CADD recallZ-1682-2025 · Smiths Medical ASD, Inc. · initiated Apr 10, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at [email protected]. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit
Recall details
- Recall number
- Z-1682-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 10, 2025
- Posted by the FDA
- May 5, 2025
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 313,676 items
- Product code
- FRN
- FDA submissions
- K111275
- Distribution
- Worldwide - US Nationwide distribution.
Product description
CADD Solis VIP Ambulatory Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE, 1/EA, Model Number: 21-2127-0105-01; 2) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED, 1/EA, Model Number: 21-2120-0105-01; 3) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH 1/EA, Model Number: 21-2120-0103-01; 4) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH 1/EA, Model Number: 21-2120-0102-51; 5) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-01; 6) PUMP KIT, CADD-SOLIS VIP, MDL 2120, CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-51; 7) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, MANUAL MODE 1/EA, Model Number: 21-2127-0104-01; 8) PUMP KIT, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER, NA ENGLISH, PHARMGUARD, 1/EA, Model Number: 21-2120-0105-00; 9) PUMP, CADD-SOLIS VIP, MDL 2120, TRIAL/LOANER CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-78; 10) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, TRIAL/LOANER 1/EA, Model Number: 21-2120-0102-78; 11) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU, NA ENGLISH, PHARMGUARD ENABLED 1/EA, Model Number: 21-2120-0104-231; 12) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249; 13) PUMP, CADD-SOLIS VIP, MDL 2120, NFHU CE GLOBAL ENGLISH 1/EA, Model Number: 21-2120-0103-249; 14) PUMP KIT, CADD-SOLIS VIP, MDL 2120, NA ENGLISH, FACTORY LIBRARY 1/EA, Model Number: 21-2125-0104-01; 15) PUMP, CADD-SOLIS VIP, MDL 2120, ENGLISH, NFHU 1/EA, Model Number: 21-2120-0102-249.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All serial numbers. 1 ) 21-2127-0105-01 , UDI-DI: 15019517150292 ; 2 ) 21-2120-0105-01 , UDI-DI: 15019517150001 ; 3 ) 21-2120-0103-01 , UDI-DI: 15019517084368 ; 4 ) 21-2120-0102-51 , UDI-DI: 10610586042829 ; 5 ) 21-2120-0104-01 , UDI-DI: 15019517126587 ; 6 ) 21-2120-0103-51 , UDI-DI: 15019517096378 ; 7 ) 21-2127-0104-01 , UDI-DI: 15019517126600 ; 8 ) 21-2120-0105-00 , UDI-DI: 15019517149999 ; 9 ) 21-2120-0103-78 , UDI-DI: 15019517096392 ; 10 ) 21-2120-0102-78 , UDI-DI: 10610586042836 ; 11 ) 21-2120-0104-231 , UDI-DI: 15019517126617 ; 12 ) 21-2120-0102-249 , UDI-DI: 10610586042843 ; 13 ) 21-2120-0103-249 , UDI-DI: 15019517096385 ; 14 ) 21-2125-0104-01 , UDI-DI: 15019517126594 ; 15 ) 21-2120-0102-249 , UDI-DI: 10610586042843 ;
Devices named in this recall
13 devices in the FDA device registry whose identifier appears in the recall's code information.
- CADD 21-2120-0105-01Icu Medical, Inc. · Pump, infusion4 recalls3 deaths5,914 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion2 recalls6 deaths4,097 reports
- CADD 21-2120-0102-51Icu Medical, Inc. · Pump, infusion4 recalls2 deaths3,555 reports
- CADD 21-2127-0105-01Icu Medical, Inc. · Pump, infusion2 recalls1,466 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion2 recalls3 deaths1,256 reports
- CADD 21-2120-0103-51Icu Medical, Inc. · Pump, infusion2 recalls1 death1,107 reports
- CADD 21-2120-0103-01Icu Medical, Inc. · Pump, infusion4 recalls585 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls2 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls0 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion4 recalls0 reports
- CADD 21-2120-0105-00Icu Medical, Inc. · Pump, infusion2 recalls97 reports
- CADD 2120Icu Medical, Inc. · Pump, infusion2 recalls4 reports
- CADD 21-2120-0102-78Icu Medical, Inc. · Pump, infusion1 recall33 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K111275 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CADD 21-2120-0102-02Icu Medical, Inc. · Pump, infusion3 recalls381 reports
- CADD 21-2111-0400-51Icu Medical, Inc. · Pump, infusion4 recalls124 reports
- CADD 21-2120-0105-231Icu Medical, Inc. · Pump, infusion1 recall4 reports
- CADD 21-0270-25Icu Medical, Inc. · Pump, infusion40 reports
- CADD 21-2125-0105-01Icu Medical, Inc. · Pump, infusion6 reports
- CADD 21-2120-0103-00Icu Medical, Inc. · Pump, infusion4 reports
- CADD 21-2120-0102-92Icu Medical, Inc. · Pump, infusion2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96624).
- Z-1681-2025 · Open, ClassifiedCADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page