HVAD Pump Implant Kit recallZ-2055-2020 · Heartware, Inc. · initiated Apr 3, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
FDA root cause category: Device Design
What the company is doing
The firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.
Recall details
- Recall number
- Z-2055-2020
- Status
- Open, Classified
- Initiated by the company
- Apr 3, 2020
- Recalling firm
- Heartware, Inc.
- Quantity
- 2614 units
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution
Product description
Heartware HVAD Pump Implant Kit, Product (REF) Number 1103
Code information (models, lots, serial numbers)
Show the full code information (24k characters)
GTIN 00888707000017, Serial Numbers: HW26899, HW27283, HW27893, HW28873, HW28999, HW29422, HW29488, HW29722, HW30166, HW30170 GTIN 00888707002639, Serial Numbers: HW29525, HW30015, HW30297, HW30477, HW31174, HW31179, HW31181, HW31246, HW31247, HW31250, HW31253, HW31257, HW31266, HW31268, hw31280, HW31281, hw31282, HW31283, HW31313, HW31385, HW31401, HW31497, HW31525, HW31526, HW31530, HW31531, HW31562, HW31567, HW31596, HW31597, HW31601, HW31602, HW31608, HW31609, HW31612, HW31624, HW31626, HW31634, HW31640, HW31641, HW31642, HW31643, HW31644, HW31645, HW31646, HW31647, HW31648, HW31649, HW31651, HW31652, HW31653, HW31655, HW31656, HW31657, HW31658, HW31659, HW31660, HW31661, HW31662, HW31663, HW31665, HW31666, HW31667, HW31668, HW31669, HW31670, HW31671, HW31672, HW31674, HW31675, HW31676, HW31677, HW31678, HW31679, HW31680, HW31681, HW31682, HW31687, HW31688, HW31689, HW31690, HW31691, HW31692, HW31693, HW31695, HW31696, HW31701, HW31705, HW31706, HW31707, HW31708, HW31709, HW31710, HW31711, HW31712, HW31713, HW31714, HW31716, HW31717, HW31719, HW31720, HW31721, HW31722, HW31723, HW31724, HW31725, HW31726, HW31728, HW31730, HW31732, HW31734, HW31736, HW31737, HW31738, HW31739, HW31740, HW31741, HW31742, HW31743, HW31744, HW31745, HW31746, HW31747, HW31748, HW31751, HW31752, HW31753, HW31754, HW31756, HW31757, HW31758, HW31760, HW31761, HW31762, HW31763, HW31764, HW31765, HW31766, HW31767, HW31768, HW31769, HW31770, HW31771, HW31772, HW31773, HW31774, HW31775, HW31776, HW31781, HW31782, HW31783, HW31784, HW31785, HW31786, HW31787, HW31788, HW31789, HW31790, HW31792, HW31793, HW31794, HW31796, HW31797, HW31798, HW31799, HW31800, HW31802, HW31803, HW31804, HW31805, HW31807, HW31816, HW31829, HW31841, HW31848, HW31850, HW31853, HW31854, HW31857, HW31861, HW31862, HW31864, HW31865, HW31866, HW31867, HW31868, HW31869, HW31870, HW31871, HW31878, HW31879, HW31880, HW31881, HW31882, HW31884, HW31885, HW31886, HW31887, HW31888, HW31889, HW31890, HW31894, HW31895, HW31896, HW31897, HW31898, HW31899, HW31900, HW31902, HW31904, HW31905, HW31906, HW31907, HW31910, HW31911, HW31913, HW31914, HW31916, HW31917, HW31918, HW31922, HW31923, HW31933, HW31935, HW31936, HW31938, HW31942, HW31943, HW31945, HW31946, HW31948, HW31949, HW31950, HW31951, HW31952, HW31953, HW31954, HW31956, HW31957, HW31959, HW31960, HW31961, HW31962, HW31963, HW31964, HW31965, HW31966, HW31967, HW31968, HW31969, HW31970, HW31971, HW31972, HW31974, HW31975, HW31977, HW31