DePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl recallZ-0707-2013 · DePuy Spine, Inc. · initiated Nov 7, 2012
Class II Terminated All recalls posted in January 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head
FDA root cause category: Component change control
What the company is doing
DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.
Recall details
- Recall number
- Z-0707-2013
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 7, 2012
- Posted by the FDA
- Jan 24, 2013
- Recalling firm
- DePuy Spine, Inc.
- Product code
- NKB
- FDA submissions
- K101993
- Distribution
- Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.
Product description
DePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
Code information (models, lots, serial numbers)
All lots
Other devices under the same FDA submission
22 devices cleared under K101993 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- EXPEDIUM 179712880Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL8 reports
- EXPEDIUM 179712980Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation6 reports
- VIPER 186715850Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease6 reports
- EXPEDIUM 179712850Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation5 reports
- EXPEDIUM 179712990Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY4 reports
- EXPEDIUM 179712199Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION2 reports
- EXPEDIUM 179712840Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation2 reports
- EXPEDIUM 179712855Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION2 reports
- EXPEDIUM 179712870Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease2 reports
- EXPEDIUM 179712875Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION2 reports
- EXPEDIUM 179712960Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION2 reports
- EXPEDIUM 179712280Medos International Sàrl · Orthosis, spondylolisthesis spinal fixation1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 63665).
- Z-0705-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
- Z-0706-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
- Z-0708-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
- Z-0709-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
- Z-0710-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
- Z-0712-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
- Z-0713-2013 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page