INLIVEN CRT-P, Model Numbers recallZ-2020-2021 · Boston Scientific Corporation · initiated Jun 3, 2021
Class I Open, Classified All recalls posted in July 2021
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-Ps) may initiate Safety Mode later in device life (i.e., prior to reaching the Explant battery indicator) when the device's battery exhibits high internal impedance. This latent battery condition puts a device at risk for system resets to occur due to temporary high-power consumption related to telemetry attempts and subsequent reversion to Safety Mode to maintain back-up pacing. Although therapy is still provided when a device is in Safety Mode, replacement is required. - Only dual chamber INGENIO family pacemakers and CRT-Ps built with the Extended Life (EL) battery are included within this advisory population
FDA root cause category: Device Design
What the company is doing
The firm issued an advisory communication on June 3, 2021 by letter to the implanting/following physicians and clinics regarding the potential device behavior. The notice included a letter to the patients to be disseminated to the patients implanted with the device at the discretion of the physician. BSC also made the patient letter available on their advisory page, found on our product performance website @ https://www.bostonscientific.com/en-US/pprc/product-advisories.html
Recall details
- Recall number
- Z-2020-2021
- Recall class
- Class I (most serious)
- Status
- Open, Classified
- Initiated by the company
- Jun 3, 2021
- Posted by the FDA
- Jul 13, 2021
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 3634 units
- Product code
- NKE
- FDA submissions
- P030005
- Distribution
- Worldwide distribution.
Product description
INLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)
Code information (models, lots, serial numbers)
Show the full code information (29k characters)
a) V284, Serial Numbers: 100002, 100000, 100003, 100005, 100009, 100010, 100011, 100012, 100014, 100015, 100016, 100017, 100018, 100019, 100022, 100023, 100027, 100029, 234350, 234352 b) V285, Serial Numbers: 246521, 100000, 100002, 100007, 100018, 100026, 100036, 100059, 100065, 100087, 100097, 100101, 100111, 100117, 100126, 100127, 100136, 100139, 100141, 100154, 100156, 100158, 100159, 100161, 100162, 100163, 246441, 246442, 246446, 246447, 246449, 246451, 246453, 246458, 246459, 246463, 246469, 246470, 246471, 246472, 246473, 246474, 246475, 246477, 246481, 246483, 246484, 246485, 246486, 246495, 246496, 246497, 246499, 246500, 246501, 246503, 246504, 246505, 246506, 246508, 246509, 246512, 246519, 246520, 246521, 246527, 246529, 246532, 246534, 246542, 246546, 246549, 246554, 246577, 246586, 246590, 246591, 246592, 246593, 246596, 246598, 246610, 246611, 246613, 246619, 246623, 246627, 246628, 246630, 246634, 246636, 246637, 246645, 246646, 246647, 246650, 246654, 246656, 246658, 246664, 246665, 246667, 246672, 246673, 246677, 246679, 246681, 246683, 246686, 246688, 246690, 246693, 246694, 246697, 246704, 246709, 246710, 246713, 246714, 246721, 246730, 246748, 246752, 246756, 246759, 246762, 246764, 246765, 246766, 246769, 100003, 100004, 100005, 100008, 100009, 100010, 100012, 100013, 100015, 100016, 100017, 100019, 100020, 100021, 100023, 100025, 100027, 100028, 100029, 100030, 100031, 100032, 100033, 100035, 100037, 100038, 100039, 100040, 100042, 100043, 100046, 100047, 100048, 100049, 100050, 100051, 100052, 100053, 100054, 100055, 100056, 100057, 100058, 100060, 100061, 100062, 100066, 100067, 100068, 100069, 100071, 100072, 100074, 100075, 100076, 100077, 100079, 100080, 100081, 100082, 100083, 100084, 100085, 100086, 100091, 100093, 100095, 100098, 100100, 100102, 100103, 100104, 100105, 100106, 100107, 100108, 100109, 100110, 100112, 100114, 100115, 100116, 100118, 100119, 100120, 100121, 100122, 100123, 100124, 100125, 100129, 100130, 100131, 100132, 100133, 100134, 100135, 100137, 100138, 100140, 100142, 100143, 100144, 100145, 100146, 100147, 100148, 100149, 100150, 100151, 100152, 100153, 100155, 100157, 100160, 100164, 246436, 246437, 246438, 246439, 246440, 246443, 246444, 246445, 246448, 246450, 246452, 246454, 246455, 246456, 246460, 246461, 246462, 246464, 246465, 246466, 246467, 246476, 246478, 246480, 246482, 246487, 246488, 246489, 