ARROW recallZ-2385-2026 · Arrow International, LLC · initiated Apr 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
Recall details
- Recall number
- Z-2385-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jun 13, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 426,041
- Product code
- DYB
- FDA submissions
- K002507, K011761, K780532, K940079
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Product description
Description/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH; PSI KIT: 9 FR X 10 CM/ASK-09903-PHP1; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-UPA; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-YNH; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09907-SL; PSI KIT: 8.5 FR X 4-1/8IN (10 CM)/NA-09803-CDC; PSI KIT: 9 FR X 4-1/8IN (10 CM)/NA-09903-CDC; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09903-ECL4; PSI KIT: 9 FR X 4 1/8 IN (10 CM)/ASK-09903-MGH3; MULTI-LUMEN/PSI KIT/AK-11142-SPCS; MAC KIT: 2-L 9 FR DISTAL X 11.5 CM/CDC-11242-1A; MULTI-LUMEN/PSI KIT/AK-11142-D; MULTILUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-11142-MGH4; MULTI-LUMEN/PSI KIT/NA-11242-CDC; MULTI-LUMEN/PSI KIT/AK-21142-SK; MAC KIT: 2L 9 FR DISTAL X 10 CM AGB/ASK-21142-LMDTG; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-RH; MULTILUMEN/PSI KIT 9 FR X 11.5 CM/ASK-21242-IMC; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/ASK-21242-LMDTG; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMD; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UU3; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-VCU; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-1A; MAC KIT: 2L 9 FR DISTAL X 11.5 CM AGB/CDC-21242-X1A; MULTI-LUMEN/PSI KIT/AK-21242-CDC; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-FH4; MULTI LUMEN/PSI KIT: 9 FR X 4 IN (10 CM)/ASK-21142-GH1; 2L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-SUU; 2-L MAC KIT: 9 FR DISTAL X 4 IN (10 CM)/ASK-21142-UPMC4; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UKL; MAC KIT: 2L 9 FR DISTAL X 11.5 CM/ASK-21242-UMF; MULTILUMEN/PSI KIT: 9 FR X 4 1/2 IN/ASK-21242-UR1; PSI KIT: 8.5 FR/AK-29803-CDC; PSI KIT: 9 FR/AK-29903-CDC; PSI KIT: 9 FR/AK-29907-CDC; PSI KIT: 8.5 FR X 4IN (10 CM)/ASK-09804-UN1; PSI KIT: 9 FR/ASK-09903-UM; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDT; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/ASK-29803-LMDTG; PSI KIT: 9 FR X 4 1/8 (10 CM)/ASK-29903-BID1; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-LMDTG; PSI KIT: 9 FR X 10 CM ANTIMICROBIAL/ASK-29903-UMD; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-29903-UR1; PSI KIT: 8.5 FR X 10 CM ANTIMICROBIAL/CDC-29803-X1A; PSI KIT: 8.5 FR X 10 CM/ASK-29803-ECL; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-29803-UPM3; PSI KIT: 9 FR X 10 CM/ASK-29903-ECL; PSI KIT: 9FR X 4IN (10 CM) ANTIMICROBIAL/ASK-29903-UPM3; PSI KIT: 9 FR X 4 IN (10 CM)/ASK-09907-TG1
Code information (models, lots, serial numbers)
Show the full code information (8k characters)
REF(Material)/UDI-DI/Lot(Batch): GU-04020/30801902118807, 10801902193275, 30801902118807, 30801902118807/33F24C0753, 33F24E0736, 33F24G0154, 33F24G0648, 33F24H0377, 33F25A0059, 33F25A0072, 33F25A0395, 33F25F0260, 33F25G0736, 33F25H0298, 33F25J0603, 33F26B0084, 33F26B0314, 33F26C0578, 33F24D0036; AK-09803-CDC/20801902121756, 10801902193275, 10801902117417/33F24B0126, 33F24B0127, 33F24C0021, 33F24D0021, 33F24E0072, 33F24G0338, 33F24G0587, 33F24J0506, 33F24K0105, 33F24L0722, 33F25A0468, 33F25C0107, 33F25C0953, 33F25E0054, 33F25E0921, 33F25G0086, 33F25H0023, 33F25J0828, 