DxC AU B86444Recalls & reported problems
Beckman Coulter, Inc. · DxC 700 AU Chemistry Analyzer (with ISE) · Class 1
All models of DxC AU (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Beckman Coulter DxC AU chemistry analyzer is a lab machine that measures blood test results. Most of the 586 reports describe results that were too low, too high, or otherwise incorrect, and a few mention results being released outside the lab. The reports include 1 death and 61 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 587
- Deaths reported
- 1
- Injuries reported
- 64
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 587 adverse event reports that list this exact device (DI 14987666542545), received between Jul 18, 2018 and Sep 17, 2026 — classified as 1 death, 64 injuries, 522 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (279 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results279
- High Test Results147
- Incorrect, Inadequate or Imprecise Result or Readings130
- Adverse Event Without Identified Device or Use Problem17
- Imprecise or Erratic Results13
- Low Readings3
- Unable to Obtain Readings2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- No Clinical Signs, Symptoms or Conditions497
- No Consequences Or Impact To Patient42
- Insufficient Information19
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available13
- Hyponatremia13
- No Code Available2
- Back pain1
- Needle Stick/Puncture1
Recent adverse event reports
The 25 most recent of 587 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
The customer reported the generation of erratic Vancomycin (Vanc) patient results on their DxC 700 AU Clinical Chemistry analyzer, One patient result was reported outside of the customer¿s laboratory and it was questioned by the physician. It was reported that the patient's vancomycin administration was modified, and an alternative antibiotic was prescribed. Patient was already under vancomycin treatment due to an infectious disease. There was no report of harm to the patient due to this event. The customer did not provide patient demographics.
Patient problems: Insufficient Information
MAUDE report 50811713 ↗InjurySep 2, 2026High Test Results
The customer reported erroneously high Hemoglobin A1c (HbA1c) results generated by the DxC 700 AU Chemistry Analyzer. The results were reported outside the laboratory, and the customer indicated that some patients may have been prescribed glucagon-like peptide-1 (GLP-1) medication based on the erroneous results. During follow-up, the customer clarified that 147 patients received falsely elevated results; however, this does not indicate that all 147 patients experienced treatment changes. The number of patients who may have been prescribed GLP-1 medication is unknown. Insufficient information was available to determine whether additional treatment changes, medical interventions, injuries, or other adverse outcomes occurred.
Patient problems: Insufficient Information
MAUDE report 50571303 ↗InjurySep 2, 2026High Test Results
The customer reported erroneously high Hemoglobin A1c (HbA1c) results generated by the DxC 700 AU Chemistry Analyzer. The results were reported outside the laboratory, and the customer indicated that one new diabetic patient was provided with diabetic supplies based on the erroneous results. There was no report of harm or death associated with event. The customer did not provide patient data or demographics.
Patient problems: Insufficient Information
MAUDE report 50571271 ↗InjuryJun 23, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR GLUCOSE, ALBUMIN, AND MAGNESIUM ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE RELEASED OUTSIDE THE LABORATORY. THE CUSTOMER INDICATED A CHANGE IN PATIENT MANAGEMENT, WHERE THE PATIENT WAS STARTED ON MAGNESIUM MEDICATION, WHICH WAS THEN DISCONTINUED AFTER CORRECTED RESULTS WERE AMENDED. BECKMAN COULTER CONTACTED THE CUSTOMER FOR MORE DETAILS, BUT THEY DECLINED TO PROVIDE ADDITIONAL PATIENT INFORMATION. HOWEVER, THEY CONFIRMED THE PATIENT WAS NOT HARMED. THERE WAS NO REPORT OF DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593517 ↗MalfunctionJun 23, 2026High Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH PATIENT RESULTS FOR SODIUM(NA) AND BICARBONATE (CO2) WITH LOW ANION GAP ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593510 ↗MalfunctionJun 23, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING ERRONEOUSLY LOW PATIENT RESULTS FOR GLUCOSE (GLU), SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) ANALYTES ON THEIR DXC700AU ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS, THE EXACT NUMBER OF PATIENT SAMPLES AFFECTED IS UNKNOWN. