Photodynamic Bone Stabilization Procedure Pack USSL-0700200Recalls & reported problems
Illuminoss Medical, Inc. · Photodynamic Bone Stabilization Procedure Pack · Class 2 · Implant
All models of Photodynamic Bone Stabilization Procedure Pack (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M986USSL07002000), received between Mar 15, 2022 and Mar 15, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 15, 2022Fracture; Fracture
ILLUMINOSS PHOTODYNAMIC BONE STABILIZATION SYSTEM IMPLANTS WERE IMPLANTED IN THE RADIUS AND ULNA ON THE SAME SIDE FOR A PATIENT WHO HAD A MIDSHAFT BREAK OF BOTH. 11-12 DAYS POST OP BOTH IMPLANTS BROKE AND REVISION SURGERY WAS PERFORMED (B)(6) 2022 ADDING PLATE AND SCREWS. FURTHER DETAILS: THE PATIENT BROKE BOTH THEIR RADIUS AND ULNA ON THE SAME SIDE, AT THE SAME POINT MIDSHAFT. ON (B)(6) 2022 TWO ILLUMINOSS IMPLANTS WERE USED TO TREAT BOTH FRACTURES (7X200MM DIAMETER IMPLANTS). THE INITIAL SURGERY WENT WELL AND THE IMPLANTS HAD GOOD CORTICAL WALL CONTACT. THE PATIENT WAS DOING WELL FOR 10-11 DAYS POST-OP WITH NO PAIN AND NO ISSUE. ON DAY 11 OR 12, BOTH IMPLANTS BROKE AT THE LOCATION OF THE ORIGINAL BONE FRACTURES. REVISION SURGERY WAS PERFORMED FOR THE BONE FRACTURES OF BOTH RADIUS AND ULNA ON (B)(6) 2022 TO ADD PLATES AND SCREWS. THE ILLUMINOSS IMPLANTS WERE LEFT IN THE BONE, AND USED TO SECURE THE ADDITIONAL PLATES AND SCREWS. THE PATIENT CONDITION WAS REPORTED AS GOOD, AFTER THE REVISION SURGERY. THE PATIENT WAS NOT GIVEN ANY POST-OPERATIVE IMMOBOBILIZATION, RESTRICTION OF MOVEMENT OR OTHER STABILIZATION INSTRUCTIONS UPON INITIAL IMPLANTATION WITH THE ILLUMINOSS IMPLANTS IN THE
Patient problems: Pain
MAUDE report 13775228 ↗
Other Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated devices
- Photodynamic Bone Stabilization Procedure Pack USSL-1700260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated2 deaths8 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213260Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 death6 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated4 deaths5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213240Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated5 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-2213160Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1700280Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated3 reports
- Photodynamic Bone Stabilization Procedure Pack USSL-1100220Illuminoss Medical, Inc. · Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page