MR 5000 SeriesRecalls & reported problems across all models

Philips Medical Systems Nederland B.V. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

8 open recalls name at least one MR 5000 Series model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls
8

We found 3 models of MR 5000 Series by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 11 adverse event reports in the FDA's MAUDE database, received between Aug 20, 2024 and Jul 3, 2026 — classified as 11 injuries.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (11 reports).

8 FDA recall records name at least one of these models by its device identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

MR 5000 Series recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 28, 2026Open, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3312-2026 · Root cause: Software design · names 3 of these models · Recall details · FDA record ↗

  2. Aug 21, 2026Open, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3241-2026 · Root cause: Software design · Recall details · FDA record ↗

  3. Jun 30, 2025Open, Classified

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

    Z-2213-2025 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  4. May 13, 2025Open, Classified

    Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.

    Z-1804-2025 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  5. Sep 3, 2024Open, Classified

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

    Z-3201-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. May 3, 2024Open, Classified

    The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

    Z-1855-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. May 3, 2024Open, Classified

    The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

    Z-1854-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Mar 12, 2024Open, Classified

    Patient support table floor plate may be incorrectly installed.

    Z-1587-2024 · Root cause: Process design · Recall details · FDA record ↗

Most reported MR 5000 Series problems

Device problems

  • Improper or Incorrect Procedure or Method11
  • Insufficient Device Problem Information8

Patient problems

  • Partial thickness (Second Degree) Burn11
  • Burn(s)1
  • Full thickness (Third Degree) Burn1

MR 5000 Series models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MR 5000 Series adverse event reports

The 11 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Philips Medical Systems Nederland B.V. products

All Philips Medical Systems Nederland B.V. devices →