Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190… recallZ-1145-2019 · Roche Diagnostics Corporation · initiated Dec 17, 2018
Class II Terminated All recalls posted in April 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
FDA root cause category: Device Design
What the company is doing
On 12/17/18, the firm distributed an Urgent Medical Device Correction letter to affected consignees. Customers were notified of the interference issues. The Limitations-interference section of the method sheets for the cobas c 311 analyzer and the cobas c 501, 502, 701, and 702 modules are being updated with the following claims: Bilirubin Total Gen.3 assay: Immunoglobulins: No significant interference from immunoglobulins up to a concentration of 28 g/L (187 mol/L) (simulated by human immunogloblulin G). Lactate Gen.2 assay (CSF samples): Ditaurobilirubin: No significant interference from ditaurobilirubin up to an approximate concentration of 102 mol/L (6 mg/dL). Tina-quant IGM-2 (Sensitive application): Hemolysis: No significant interference up to an H index of 600 (approximate hemoglobin concentration of 373 mol/L or 600 mg/dL). Lipemia (Intralipid): No significant interference up to an L index of 600 Customers were asked to do the following: " Be aware of the new interference claims listed in this Urgent Medical Device Correction (UMDC). This UMDC serves as labeling until the updated method sheets are available. The method sheets for the MODULAR ANALYTICS P/D modules will not be updated. " Manually update the H and L indexes for the IGMP2 sensitive application on your cobas c 311 analyzer or cobas c 501, 502, 701, and 702 modules following the instructions provided in this UMDC. " Complete all sections of the enclosed fax form (TP-00507) and fax it to 1-xxx-xxx-xxx or email it to [email protected]. " File this UMDC for future reference. When the method sheets are updated, they will be available on the diagnostics.roche.com Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions. website and on the cobas e-library. ***UPDATE: Firm sent updated communication to customers dated 5/9/2019 informing them that L and H index values
Recall details
- Recall number
- Z-1145-2019
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 17, 2018
- Posted by the FDA
- Apr 13, 2019
- Recalling firm
- Roche Diagnostics Corporation
- Quantity
- 5292 analytical units
- Product code
- CIG
- FDA submissions
- K131544
- Distribution
- US Nationwide distribution.
Product description
Bilirubin Total Gen.3, Cat. Nos. 05795397190, 05795419190, 05795320160, 05795338160, 05795346160 ***added as of 5/10/19***05795397190, 05795648190 Product Usage: In vitro test for the quantitative determination of total bilirubin in serum and plasma of adults and neonates on Roche/Hitachi cobas c systems.
Code information (models, lots, serial numbers)
n/a
Other devices under the same FDA submission
7 devices cleared under K131544 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Bilirubin Total Gen3 5795397190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN16 reports
- Bilirubin Total Gen.3 08056960190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN14 reports
- Bilirubin Total Gen3 5795419190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN5 reports
- Bilirubin Total Gen.3 10421436190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN4 reports
- Bilirubin Total Gen3 5795648190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN1 report
- Bilirubin Total Gen.3 10421410190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN1 report
- Bilirubin Total Gen.3 10421428190Roche Diagnostics GmbH · DIAZO COLORIMETRY, BILIRUBIN1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 81859).
- Z-1146-2019 · TerminatedLactate Gen.2, Catalog Nos. 03183700190, 05171881190, 11822837190, ***added as of 5/10/19*** 03183700190 Product Usage: An In vitro diagnostic reagent system intended for u…
- Z-1147-2019 · TerminatedTina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quantitative in vitro determination of Ig…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page