33F25K0255, 33F25K0364, 33F25L0328, 33F24A0239; AK-09903-CDC/10801902117417, 10801902193275, 20801902121916/33F24A0356, 33F24B0128, 33F24B0129,33F24C0022, 33F24D0022, 33F24E0074, 33F24E0075, 33F24G0340, 33F24G0588, 33F24J0507, 33F24K0107, 33F24L0676, 33F24L0759, 33F25A0470, 33F25C0108, 33F25C0730, 33F25E0056, 33F25E0923, 33F25H0025, 33F25J0471, 33F25K0031, 33F25L0237, 33F25M0920, 33F26A1053, 33F24A0386; ASK-09810-PCMH1/10801902193275/33F25K0051, 33F24J0109; ASK-09903-KH/10801902119350, 10801902193275/33F24D0713, 33F24K0385, 33F24L1060, 33F25D0340, 33F25E0778, 33F25H0567, 33F25M0163, 33F26A1407, 33F25E1184; ASK-09903-PHP1/10801902193275, 10801902115222/33F24J0115, 33F25H0497, 33F26A0410, 33F24F0625; ASK-09903-UPA/10801902193275, 10801902118193/33F24K0387, 33F24L0855, 33F24A0787; ASK-09903-YNH/10801902118193, 10801902193275, 10801902118209/33F24E0527, 33F24H0478, 33F24L0682, 33F25H0135, 33F25J0915, 33F24A0110; ASK-09907-SL/10801902118209, 10801902193275/33F24D0053, 33F24D0967, 33F24G0265, 33F25E0215, 33F25E0454, 33F25G0680, 33F25J0551, 33F25K0328, 33F25L0206, 33F26C0312, 33F25B0188; NA-09803-CDC/10801902119930, 10801902193275, 10801902119947/33F24C0756, 33F24E0739, 33F25A0342, 33F25J0615, 33F26B0028, 33F24A1327; NA-09903-CDC/10801902119947, 10801902193275, 10801902119954/33F24E0740, 33F24H0381, 33F24J0199, 33F24J0361, 33F24L0020, 33F25B0278, 33F25C0551, 33F25E0340, 33F25E0400, 33F25E0468, 33F25F0419, 33F25H0708, 33F26A1655, 33F26B0023, 33F24A0113; ASK-09903-ECL4/10801902193275, 10801902224344/33F26A1582, 33F26B0399, 33F25E0534; ASK-09903-MGH3/10801902193275, 10801902189841, 10801902212006/33F25C0264, 33F25G0338, 33F25J0360, 33F25K0130, 33F26B0126, 33F26C0004, 33F24D0780; AK-11142-SPCS/10801902196412, 10801902193275, 60801902123093/33F24A0382, 33F24A1019, 33F24C0024, 33F24E0078, 33F24E0154, 33F24G0343, 33F24J0510, 33F24K0110, 33F24L0673, 33F25A0472, 33F25C0381, 33F25C0389, 33F25C0952, 33F25E0059, 33F25H0027, 33F25J0831, 33F25K0302, 33F25L0365, 33F26B0514, 33S24J0510, 33F24D0075; CDC-11242-1A/60801902116736, 10801902193275, 40801902111768/33F24D0431, 33F24E0213, 33F24F0140, 33F24F0220, 33F24G0103, 33F24H0392, 33F24H0425, 33F24H0450, 33F24L0011, 33F25B0018, 33F25B0096, 33F25D0705, 33F25E1120, 33F25F0024, 33F25F0414, 33F25G0043, 33F25G0486, 33F25J0016, 33F25J0992, 33F25K0308, 33F25M0481, 33F24C1001; AK-11142-D/20801902121916, 10801902193275, 10801902196412/33F24A1018, 33F24C0023,33F24E0077, 33F24G0342,33F24J0509, 33F24K0109, 33F24L0615, 33F25A0471, 33F25C0842, 33F25E0058, 33F25G0088, 33F25J0472, 33F25K0301, 33F25L0467, 33F26B0593, 33F26B0958, 33F24A0252; ASK-11142-MGH4/10801902193275, 10801902189865, 10801902182545/33F25D0466, 33F25F0113, 33F25L0130, 33F25M0398, 33F26A1539, 33F26B0680, 33F24D0736; NA-11242-CDC/10801902119954, 10801902193275, 10801902119978/33F24A0817, 33F24A1218, 33F24B0399, 33F24C0270, 33F24G0554, 33F24J0663, 33F24L0743, 33F25C0197, 33F25C0870, 33F25E0273, 33F25H0179, 33F25H0761, 33F25L0419, 33F26B0562, 33F24C0271; AK-21142-SK/10801902193275, 30801902122248, 30801902122255/33F24G0356, 33F24G0589, 33F24J0520, 33F24K0122, 33F24L0670, 33F25A0480, 33F25C0720, 33F25E0071, 33F25H0030, 33F25J0837, 33F25L0473, 33F25M0964, 33F26A1097, 33F26B0956, 33R24L0670, 33F24D0023; ASK-21142-LMDTG/10801902193275/33F25K0417, 33F25F0007; ASK-21142-RH/10801902193275/10801902129953/33F24J0108, 33F24L0165, 33F24M0172, 33F25B0422, 33F25E1172, 33F25H0776, 33F25K0424, 33F26C0337, 33F25A0245; ASK-21242-IMC/10801902193275, 10801902118407/33F24G0500, 33F24J0631, 33F24K0255, 33F24L0609, 33F25A0572, 33F25E0223, 33F25H0142, 33F25J0921, 33F26B0543, 