Skip to content

Vyaire V-892380Recalls & reported problems

Jaeger Medical GmbH · MicroGard II, bacterial/viral filter · Class 2

All models of Vyaire (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 14250892903115), received between Oct 25, 2023 and Apr 17, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Component Misassembled” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Misassembled2
  • Failure to Calibrate2
  • Component Missing1
  • Contamination1
  • Sharp Edges1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Aspiration/Inhalation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 17, 2025Sharp Edges; Sharp Edges

    THE CUSTOMER REPORTED THAT THE MICROGARD FILTER HAS SHARP EDGES. THERE WAS NO PATIENT INVOLVEMENT REPORTED SINCE THE EVENT OCCURRED DURING AN INCOMING INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21856584 ↗
  • InjuryJan 15, 2025Contamination; Contamination

    THE CUSTOMER REPORTED TO VYAIRE ON (B)(6)2024 THAT A LUNG INJURY OCCURRED WHILE INHALING DURING A SPIROMETRY MEASUREMENT. DURING A FORCED SPIROMETRY MEASUREMENT FOR A NOCION CHRONIC COUGH STUDY USING THE MICROGARD II FILTER, THE TEST PATIENT FELT LIKE SHE HAD INHALED SOMETHING FROM THE DEVICE. AS A RESULT, THE PATIENT HAD TO COUGH AND THE VOICE BECAME UNUSUAL AND IRRITATED. A DOCTOR EXAMINED HER THROAT AND LISTENED TO HER LUNGS AND SUGGESTED THAT SHE GO TO THE ER FOR A CT SCAN BECAUSE HE BELIEVED THAT SHE HAD A TRACHEAL TEAR. A CT SCAN SHOWED THAT SHE DID HAVE A TRACHEAL TEAR. A SURGEON WAS CONSULTED AND AFTER TWO DAYS OF OBSERVATION, IT WAS DETERMINED THAT SHE WOULD NOT NEED SURGERY. AFTER TWO DAYS THE TEST PATIENT WAS DISCHARGED FROM THE HOSPITAL. THE SITE COORDINATOR CONDUCTING THE TEST INSPECTED THE MACHINE AND ALL DEVICES THAT WERE ATTACHED BUT DID NOT OBSERVE ANYTHING UNUSUAL.

    Patient problems: Aspiration/Inhalation

    MAUDE report 21158669 ↗
  • MalfunctionJun 24, 2024Failure to Calibrate; Component Misassembled; Failure to Calibrate; Component Misassembled

    THE CUSTOMER COMPLAINED THAT THE FILTER-LINING OF THE MICROGARD FILTER IS MISPLACED. THIS WAS NOTICED DURING AN INCOMING INSPECTION. THERE WAS NO INDICATION THAT DEFECTIVE FILTERS WERE USED FOR PATIENT MEASUREMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19592379 ↗
  • MalfunctionJun 19, 2024Failure to Calibrate; Component Misassembled; Failure to Calibrate; Component Misassembled

    THE CUSTOMER COMPLAINED THAT THE FILTER-LINING OF THE MICROGARD FILTER IS MISPLACED. THIS WAS NOTICED DURING AN INCOMING INSPECTION. THERE WAS NO INDICATION THAT DEFECTIVE FILTERS WERE USED FOR PATIENT MEASUREMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19565041 ↗
  • MalfunctionOct 25, 2023Component Missing; Component Missing

    IT WAS REPORTED TO VYAIRE THAT THE CUSTOMER (B)(6) GMBH RECEIVED A DELIVERY FROM VYAIRE WITH TWO MICROGARD FILTERS THAT WERE INCOMPLETE. IT WAS STATED THAT THE FILTER INLAY WAS MISSING FOR ONE FILTER, AND THE OTHER WAS MISSING 1/2 OF THE FILTER HOUSING. THE FILTERS WERE IDENTIFIED AND RETURNED TO VYAIRE FOR ANALYSIS. THE FILTERS WERE RETURNED INSIDE THEIR ORIGINAL & SEALED PACKAGING; THERE IS NO INDICATION THAT INCOMPLETE FILTERS WERE USED FOR PATIENT MEASUREMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18008557 ↗

Other Filter, Bacterial, Breathing-Circuit devices