246490, 246491, 246492, 246493, 246494, 246498, 246502, 246507, 246510, 246511, 246514, 246515, 246516, 246518, 246522, 246524, 246525, 246526, 246528, 246530, 246535, 246541, 246543, 246544, 246545, 246548, 246550, 246551, 246552, 246555, 246556, 246558, 246559, 246560, 246561, 246564, 246566, 246567, 246568, 246569, 246570, 246571, 246572, 246573, 246574, 246575, 246576, 246578, 246579, 246580, 246581, 246582, 246583, 246584, 246585, 246587, 246588, 246589, 246594, 246595, 246597, 246602, 246603, 246604, 246606, 246607, 246608, 246609, 246614, 246615, 246616, 246617, 246618, 246620, 246621, 246622, 246624, 246625, 246626, 246629, 246631, 246633, 246635, 246640, 246641, 246642, 246643, 246644, 246649, 246651, 246652, 246653, 246655, 246657, 246659, 246660, 246661, 246662, 246663, 246668, 246670, 246675, 246676, 246678, 246680, 246682, 246684, 246685, 246687, 246691, 246692, 246695, 246696, 246698, 246699, 246700, 246701, 246702, 246703, 246705, 246706, 246708, 246711, 246712, 246715, 246717, 246718, 246719, 246720, 246723, 246724, 246725, 246726, 246728, 246729, 246731, 246732, 246733, 246734, 246736, 246738, 246739, 246740, 246741, 246742, 246743, 246744, 246745, 246746, 246749, 246750, 246751, 246755, 246757, 246758, 246767 c) W274, Serial Numbers: 111154, 111189, 111247, 111112, 111113, 111114, 111115, 111118, 111119, 111120, 111121, 111122, 111124, 111125, 111127, 111128, 111131, 111133, 111134, 111136, 111138, 111139, 111140, 111142, 111143, 111144, 111145, 111147, 111148, 111149, 111150, 111151, 111152, 111155, 111158, 111159, 111163, 111164, 111165, 111166, 111167, 111168, 111169, 111171, 111173, 111174, 111175, 111176, 111177, 111180, 111181, 111184, 111185, 111186, 111187, 111188, 111191, 111192, 111193, 111195, 111196, 111197, 111198, 111199, 111203, 111204, 111205, 111206, 111208, 111209, 111211, 111213, 111214, 111215, 111216, 111220, 111221, 111222, 111225, 111226, 111227, 111230, 111231, 111232, 111233, 111234, 111236, 111237, 111241, 111242, 111244, 111249, 111250, 111251, 111256, 111257, 111261, 111262, 111264, 111267, 111268, 111270, 111271, 111272, 111275, 111276, 111278, 111279, 111281, 111282, 111284, 111285, 111288, 111289, 111293, 111294, 111299, 111300, 111301, 111302, 111304, 111305, 111307, 111312, 111314, 111316, 111317, 111323, 111327, 111328, 111332, 111334, 111336, 111337, 111338, 111341, 111343, 111345, 111346, 111347, 111348, 111349, 111350, 111351, 111356, 111357, 111358, 111359, 111360, 111362, 111363, 111365, 111366, 111367, 111368, 111369, 111370, 111372, 111373, 111374, 111378, 111380, 111381, 111382, 111385, 111387, 111388, 111389, 111390, 111391, 300711, 300713, 300714, 300716, 300717, 300718, 300719, 300731 d) W275, Serial Numbers: 111149, 111152, 111164, 111166, 111178, 111189, 111190, 111207, 111217, 111218, 111230, 111256, 111287, 111310, 111321, 111363, 111373, 111379, 111394, 111397, 111405, 111407, 111417, 111423, 111441, 111466, 111472, 111484, 111486, 111489, 111516, 111519, 111524, 111536, 111545, 111559, 111566, 111593, 111609, 111613, 111692, 111703, 111709, 111765, 111773, 111820, 111826, 111842, 111883, 111959, 111974, 111991, 112097, 112159, 112165, 112194, 112226, 112284, 112367, 112459, 112469, 112473, 112497, 112500, 112574, 112578, 112642, 112644, 112685, 112691, 112710, 112711, 112724, 112958, 112982, 112994, 112999, 113042, 113059, 113061, 113126, 113137, 113144, 113175, 113192, 113295, 113297, 113302, 113322, 113377, 113416, 113444, 113550, 300827, 300843, 300955, 301017, 301059, 301132, 301155, 301185, 301256, 301259, 301261, 301263, 301264, 301265, 301267, 301270, 301271, 301272, 301273, 301275, 301277, 301281, 301282, 301284, 301289, 301293, 301296, 301300, 301302, 301305, 301306, 301363, 301364, 301371, 301377, 301398, 301478, 301480, 301491, 301494, 111119, 111120, 111121, 111122, 111123, 111124, 111125, 111127, 111128, 111129, 111130, 111131, 111132, 111133, 111135, 111136, 111137, 111138, 111139, 111140, 111141, 111142, 111143, 111144, 111145, 111146, 111147, 111148, 111150, 111151, 111153, 111154, 111155, 111156, 111157, 111158, 111159, 111160, 111161, 111162, 111163, 111165, 111167, 111168, 111169, 111170, 111171, 111173, 111174, 111175, 111176, 111177, 111179, 111180, 111181, 111182, 111183, 111184, 111186, 111187, 111188, 111191, 111193, 111194, 111195, 