Olympus WA22751SRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · HF-resection electrode PlasmaRoller roller, 24 Fr., 12°-30°, ESG TURis · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 14042761085486), received between Apr 14, 2021 and Apr 30, 2026 — classified as 5 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Detachment of Device or Device Component3
- Thermal Decomposition of Device2
- Material Fragmentation1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2026Material Twisted/Bent
IT WAS REPORTED THE RESECTION ELECTRODE HAD A BENT FORK TUBE. THERE WERE NO REPORTS OF PATIENT INJURY OR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25052819 ↗MalfunctionAug 18, 2025Break; Break
IT WAS REPORTED THE SUBJECT DEVICE BROKE DURING A TRANSURETHRAL RESECTION OF BLADDER TUMOR (TURBT) PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22826354 ↗InjuryNov 7, 2024Break; Thermal Decomposition of Device; Detachment of Device or Device Component; Break; Thermal Decomposition of Device; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20635252 ↗InjuryNov 7, 2024Break; Thermal Decomposition of Device; Detachment of Device or Device Component; Break; Thermal Decomposition of Device; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BALL AT THE TIP OF THE RESECTION ELECTRODE DEVICE BROKE OFF INSIDE THE PATIENT. THE PHYSICIAN DESCRIBED THE EVENT AS IT "BURNED OFF." THE EVENT OCCURRED AFTER APPROXIMATELY 20 MINUTES OF USE DURING A THERAPEUTIC PROSTATE VAPORIZATION. THE BALL WAS RETRIEVED INTACT. THE PROCEDURE LASTED FOR APPROXIMATELY 70 MINUTES, WITH AN ADDITIONAL 1-2 MINUTES IN RETRIEVING THE BROKEN PIECE AND WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT OR USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20632381 ↗InjuryJun 17, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THE LOOP BROKE INSIDE PATIENT DURING A THERAPEUTIC PROSTATE RESECTION PROCEDURE. SURGEON WAS ABLE TO RETRIEVE THE PIECE SUCCESSFULLY. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE WITHOUT DELAY. THERE WERE NO REPORTS OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19553652 ↗InjuryJan 3, 2024Break; Break
A OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A UNKNOWN THERAPEUTIC PROCEDURE THE TIP OF HF-RESECTION ELECTRODE "PLASMAROLLER", ROLLER, 24 FR., 12°-30°, ESG TURIS BROKE OFF INSIDE THE PATIENT. THE BROKEN PIECE AS RECOVERED AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18434300 ↗InjuryJan 3, 2024Break; Break
A OLYMPUS REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, DURING A UNKNOWN THERAPEUTIC PROCEDURE THE TIP OF HF-RESECTION ELECTRODE "PLASMAROLLER", ROLLER, 24 FR., 12°-30°, ESG TURIS BROKE OFF INSIDE THE PATIENT. THE BROKEN PIECE AS RECOVERED AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18434272 ↗MalfunctionApr 14, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED TO OLYMPUS THAT THE ROLLER BALL FELL OFF THE ROLLER DURING A PROCEDURE. IT WAS NOT DETERMINED IF THE BALL FELL INTO THE PATIENT OR ON THE GROUND. THE PATIENT HAD X-RAYS, HOWEVER, THE ROLLER BALL WAS NOT LOCATED IN THE PATIENT. THE FACILITY WILL MONITOR THE PATIENT. NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11665110 ↗
FDA clearance and classification
The FDA regulates product code FAS (electrode, electrosurgical, active, urological) under 21 CFR 876.4300 as a Class II device. 21 CFR 876.4300 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Olympus WA22751S
Has the Olympus WA22751S been recalled?
No FDA recall record in our data names this exact device identifier (DI 14042761085486), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Olympus WA22751S?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Apr 14, 2021 and Apr 30, 2026. In those reports, the most common specific device problems are “Break” (6), “Detachment of Device or Device Component” (3) and “Thermal Decomposition of Device” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Olympus WA22751S?
None of the reports are classified as deaths, 5 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Olympus WA22751S, and is it still sold?
The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model WA22751S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Electrode, Electrosurgical, Active, Urological devices
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Olympus WA22603SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall7 reports
- Olympus WA22621SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall4 reports
- Olympus WA22305SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall3 reports
- Olympus WA22657SOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22332DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
- Olympus WA22301DOLYMPUS Winter & Ibe GmbH · Electrode, Electrosurgical, Active, Urological1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page