33F25C0696; ASK-21242-LMDTG/10801902193275, 10801902220827/33F25G0202, 33F25H0414, 33F25K0382, 33F25M0130, 33F26C0378, 33F24H0123; ASK-21242-UMD/10801902193275/33F24K0476, 33F24L0315, 33F25B0196, 33F25C0279, 33F25D0476, 33F25H0492, 33F25J0378, 33F25M0150, 33F26A1365, 33F25B0551; ASK-21242-UU3/10801902193275/33F24G0503, 33F24L0652, 33F25C0181, 33F25E0226, 33F25H0144, 33F25J0556, 33F24K0077; ASK-21242-VCU/10801902193275, 10801902112962/33F24K0694, 33F25G0171, 33F25H0495, 33F25J0379, 33F26C0494, 33F25C0439; CDC-21242-1A/50801902116821, 10801902193275/33F24D0486, 33F24D0981, 33F24E0227, 33F24F0175, 33F24G0104, 33F24H0296, 33F24H0305, 33F24H0331, 33F24H0357, 33F24H0393, 33F24H0427, 33F24M0019, 33F24M0055, 33F25A0131, 33F25A0367, 33F25B0019, 33F25C0511, 33F25D0706, 33F25E0417, 33F25E1122, 33F25F0416, 33F25G0066, 33F25G0476, 33F25G0711, 33F25H0852, 33F25J0964, 33F25K0352, 33F25L0229, 33F26A1614, 33F26A1624, 33F26A1649, 33F24H0332; CDC-21242-X1A/10801902193275, 30801902116834, 10801902196511/33F25J0291, 33F25J0672, 33F25J0678, 33F25J0690, 33F25J0700, 33F25J0708, 33F25J0724, 33F25L0067, 33F26A1616, 33F26A1625, 33F26B0642, 36F25B0014, 36F25E0020, 36F25G0029, 36F25H0016, 36F25J0005, 33F24C0518; AK-21242-CDC/30801902122255, 10801902193275, 30801902122262/33F24E0082, 33F24E0083, 33F24J0521, 33F24K0123, 33F24L0618, 33F25A0481, 33F25C0946, 33F25E0072, 33F25H0031, 33F25K0037, 33F25K0414, 33F25L0372, 33F26B0520, 33F26C0626, 33F24E0352; ASK-21142-FH4/10801902193275/33F24L0683; ASK-21142-GH1/10801902193275/33F25C0180, 33F25E1023, 33F25H0141, 33F25H0752, 33F25J0920, 33F25L0455, 33F25B0276; ASK-21142-SUU/10801902193275/33F25J0096, 33F25C0275; ASK-21142-UPMC4/10801902193275/33F25A0247; ASK-21242-UKL/10801902114065, 10801902193275, 10801902136487/33F24F0710, 33F24K0467, 33F24L0314, 33F25B0195, 33F25C0278, 33F25D0475, 33F25G0412, 33F25J0149, 33F25M0202, 33F26A1551, 33F26C0010, 33F26C0759, 33F24F0849; ASK-21242-UMF/10801902193275, 10801902129502/33F25D0477, 33F25H0487, 33F25J0150, 33F25K0096, 33F24D0715; ASK-21242-UR1/10801902193275, 10801902119435, 10801902118902/33F25E0225, 33F25J0555, 33F25L0427, 33F25M0928, 33F26C0261, 33F24E0542; AK-29803-CDC/10801902193275, 20801902117506/33F24G0378, 33F24J0283, 33F24J0533, 33F24K0141, 33F24L0620, 33F25A0488, 33F25C0901, 33F25E0090, 33F25G0095, 33F25H0039, 33F25J0487, 33F25L0389, 33F25M0893, 33F26A1021, 33F26B0537, 33F26C0262, 33F24G0379; AK-29903-CDC/10801902193275, 10801902117516, 10801902117523/33F24J0345, 33F24J0534, 33F24K0142,33F24L0672, 33F25A0489, 33F25C0904, 33F25E0091, 33F25G0096, 33F25H0040, 33F25J0488, 33F25L0382, 33F25M0889, 33F26B0533, 33F24D0091; AK-29907-CDC/10801902117523, 10801902193275/33F24E0377, 33F24J0535, 33F24L0606, 33F25C0122, 33F25C0854, 33F25E0092, 33F25H0041, 33F25J0848, 33F26B0503, 33F25A0146; ASK-09804-UN1/10801902193275, 10801902145847/33F25H0485, 33F26A0248, 33F24B0018; ASK-09903-UM/10801902193275, 10801902118186/33F25B0432, 33F25F0361, 33F25L0136, 33F24D0545; ASK-29803-LMDT/10801902193275, 10801902220810/33F25H0367, 33F25J0153, 33F25K0258, 33F25K0418, 33F26A1354, 33F26C0357, 33F25G0464; ASK-29803-LMDTG/10801902193275, 10801902220858/33F25H0413, 33F25H0585, 33F25K0207, 33F25K0419, 33F25M0198, 33F25M0440, 33F26C0170, 33F24G0607; ASK-29903-BID1/10801902193275/33F25K0066, 33F26A0251, 33F24H0139; ASK-29903-LMDTG/10801902193275, 10801902220735/33F25L0025, 33F25M0338, 33F26C0021, 33F24H0124; ASK-29903-UMD/10801902193275/33F25C0583, 33F26A0409, 33F24K0402; ASK-29903-UR1/10801902193275/33F25J0796, 33F24J0107; CDC-29803-X1A/10801902193275, 30801902117022, 10801902193114/33F25F0334, 33F25H0878, 33F25J0292, 33F25K0016, 33F25K0315, 33F25M0088, 33F26A1415, 33F26B0473, 36F25E0019, 36F25G0020, 36F25H0013, 33F24E0214; ASK-29803-ECL/10801902193275, 10801902200478/33F26A1576, 33F24D0625; ASK-29803-UPM3/10801902193275/33F25B0236; ASK-29903-ECL/10801902193275/33F26A1580, 33F25D0235; ASK-29903-UPM3/10801902193275/33F25J0558; ASK-09907-TG1/10801902118216/33F24D1025
