Coated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product… recallZ-0989-2020 · Ethicon, Inc. · initiated Jan 3, 2020
Class II Terminated All recalls posted in February 2020
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some sutures in the lot may exhibit a small section of braid fraying, which can lead to reduced tensile strength at that point.
FDA root cause category: Process control
What the company is doing
The firm notified customers of the recall on January 3, 2020. Customers were informed to immediately cease usage or distribution of the recalled lot. All unused product subject to the recall is to be returned to the firm. If any recalled product was forwarded to another facility, contact that facility to arrange return. Customers were also asked to complete and return the included business reply form to the recalling firm even if they did not have product subject to this recall. In addition, customers are to keep the recall notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. If you require any assistance with returning product, please contact Stericycle at 877-550-0701 and reference Event # 3453. Health care practitioners who have treated patients using this product should follow those patients post-operatively in the usual manner with no additional action required. If you have additional questions regarding this recall or to report any customer complaints, please contact Ethicon Customer Support Center at 1-877-ETHICON (1-877-384-4266). The Customer Support Center is open Monday through Friday, 7:30 AM to 6:30 PM ET.
Recall details
- Recall number
- Z-0989-2020
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 3, 2020
- Posted by the FDA
- Feb 4, 2020
- Recalling firm
- Ethicon, Inc.
- Quantity
- 23688 units
- Product code
- GAM
- FDA submissions
- K022269
- Distribution
- Domestic distribution nationwide. US and Puerto Rico only.
Product description
Coated VICRYL (Polyglactin 910) Suture, 18" (45cm) Product Code J497H. The device is a synthetic absorbable sterile surgical suture.
Code information (models, lots, serial numbers)
lot PGZ908
Other devices under the same FDA submission
145 devices cleared under K022269 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Coated VICRYL J433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID255 reports
- Coated VICRYL J570GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID170 reports
- Coated VICRYL J258HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID155 reports
- VICRYL V960GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall55 reports
- Coated VICRYL J494GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID86 reports
- Coated VICRYL JB840Ethicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID81 reports
- Coated VICRYL J603HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID80 reports
- Coated VICRYL J602HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID76 reports
- VICRYL V449GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID74 reports
- Coated VICRYL J311HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID71 reports
- Coated VICRYL J490GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID67 reports
- Coated VICRYL J493GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID63 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page