Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 20653405046162), received between Dec 17, 2015 and Jun 7, 2016 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated5
  • Component Falling3
  • Detachment Of Device Component1
  • Gas/Air Leak1

Patient problems

  • No Known Impact Or Consequence To Patient4
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 7, 2016Component Falling; Device Dislodged or Dislocated

    N/A

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5703753 ↗
  • MalfunctionJun 7, 2016Detachment Of Device Component

    THE END-USER FACILITY REPORTED THAT DURING USE OF A CD775 CORE ENTRɅ II 5-12MM VALVE/REDUCER IN A LAPAROSCOPIC HERNIA REPAIR, THE VALVE BECOME DISLODGED AND FELL INTO THE PATIENT'S PERITONEAL CAVITY. THE VALVE WAS IMMEDIATELY RETRIEVED AND REMOVED FROM THE PATIENT'S PERITONEAL CAVITY. THE SURGEON REPLACED THE DEVICE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. AS REPORTED, THE PROCEDURE WAS COMPLETED WITH NO PATIENT INJURY AS A RESULT OF THE REPORTED EVENT. TO DATE, THERE HAS BEEN NO ADDITIONAL INFORMATION RECEIVED REGARDING THE PATIENT'S LATEST CONDITION OR ANY INDICATION THAT ANY LONG TERM ADVERSE EFFECT HAS OCCURRED.. THIS HAPPENED IN THREE LAPAROSCOPIC HERNIA REPAIR PROCEDURES ON THREE SEPARATE DAYS (DATES OF ACTUAL EVENTS UNKNOWN). THE DEVICES WERE ALL FROM THE SAME LOT. PER THE REPORTER THE PATIENT DEMOGRAPHICS (AGE, GENDER, WEIGHT) FOR EACH CASE IS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5703714 ↗
  • MalfunctionJun 7, 2016Component Falling; Device Dislodged or Dislocated

    N/A

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5703631 ↗
  • MalfunctionMay 10, 2016Component Falling; Device Dislodged or Dislocated

    THE END-USER FACILITY REPORTED THAT DURING USE OF A CD775 CORE ENTREE 5-12MM VALVE/REDUCER IN A LAPAROSCOPIC PROCEDURE, THE INNER PLASTIC VALVE BECOME DISLODGED AND FELL INTO THE PATIENT'S PERITONEAL CAVITY. THE VALVE WAS IMMEDIATELY RETRIEVED AND REMOVED FROM THE PATIENT'S PERITONEAL CAVITY. THE SURGEON REPLACED THE DEVICE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. AS REPORTED, THE PROCEDURE WAS COMPLETED WITH NO PATIENT INJURY AS A RESULT OF THE REPORTED EVENT. TO DATE, THERE HAS BEEN NO ADDITIONAL INFORMATION RECEIVED REGARDING THE PATIENT'S LATEST CONDITION OR ANY INDICATION THAT ANY LONG TERM ADVERSE EFFECT HAS OCCURRED..

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5644559 ↗
  • MalfunctionMay 3, 2016Device Dislodged or Dislocated

    THE END-USER FACILITY REPORTED THAT DURING USE OF A CD775 CORE ENTRÉE II 5-12MM VALVE/REDUCER IN A LAPAROSCOPIC PROCEDURE, THE INNER PLASTIC VALVE BECOME DISLODGED AND FELL INTO THE PATIENT'S PERITONEAL CAVITY. THE VALVE WAS IMMEDIATELY RETRIEVED AND REMOVED FROM THE PATIENT'S PERITONEAL CAVITY. THE SURGEON REPLACED THE DEVICE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. AS REPORTED, THE PROCEDURE WAS OTHERWISE COMPLETED WITH NO FURTHER COMPLICATIONS OR PATIENT INJURY AS A RESULT OF THE REPORTED EVENT. PATIENT DEMOGRAPHICS REGARDING THIS INCIDENT (AGE, GENDER, WEIGHT) WERE UNAVAILABLE FROM THE END-USER FACILITY. IT WAS ALSO REPORTED THAT THIS SAME INCIDENT OCCURRED AGAIN IN A LAPAROSCOPIC PROCEDURE BY THE SAME SURGEON PROVIDER A WEEK SUBSEQUENT TO THIS REPORTED EVENT (SEE ASSOCIATED MEDWATCH 1320894-2016-00035).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5627976 ↗
  • MalfunctionMay 3, 2016Device Dislodged or Dislocated

