DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522Recalls & reported problems

Beckman Coulter, Inc. · The DxC 500i Clinical Analyzer combines the DxC 500 AU Clinical Chemistry Analyzer and the Access 2 Immunoassay System into a single instrument presentation. The system is for in vitro diagnostic use only. The chemistry module of the DxC 500i Clinical Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. The immunoassay module of the DxC 500i Clinical Analyzer is an in-vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids. C63522 includes an ISE module. · Class 1

All models of DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

4 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
4

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 14987666545089), received between Sep 5, 2025 and May 14, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

4 FDA recall records name this device's identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 2, 2026Class IIOpen, Classified

    Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

    Z-0083-2027 · Root cause: Software design · Recall details · FDA record ↗

  2. Jul 15, 2026Class IIOpen, Classified

    a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

    Z-2973-2026 · Root cause: Software design (manufacturing process) · Recall details · FDA record ↗

  3. Sep 19, 2025Class IIOpen, Classified

    Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

    Z-0215-2026 · Root cause: Software design · Recall details · FDA record ↗

  4. Aug 4, 2025Class IIOpen, Classified

    On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

    Z-0032-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • High Test Results2
  • Adverse Event Without Identified Device or Use Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2026Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED OBTAINING ERRATIC PATIENT RESULTS FOR ION SELECTIVE ELECTRODE ON SODIUM, POTASSIUM AND CHLORIDE ON THEIR DXC 500I CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25178441 ↗
  • InjuryMar 13, 2026High Test Results

    THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH LACTATE PATIENT RESULTS ON THEIR DXC 500I CHEMISTRY ANALYZER. THE CUSTOMER STATED THAT TWO (2) PATIENTS RECEIVED FLUID BOLUS TREATMENT BECAUSE OF THE ELEVATED RESULTS. THERE WAS NO REPORT OF DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT PROVIDED THE RESULTS FOR THIS PATIENT. MDR-2 ADDRESSES A SECOND PATIENT WHO RECEIVED TREATMENT. MDR-1, ADDRESSES FIRST PATIENT (PATIENT 1), IS REFERENCED IN MDR 9612296-2026-00052.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24599211 ↗
  • InjuryFeb 4, 2026High Test Results

    THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH LACTATE PATIENT RESULTS ON THEIR DXC 500I CHEMISTRY ANALYZER. THE CUSTOMER STATED THAT TWO (2) PATIENTS RECEIVED FLUID BOLUS TREATMENT BECAUSE OF THE ELEVATED RESULTS. THERE WAS NO REPORT OF DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT PROVIDED RESULTS FOR THIS PATIENT. MDR-1: FOR FIRST PATIENT (PATIENT 1) IS REFERENCED IN MDR 9612296-2026-00052. MDR-2: FOR A SECOND PATIENT (PATIENT 2) IS REFERENCED IN A NEW MDR 9612296-2026-00110.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260135 ↗
  • MalfunctionDec 1, 2025Low Test Results; Low Test Results

    THE CUSTOMER QUESTIONED LOW SODIUM PATIENT RESULTS GENERATED ON THEIR DXC500AU CLINICAL CHEMISTRY ANALYZER (P/N C63522 SN (B)(6)). THERE WAS NO REPORT OF INJURY, DEATH OR DELAYED/CHANGE IN TREATMENT OR DIAGNOSIS ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23688669 ↗
  • InjurySep 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED THERE WAS A DELAY IN RESULTS AND DIAGNOSIS AND TREATMENT FOR ONE (1) PATIENT INVOLVING THEIR DXC 500I CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THEY RECEIVED TROPONIN (HSTNI) AND TSH (THYROID STIMULATING HORMONE) PATIENT RESULTS 1-2 HOURS LATE AND B-TYPE NATRIURETIC PEPTIDE (BNP) RESULTS WERE DELAYED FOR 6 HOURS. THE CUSTOMER STATED THAT THE DELAY IN RESULTS WAS DUE TO NOT HAVING QUALITY CONTROL (QC) AND THEY COULD NOT REPORT PATIENT RESULTS. THE PATIENT HEART ATTACK DIAGNOSIS AND TREATMENT WERE DELAYED DUE TO DELAY IN RESULTS. AFTER OBTAINING RESULTS, AN ELECTROCARDIOGRAM (ECG) WAS PERFORMED, TROPONIN LEVELS TESTING WAS ORDERED AND PERFORMED WHICH CONFIRMED ELEVATED RESULTS. THE PATIENT WAS ADMITTED TO THE HOSPITAL AND WAS PROVIDED TREATMENT. THERE IS NO DEATH ASSOCIATED WITH THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22978325 ↗

FDA clearance and classification

The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522

Has the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522 been recalled?

Yes. 4 FDA recall records name this device's identifier: 4 still open. The most recent, Z-0083-2027 (Class II), began on Sep 2, 2026: Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Sep 5, 2025 and May 14, 2026. In those reports, the most common specific device problems are “High Test Results” (2), “Incorrect, Inadequate or Imprecise Result or Readings” (1) and “Low Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522?

None of the reports are classified as deaths, 3 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522 FDA approved?

Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.

Who makes the DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module C63522, and is it still sold?

The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module, model C63522). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices

All Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page