Heartware HVAD Pump Implant Kit, Product (REF) Number 1104 recallZ-2056-2020 · Heartware, Inc. · initiated Apr 3, 2020
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heartware, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump.
FDA root cause category: Device Design
What the company is doing
The firm disseminated a Medical Device Safety Alert by letter beginning on 04/03/2020. The notice explained he potential for damage that can occur during the procedure. Specifically, the Outflow Graft may be subject to tears and the Strain Relief screw may break during the pre-implant pump assembly and attachment to the HVAD Pump. The letter also provided additional steps for assembly and attachment to reduce the risk of damage and tearing during the assembly procedure which was provided in Appendix A.
Recall details
- Recall number
- Z-2056-2020
- Status
- Open, Classified
- Initiated by the company
- Apr 3, 2020
- Recalling firm
- Heartware, Inc.
- Quantity
- 2225 Units
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution
Product description
Heartware HVAD Pump Implant Kit, Product (REF) Number 1104
Code information (models, lots, serial numbers)
Show the full code information (21k characters)
GTIN 00888707002028, Serial Numbers: HW28848, hw29011, HW29154, HW29211, HW29248, HW29249, HW29273, HW29329, HW29426, HW29465, HW29472, HW29522, HW29534, HW29548, HW29549, HW29567, HW29588, HW29593, HW29602, HW29612, HW29667, HW29700, HW29786, HW29815, HW29915, HW29928, HW29949, HW30039, HW30082, HW30083, HW30150, HW30153, HW30208, HW30211, HW30226, HW30320, HW30342, HW30358, HW30360, HW30363, HW30378, HW30380, HW30382, HW30390, HW30398, HW30400, HW30402, HW30406, HW30410, HW30430, HW30475, HW30484, HW30494, HW30498, HW30595, HW30596, HW30597, HW30601, HW30604, HW30651, HW30768, HW30769, HW30770, HW30771, HW30814, HW30821, HW30829, HW30863, HW30865, HW30867, HW30871, HW30953, HW30961, HW30962, HW30966, HW30971, HW30972, HW30974, HW31020, HW31076, HW31101, HW31103, HW31104, HW31107, HW31113, HW31115, HW31117, HW31119, HW31128, HW31132, HW31134, HW31192, HW31219, HW31223, HW31232, HW31235, HW31242, HW31244, HW31249, HW31254, HW31255, HW31260, HW31261, HW31262, HW31267, HW31383, HW31387, HW31388, HW31399, HW31409, HW31410, HW31411, HW31412, HW31414, hw31445, HW31446, hw31447, hw31448, hw31449, hw31451, HW31452, hw31453, hw31454, HW31455, HW31456, HW31457, HW31458, HW31541, HW31543, HW31544, HW31545, HW31547, HW31548, HW31549, HW31550, HW31552, HW31559, HW31564, HW31571, HW31573, HW31578, HW31579, HW31686, HW31694, HW31697, HW31731, HW31749, HW31750, HW31777, HW31778, HW31810, HW31819, HW31830, HW31831, HW31832, HW31834, HW31835, HW31836, HW31838, HW31839, HW31840, HW31842, HW31843, HW31844, HW31845, HW31846, HW31847, HW31849, HW31855, HW31858, HW31876, HW31883, HW31944, HW31947, HW31979, HW31981, HW31986, HW31987, HW31988, HW31989, HW31991, HW32068, HW32116