Skip to content

BD BBL™ Sensi-Disc™ Erythromycin 15 µg 231290Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Sensi-Disc™ Erythromycin 15 µg · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30382902312902), received between Feb 18, 2025 and Feb 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BBL¿ SENSI-DISC¿ ERYTHROMYCIN 15G A PATIENT ISOLATE TESTED SENSITIVE, SHOWING A CLEAR NO-GROWTH ZONE AROUND THE DISC FOR THE DRUG ERYTHROMYCIN, BUT WHEN COMPARED TO THE INSTRUMENTATION RESULT IT WAS RESISTANT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21395932 ↗
  • MalfunctionFeb 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING THE BD BBL¿ SENSI-DISC¿ ERYTHROMYCIN 15G A PATIENT ISOLATE TESTED SENSITIVE, SHOWING A CLEAR NO-GROWTH ZONE AROUND THE DISC FOR THE DRUG ERYTHROMYCIN, BUT WHEN COMPARED TO THE INSTRUMENTATION RESULT IT WAS RESISTANT. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21395322 ↗

Other SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices

All SUSCEPTIBILITY TEST DISCS, ANTIMICROBIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page