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593509 ↗MalfunctionJun 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED OBTAINING ERRATIC THREE (3) PATIENT RESULTS FOR SODIUM (NA), AND CHLORIDE (CL) DUE TO NEGATIVE ANION GAP VALUES, ON THE DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT PROVIDED THE RESULTS FOR ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25522663 ↗MalfunctionJun 11, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED OBTAINING ERRATIC PATIENT RESULTS ON ION SELECTIVE ELECTRODE FOR SODIUM, POTASSIUM AND CHLORIDE DUE TO LOW ANION GAP VALUES, INVOLVING THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE RELEASED OUTSIDE THE LABORATORY. HOWEVER, THE CUSTOMER INDICATED THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25479238 ↗MalfunctionJun 11, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR MAGNESIUM, PHOSPHORUS, GLUCOSE AND SODIUM ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25477146 ↗MalfunctionJun 8, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING FALSE LOW TOTAL BILIRUBIN (TBIL) PATIENT RESULT FOR ONE PATIENT, ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER (PART NUMBER (PN) B86444, SERIAL NUMBER (SN) (B)(6). THE CUSTOMER REPEATED THE SAMPLE AND OBTAINED A RESULT CONSIDERED CORRECT BY THE CUSTOMER. THERE WAS NO INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25436085 ↗MalfunctionJun 8, 2026Imprecise or Erratic Results
THE CUSTOMER REPORTED OBTAINING FALSE HIGH SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) PATIENT RESULTS GENERATED WITH LOW ANION GAP CALCULATION VALUES ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER (PART NUMBER (PN) B86444, SERIAL NUMBER (SN) (B)(6). THE CUSTOMER REPEATED THE SAMPLE AND OBTAINED A RESULT CONSIDERED CORRECT BY THE CUSTOMER. THERE WAS NO INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25433844 ↗MalfunctionJun 5, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR MULTIPLE CHEMISTRIES INCLUDING SODIUM, CHLORIDE, TOTAL PROTEIN, ASPARTATE AMINOTRANSFERASE, MAGNESIUM, BLOOD UREA NITROGEN, GLUCOSE AND PHOSPHORUS ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE PATIENT SAMPLE WAS REPEATED ON A DIFFERENT AU ANALYZER WITH NORMAL RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25416298 ↗MalfunctionJun 4, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED OBTAINING ERRATIC SODIUM PATIENT RESULTS ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THOSE ERRATIC RESULTS WERE NOT RELEASED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25399621 ↗MalfunctionMay 29, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW ION-SELECTIVE ELECTROLYTES (ISE) PATIENT RESULTS, INCLUDING SODIUM (NA), CHLORIDE (CL) AND POTASSIUM (K), ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25333986 ↗MalfunctionMay 28, 2026High Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH ION-SELECTIVE ELECTROLYTES (ISE) PATIENT RESULTS, INCLUDING SODIUM (NA), CHLORIDE (CL) AND POTASSIUM (K), ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25320842 ↗MalfunctionMay 26, 2026Imprecise or Erratic Results
THE CUSTOMER QUESTIONED THE GENERATION OF SODIUM (NA) PATIENT RESULTS ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER (PRODUCT NUMBER: B86444, SERIAL NUMBER: (B)(6). THE CUSTOMER REPORTED THAT NA PATIENT RESULTS RANGED BETWEEN 134-142 MMOL/L AND REPEATED +/- 5 MMOL/L ON RERUN TESTS. THERE WERE NO QUESTIONED NA PATIENT RESULTS RELEASED OUTSIDE OF THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25295414 ↗MalfunctionMay 22, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF FALSE LOW PATIENT RESULTS FOR SODIUM (NA) AND CHLORIDE (CL) WITH LOW ANION GAPS AND ALSO NOTED THAT MIX BAR MAKING NOISE ON THEIR DXC 700 AU CHEMISTRY ANALYZER. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25262529 ↗MalfunctionMay 22, 2026High Test Results
THE CUSTOMER REPORTED OBTAINING HIGH SODIUM (NA) PATIENT RESULTS WITH AN ¿H¿ FLAG ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER RELEASED EIGHT PATIENTS RESULTS OUTSIDE THE LABORATORY; HOWEVER, THOSE RESULTS WERE AMENDED UPON REPEAT TESTING. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT ONLY SHARED VERBAL RESULTS FOR TWO PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257651 ↗MalfunctionMay 22, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW ION-SELECTIVE ELECTROLYTES (ISE) PATIENT RESULTS, INCLUDING SODIUM (NA), CHLORIDE (CL) AND POTASSIUM (K), ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25256897 ↗InjuryMay 21, 2026Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REPORTED ON THEIR DXC 700 ANALYZER WHERE THE WASTE DRAIN GENERATED FOAM AND OVERFLOWED, MAKING THE FLOOR SLIPPERY. AS A RESULT, A LABORATORY OPERATOR SLIPPED, FELL AND SUSTAINED INJURIES TO THEIR BACK AND LEG. AT THE TIME OF THE INCIDENT, THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE). THE OPERATOR SOUGHT URGENT CARE AND BEGAN PHYSICAL THERAPY. IT IS UNKNOWN IF MEDICATION WAS PRESCRIBED. THE OPERATOR RETURNED TO WORK ON (B)(6) 2026, WITH WORK RESTRICTIONS. NO ADDITIONAL DETAILS WERE PROVIDED. THERE WAS NO REPORT OF DEATH ASSOCIATED WITH THIS EVENT.