978, HW31980, HW31990, HW32003, HW32004, HW32005, HW32007, HW32016, HW32019, HW32020, HW32031, HW32033, HW32037, HW32041, HW32044, HW32045, HW32047, HW32048, HW32049, HW32051, HW32052, HW32054, HW32058, HW32059, HW32060, HW32061, HW32063, HW32066, HW32067, HW32069, HW32070, HW32072, HW32073, HW32074, HW32075, HW32076, HW32077, HW32078, HW32079, HW32080, HW32082, HW32083, HW32084, HW32087, HW32088, HW32089, HW32090, HW32091, HW32092, HW32093, HW32095, HW32096, HW32097, HW32098, HW32099, HW32100, HW32101, HW32104, HW32105, HW32106, HW32107, HW32110, HW32111, HW32114, HW32115, HW32117, HW32119, HW32120, HW32122, HW32126, HW32127, HW32128, HW32129, HW32130, HW32133, HW32134, HW32136, HW32137, HW32139, HW32140, HW32142, HW32145, HW32146, HW32148, HW32150, HW32151, HW32154, HW32155, HW32157, HW32158, HW32168, HW32169, HW32171, HW32173, HW32174, HW32176, HW32177, HW32178, HW32184, HW32185, HW32187, HW32188, HW32191, HW32196, HW32203, HW32208, HW32213, HW32225, HW32227, HW32233, HW32235, HW32237, HW32247, HW32249, HW32251, HW32252, HW32254, HW32262, HW32266, HW32267, HW32271, HW32272, HW32274, HW32275, HW32276, HW32277, HW32278, HW32280, HW32283, HW32284, HW32287, HW32288, HW32291, HW32293, HW32295, HW32296, HW32298, HW32299, HW32300, HW32302, HW32303, HW32304, HW32305, HW32306, HW32307, HW32308, HW32310, HW32311, HW32312, HW32313, HW32314, HW32316, HW32317, HW32318, HW32319, HW32320, HW32321, HW32322, HW32323, HW32324, HW32325, HW32326, HW32327, HW32328, HW32329, HW32330, HW32331, HW32332, HW32333, HW32334, HW32337, HW32338, HW32339, HW32340, HW32341, HW32343, HW32344, HW32345, HW32346, HW32348, HW32349, HW32350, HW32351, HW32352, HW32354, HW32355, HW32358, HW32359, HW32360, HW32361, HW32362, HW32363, HW32364, HW32365, HW32366, HW32367, HW32368, HW32370, HW32371, HW32372, HW32374, HW32375, HW32376, HW32377, HW32378, HW32379, HW32380, HW32381, HW32382, HW32383, HW32384, HW32387, HW32389, HW32390, HW32391, HW32392, HW32395, HW32396, HW32397, HW32398, HW32399, HW32401, HW32402, HW32403, HW32404, HW32405, HW32406, HW32409, HW32411, HW32412, HW32413, HW32414, HW32417, HW32418, HW32421, HW32422, HW32424, HW32425, HW32428, HW32429, HW32430, HW32431, HW32432, HW32433, HW32434, HW32436, HW32438, HW32439, HW32440, HW32441, HW32442, HW32443, HW32444, HW32446, HW32447, HW32448, HW32449, HW32450, HW32451, HW32452, HW32453, HW32454, HW32455, HW32456, HW32460, HW32461, HW32462, HW32463, HW32466, HW32467, HW32469, HW32475, HW32476, HW32477, HW32478, HW32480, HW32481, HW32482, HW32484, HW32485, HW32486, HW32487, HW32488, HW32489, HW32490, HW32491, HW32492, HW32493, HW32494, HW32495, HW32496, HW32497, HW32498, HW32500, HW32501, HW32502, HW32503, HW32512, HW32526, HW32527, HW32528, HW32529, HW32530, HW32531, HW32532, HW32533, HW32535, HW32536, HW32537, HW32538, HW32539, HW32540, HW32541, HW32542, HW32543, HW32544, HW32545, HW32547, HW32549, HW32550, HW32551, HW32552, HW32553, HW32555, HW32556, HW34011, HW34013, HW34014, HW34016, HW34032, HW34055, HW34057, HW34058, HW34059, HW34061, HW34063, HW34064, HW34065, HW34067, HW34070, HW34071, HW34073, HW34074, HW34076, HW34077, HW34079, HW34080, HW34081, HW34082