Skip to content

CADD 21-2120-0105-00Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of CADD (123)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This CADD infusion pump from ICU Medical has 95 reports, all describing malfunctions, with no deaths and no injuries. The reports mostly describe the pump failing to deliver fluid, sensing and alarm problems, and air in the line, and several recent ones mention events during testing. Two recalls are open: one over possible overheating damage in some CADD-Solis pumps, and the other linked to a software problem.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
97
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 97 adverse event reports that list this exact device (DI 15019517149999), received between Aug 9, 2022 and Sep 21, 2026 — classified as 97 malfunctions.

The most frequently reported specific device problem is “Failure to Infuse” (26 reports).

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K111275); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 4 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 8 reportsNov 2024: 6 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 6 reportsJul 2025: 5 reportsAug 2025: 9 reportsSep 2025: 3 reportsOct 2025: 4 reportsNov 2025: 9 reportsDec 2025: 1 reportsJan 2026: 7 reportsFeb 2026: 4 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2025Class IOpen, Classified

    There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.

    Z-1682-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 27, 2024Class IOpen, Classified

    The CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.

    Z-2421-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Infuse26
  • Device Sensing Problem15
  • Failure to Sense13
  • Device Alarm System11
  • Degraded9
  • Obstruction of Flow7
  • Air/Gas in Device6
  • Detachment of Device or Device Component5

Patient problems

  • No Clinical Signs, Symptoms or Conditions95
  • Abdominal Pain1
  • Fatigue1
  • Sleep Dysfunction1

Recent adverse event reports

The 25 most recent of 97 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Insufficient Flow or Under Infusion; Failure to Infuse

    It was reported that the pump alarmed with a ¿cassette not attached¿ error during infusion, resulting in an extended treatment time and under-delivery. The programmed rate was 2.2 mL/hr, with 92 mL remaining. There was a patient involvement and there were no additional adverse consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50842605 ↗
  • MalfunctionSep 10, 2026Failure to Sense

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722851 ↗
  • MalfunctionAug 17, 2026Excess Flow or Over-Infusion

    It was reported that the device may have over-delivered Adrucil. The patient expected a 2-day infusion; however, the device was labeled and programmed for a 24-hour infusion at 4.6 mL/hr. The pump was found empty after approximately 1 day, with 0 mL remaining. The patient was advised to contact the physician immediately due to the discrepancy in infusion duration. There was a patient involvement and there were no additional adverse consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50288070 ↗
  • MalfunctionJul 27, 2026Device Sensing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50024795 ↗
  • MalfunctionJul 27, 2026Device Sensing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50022909 ↗
  • MalfunctionJul 13, 2026Failure to Infuse

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25804236 ↗
  • MalfunctionMay 11, 2026Air/Gas in Device

    IT WAS REPORTED THAT AIR WAS PRESENT IN THE LINE. THE MEDICATION WAS INFUSED AT A PROGRAMMED RATE OF 5.4 ML PER HOUR, WITH A TOTAL PROGRAMMED VOLUME OF 250 ML. THE REMAINING RESERVOIR VOLUME WAS 1.2 ML, AND THE VOLUME DELIVERED WAS 248.85 ML. THE AIR DETECTOR WAS ENABLED (LOW SENSITIVITY), AND THE UPSTREAM SENSOR WAS ALSO ENABLED. IT WAS NOTED THAT THE PUMP SCREEN STILL INDICATED 1.2 ML REMAINING. UPON OBSERVING THE AIR IN THE LINE, THE PUMP WAS IMMEDIATELY STOPPED. THE EVENT OCCURRED WHILE THE DEVICE WAS IN USE WITH A PATIENT AND THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136895 ↗
  • MalfunctionMar 23, 2026Excess Flow or Over-Infusion

    IT WAS REPORTED THAT THE PUMP EXHIBITED AN OVER INFUSION DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24666391 ↗
  • MalfunctionFeb 23, 2026Device Alarm System; Air/Gas in Device

    IT WAS REPORTED THAT DURING TESTING THE PUMP EXPERIENCED AIR IN LINE ERROR. THERE WERE NO PATIENT INVOLVEMENT AND PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24423138 ↗
  • MalfunctionFeb 17, 2026Failure to Sense

    IT WAS REPORTED THAT PUMP CASSETTE WAS LOCKED AND THERE WAS A LATCHED ERROR. THE EVENT HAD OCCURRED DURING TESTING. EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORTED ISSUE AND IDENTIFIED A FAULTY LATCH/LOCK FLEX SENSOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24369468 ↗
  • MalfunctionFeb 17, 2026Failure to Deliver