Patient problems: Back pain
MAUDE report 25250169 ↗MalfunctionMay 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED OBTAINING ERRATIC PATIENT RESULTS ON SODIUM (NA) AND CHLORIDE (CL) DUE TO LOW ANION GAP VALUES INVOLVING THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE ERRATIC RESULTS WERE NOT RELEASED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25247135 ↗MalfunctionMay 21, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUS LOW SODIUM PATIENT RESULTS ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25244650 ↗MalfunctionMay 19, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS WITH A ¿G¿ FLAG FOR MULTIPLE CHEMISTRIES INCLUDING SODIUM (NA), POTASSIUM (K), CHLORIDE (CL), BICARBONATE (CO2), GLUCOSE (GLU), TOTAL PROTEIN (TP), TOTAL BILIRUBIN (TBIL), ALBUMIN (ALB), CREATININE (CREAT), CALCIUM (CA), BLOOD UREA NITROGEN (BUN), ASPARTATE AMINOTRANSFERASE (AST), ALKALINE PHOSPHATASE (ALP), AND ALANINE AMINOTRANSFERASE (ALT) INVOLVING THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE LOW RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT SHARED ONLY THE INITIAL RESULTS FOR TWO PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25224197 ↗MalfunctionMay 18, 2026Imprecise or Erratic Results
THE CUSTOMER REPORTED OBTAINING ONGOING ERRONEOUS ZERO RESULTS WITH ¿G¿ FLAGS FOR PATIENT SAMPLES ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER (PRODUCT NUMBER B86444, SERIAL NUMBER (B)(6). UPON FOLLOW-UP, THE CUSTOMER SPECIFIED ALL ASSAYS WERE IMPACTED. THERE WAS NO ERRONEOUS PATIENT RESULT REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA AND/OR PATIENT DEMOGRAPHICS. NOTE: PER DXC 700 AU CHEMISTRY ANALYZER INSTRUCTIONS FOR USE (IFU) GUIDE B71495AF: O FLAG ¿G¿: RESULT IS LOWER THAN THE ANALYTICAL MEASURING RANGE. O FLAG ¿/¿: TEST PENDING OR NOT ANALYZED. O FLAG ¿?¿: UNABLE TO CALCULATE A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25204175 ↗MalfunctionMay 14, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR MULTIPLE CHEMISTRIES ON THEIR DXC 700 AU CLINICAL CHEMISTRY ANALYZER. THE AFFECTED CHEMISTRIES WERE NOT SPECIFIED. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25178741 ↗
FDA clearance and classification
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DxC AU B86444
Has the DxC AU B86444 been recalled?
No FDA recall record in our data names this exact device identifier (DI 14987666542545), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DxC AU B86444?
The FDA's MAUDE database lists 587 adverse event reports for this device, received between Jul 18, 2018 and Sep 17, 2026. In those reports, the most common specific device problems are “Low Test Results” (279), “High Test Results” (147) and “Incorrect, Inadequate or Imprecise Result or Readings” (130); the most common patient problems are “Hyponatremia” (13), “Back pain” (1) and “Needle Stick/Puncture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DxC AU B86444?
1 report is classified as a death, 64 as injuries and 522 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the DxC AU B86444 FDA approved?
Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.
Who makes the DxC AU B86444, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand DxC AU, model B86444). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls677 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death645 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE680 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page