, HW34083, HW34085, HW34087, HW34089, HW34097, HW34098, HW34099, HW34101, HW34102, HW34103, HW34105, HW34108, HW34112, HW34114, HW34116, HW34118, HW34119, HW34127, HW34128, HW34130, HW34131, HW34133, HW34134, HW34135, HW34136, HW34137, HW34138, HW34139, HW34140, HW34141, HW34142, HW34143, HW34144, HW34145, HW34146, HW34147, HW34153, HW34154, HW34155, HW34156, HW34157, HW34158, HW34159, HW34160, HW34162, HW34163, HW34164, HW34165, HW34166, HW34167, HW34168, HW34170, HW34171, HW34172, HW34173, HW34174, HW34175, HW34176, HW34177, HW34178, HW34183, HW34185, HW34188, HW34189, HW34190, HW34194, HW34195, HW34198, HW34199, HW34200, HW34201, HW34203, HW34212, HW34213, HW34214, HW34215, HW34216, HW34217, HW34218, HW34219, HW34220, HW34221, HW34222, HW34223, HW34224, HW34225, HW34227, HW34228, HW34230, HW34231, HW34233, HW34234, HW34235, HW34236, HW34238, HW34239, HW34240, HW34241, HW34242, HW34243, HW34244, HW34245, HW34246, HW34247, HW34248, HW34249, HW34250, HW34253, HW34254, HW34255, HW34256, HW34257, HW34258, HW34259, HW34260, HW34261, HW34262, HW34263, HW34264, HW34265, HW34267, HW34276, HW34279, HW34282, HW34284, HW34285, HW34286, HW34287, HW34288, HW34289, HW34291, HW34294, HW34295, HW34297, HW34298, HW34299, HW34300, HW34302, HW34303, HW34304, HW34305, HW34306, HW34307, HW34308, HW34310, HW34311, HW34312, HW34313, HW34314, HW34315, HW34316, HW34318, HW34320, HW34321, HW34322, HW34323, HW34325, HW34332, HW34333, HW34334, HW34335, HW34338, HW34339, HW34353, HW34360, HW34385, HW34386, HW34387, HW34389, HW34390, HW34391, HW34392, HW34393, HW34394, HW34396, HW34397, HW34399, HW34400, HW34402, HW34403, HW34404, HW34405, HW34406, HW34407, HW34409, HW34410, HW34411, hw34413, hw34416, hw34417, HW34422, HW34423, HW34424, HW34430, HW34435, HW34441, HW34445, HW34447, HW34450, HW34451, HW34453, HW34455, HW34457, HW34458, HW34459, HW34460, HW34462, HW34463, HW34464, HW34465, HW34466, HW34467, HW34468, HW34469, HW34470, HW34471, HW34472, HW34473, HW34475, HW34476, HW34477, HW34479, HW34480, HW34481, HW34482, HW34483, HW34485, HW34486, HW34487, HW34488, HW34489, HW34492, HW34493, HW34494, HW34496, HW34498, HW34499, HW34500, HW34501, HW34502, HW34503, HW34504, HW34505, HW34506, HW34515, HW34525, HW34526, HW34527, HW34528, HW34529, HW34530, HW34531, HW34532, HW34533, HW34534, HW34535, HW34537, HW34538, HW34540, HW34541, HW34542, HW34543, HW34544, HW34545, HW34548, HW34549, HW34552, HW34553, HW34554, HW34558, HW34562, HW34564, HW34565, HW34566, HW34567, HW34568, HW34570, HW34571, HW34572, HW34573, HW34574, HW34575, HW34576, HW34577, HW34578, HW34579, HW34580, HW34581, HW34583, HW34584, HW34585, HW34587, HW34589, HW34592, HW34593, HW34597, HW34598, HW34599, HW34600, HW34601, HW34602, HW34604, HW34606, HW34608, HW34611, HW34612, HW34613, HW34614, HW34616, HW34626, HW34627, HW34632, HW34633, HW34634, HW34635, HW34638, HW34639, HW34640, HW34642, HW34644, HW34645, HW34646, HW34647, HW34648, HW34649, HW34650, HW34651, HW34652, HW34653, HW34654, HW34655, HW34657, HW34658, HW34659, HW34660, HW34661, HW34662, HW34663, HW34665, HW34667, HW34668, HW34669, HW34671, HW34672, HW34674, HW34675, HW34676, HW34677