, HW32121, HW32141, HW32143, HW32156, HW32159, HW32161, HW32162, HW32164, HW32167, HW32172, HW32179, HW32189, HW32190, HW32192, HW32193, HW32194, HW32197, HW32199, HW32201, HW32204, HW32206, HW32209, HW32210, HW32214, HW32215, HW32216, HW32218, HW32220, HW32222, HW32223, HW32224, HW32228, HW32229, HW32230, HW32231, HW32248, HW32258, HW32259, HW32260, HW32261, HW32273, HW32279, HW32281, HW32285, HW32286, HW32290, HW32292, HW32315, HW32335, HW32357, HW32386, HW32393, HW32400, HW32419, HW32423, HW32426, HW32427, HW32457, HW32458, HW32459, HW32464, HW32465, HW32468, HW32470, HW32471, HW32472, HW32474, HW32479, HW32499, HW32504, HW32508, HW32509, HW32510, HW32511, HW32516, HW32534, HW34001, HW34002, HW34025, HW34034, HW34035, HW34036, HW34037, HW34038, HW34040, HW34041, HW34042, HW34043, HW34044, HW34045, HW34046, HW34047, HW34048, HW34051, HW34066, HW34096, HW34100, HW34104, HW34106, HW34109, HW34110, HW34111, HW34113, HW34117, HW34121, HW34126, HW34151, HW34169, HW34179, HW34181, HW34186, HW34187, HW34192, HW34193, HW34196, HW34197, HW34202, HW34204, HW34205, HW34207, HW34209, HW34211, HW34226, HW34317, HW34319, HW34326, HW34328, HW34329, HW34330, HW34331, HW34336, HW34342, HW34344, HW34347, HW34368, HW34369, HW34371, HW34372, HW34373, HW34375, HW34376, HW34377, HW34378, HW34379, HW34425, HW34427, HW34428, HW34429, HW34432, HW34446, HW34495, HW34519, HW34520, HW34521, HW34522, HW34523, HW34546, HW34547, HW34550, HW34551, HW34555, HW34556, HW34588, HW34590, HW34596, HW34603, HW34605, HW34609, HW34610, HW34643, HW34673, HW34752, HW34768, HW34805, HW34806, HW34808, HW34810, HW34824, HW34840, HW34842, HW34845, HW34866, HW34868, HW34877, HW34902, HW34903, HW34904, HW34908, HW34910, HW34913, HW34914, HW34923, HW34924, HW34925, HW34928, HW34933, HW34934, HW34938, HW34939, HW34944, HW34946, HW34947, HW34948, HW34949, HW34957, HW34979, HW34990, HW34997, HW35012, HW35032, HW35099, HW35122, HW35123, HW35124, HW35125, HW35127, HW35128, HW35131, HW35133, HW35134, HW35136, HW35138, HW35141, HW35142, HW35143, HW35161, HW35189, HW35191, HW35204, HW35205, HW35209, HW35211, HW35212, HW35213, HW35214, HW35215, HW35220, HW35284, HW35305, HW35317, HW35319, HW35321, HW35324, HW35325, HW35326, HW35332, HW35334, HW35337, HW35339, HW35340, HW35341, HW35357, HW35360, HW35361, HW35367, HW35385, HW35464, HW35477, HW35478, HW35479, HW35480, HW35481, HW35482, HW35483, HW35487, HW35488, HW35489, HW35492, HW35493, HW35494, HW35495, HW35499, HW35500, HW35501, HW35502, HW35504, HW35506, HW35508, HW35534, HW35535, HW35539, HW35540, HW35541, HW35542, HW35543, HW35545, HW35548, HW35563, HW35577, HW35578, HW35580, HW35582, HW35583, HW35584, HW35586, HW35587, HW35588, HW35624, HW35625, HW35626, HW35627, HW35628, HW35629, HW35630, HW35631, HW35632, HW35633, HW35634, HW35635, HW35636, HW35637, HW35639, HW35641, HW35644, HW35646, HW35647, HW35668, HW35669, HW35670, HW35671, HW35673, HW35674, HW35676, HW35677, HW35678, HW35680, HW35681, HW35682, HW35683, HW35684, HW35686, HW35687, HW35688, HW35689, HW35690, HW35691, HW35692, HW35693, HW35695, HW35696, HW35697, HW35698, HW35701, HW35702, HW35703, HW35704, HW35705, HW35707, HW35708, HW35730, HW35731 