Skip to content

DePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl recallZ-0709-2013 · DePuy Spine, Inc. · initiated Nov 7, 2012

Class II Terminated All recalls posted in January 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Polyaxial screws 8x 80 and larger may disassociate when inserting in the Ilium., VIPER large diameter long screws in high-torque circumstances, the screw shank has the potential to separate from the polyaxial head

FDA root cause category: Component change control

What the company is doing

DePuy Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated November 7, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to examine inventory, quarantine the material and return remaining inventory. Customers were advised to complete the enclosed Business Reply Form and fax back to 508-828-3762 or email to [email protected]. For questions call Linda Regan @508-977-6606.

Recall details

Recall number
Z-0709-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 7, 2012
Posted by the FDA
Jan 24, 2013
Recalling firm
DePuy Spine, Inc.
Quantity
34 uits
Product code
NKB
FDA submissions
K101993
Distribution
Worldwide Distribution - US (nationwide) including the states of: AZ, CA, FL, HI, IL, MD,MI, MO,NC, TN, TX and the country of Australia.

Product description

DePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.

Code information (models, lots, serial numbers)

All lots

Other devices under the same FDA submission

22 devices cleared under K101993 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 63665).

Sources and updates