, HW34678, HW34679, HW34680, HW34681, HW34683, HW34684, HW34685, HW34686, HW34687, HW34688, HW34689, HW34690, HW34695, HW34696, HW34697, HW34698, HW34699, HW34701, HW34702, HW34705, HW34706, HW34707, HW34708, HW34709, HW34712, HW34714, HW34715, HW34716, HW34718, HW34721, HW34722, HW34725, HW34728, HW34729, HW34730, HW34731, HW34733, HW34735, HW34736, HW34740, HW34741, HW34744, HW34747, HW34754, HW34755, HW34756, HW34760, HW34763, HW34765, HW34766, HW34770, HW34771, HW34772, HW34773, HW34774, HW34775, HW34777, HW34778, HW34780, HW34781, HW34782, HW34783, HW34784, HW34786, HW34787, HW34788, HW34789, HW34790, HW34793, HW34794, HW34795, HW34796, HW34797, HW34798, HW34799, HW34800, HW34804, HW34807, HW34809, HW34811, HW34812, HW34814, HW34815, HW34816, HW34819, HW34820, HW34821, HW34822, HW34823, HW34825, HW34826, HW34827, HW34828, HW34830, HW34831, HW34832, HW34835, HW34843, HW34852, HW34854, HW34856, HW34857, HW34858, HW34859, HW34860, HW34861, HW34863, HW34865, HW34869, HW34870, HW34871, HW34872, HW34874, HW34878, HW34883, HW34887, HW34891, HW34896, HW34899, HW34917, HW34918, HW34921, HW34922, HW34926, HW34927, HW34929, HW34930, HW34931, HW34935, HW34937, HW34942, HW34943, HW34952, HW34956, HW34958, HW34961, HW34962, HW34964, HW34967, HW34969, HW34970, HW34971, HW34972, HW34973, HW34976, HW34978, HW34980, HW34981, HW34983, HW34984, HW34987, HW34988, HW34989, HW34993, HW34995, HW34998, HW34999, HW35001, HW35003, HW35004, HW35005, HW35006, HW35007, HW35008, HW35009, HW35010, HW35011, HW35013, HW35014, HW35015, HW35017, HW35018, HW35020, HW35022, HW35023, HW35026, HW35027, HW35029, HW35030, HW35036, HW35037, HW35042, HW35044, HW35045, HW35046, HW35047, HW35048, HW35050, HW35051, HW35053, HW35055, HW35058, HW35059, HW35060, HW35063, HW35066, HW35069, HW35071, HW35072, HW35073, HW35074, HW35075, HW35076, HW35077, HW35078, HW35079, HW35081, HW35082, HW35084, HW35087, HW35088, HW35089, HW35092, HW35096, HW35098, HW35100, HW35101, HW35103, HW35104, HW35105, HW35106, HW35107, HW35109, HW35110, HW35111, HW35112, HW35113, HW35115, HW35116, HW35117, HW35118, HW35126, HW35129, HW35130, HW35132, HW35135, HW35140, HW35152, HW35153, HW35159, HW35160, HW35163, HW35165, HW35166, HW35167, HW35168, HW35169, HW35170, HW35172, HW35173, HW35174, HW35175, HW35176, HW35177, HW35178, HW35179, HW35180, HW35181, HW35182, HW35183, HW35184, HW35190, HW35192, HW35193, HW35194, HW35195, HW35198, HW35199, HW35203, HW35206, HW35219, HW35221, HW35222, HW35223, HW35224, HW35225, HW35226, HW35227, HW35229, HW35231, HW35232, HW35233, HW35234, HW35235, HW35236, HW35238, HW35239, HW35240, HW35241, HW35242, HW35243, HW35244, HW35245, HW35246, HW35247, HW35249, HW35253, HW35254, HW35255, HW35256, HW35257, HW35258, HW35259, HW35261, HW35262, HW35263, HW35264, HW35265, HW35266, HW35267, HW35268, HW35269, HW35270, HW35271, HW35273, HW35274, HW35276, HW35277, HW35280, HW35285, HW35286, HW35287, HW35288, HW35289, HW35291, HW35292, HW35293, HW35294, HW35295, HW35296, HW35299, HW35301, HW35302, HW35347, HW35354, HW35358, HW35399, HW35403, HW35406, HW35410, HW35413, HW35414, HW35415, HW35416, HW35417, HW35420, HW35422, HW35424, HW35425, HW35426, HW35427, HW35428, HW35429, HW35430, HW35431, HW35433, HW35434, HW35435, HW35436, HW35437, HW35438, HW35439, HW35440, HW35441, HW35442, HW35443, HW35444, HW35445, HW35446, HW35447, HW35448, HW35449, HW35450, HW35451, HW35452, HW35453, HW35454, HW35455, HW35456, HW35457, HW35459, HW35460, HW35461, HW35462, HW35463, HW35466, HW35469, HW35471, HW35472, HW35473, HW30051, HW30066, HW30169, HW30330, HW30573, HW30605, HW30613, HW30628, HW30677, HW30700, HW30703, HW30884, HW30904, HW30912, HW30992, HW30999, HW31004, HW31013, HW31038, HW31046, HW31053, HW31060, HW31149, HW31151, HW31199, HW31225, HW31230, HW31321, HW31367, HW31407, HW31463, HW31537 GTIN 00888707003261, Serial Numbers: HW40002, HW40004, HW40006, HW40008, HW40009, HW40010, HW40012, HW40013, HW40015, HW40016, HW40017, HW40018, HW40019, HW40021, HW40022, HW40023, HW40024, HW40025, HW40026, HW40028, HW40030, HW40032, HW40033, HW40035, HW40036, HW40037, HW40038, HW40039, HW40041, HW40042, HW40043, HW40044, HW40045, HW40046, HW40047, HW40048, HW40049, HW40050, HW40051, HW40054, HW40055, HW40057, HW40058, HW40059, HW40060, HW40061, HW40062, HW40063, HW40064, HW40065, HW40066, HW40067, HW40068, HW40069, HW40070, HW40071, HW40072, HW40073, HW40074, HW40075, HW40076, HW40077, HW40078, HW40079, HW40080, HW40081, HW40082, HW40083, HW40084, HW40085, HW40086, HW40087, HW40088, HW40089, HW40090, HW40091, HW40092, HW40093, HW40094, HW40095, HW40096, HW40097, HW40098, HW40099, HW40100, HW40102, HW40103, HW40105, HW40106, HW40107, HW40108, HW40109, HW40110, HW40111, HW40112, HW40113, HW40114, HW40115, HW40116, HW40117, HW40118, HW40119, HW40120, HW40121, HW40122, HW40123, HW40124, HW40125, HW40126, HW40127, HW40128, HW40129, HW40130, HW40131, HW40132, HW40135, HW40136, HW40137, HW40138, HW40139, HW40140, HW40142, HW40143, HW40144, HW40145, HW40146, HW40147, HW40148, HW40149, HW40150, HW40151, HW40152, HW40153, HW40154, HW40155, HW40156, HW40157, HW40158, HW40159, HW40160, HW40161, HW40163, HW40164, HW40165, HW40166, HW40167, HW40168, HW40169, HW40170, HW40171, HW40172, HW40174, HW40175, HW40176, HW40177, HW40178, HW40179, HW40180, HW40181, HW40182, HW40184, HW40185, HW40186, HW40187, HW40188, HW40189, HW40190, HW40191, HW40192, HW40193, HW40194, HW40195, HW40196, HW40197, HW40198, HW40199, HW40200, HW40201, HW40202, HW40203, HW40204, HW40206, HW40207, HW40208, HW40209, HW40210, HW40211, HW40212, HW40213, HW40214, HW40215, HW40217, HW40218, HW40219, HW40220, HW40222, HW40223, HW40224, HW40225, HW40226, HW40227, HW40228, HW40229, HW40230, HW40231, HW40232, HW40233, HW40234, HW40235, HW40236, HW40237, HW40238, HW40239, HW40240, HW40241, HW40242, HW40243, HW40244, HW40245, HW40246, HW40247, HW40248, HW40249, HW40250, HW40251, HW40252, HW40253, HW40254, HW40255, HW40256, HW40257, HW40258, HW40259, HW40260, HW40261, HW40262, HW40263, HW40264, HW40265, HW40266, HW40267, HW40268, HW40269, HW40270, HW40271, HW40272, HW40273, HW40274, HW40275, HW40276, HW40277, HW40278, HW40279, HW40280, HW40281, HW40282, HW40283, HW40284, HW40285, HW40286, HW40287, HW40288, HW40289, HW40290, HW40291, HW40292, HW40293, HW40294, HW40295, HW40296, HW40297, HW40298, HW40299, HW40300, HW40301, HW40302, HW40303, HW40304, HW40305, HW40306, HW40307, HW40308, HW40309, HW40310, HW40311, HW40312, HW40313, HW40314, HW40315, HW40316, HW40317, HW40318, HW40319, HW40320, HW40321, HW40322, HW40323, HW40324, HW40325, HW40326, HW40327, HW40328, HW40329, HW40330, HW40331, HW40332, HW40333, HW40334, HW40335, HW40337, HW40338, HW40339, HW40340, HW40341, HW40359, HW40360, HW40361, HW40362, HW40364, HW40371, HW40372, HW40373, HW40375, HW40376, HW40377, HW40378, HW40379, HW40380, HW40381, HW40382, HW40384, HW40385, HW40386, HW40387, HW40388, HW40389, HW40390, HW40391, HW40392, HW40393, HW40394, HW40395, HW40396, HW40397, HW40399, HW40400, HW40401, HW40402, HW40403, HW40404, HW40405, HW40406, HW40407, HW40408, HW40409, HW40410, HW40411, HW40432, HW40433, HW40434, HW40435, HW40436, HW40437, HW40438, HW40440, HW40441, HW40442, HW40443, HW40444, HW40445, HW40446, HW40447, HW40448, HW40449, HW40451, HW40452, HW40453, HW40454, HW40455, HW40456, HW40457, HW40460, HW40461, HW40462, HW40463, HW40464, HW40466, HW40467, HW40468, HW40469, HW40470, HW40481, HW40482, HW40483, HW40484, HW40485, HW40486, HW40487, HW40488, HW40489, HW40490, HW40491, HW40492, HW40493, HW40494, HW40496, HW40497, HW40498, HW40499, HW40500, HW40501, HW40502, HW40503, HW40504, HW40505, HW40506, HW40507, HW40508, HW40509, HW40510, HW40511, HW40512, HW40513, HW40514, HW40515, HW40516, HW40517, HW40518, HW40519, HW40520, HW40521, HW40522, HW40523, HW40524, HW40525, HW40526, HW40527, HW40528, HW40529, HW40530, HW40531, HW40532, HW40533, HW40534, HW40535, HW40536, HW40537, HW40538, HW40539, HW40540, HW40541, HW40542, HW40543, HW40544, HW40545, HW40546, HW40547, HW40548, HW40549, HW40550, HW40551, HW40558, HW40559, HW40560, HW40561, HW40562, HW40563, HW40566, HW40567, HW40568, HW40569, HW40570, HW40571, HW40572, HW40573, HW40574, HW40575, HW40576, HW40577, HW40578, HW40581, HW40582, HW40583, HW40584, HW40585, HW40589, HW40590, HW40591, HW40592, HW40593, HW40594, HW40595, HW40596, HW40597, HW40598, HW40601, HW40605, HW40606, HW40610, HW40616, HW40617, HW40620, HW40621, HW40622, HW40623, HW40627, HW40628, HW40629, HW40630, HW40631, HW40633, HW40634, HW40635, HW40636, HW40637, HW40638, HW40639, HW40640, HW40641, HW40642, HW40643, HW40644, HW40645, HW40646, HW40647, HW40648, HW40671, HW40672, HW40673, HW40674, HW40676, HW40677, HW40678, HW40679, HW40680, HW40681, HW40682, HW40683, HW40684, HW40685, HW40686, HW40687, HW40688, HW40689, HW40690, HW40691, HW40692, HW40693, HW40694, HW40695, HW40696, HW40697, HW40698, HW40699, HW40700, HW40703, HW40704, HW40705, HW40706, HW40707, HW40708, HW40709, HW40710, HW40711, HW40712, HW40713, HW40714, HW40715, HW40716, HW40717, HW40718, HW40719, HW40720, HW40721, HW40722, HW40723, HW40724, HW40725, HW40726, HW40727, HW40728, HW40729, HW40730, HW40731, HW40732, HW40733, HW40734, HW40735, HW40736, HW40737, HW40738, HW40739, HW40740, HW40741, HW40742, HW40743, HW40744, HW40745, HW40746, HW40747, HW40748, HW40749, HW40750, HW40751, HW40752, HW40753, HW40754, HW40755, HW40756, HW40757, HW40758, HW40759, HW40760, HW40761, HW40762, HW40763, HW40764, HW40766, HW