DxC 500i Clinical Analyzer, DxC 500i AU Chemistry ModuleRecalls & reported problems across all models

Beckman Coulter, Inc. · 2 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

4 open recalls name at least one DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls
4

We found 2 models of DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module by Beckman Coulter, Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Sep 5, 2025 and May 14, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

4 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 2, 2026Open, Classified

    Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

    Z-0083-2027 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  2. Jul 15, 2026Class IIOpen, Classified

    a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

    Z-2973-2026 · Root cause: Software design (manufacturing process) · names 2 of these models · Recall details · FDA record ↗

  3. Sep 19, 2025Class IIOpen, Classified

    Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

    Z-0215-2026 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  4. Aug 4, 2025Class IIOpen, Classified

    On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

    Z-0032-2026 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗

Most reported DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module problems

Device problems

  • High Test Results2
  • Adverse Event Without Identified Device or Use Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1

DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent DxC 500i Clinical Analyzer, DxC 500i AU Chemistry Module adverse event reports

The 5 most recent reports across all models, in the reporter's own words as published by the FDA.