    THE END-USER FACILITY REPORTED THAT DURING USE OF A CD775 CORE ENTRɅ II 5-12MM VALVE/REDUCER IN A LAPAROSCOPIC PROCEDURE, THE INNER PLASTIC VALVE BECOME DISLODGED AND FELL INTO THE PATIENT'S PERITONEAL CAVITY. THE VALVE WAS IMMEDIATELY RETRIEVED AND REMOVED FROM THE PATIENT'S PERITONEAL CAVITY. THE SURGEON REPLACED THE DEVICE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. AS REPORTED, THE PROCEDURE WAS OTHERWISE COMPLETED WITH NO FURTHER COMPLICATIONS OR PATIENT INJURY AS A RESULT OF THE REPORTED EVENT. PATIENT DEMOGRAPHICS REGARDING THIS INCIDENT (AGE, GENDER, WEIGHT) WERE UNAVAILABLE FROM THE END-USER FACILITY. IT WAS ALSO REPORTED THAT THIS SAME INCIDENT OCCURRED EARLIER IN A LAPAROSCOPIC PROCEDURE BY THE SAME SURGEON PROVIDER A WEEK PRIOR TO THIS (SEE ASSOCIATED MEDWATCH 1320894-2016-00036).

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5627789 ↗
  • MalfunctionDec 17, 2015Gas/Air Leak

    THE END-USER FACILITY REPORTED THAT DURING USE OF A CD775 CORE ENTRɅ II 5-12MM VALVE/REDUCER IN THE SURGERY, WHEN INSUFFLATION GAS LEAKED WAS NOTED. SUBSEQUENTLY, THE DEVICE WAS EXAMINED AND COMPLETE DETACHMENT OF THE MAIN BODY SEAL WAS DETECTED. THERE WAS NO REPORT OF SURGICAL COMPLICATION AND NO REPORT OF THE DETACHED SEAL FELL INTO THE SURGICAL SITE RECEIVED FROM THE USER FACILITY. AS REPORTED, THE PROCEDURE WAS OTHERWISE COMPLETED WITH NO FURTHER COMPLICATIONS OR PATIENT INJURY. DESPITE MULTIPLE FOLLOW-UP REQUESTS, NO ADDITIONAL INFORMATION RECEIVED IN REGARDS TO THE PATIENT'S DEMOGRAPHIC INFORMATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5308294 ↗

FDA clearance and classification

Cleared through 510(k) K950457 on May 5, 1995 (decision: Substantially Equivalent). Submitted by Core Dynamics, Inc. as “Cannual And Trocar, Superpubic”. The FDA review took 91 days. FDA record ↗

The FDA regulates product code HET (laparoscope, gynecologic (and accessories)) under 21 CFR 884.1720 as a Class II device. 21 CFR 884.1720 ↗

Common questions about the CORE DYNAMICS CD775

Has the CORE DYNAMICS CD775 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20653405046162), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CORE DYNAMICS CD775?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Dec 17, 2015 and Jun 7, 2016. In those reports, the most common specific device problems are “Device Dislodged or Dislocated” (5), “Component Falling” (3) and “Detachment Of Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CORE DYNAMICS CD775?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CORE DYNAMICS CD775 FDA approved?

It was cleared through 510(k) K950457 on May 5, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CORE DYNAMICS CD775, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand CORE DYNAMICS, model CD775). Its record shows it as in commercial distribution.