GTIN 00888707002387, Serial Numbers: HW28851, HW29783, HW29793, HW29794, HW29827, HW29870, HW29918, HW30045, HW30046, HW30505, HW30765, HW30842, HW30844, HW30845, HW30846, hw30978, HW31075, HW31077, HW31080, hw31089, hw31091, HW31095, HW31161, HW31342, HW31348, HW31555, HW31556, HW31557, HW31683, HW31860, HW31903, HW31920, HW31924, HW31926, HW32026, HW32027, HW32028, HW32149, HW32153, HW32163, HW32282, HW32513, HW32514, HW32562, HW32563, HW34004, HW34006, HW34091, HW34092, HW34125, HW34272, HW34273, HW34274, HW34349, HW34350, HW34351, HW34352, HW34357, HW34358, HW34359, HW34366, HW34374, HW34436, HW34953, HW34968, HW35080, HW35144, HW35145, HW35148, HW35327, HW35342, HW35344, HW35537, HW35596, HW35603, HW35604, HW35605, HW35606, HW35607, HW35713, HW35714, HW35717, HW35718, HW35721, HW35723 GTIN 00888707001649, Serial Numbers: HW29132, HW29232, hw29653, HW29682, HW29684, HW29705, HW29864, HW29985, HW29991, HW30096, HW30100, HW30145, HW30146, HW30373, HW30374, HW30421, HW30525, HW30529, HW30726, HW30738, HW30836, HW30840, HW30862, HW30939, hw30949, hw30950, HW31048, HW31071, HW31082, HW31088, hw31133, hw31394, HW31415, HW31417, HW31420, HW31421, HW31423, hw31483, hw31505, hw31506, hw31507, hw31510, HW31511, hw31512, hw31513, hw31515, hw31516, hw31517, HW31546, HW31551, HW31554, HW31566, HW31587, HW31588, HW31589, HW31590, HW31592, HW31632, HW31639, HW31685, HW31795, HW31801, HW31806, HW31808, HW31809, HW31811, HW31812, HW31813, HW31815, HW31891, HW31892, HW31893, HW31919, HW31921, HW31973, HW31982, HW31985, HW32000, HW32001, HW32014, HW32015, HW32018, HW32021, HW32022, HW32024, HW32042, HW32050, HW32055, HW32056, HW32064, HW32081, HW32094, HW32112, HW32123, HW32138, HW32147, HW32186, HW32245, HW32255, HW32264, HW32265, HW32268, HW32270, HW32301, HW32394, HW32410, HW32420, HW32558, HW32559, HW32560, HW32561, HW34029, HW34031, HW34093, HW34094, HW34132, HW34148, HW34180, HW34266, HW34268, HW34269, HW34270, HW34271, HW34296, HW34301, HW34324, HW34327, HW34340, HW34341, HW34343, HW34345, HW34346, HW34348, HW34418, HW34420, HW34431, HW34507, HW34508, HW34641, HW34692, HW34745, HW34764, HW34836, HW34906, HW34932, HW34936, HW34940, HW34941, HW34963, HW34977, HW34992, HW34994, HW35019, HW35021, HW35033, HW35035, HW35038, HW35041, HW35065, HW35068, HW35085, HW35090, HW35094, HW35114, HW35147, HW35157, HW35196, HW35197, HW35210, HW35237, HW35250, HW35290, HW35308, HW35309, HW35310, HW35311, HW35312, HW35313, HW35314, HW35315, HW35318, HW35320, HW35328, HW35329, HW35330, HW35333, HW35338, HW35349, HW35365, HW35366, HW35368, HW35370, HW35371, HW35372, HW35373, HW35375, HW35378, HW35402, HW35404, HW35405, HW35