ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002Recalls & reported problems
Applied Medical Resources Corporation · Kit Contains: B1005, ACUCISE® ENDOPYELOTOMY SYSTEM B3301, STENT, C-FLEX, 7/10F X 26CM B7016, FORTE AxP SET 12/14/18F 35CM 6/BX · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20607915128122), received between Feb 1, 2016 and Feb 1, 2016 — classified as 1 malfunction.
The most frequently reported specific device problem is “Inflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Inflation Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 1, 2016Inflation Problem; Inflation Problem
ACUCSIE ENDOPYELOTOMY - "DR. (B)(6) PERFORMED THE PROCEDURE. THE BALLOON DID NOT STAY INFLATED FOLLOWING ACTIVATION OF THE CUTTING WIRE. WIRE WAS ACTIVATED FOR 5 SECONDS AT 75W PURE CUT. DEVICE WAS OBSERVED TO HOLD INFLATION FOR CUTTING STEP OF PROCEDURE, BUT WHEN FLUOROSCOPY WAS PERFORMED MINUTES AFTER, NO CONTRAST SOLUTION REMAINED IN THE BALLOON. UPON REMOVAL OF DEVICE, DAMAGE TO THE BALLOON WAS OBSERVED. AFTER REMOVAL OF THE DEVICE, A 2CC SYRINGE OF STERILE WATER WAS INJECTED INTO THE BALLOON INFLATION PORT AND LEAKAGE WAS CONFIRMED. THIS B1005 WAS PART OF THE BK002 KIT LOT 1248705." PATIENT STATUS - "NA."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5402510 ↗
FDA clearance and classification
The FDA regulates product code KNY (accessories, catheter, g-u) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002
Has the ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002 been recalled?
No FDA recall record in our data names this exact device identifier (DI 20607915128122), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 1, 2016. That report lists device problem “Inflation Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A BK002, and is it still sold?
The FDA's device database lists Applied Medical Resources Corporation as the company responsible for it (brand ACUCISE® PROCEDURAL PACK UNIVERSAL PACK A, model BK002). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Accessories, Catheter, G-U devices
- UroPass 61238BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall10 reports
- UroPass 61138BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61024BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall5 reports
- UroPass 61224BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61146BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61254BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61046BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
- UroPass 61054BXGyrus ACMI, LLC · ACCESSORIES, CATHETER, G-U1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page