Affinity Fusion recallZ-1024-2023 · Medtronic Perfusion Systems · initiated Dec 12, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
FDA root cause category: Process change control
What the company is doing
Medtronic initially notified consignees on 12/12/2022 and followed up with an updated notification on 12/21/2022. Consignees were advised of two options: 1. Continue to use the Affinity Fusion Oxygenator without using the TMA (use other conventional perfusion circuit temperature monitoring methods); or 2. Continue to use the Affinity Fusion Oxygenator and the TMA for arterial temperature monitoring. Ensure there is minimal torque applied to the oxygenator TMA when attaching or detaching the Temperature Probe (Product Number ATP210). Also, minimize manipulation of the TMA-probe connection during the clinical procedure. Consignees were also advised to complete and return the Customer Confirmation Form, and notify personnel or facilities were affected units are located. On about 02/10/2023, Medtronic initiated an expansion and notified customers. Instructions in the customer letter remained the same.
Recall details
- Recall number
- Z-1024-2023
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2022
- Posted by the FDA
- Jan 27, 2023
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 90,220 units
- Product code
- DTZ
- FDA submissions
- K203111
- Distribution
- Worldwide distribution. US Nationwide, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Belgium, Plurinational State of Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Honduras, Hungary, Iceland, Indonesia, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Latvia, Lithuania, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Slovenia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Turkey, Turkmenistan, Uganda, United Arab Emirates, United Kingdom, Uruguay, Viet Nam, and Yemen.
Product description
Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, OXY/CVR COMBO BB841 FUSION W/BALANCE 26L, REF BB841. Used in Cardiopulmonary bypass.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Model BB841; Serial Number Range: 8111483548 - 8113999999; GTIN Number: 00643169354869 (Lot Numbers: 13384637, 13385152, 13385470, 13385745, 13385981, 13386194, 13386335, 13386747, 13386941, 13387068, 13387247, 13387298, 13387483, 13387670, 13387792, 13388008, 13388227, 13389074, 13389305, 13389823, 13389827, 13390010, 13390330, 13390505, 13390516, 13390751, 13391244, 13391264, 13391564, 13392146, 13392253, 13392727, 13393000, 13393036, 13393583, 13393871, 13394240, 13394370, 13394371, 13394424, 13394642, 13394972, 13395023, 13395803, 13395855, 13395924, 13396141, 13396262, 223431961, 223431962, 223476434, 223476435, 223525108, 223525109, 223525110, 223662140, 223662141, 223662142, 223662143, 223662144, 223746776, 223748866, 223751647, 223753656, 223753657, 223753658, 223753659, 223779522, 223782425, 223785172, 223790984, 223799185, 223800819, 223800886, 223806602, 223808121, 223812498, 223834087, 223834088, 223834089, 223834090, 223923171, 224046843, 224049195, 224060623, 224062467, 224071460, 224078980, 224087129, 224097013, 224169153, 224173725, 224180684, 224189083, 224198472, 224207770, 224217125, 224229400, 224229401, 224351135, 224384707, 224391477, 224401622, 224414139, 224428547, 224434720, 224459327, 224465687, 224469080, 224555817, 224562193, 224568610, 224571733, 224582640, 224591854, 224600536, 224609441, 224620065, 224629824, 224673589, 224673593, 224694993, 224702972, 224715756, 224722879, 224730837, 224736123, 224746679, 224754722, 224772850, 224772851, 224772852, 224772853, 224772854, 224772855, 224772856, 224772857, 224772858, 224772859, 224772860, 224772861, 224772862, 224772863, 224772864, 224772865, 224772886, 224772887, 224772889, 224772890, 224772891, 224772892, 224772893, 224772895, 224772896, 224772897, 225020112, 225209644, 225209645, 225210026, 225332480, 225332481, 225332482, 225332483); GTIN Number: 00643169354883 (Lot Numbers: 13384723, 13384730, 13384881, 13384976, 13385583, 13385767, 13385867, 13386228, 13386478, 13388449, 13388452, 13388544, 13388598, 13389191, 13389598, 13389870, 13390168, 13390640, 13390684, 13390874, 13390984, 13391365, 13391390, 13391663, 13391937, 13392221, 13392464, 13392614, 13392903, 13392975, 13393065, 13393176, 13393342, 13393342, 13393658, 13393890, 13393921, 13394029, 13394151, 13394821, 13395330, 13395332, 13395410, 13395461, 13395524, 13395771, 223388884, 223392415, 223392423, 223392435, 223431964, 223431965, 223431986, 223476436, 223476437, 223476438, 223476439, 223476439, 223476440, 223523220, 223523221, 223523222, 223523223, 223523224, 223523225, 223525106, 223525107, 223637178, 223641990, 223650772, 223660380, 223662145, 223662186, 223662187, 223662188, 223662189, 223662190, 223662191, 223723553, 223738421, 223750562, 223753660, 223753661, 223753662, 223753663, 223753664, 223753967, 223755426, 223755427, 223755428, 223833251, 223833256, 223834091, 223834092, 223834093, 223834094, 223834095, 223834096, 223872428, 223878290, 223878291, 223883408, 223893771, 223898271, 223903420, 223905400, 223913345, 223914588, 223928428, 223936394, 223939036, 223947101, 223953209, 223961794, 223964866, 223972811, 223981604, 223995093, 223995666, 224000958, 224006521, 224006951, 224011284, 224011625, 224011646, 224011647, 224011648, 224011649, 224011650, 224101106, 224236728, 224253937, 224312807, 224323095, 224331329, 224334991, 224346525, 224356382, 224356383, 224372815, 224475835, 224478783, 224484698, 224494813, 224506201, 224511441, 224518515, 224526165, 224526860, 224532203, 224535429, 224544419, 224630603, 224630604, 224644422, 224673817, 224686850, 224763409, 224771068, 224772898, 224772899, 224772900, 224772901, 224772902, 224772903, 224772904, 224772905, 224772926, 224772927, 224772928, 224772929, 224772930, 224772931, 224772932, 224772933, 224772935, 224772936, 224772937, 224772940, 224772942, 224772943, 224772944, 224772945, 224772986, 225090558, 225090559, 225090560, 225162245, 225162946, 225174904, 225174905, 225176086, 225294537, 225294539, 225294541, 225294543, 225294544, 225332485, 225332486, 225332487, 225332488, 225332489); GTIN Number: 00643169588660(Lot Numbers: 13385981, 13387247, 13387483, 13387792, 13389305, 13389823, 13389827, 13390516, 13390751, 13391244, 13391564, 13394370, 13394424, 13395924, 223476434, 223662140, 223662144, 223961794, 224087129, 224401622, 224526860, 224772929, 225020112, 225090559, 225162245, 225162946)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Affinity Fusion® BB841Medtronic, Inc. · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS2 recalls36 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K203111 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Affinity Fusion® BB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- Affinity Fusion™ CB811Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS1 recall4 reports
- Affinity Pixie™ BBP241Medtronic, Inc. · RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS1 recall2 reports
- Affinity Fusion™ CB841Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS1 recall0 reports
- Affinity Pixie™ BBP211Medtronic, Inc. · OXYGENATOR, CARDIOPULMONARY BYPASS12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91343).
- Z-1023-2023 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF BB811. Used in Cardiopulmonary bypass.
- Z-1025-2023 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB811. Used in Cardiopulmonary bypass.
- Z-1026-2023 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass.
- Z-1027-2023 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF BB811. Used in Cardiopulmonary bypass.
- Z-1028-2023 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF CB811. Used in Cardiopulmonary bypass.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.