uCentum recallZ-1126-2019 · ulrich medical USA Inc · initiated Mar 8, 2019
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ulrich medical USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.
FDA root cause category: Storage
What the company is doing
On March 13 , 2019 the firm sent letters to their consignees stating the following: The purpose of this letter is to advise you that ulrich medical USA is voluntarily recalling certain Torque Limiting Handles. Reason for the voluntary recall: Certain Torque Limiting Handles in the field are beyond their 3 year calibration life. The calibration issue experienced to date has been limited to select instruments and not all of these instruments have been affected. Due to the nature of how these instruments are used the risk to the patient is low. No injury or death has been reported. Risk to health: Over or under tightening of the set screw could cause the construct to fatigue or loosen. The risk to the patient is low. Actions to be taken by the user: Return the torque limiting handles to ulrich medical USA at ulrichs expense, and a new torque limiting handle will be shipped back. Instructions for acknowledgement of receipt include FedEx signature required. Actions to be taken by the Company: All affected instruments will be replaced with new torque limiting handle upon return. Contact information: If you have any questions or concerns please contact: Louis Milos Sr. Manager Business and Quality Systems 18221 Edison Ave. Chesterfield, MO 63005 636-519-0268 ext. 105 [email protected] www.ulrichmedicalusa.com Available Monday through Friday 8:00am to 5:00pm CST
Recall details
- Recall number
- Z-1126-2019
- Status
- Terminated
- Initiated by the company
- Mar 8, 2019
- Recalling firm
- ulrich medical USA Inc
- Quantity
- 25
- Product code
- LXL
- FDA submissions
- K102853, K123717, K161032
- Distribution
- AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.
Product description
ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Code information (models, lots, serial numbers)
Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- uCentum CS 7146Ulrich GmbH & Co. KG · Orthopedic Manual Surgical Instrument1 recall0 reports
- uCentum CS 7148Ulrich GmbH & Co. KG · Orthopedic Manual Surgical Instrument1 recall0 reports
- mambo CS 7134Ulrich GmbH & Co. KG · Orthopedic Manual Surgical Instrument1 recall0 reports
Product lines
Other devices under the same FDA submission
39 devices cleared under K102853, K123717, K161032 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- uCentum™ comprehensive posterior system CS 3802-065-050Ulrich GmbH & Co. KG · Orthosis, spinal pedicle fixation, for degenerative disc disease16 reports
- uCentum™ comprehensive posterior system CS 3801-01Ulrich GmbH & Co. KG · Orthosis, spinal pedicle fixation, for degenerative disc disease12 reports
- uCentum™ comprehensive posterior system CS 3810-01Ulrich GmbH & Co. KG · ORTHOSIS, SPINAL PEDICLE FIXATION7 reports
- uCentum™ comprehensive posterior system CS 3802-055-040Ulrich GmbH & Co. KG · Orthosis, spondylolisthesis spinal fixation6 reports
- neon³ CT spinal stabilization CS 3906-40-30Ulrich GmbH & Co. KG · APPLIANCE, FIXATION, SPINAL INTERLAMINAL6 reports
- uCentum™ comprehensive posterior system CS 3802-065-045Ulrich GmbH & Co. KG · Orthosis, spinal pedicle fixation, for degenerative disc disease5 reports
- neon³ CT spinal stabilization CS 3902-35-14Ulrich GmbH & Co. KG · Orthosis, cervical pedicle screw spinal fixation5 reports
- neon³™ OCT spinal stabilization CS 3913-05Ulrich GmbH & Co. KG · Orthosis, cervical pedicle screw spinal fixation5 reports
- uCentum™ comprehensive posterior system CS 3804-065-045Ulrich GmbH & Co. KG · ORTHOSIS, SPINAL PEDICLE FIXATION4 reports
- N/A CS 3825-01Ulrich GmbH & Co. KG · Orthosis, spondylolisthesis spinal fixation4 reports
- neon³ CT spinal stabilization CS 3902-35-16Ulrich GmbH & Co. KG · APPLIANCE, FIXATION, SPINAL INTERLAMINAL4 reports
- uCentum™ comprehensive posterior system CS 3802-075-045Ulrich GmbH & Co. KG · Orthosis, spinal pedicle fixation, for degenerative disc disease3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page