40767, HW40768, HW40769, HW40770, HW40771, HW40772, HW40773, HW40774, HW40775, HW40776, HW40777, HW40778, HW40779, HW40780, HW40781, HW40783, HW40784, HW40785, HW40786, HW40790, HW40791, HW40792, HW40793, HW40794, HW40795, HW40796, HW40797, HW40798, HW40799, HW40800, HW40801, HW40802, HW40803, HW40804, HW40805, HW40806, HW40807, HW40808, HW40809, HW40810, HW40811, HW40812, HW40813, HW40814, HW40815, HW40817, HW40818, HW40819, HW40820, HW40821, HW40822, HW40823, HW40824, HW40825, HW40826, HW40827, HW40828, HW40829, HW40830, HW40831, HW40832, HW40833, HW40834, HW40837, HW40838, HW40839, HW40840, HW40841, HW40857, HW40858, HW40859, HW40860, HW40861, HW40862, HW40863, HW40864, HW40865, HW40866, HW40867, HW40868, HW40869, HW40870, HW40872, HW40873, HW40874, HW40875, HW40876, HW40877, HW40878, HW40879, HW40880, HW40882, HW40883, HW40884, HW40885, HW40886, HW40887, HW40888, HW40889, HW40890, HW40891, HW40892, HW40893, HW40894, HW40896, HW40897, HW40899, HW40902, HW40903, HW40904, HW40905, HW40906, HW40907, HW40908, HW40910, HW40911, HW40912, HW40913, HW40914, HW40915, HW40916, HW40917, HW40918, HW40919, HW40920, HW40921, HW40922, HW40923, HW40924, HW40925, HW40926, HW40927, HW40928, HW40929, HW40930, HW40931, HW40932, HW40933, HW40935, HW40936, HW40937, HW40939, HW40940, HW40941, HW40942, HW40943, HW40944, HW40946, HW40947, HW40948, HW40949, HW40962, HW40963, HW40964, HW40965, HW40966, HW40967, HW40968, HW40969, HW40970, HW40971, HW40972, HW40973, HW40974, HW40975, HW40976, HW40977, HW40979, HW40982, HW40983, HW40984, HW40985, HW40986, HW40987, HW40988, HW40989, HW40990, HW40991, HW40992, HW40993, HW40994, HW40995, HW40996, HW40997, HW40998, HW40999, HW41001, HW41002, HW41004, HW41005, HW41006, HW41007, HW41008, HW41009, HW41010, HW41011, HW41012, HW41014, HW41015, HW41016, HW41017, HW41018, HW41019, HW41020, HW41021, HW41022, HW41023, HW41024, HW41025, HW41027, HW41029, HW41030, HW41031, HW41032, HW41033, HW41034, HW41035, HW41037, HW41038, HW41039, HW41040, HW41042, HW41043, HW41044, HW41046, HW41051, HW41052, HW41053, HW41054, HW41055, HW41057, HW41058, HW41059, HW41060, HW41061, HW41062, HW41063, HW41064, HW41065, HW41066, HW41067, HW41068, HW41069, HW41070, HW41071, HW41072, HW41073, HW41075, HW41076, HW41077, HW41078, HW41080, HW41081, HW41084, HW41086, HW41088, HW41089, HW41090, HW41091, HW41092, HW41093, HW41097, HW41098, HW41099, HW41100, HW41102, HW41103, HW41104, HW41105, HW41106, HW41107, HW41108, HW41111, HW41112, HW41115, HW41118, HW41119, HW41120, HW41121, HW41122, HW41123, HW41124, HW41125, HW41126, HW41127, HW41128, HW41130, HW41131, HW41132, HW41133, HW41135, HW41136, HW41137, HW41138, HW41139, HW41140, HW41141, HW41142, HW41143, HW41144, HW41145, HW41146, HW41147, HW41148, HW41149, HW41150, HW41152, HW41153, HW41154, HW41155, HW41157, HW41158, HW41159, HW41160, HW41161, HW41162, HW41164, HW41165, HW41166, HW41167, HW41168, HW41169, HW41170, HW41171, HW41172, HW41173, HW41174, HW41176, HW41177, HW41178, HW41179, HW41180, HW41181, HW41182, HW41184, HW41185, HW41186, HW41187, HW41188, HW41189, HW41190, HW41193, HW41194, HW41195, HW41196, HW41197, HW41198, HW41199, HW41