408, HW35409, HW35411, HW35412, HW35418, HW35421, HW35423, HW35465, HW35470, HW35475, HW35476, HW35485, HW35486, HW35546, HW35547, HW35550, HW35551, HW35552, HW35553, HW35554, HW35555, HW35556, HW35557, HW35558, HW35559, HW35560, HW35564, HW35598, HW35599, HW35602, HW35640, HW35657, HW35658, HW35659, HW35662, HW35663, HW35664, HW35665, HW35666, HW35667, HW35672, HW35694, HW35719, HW35750, HW35753, HW36121, HW36154, HW36165, HW36168 GTIN 00888707000581, Serial Numbers: HW29203, HW29425, HW29427, HW29428.HW29430, HW29439, HW29517, HW29679, HW29680, HW30589, HW30638, HW30706, HW30876, HW31144, HW31187, hw31376, hw31380, HW31442, HW31727, HW31998, HW32023, HW32416, HW34017 GTIN 00888707002417, Serial Numbers: HW29478, HW30381, HW30831, HW34380, HW34421, HW35348, HW35613, HW35756, HW35761, HW35763 GTIN 00888707002493, Serial Numbers: HW29483, HW29697, HW29698, hw31178, HW31520, HW31821, HW31823, HW31827, HW32086, HW32242, HW32246, HW32518, HW32519, HW34021, HW34365, HW34367, HW34792, HW34915, HW35155, HW35158, HW35252, HW35343, HW35345, HW35346, HW35350, HW35400, HW35401, HW35529, HW35530, HW35574, HW35591, HW35660, HW35661, HW35785 GTIN 00888707002400, Serial Numbers: HW29518, HW29693, HW30092, HW32257, HW34012, HW34210, HW35034, HW35154, HW35573, HW35653 GTIN 00888707002547, Serial Numbers: HW29561, HW29601, HW29931, HW30502, HW30593, HW30619, HW30632, HW30783, HW30796, HW30819, HW30826, HW31136, HW31140, HW31141, HW31156, HW31157, HW31158, HW31382, HW31393, HW31406, HW31408, HW31413, HW31416, HW31426, HW31429, HW31553, HW31791, HW31937, HW31994, HW32118, HW32269, HW32309, HW32336, HW32347, HW32373, HW32385, HW32522, HW32523, HW32524, HW32554, HW32557, HW34003, HW34007, HW34008, HW34049, HW34050, HW34053, HW34054, HW34382, HW34383, HW34513, HW34514, HW34516, HW34517, HW34518, HW35251, HW35275 GTIN 00888707000680, Serial Numbers: HW29571, HW29578, HW29845, HW30635, HW30639, HW30711, HW31150 GTIN 00888707002394, Serial Numbers: HW30469, HW30506, HW30508, HW30509, HW30510, HW30511, HW30512, HW30655, HW31580, HW31582, HW31583, HW31584, HW31992, HW31993, HW32200, HW32207, HW32217, HW32238, HW32239, HW32240, HW32515, HW32517, HW34022, HW34023, HW34395, HW34461, HW34557, HW34561, HW34582, HW34751, HW34950, HW34955, HW34991, HW34996, HW35016, HW35149, HW35150, HW35186, HW35336, HW35386, HW35397, HW35484, HW35509, HW35510, HW35538, HW35566, HW35609, HW35610, HW35611, HW35648, HW35649, HW35651, HW35679, HW35699, HW35700, HW35706, HW35776, HW35779 GTIN 00888707000239, Serial Numbers: HW30634, HW30645, HW30877, HW34737 GTIN 00888707002530, Serial Numbers: HW30797, HW31108, HW31440, HW31828, HW31928, HW32182, HW32253, HW32263, HW34559, HW35000, HW35162, HW35532, HW35533, HW35590 GTIN 00888707002523, Serial Numbers: HW30859, HW31703, HW31704, HW31859, HW32030, HW32036, HW32040, HW32102, HW32520, HW32521, HW34018, HW34020, HW34027, HW34028, HW34986, HW35040, HW35043, HW35049, HW35216, HW35217, HW35218, HW35490, HW35531, HW35619, HW35786 