Devices named in this recall
44 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN036821Teleflex Incorporated · INTRODUCER, CATHETER2 recalls8 reports
- ARROW IPN036830Teleflex Incorporated · INTRODUCER, CATHETER2 recalls3 reports
- ARROW IPN036824Teleflex Incorporated · INTRODUCER, CATHETER2 recalls2 reports
- ARROW IPN037022Teleflex Incorporated · INTRODUCER, CATHETER2 recalls1 report
- ARROW IPN036838Teleflex Incorporated · INTRODUCER, CATHETER2 recalls1 report
- ARROW IPN036446Teleflex Incorporated · INTRODUCER, CATHETER2 recalls1 report
- ARROW IPN037258Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN037025Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN037016Teleflex Incorporated · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls0 reports
- ARROW IPN037610Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN039798Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN056804Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN921749Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN926611Teleflex Incorporated · Introducer, catheter2 recalls0 reports
- ARROW IPN036186Teleflex Incorporated · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls0 reports
- ARROW IPN036826Teleflex Incorporated · INTRODUCER, CATHETER1 recall6 reports
- ARROW IPN036444Teleflex Incorporated · INTRODUCER, CATHETER1 recall5 reports
- ARROW IPN036335Teleflex Incorporated · INTRODUCER, CATHETER1 recall3 reports
- ARROW IPN036827Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall3 reports
- ARROW IPN036351Teleflex Incorporated · INTRODUCER, CATHETER1 recall2 reports
- ARROW IPN037011Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN036823Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN036682Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036863Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036858Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036931Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037030Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036956Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037162Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037232Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037164Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037139Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN038093Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN919669Teleflex Incorporated · Introducer, catheter1 recall0 reports
- ARROW IPN919671Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN929716Teleflex Incorporated · Introducer, catheter1 recall0 reports
- ARROW IPN929720Teleflex Incorporated · Introducer, catheter1 recall0 reports
- ARROW IPN929721Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN929737Teleflex Incorporated · Introducer, catheter1 recall0 reports
- ARROW IPNB00136Teleflex Incorporated · Introducer, catheter1 recall0 reports
- ARROW IPN036471Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036861Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN036449Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
- ARROW IPN036969Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall0 reports
Product lines
Other devices under the same FDA submission
