ARROW recallZ-2377-2026 · Arrow International, LLC · initiated Apr 21, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.

Recall details

Recall number
Z-2377-2026
Status
Open, Classified
Initiated by the company
Apr 21, 2026
Posted by the FDA
Jun 13, 2026
Recalling firm
Arrow International, LLC
Quantity
126,530
Product code
DQY
FDA submissions
K021120, K171146, K810962, K820009, K862056
Distribution
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Product description

Description/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSEL CATH KIT: 20GA X 5IN/AK-04550-S CVC KIT: 1-LUMEN 5 FR X 20 CM/AK-04301-SP CVC KIT: 16 GA X 6IN (16 CM)/AK-04306-SP CVC KIT: 16GA X 6 IN/AK-04800-S CVC KIT: 2-LUMEN 4 FR X 5 CM/AK-12402 CVC KIT: 3-LUMEN 7 FR X 16 CM/AK-12703-CDC CVC KIT: 2L 4 FR X 13 CM/AK-14402 CVC KIT: 2-LUMEN 5 FR X 13 CM/AK-14502 CVC KIT: 3-LUMEN 5.5 FR X 30 CM/AK-14553 CVC KIT: 3-LUMEN 7 FR X 30 CM/AK-14703-SP CVC KIT: 2-LUMEN 4 FR X 8 CM/AK-15402 CVC KIT: 3-LUMEN 5.5 FR X 8 CM/AK-15553 CVC KIT: 2-LUMEN 4 FR X 30 CM/AK-16402 CVC KIT: 3-LUMEN 5.5 FR X 13 CM/AK-16553 CVC KIT: 5 FR X 20 CM/ASK-04301-HH2 CVC KIT 3L: 7 FR X 6 IN (16 CM)/ASK-12703-MGH3

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

REF(Material)/UDI-DI/Lot(Batch): ASK-04510-MM6/10801902193275, 10801902221909/33F25E0590, 33F25G0196, 33F25M0005, 33F25M0149, 33F26A0302, 33F26C0723, 33F24K0212; NA-04020-X1ASP/10801902193275/33F25J0575, 33F25J0756, 33F26B0065, 33F26B0027; ASK-04001-MS/10801902193275/33F24K0706, 33F25F0375, 33F25F0005; ASK-00820-UM1/10801902193275/33F24G0150; AK-04550-S/10801902196719, 10801902193275, 60801902121709/33F24A0029, 33F24A0355, 33F24A1014, 33F24A1619, 33F24B0112, 33F24C0548, 33F24D0017, 33F24E0174, 33F24G0328, 33F24G0576, 33F24J0276, 33F24J0494, 33F24K0092, 33F24L0720, 33F25A0458, 33F25C0950 33F25E0041, 33F25E0911, 33F25G0080, 33F25H0013, 33F25J0820, 33F25L0402, 33F26B0550, 33F26C0267, 33F23M0175; AK-04301-SP/10801902121650, 10801902193275, 10801902121674/33F24A1010, 33F24C0093, 33F24D0068, 33F24E0325, 33F24G0325, 33F24J0274, 33F24K0090, 33F24L0809, 33F25C0757, 33F25E0038, 33F25G0760, 33F25H0721, 33F25J0462, 33F25L0239, 33F25M0888, 33F24A0146; AK-04306-SP/10801902121674, 10801902193275, 10801902192230/33F24A0846, 33F24B0174, 33F24C0094, 33F24E0327, 33F24E0329, 33F24J0275, 33F24J0492, 33F24L0717, 33F25B0249, 33F25C0783, 33F25E0039, 33F25J0463, 33F25L0054, 33F25L0345, 33F26B0684, 33F23J0175; AK-04800-S/60801902121709, 10801902193275, 10801902209501/33F24B0437, 33F24C0735, 33F24E0708, 33F24G0120, 33F24G0647, 33F24H0415, 33F24J0448, 33F24L0037, 33F24M0128, 33F25A0396, 33F25B0111, 33F25D0028, 33F25G0509, 33F25H0326, 33F25J0997, 33F26C0274, 33F24C0013; AK-12402/10801902168471, 10801902193275, 10801902168488/33F24E0342, 33F24F0325, 33F24G0347, 33F24J0512, 33F24K0113, 33F24L0750, 33F25C0109, 33F25C0964, 33F25E0063, 33F25E0926, 33F25J0474, 33F24A0408; AK-12703-CDC/10801902168488, 10801902193275, 10801902168495/33F24A1020, 33F24B0187, 33F24C0104, 33F24D0078, 33F24G0348, 33F24J0513, 33F24K0114, 33F24L0596, 33F25C0110, 33F25C0759, 33F25J0834, 33F25M0948, 33F24A0401; AK-14402/10801902168495, 10801902193275, 10801902168495, 10801902168518/33F24E0177, 33F24E0343, 33F24F0328, 33F24G0349, 33F24J0514, 33F24K0115, 33F25B0369, 33F25C0875, 33F25E0064, 33F25E0928, 33F25H0029, 33F25J0475, 33F26B0564, 33F24A0813; AK-14502/10801902168518, 10801902193275, 30801902122040/33F24E0344, 33F24G0350, 33F24K0116, 33F24L0656, 33F25C0878, 33F25E0065, 33F25G0089, 33F25J0476, 33F26B0568, 33F24A0849; AK-14553/30801902122040, 10801902193275, 10801902122077/33F24B0190, 33F24C0107, 33F24E0345, 33F24E0346, 33F24F0332, 33F24G0351, 33F24K0117, 33F24L0689, 33F25C0112, 33F25C0848, 33F25G0090, 33F25J0835, 33F24B0191; AK-14703-SP/10801902122077, 10801902193275, 10801902197136/33F24E0347, 33F24G0352, 33F24J0516, 33F24K0118, 33F24L0737, 33F25A0476, 33F25C0887, 33F25E0067, 33F25E0929, 33F25J0477, 33F25L0399, 33F25M0905, 33F24D0079; AK-15402/10801902168532, 10801902193275, 10801902168532, 10801902168549/33F24B0194, 33F24E0350, 33F24G0354, 33F24H0034, 33F24J0518, 33F24K0120, 33F24L0751, 33F25A0479, 33F25C0799, 33F25E0069, 33F25E0932, 33F25H0723, 33F25J0836, 33F25L0348, 33F26C0317, 33F24A1268; AK-15553/10801902168549, 10801902193275, 10801902168549, 10801902168563/33F24E0714, 33F24G0226, 33F24G0687, 33F24H0249, 33F24J0353, 33F25B0332, 33F25D0195, 33F25H0796, 33F25J0081, 33F26B0287, 33F24A1269; AK-16402/10801902168563, 10801902193275, 10801902168563, 10801902168570/33F24C0738, 33F24E0715, 33F24H0065, 33F24H0416, 33F24J0212, 33F24J0426, 33F25A0046, 33F25C0087, 33F25C0604, 33F25G0458, 33F25J0082, 33F26A1658, 33F26B0069, 33F26C0446, 33F24A0796; AK-16553/10801902168570, 10801902193275, 30801902122248/33F24E0178, 33F24E0351, 33F24G0355, 33F24J0519, 33F24K0121, 33F24L0790, 33F25C0980, 33F25E0070, 33F25J0478, 33F24C0025; ASK-04301-HH2/10801902193275, 20801902114987/33F25H0640, 33F25K0070, 33F25K0090, 33F24D0733; ASK-12703-MGH3/10801902193275, 10801902115260/33F25D0263, 33F24F0838

Devices named in this recall

18 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

152 devices cleared under K021120, K171146, K810962, K820009, K862056 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98793).