, Urological devices
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22332DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22301DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.