200, HW41201, HW41202, HW41203, HW41204, HW41205, HW41206, HW41207, HW41208, HW41210, HW41212, HW41213, HW41215, HW41216, HW41217, HW41218, HW41219, HW41220, HW41221, HW41223, HW41224, HW41225, HW41226, HW41227, HW41228, HW41229, HW41230, HW41231, HW41232, HW41233, HW41234, HW41235, HW41236, HW41237, HW41238, HW41239, HW41241, HW41242, HW41243, HW41244, HW41245, HW41246, HW41247, HW41248, HW41249, HW41250, HW41267, HW41268, HW41272, HW41273, HW41274, HW41275, HW41276, HW41277, HW41278, HW41279, HW41280, HW41281, HW41282, HW41283, HW41284, HW41285, HW41286, HW41287, HW41288, HW41289, HW41291, HW41293, HW41294, HW41295, HW41296, HW41297, HW41298, HW41299, HW41300, HW41301, HW41302, HW41303, HW41304, HW41305, HW41307, HW41310, HW41312, HW41318, HW41319, HW41320, HW41321, HW41322, HW41323, HW41324, HW41325, HW41326, HW41327, HW41329, HW41330, HW41331, HW41332, HW41333, HW41334, HW41336, HW41337, HW41338, HW41339, HW41341, HW41343, HW41344, HW41345, HW41346, HW41347, HW41348, HW41349, HW41350, HW41351, HW41352, HW41354, HW41355, HW41357, HW41358, HW41359, HW41360, HW41361, HW41362, HW41363, HW41364, HW41368, HW41369, HW41371, HW41372, HW41373, HW41374, HW41376, HW41377, HW41378, HW41379, HW41380, HW41381, HW41382, HW41383, HW41384, HW41385, HW41386, HW41387, HW41388, HW41389, HW41390, HW41391, HW41392, HW41393, HW41394, HW41395, HW41396, HW41398, HW41399, HW41400, HW41401, HW41402, HW41404, HW41405, HW41406, HW41407, HW41408, HW41409, HW41410, HW41412, HW41413, HW41414, HW41415, HW41416, HW41417, HW41418, HW41419, HW41420, HW41422, HW41423, HW41424, HW41425, HW41426, HW41427, HW41428, HW41429, HW41430, HW41431, HW41432, HW41433, HW41434, HW41435, HW41436, HW41437, HW41438, HW41439, HW41440, HW41441, HW41442, HW41443, HW41444, HW41445, HW41446, HW41447, HW41448, HW41449, HW41450, HW41452, HW41453, HW41454, HW41455, HW41456, HW41458, HW41459, HW41460, HW41461, HW41462, HW41463, HW41465, HW41466, HW41467, HW41468, HW41469, HW41470, HW41471, HW41472, HW41473, HW41474, HW41475, HW41476, HW41477, HW41478, HW41479, HW41480, HW41481, HW41482, HW41483, HW41484, HW41486, HW41488, HW41489, HW41490, HW41491, HW41492, HW41493, HW41494, HW41495, HW41496, HW41497, HW41498, HW41499, HW41500, HW41502, HW41503, HW41504, HW41505, HW41506, HW41507, HW41508, HW41509, HW41644, HW41645, HW41646, HW41648, HW41649, HW41650, HW41651, HW41653, HW41654
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
Product lines
Other devices under the same FDA submission
31 devices cleared under P100047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass4 recalls1 report
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall150 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85397).
- Z-2056-2020 · Open, ClassifiedHeartware HVAD Pump Implant Kit, Product (REF) Number 1104
- Z-2057-2020 · Open, ClassifiedHeartware HVAD Pump Implant Kit, Product (REF) Number 1104JP
- Z-2058-2020 · Open, ClassifiedHeartware HVAD Pump Outflow Graft, Product (REF) Number 1125
- Z-2059-2020 · Open, ClassifiedHeartware HVAD Pump Implant accessories, Product (REF) Number 1153
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page