    IT WAS REPORTED THAT ALARM 46876 OCCURRED. UPON POWERING THE DEVICE BACK ON, A ¿REMOVE AND REATTACH CASSETTE¿ ALARM WAS DISPLAYED, FOLLOWED BY A ¿PATIENT DATA LOST¿ ALARM. TROUBLESHOOTING WAS ATTEMPTED, HOWEVER; THE PATIENT DATA REMAINED LOST. THE EVENT OCCURRED AT THE PATIENT¿S HOME, AND THE DEVICE WAS HANDLED BY A HEALTHCARE PROFESSIONAL. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24368557 ↗
  • MalfunctionFeb 12, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PUMP EXHIBITED A CASSETTE NOT ATTACHED ERROR DURING TESTING. DURING PHYSICAL INSPECTION, IT WAS FOUND THAT THERE WAS A DETACHED DOWNSTREAM OCCLUSION SENSOR. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24327165 ↗
  • MalfunctionJan 28, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE DEVICE'S DOWNSTREAM OCCLUSION SEAL WAS RIPPED AND BUBBLED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24203320 ↗
  • MalfunctionJan 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE DOWNSTREAM OCCLUSION SEAL RIPPED AND IS BUBBLED. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24171750 ↗
  • MalfunctionJan 23, 2026Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE DOWNSTREAM OCCLUSION (DSO) SEAL WAS RIPPED AND BUBBLED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24171635 ↗
  • MalfunctionJan 23, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT THE DOWNSTREAM OCCLUSION SEAL IS RIPPED AND BUBBLED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24171473 ↗
  • MalfunctionJan 21, 2026Failure to Infuse

    IT WAS REPORTED THAT THE PUMP HAD DISPLAYED AN ERROR MESSAGE DOWNSTREAM OCCLUSION. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24136424 ↗
  • MalfunctionJan 21, 2026Failure to Infuse

    IT WAS REPORTED THAT THE PUMP HAD ALARMED FOR "CHECK FOR EMPTY TUBING OR RESERVOIR" ERROR MESSAGE. PER REPORTER, THERE WAS PATIENT INVOLVEMENT AND NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24136066 ↗
  • MalfunctionJan 6, 2026Failure to Sense

    IT WAS REPORTED THAT THE PUMP EXHIBITED A DOWNSTREAM OCCLUSION FAILURE. DURING PHYSICAL INSPECTION, IT WAS FOUND THAT THERE WAS A DOWNSTREAM OCCLUSION SENSOR ISSUE. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23981114 ↗
  • MalfunctionDec 4, 2025Obstruction of Flow; Obstruction of Flow

    IT WAS REPORTED THAT THE PUMP STATED DOWNSTREAM OCCLUSION, BUT THERE WAS NO OCCLUSION. NO PATIENT HARM OR NO PATIENT INJURY. THE EVENT OCCURRED WHILE IN USE WITH PATIENT AT PATIENT'S HOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23717482 ↗
  • MalfunctionNov 19, 2025Device Sensing Problem; Device Sensing Problem

    THE DEVICE WAS RETURNED DUE TO A CASSETTE SENSOR ERROR. THE RESULTS OF THE INVESTIGATION FOUND THAT THE DEVICE'S DOWNSTREAM OCCLUSION (DSO) SENSOR WAS MALFUNCTIONING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23593583 ↗
  • MalfunctionNov 17, 2025Degraded; Failure to Sense; Degraded; Failure to Sense

    THE DEVICE WAS RETURNED DUE TO THE FACT THAT IT WAS REPORTED THAT THE DEVICE HAD A DOWNSTREAM OCCLUSION FAILED CALIBRATION. THE RESULTS OF THE INVESTIGATION FOUND THAT THE DEVICE'S DOWNSTREAM OCCLUSION (DSO) SENSOR FAILED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23573338 ↗
  • MalfunctionNov 17, 2025Failure to Sense; Failure to Sense

    THE DEVICE WAS RETURNED DUE TO THE CASSETTE SENSOR REPORTEDLY BEING DEFECTIVE. THE RESULTS OF THE INVESTIGATION FOUND AND CONFIRMED THAT THE DEVICE HAD A BAD DOWNSTREAM OCCLUSION (DSO) SENSOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23572909 ↗
  • MalfunctionNov 17, 2025Crack; Crack

    THE DEVICE WAS RETURNED DUE TO THE DEVICE'S DOWNSTREAM OCCLUSION (DSO) WAS NOT CALIBRATING. THE RESULTS OF THE INVESTIGATION FOUND THAT THE DSO SEAL WAS CRACKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23572350 ↗
  • MalfunctionNov 17, 2025Loss of Power; Battery Problem; Loss of Power; Battery Problem

    THE DEVICE WAS RETURNED DUE TO A REPORT THAT THE PUMP HAD READ AS LOW BATTERY AND HAD THEN SHUT OFF. THE INVESTIGATION IDENTIFIED THAT THERE WAS A LOSS OF POWER DUE TO DEPLETED BATTERY. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM/ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23572301 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111275. One submission often covers many models, so check the product and lot details in each record.

  • Z-1668-2025 · Apr 10, 2025 · Open, Classified
    CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0
  • Z-0222-2025 · Jul 16, 2024 · Open, Classified
    CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the following CADD Solis Ambulatory Infusion Systems: 21-2101-0100-50, 21-2101-
  • Z-0466-2025 · Jun 26, 2024 · Open, Classified
    smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps
  • Z-0107-2022 · Mar 24, 2020 · Terminated
    CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 2

Other Pump, infusion devices

All Pump, infusion devices →

Sources and updates