Superior Approach Resection Guide Assembly recallZ-1548-2013 · Biomet, Inc. · initiated May 15, 2013

Class II Terminated All recalls posted in June 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.

FDA root cause category: Device Design

What the company is doing

On 5/15/2013 an URGENT MEDICAL DEVICE RECALL notification was sent to all consignees detailing to immediately locate and discontinue use of the product and return to Biomet. The following actions are REQUIRED: -Immediately locate and remove the identified device(s) listed below from circulation . -Carefully follow the instructions on the enclosed "FAX Back Response Form" . -Fax a copy of the Response Form to 574-372-1683 prior to return of product. -Use priority carrier for your shipment. - If you have further distributed this product. you MUST notify hospital personnel of this action via the enclosed "Dear O.R. Manager" notice. This letter MUST be given to hospital personnel responsible for receiving recall notices.

Recall details

Recall number
Z-1548-2013
Recall class
Class II
Status
Terminated
Initiated by the company
May 15, 2013
Posted by the FDA
Jun 17, 2013
Recalling firm
Biomet, Inc.
Quantity
17
Product code
PAO
FDA submissions
K080642
Distribution
Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.

Product description

Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.

Code information (models, lots, serial numbers)

Catalog number: 010001814 and lot numbers: 318520 and 644720.

Other devices under the same FDA submission

46 devices cleared under K080642 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.