93 devices cleared under K002507, K011761, K780532, K940079 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARROW IPN036452Teleflex Incorporated · INTRODUCER, CATHETER2 recalls3 reports
- ARROW IPN036448Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN029661Teleflex Incorporated · INTRODUCER, CATHETER2 recalls0 reports
- ARROW IPN927550Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN929229Teleflex Incorporated · Introducer, catheter2 recalls0 reports
- ARROW IPN036456Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN929678Teleflex Incorporated · Introducer, catheter1 recall3 reports
- ARROW IPN037890Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN038521Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN036450Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN036336Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
- ARROW IPN037637Teleflex Incorporated · INTRODUCER, CATHETER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98793).
- Z-2363-2026 · Open, ClassifiedPARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU
- Z-2364-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;…
- Z-2365-2026 · Open, ClassifiedDescription/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU
- Z-2366-2026 · Open, ClassifiedDescription/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM
- Z-2367-2026 · Open, ClassifiedDescription/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) V…
- Z-2368-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018…
- Z-2369-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 …
- Z-2370-2026 · Open, ClassifiedDescription/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT…
- Z-2371-2026 · Open, ClassifiedDescription/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION K…
- Z-2372-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A
- Z-2373-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYS…
- Z-2374-2026 · Open, ClassifiedDescription/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-1…
- Z-2376-2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-…
- Z-2377-2026 · Open, ClassifiedDescription/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE…
- Z-2378-2026 · Open, ClassifiedDescription/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR …
- Z-2379-2026 · Open, ClassifiedDescription/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JA…
- Z-2380-2026 · Open, ClassifiedDescription/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000
- Z-2381-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
- Z-2382-2026 · Open, ClassifiedDescription/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP,…
- Z-2383-2026 · Open, ClassifiedDescription/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU
- Z-2384-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUME…
- Z-2387-2026 · Open, ClassifiedDescription/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
- Z-2388-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 C…
- Z-2389-2026 · Open, ClassifiedDescription/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINA…
- Z-2390-2026 · Open, ClassifiedDescription/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B
- Z-2391-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHIL…
- Z-2393-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; …
- Z-2394-2026 · Open, ClassifiedDescription/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MI…
- Z-2396-2026 · Open, ClassifiedDescription/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
- Z-2397-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page