GTIN 00888707002431, Serial Numbers: hw30979, hw31123, HW32029, HW32180, HW34024, HW34381, HW34615, HW35119, HW35733 GTIN 00888707002585, Serial Numbers: hw31079, hw31081, hw31168, hw31170, HW35095, HW35164, HW35351, HW35353, HW35355, HW35496, HW35497, HW35575, HW35576 GTIN 00888707002516, Serial Numbers: hw31147, hw31162, HW32507, HW34524, HW35474 GTIN 00888707006330, Serial Numbers: HW31419, HW31699, HW31733, HW32505, HW35393, HW35396, HW35726, HW35749, HW35766, HW35769, HW35784, HW35829, HW35872, HW36005, HW36007, HW36008, HW36009, HW36010, HW36013, HW36014, HW36015, HW36018, HW36019, HW36020, HW36021, HW36067, HW36068, HW36069, HW36070, HW36071, HW36131, HW36132, HW36138, HW36174, HW36175, HW36176, HW36177, HW36178, HW36179, HW36180, HW36182, HW36184, HW36185, HW36186, HW36192, HW36193, HW36194, HW36195, HW36196, HW36197, HW36198, HW36199, HW36200, HW36201, HW36224, HW36225, HW36226, HW36227, HW36228, HW36229, HW36230, HW36231, HW36232, HW36233, HW36248, HW36249, HW36251, HW36252, HW36253, HW36254, HW36255, HW36256, HW36258, HW36259, HW36260, HW36276, HW36298, HW36299, HW36300, HW36303, HW36305, HW36306, HW36307, HW36309, HW36310, HW36313, HW36314, HW36316, HW36317, HW36318, HW36319, HW36320, HW36321, HW36323, HW36324, HW36325, HW36326, HW36328, HW36332, HW36333, HW36334, HW36335, HW36337, HW36354, HW36355, HW36356, HW36357, HW36358, HW36359, HW36360, HW36361, HW36362, HW36363, HW36396, HW36397, HW36398, HW36399, HW36400, HW36401, HW36403, HW36404, HW36405, HW36406, HW36407, HW36408, HW36409, HW36410, HW36411, HW36412, HW36419, HW36420, HW36428, HW36429, HW36430, HW36432, HW36433, HW36434, HW36435, HW36436, HW36437, HW36438, HW36500, HW36501, HW36502, HW36503, HW36504, HW36505, HW36506, HW36507, HW36508, HW36509, HW36510, HW36511, HW36513, HW36514, HW36515, HW36530, HW36553, HW36554, HW36555, HW36556, HW36557, HW36558, HW36559, HW36560, HW36561, HW36562, HW36563, HW36564, HW36595, HW36596, HW36597, HW36598, HW36599, HW36600, HW36601, HW36602, HW36603, HW36624, HW36625, HW36626, HW36628, HW36630, HW36631, HW36632, HW36633, HW36636, HW36637, HW36655, HW36659, HW36660, HW36662, HW36663, HW36664, HW36669, HW36670, HW36671, HW36672, HW36703, HW36704, HW36705, HW36708, HW36709, HW36710, HW36711, HW36712, HW36713, HW36853, HW36854, HW36855, HW36898, HW36899, HW36900, HW36903, HW36919, HW36921 GTIN 00888707002486, Serial Numbers: HW31444, HW31569, HW31673, HW31814, HW31817, HW31818, HW31820, HW31824, HW31825, HW31826, HW31872, HW31873, HW31875, HW31877, HW31927, HW31932, HW31934, HW31939, HW31940, HW31941, HW31995, HW31996, HW31997, HW31999, HW32002, HW32008, HW32009, HW32010, HW32011, HW32013, HW32032, HW32034, HW32038, HW32046, HW32053, HW32057, HW32062, HW32065, HW32071, HW32085, HW32109, HW32165, HW32166, HW32195, HW32205, HW32211, HW32221, HW32236, HW32243, HW32250, HW32506, HW34015, HW34122, HW34123, HW34124, HW34251, HW34252, HW34275, HW34277, HW34280, HW34354, HW34355, HW34356, HW34361, HW34364, HW34433, HW34434, HW34437, HW34439, HW34440, HW34442, HW34443, HW34444, HW34452, HW34454, HW34619, HW34620, HW34621, HW34622, HW34623, HW34624, HW34629, HW34630, HW34631, HW34636, HW34656, HW34719, HW34738, HW34739, HW34742, HW34748, HW34749, HW34750, HW34753, HW34757, HW34758, HW34761, HW34762, HW34767, HW34769, HW34779, HW34818, HW34853, HW34867, HW34873, HW34879, HW34880, HW34882, HW34885, HW34888, HW34890, HW34897, HW34898, HW34900, HW34905, HW34909, HW34911, HW34912, HW34945, HW34954, HW34960, HW35171, HW35228, HW35230, HW35278, HW35282, HW35283, HW35297, HW35298, HW35300, HW35303, HW35304, HW35306, HW35307, HW35512, HW35513, HW35514, HW35516, HW35517, HW35518, HW35519, HW35520, HW35521, HW35522, HW35523, HW35525, HW35526, HW35527, HW35528, HW35565, HW35567, HW35568, HW35569, HW35570, HW35593, HW35600, HW35612, HW35614, HW35615, HW35616, HW35617, HW35618, HW35645, HW35650, HW35652, HW35685, HW35725, HW35734, HW35737, HW35738, HW35740, HW35742, HW35743, HW35744, HW35745, HW35746, HW35747, HW35748, HW35751, HW35752, HW35760, HW35762, HW35765 GTIN 00888707002462, Serial Numbers: HW31586, HW31700, HW32035 GTIN 00888707002424, Serial Numbers: HW31698, HW31851 GTIN 00888707002448, Serial Numbers: HW31779, HW31780, HW32546, HW34095, HW34617, HW35398, HW35511, HW35589, HW35597 GTIN 00888707002455, Serial Numbers: HW31852, HW31856, HW32025, HW32198, HW34026, HW34637, HW35544 GTIN 00888707006019, Serial Numbers: HW31983, HW34720, HW34974, HW35376, HW35377, HW35849, HW35866, HW35869, HW35871, HW35873, HW35877, HW35878, HW35879, HW35880, HW35998, HW36000, HW36056, HW36057, HW36058, HW36059, HW36060, HW36061, HW36062, HW36063, HW36113, HW36114, HW36115, HW36116, HW36117, HW36118, HW36119, HW36120, HW36145, HW36146, HW36155, HW36156, HW36157, HW36158, HW36159, HW36167, HW36221, HW36222, HW36223, HW36350, HW36351, HW36352 GTIN 00888707006408, Serial Numbers: HW34292, HW35932, HW36083, HW36084, HW36087, HW36129, HW36130, HW36171, HW36172, HW36173, HW36212, HW36247, HW36250, HW36261, HW36263, HW36264, HW36265, HW36266, HW36348, HW36349, HW36413, HW36414, HW36415, HW36416, HW36594, HW36638 GTIN 00888707002554, Serial Numbers: HW34337, HW34362, HW34484, HW34591, HW34594, HW34625, HW34682, HW34693, HW34732, HW34833, HW34838, HW34846, HW34876, HW34884, HW34889, HW35052, HW35057, HW35061, HW35067, HW35316, HW35323, HW35359, HW35362, HW35363, HW35364, HW35458, HW35536, HW35585, HW35620, HW35622, HW35623, HW35709, HW35710, HW35728, HW35732, HW35758, HW35759, HW35764, HW35767, HW35771, HW35772, HW35775, HW35777, HW35780, HW35782 GTIN 00888707004558, Serial Numbers: HW34759, HW34785, HW35655, HW35656 GTIN 00888707002141, Serial Numbers: HW34848, HW34849, HW35711 GTIN 00888707006323, Serial Numbers: HW34982, HW35031, HW35379, HW35380, HW35381, HW35382, HW35383, HW35384, HW35388, HW35389, HW35390, HW35391, HW35392, HW35621, HW35642, HW35654, HW35712, HW35722, HW35727, HW35729, HW35736, HW35755, HW35770, HW35781, HW35783, HW35977, HW36024, HW36025, HW36027, HW36028, HW36029, HW36030, HW36032, HW36033, HW36034, HW36042, HW36043, HW36044, HW36045, HW36046, HW36064, HW36065, HW36072, HW36073, HW36074, HW36075, HW36076, HW36077, HW36078, HW36079, HW36080, HW36081, HW36082, HW36123, HW36124, HW36125, HW36126, HW36127, HW36128, HW36133, HW36134, HW36135, HW36136, HW36137, HW36139, HW36140, HW36141, HW36142, HW36143, HW36144, HW36147, HW36148, HW36149, HW36150, HW36151, HW36152, HW36153, HW36187, HW36188, HW36189, HW36190, HW36191, HW36202, HW36203, HW36204, HW36206, HW36207, HW36208, HW36209, HW36210, HW36234, HW36235, HW36236, HW36237, HW36238, HW36239, HW36242, HW36244, HW36245, HW36267, HW36268, HW36271, HW36272, HW36274, HW36275, HW36284, HW36286, HW36290, HW36291, HW36292, HW36293, HW36295, HW36296, HW36297, HW36339, HW36340, HW36342, HW36343, HW36344, HW36346, HW36364, HW36366, HW36367, HW36375, HW36378, HW36379, HW36380, HW36439, HW36440, HW36441, HW36443, HW36444, HW36445, HW36446, HW36450, HW36451, HW36452, HW36453, HW36454, HW36455, HW36456, HW36520, HW36521, HW36522, HW36524, HW36528, HW36529, HW36532, HW36533, HW36534, HW36535, HW36537, HW36539, HW36548, HW36549, HW36550, HW36551, HW36552, HW36604, HW36605, HW36606, HW36607, HW36608, HW36639, HW36640, HW36641, HW36645, HW36649, HW36651, HW36661, HW36674, HW36675, HW36676, HW36687, HW36714, HW36729, HW36731, HW36732, HW36733, HW36734, HW36735, HW36737, HW36738, HW36739, HW36744, HW36908, HW36909, HW36922, HW36923, HW36925, HW36932 GTIN 00888707005951, Serial Numbers: HW35279, HW35331, HW35503, HW35524, HW35581, HW35675, HW35715, HW35720, HW35739, HW35768, HW35787, HW35788, HW35789, HW35790, HW35791, HW35792, HW35793, HW35794, HW35795, HW35796, HW35797, HW35798, HW35799, HW35821, HW35822, HW35823, HW35824, HW35825, HW35826, HW35827, HW35831, HW35833, HW35834, HW35835, HW35836, HW35837, HW35838, HW35839, HW35840, HW35841, HW35842, HW35843, HW35844, HW35845, HW35846, HW35883, HW35887, HW35888, HW35889, HW35890, HW35891, HW35892, HW35893, HW35894, HW35895, HW35896, HW35897, HW35898, HW35900, HW35905, HW35906, HW35907, HW35921, HW35925, HW35926, HW35928, HW35930, HW35933, HW35934, HW35945, HW35946, HW35947, HW35948, HW35949, HW35950, HW35951, HW35952, HW35953, HW35954, HW35955, HW35956, HW35957, HW35959, HW35960, HW35961 GTIN 00888707002264, Serial Numbers: HW35741, HW35773 GTIN 00888707006316, Serial Numbers: HW35754, HW35774, HW35865, HW35868, HW35874, HW35940, HW35958, HW35962, HW35963, HW35964, HW35965 GTIN 00888707005968, Serial Numbers: HW35801, HW35802, HW35803, HW35804, HW35805, HW35806, HW35807, HW35808, HW35809, HW35810, HW35811, HW35812, HW35813, HW35814, HW35815, HW35816, HW35817, HW35818, HW35819, HW35820, HW35847, HW35848, HW35850, HW35851, HW35852, HW35853, HW35854, HW35855, HW35856, HW35857, HW35858, HW35859, HW35860, HW35861, HW35862, HW35863, HW35864, HW35867, HW35870, HW35875, HW35923, HW35924, HW35927, HW35929, HW35931, HW35935, HW35936, HW35937, HW35938, HW35939, HW35966, HW35967, HW35968, HW35969, HW35970, HW35971, HW35972, HW35973, HW35975, HW35976, HW35978, HW35988, HW35989, HW35990, HW35991, HW35993, HW35994, HW35995, HW35996, HW35997 GTIN 00888707006309, Serial Numbers: HW35876, HW35882, HW35884, HW35885, HW35886, HW35941, HW35942, HW35999 GTIN 00888707006347, Serial Numbers: HW35902, HW35903, HW35908, HW35909, HW35910, HW35911, HW35912, HW35913, HW35914, HW35915, HW35918, HW35920, HW36001, HW36002, HW36003, HW36016, HW36035, HW36036, HW36037, HW36038, HW36039, HW36040, HW36048, HW36049, HW36050, HW36051, HW36052, HW36053, HW36054, HW36055, HW36088, HW36089, HW36092, HW36093, HW36094, HW36095, HW36096, HW36097, HW36100, HW36101, HW36102, HW36103, HW36105, HW36106, HW36107, HW36108, HW36109, HW36110, HW36111, HW36112, HW36170, HW36540, HW36541, HW36542, HW36543, HW36544, HW36565, HW36566, HW36568, HW36569, HW36570, HW36571, HW36572, HW36573, HW36574, HW36575, HW36576, HW36577, HW36578, HW36579, HW36581, HW36582, HW36583, HW36584, HW36585, HW36586, HW36611, HW36612, HW36615, HW36616, HW36618, HW36619, HW36620, HW36621, HW36622, HW36623 GTIN 00888707006446, Serial Numbers: HW35979, HW35980, HW35982, HW35984, HW36160, HW36161, HW36162, HW36163, HW36164, HW36278, HW36287, HW36381, HW36382, HW36383, HW36384, HW36385, HW36386, HW36387, HW36388, HW36460, HW36462, HW36464, HW36465, HW36469, HW36470, HW36471, HW36472, HW36480, HW36481, HW36482, HW36483, HW36484, HW36486, HW36487, HW36489, HW36491, HW36492, HW36494, HW36495 GTIN 00888707006392, Serial Numbers: HW35986, HW35987, HW36066, HW36085, HW36086, HW36213, HW36214, HW36215, HW36216, HW36217, HW36218, HW36219, HW36220, HW36289, HW36311, HW36417, HW36418, HW36517, HW36587, HW36588, HW36589, HW36590, HW36652, HW36656, HW36860 GTIN 00888707007290, Serial Numbers: HW36421, HW36422 GTIN 00888707006354, Serial Numbers: HW36838, HW36880 No GTIN, Serial Numbers: HW26099, HW26126, HW26330, HW26660, HW26939, HW26940, HW27099, HW27452, HW27678, HW28063, HW28199, HW28373, HW28418, HW28895, HW28973, HW29073, HW29214, HW29298, HW29318, HW29351, HW29353, HW29354, HW29359
Other devices under the same FDA submission
33 devices cleared under P100047 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 85397).
- Z-2055-2020 · Open, ClassifiedHeartware HVAD Pump Implant Kit, Product (REF) Number 1103
- Z-2057-2020 · Open, ClassifiedHeartware HVAD Pump Implant Kit, Product (REF) Number 1104JP
- Z-2058-2020 · Open, ClassifiedHeartware HVAD Pump Outflow Graft, Product (REF) Number 1125
- Z-2059-2020 · Open, ClassifiedHeartware HVAD Pump Implant accessories, Product (REF) Number 1153
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page