111196, 111197, 111198, 111199, 111200, 111201, 111202, 111203, 111205, 111206, 111208, 111209, 111210, 111211, 111212, 111214, 111215, 111216, 111219, 111220, 111221, 111222, 111223, 111224, 111225, 111226, 111227, 111228, 111229, 111231, 111233, 111234, 111235, 111236, 111237, 111238, 111239, 111240, 111241, 111242, 111243, 111244, 111245, 111246, 111247, 111248, 111249, 111250, 111251, 111252, 111253, 111255, 111257, 111258, 111260, 111263, 111265, 111267, 111268, 111269, 111270, 111271, 111272, 111273, 111274, 111275, 111276, 111277, 111279, 111281, 111283, 111284, 111285, 111286, 111289, 111291, 111293, 111294, 111296, 111297, 111298, 111300, 111301, 111303, 111304, 111307, 111308, 111309, 111311, 111312, 111313, 111315, 111316, 111317, 111318, 111319, 111320, 111322, 111323, 111325, 111326, 111327, 111328, 111329, 111331, 111332, 111333, 111334, 111335, 111336, 111337, 111338, 111339, 111342, 111343, 111344, 111345, 111346, 111348, 111349, 111350, 111351, 111352, 111353, 111354, 111355, 111356, 111357, 111359, 111360, 111361, 111362, 111364, 111366, 111367, 111368, 111369, 111370, 111371, 111372, 111376, 111378, 111380, 111381, 111382, 111383, 111384, 111386, 111387, 111388, 111389, 111390, 111391, 111392, 111396, 111398, 111399, 111400, 111401, 111403, 111404, 111406, 111408, 111409, 111410, 111411, 111412, 111414, 111415, 111416, 111419, 111420, 111421, 111422, 111424, 111425, 111427, 111428, 111429, 111430, 111431, 111432, 111433, 111434, 111435, 111436, 111437, 111439, 111440, 111442, 111443, 111444, 111445, 111446, 111447, 111448, 111449, 111450, 111451, 111452, 111453, 111454, 111455, 111456, 111457, 111459, 111460, 111461, 111462, 111463, 111464, 111465, 111467, 111468, 111469, 111471, 111474, 111475, 111476, 111477, 111478, 111479, 111480, 111481, 111483, 111485, 111488, 111490, 111491, 111492, 111493, 111495, 111496, 111498, 111499, 111500, 111501, 111503, 111504, 111505, 111506, 111507, 111508, 111509, 111510, 111511, 111512, 111513, 111514, 111515, 111517, 111518, 111520, 111521, 111522, 111523, 111525, 111528, 111529, 111530, 111532, 111533, 111534, 111535, 111539, 111541, 111542, 111543, 111544, 111546, 111547, 111548, 111549, 111550, 111551, 111552, 111553, 111554, 111555, 111556, 111557, 111558, 111560, 111561, 111562, 111564, 111565, 111567, 111568, 111569, 111570, 111571, 111572, 111574, 111576, 111578, 111579, 111580, 111581, 111582, 111583, 111585, 111586, 111587, 111588, 111590, 111591, 111594, 111595, 111596, 111597, 111598, 111599, 111600, 111601, 111602, 111604, 111605, 111606, 111607, 111608, 111610, 111612, 111614, 111618, 111619, 111620, 111621, 111622, 111623, 111624, 111626, 111627, 111628, 111629, 111630, 111631, 111632, 111633, 111635, 111636, 111637, 111638, 111639, 111640, 111641, 111643, 111644, 111645, 111647, 111648, 111649, 111652, 111653, 111654, 111655, 111656, 111658, 111659, 111660, 111661, 111662, 111663, 111664, 111665, 111666, 111668, 111669, 111670, 111671, 111672, 111673, 111675, 111676, 111677, 111678, 111679, 111680, 111682, 111684, 111685, 111686, 111687, 111688, 111689, 111690, 111693, 111695, 111696, 111697, 111699, 111700, 111701, 111704, 111705, 111706, 111707, 111710, 111711, 111712, 111713, 111714, 111715, 111716, 111717, 111718, 111719, 111720, 111721, 111722, 111723, 111724, 111725, 111726, 111727, 111729, 111730, 111731, 111732, 111733, 111734, 111735, 111736, 111737, 111738, 111739, 111740, 111741, 111742, 111743, 111744, 111745, 111746, 111747, 111748, 111749, 111750, 111751, 111752, 111753, 111754, 111755, 111756, 111757, 111758, 111759, 111760, 111761, 111762, 111763, 111764, 111766, 111767, 111768, 111770, 111771, 111772, 111774, 111775, 111776, 111777, 111778, 111779, 111780, 111781, 111782, 111783, 111784, 111785, 111787, 111788, 111789, 111790, 111792, 111793, 111794, 111795, 111796, 111798, 111799, 111800, 111801, 111802, 111803, 111804, 111805, 111806, 111808, 111809, 111810, 111811, 111812, 111813, 111814, 111815, 111816, 111817, 111818, 111819, 111821, 111822, 111823, 111824, 111825, 111827, 111828, 111829, 111830, 111831, 111833, 111834, 111835, 111836, 111837, 111838, 111839, 111840, 111841, 111843, 111845, 111846, 111847, 111849, 111850, 111851, 111852, 111853, 111855, 111856, 111857, 111859, 111860, 111861, 111862, 111863, 111864, 111866, 111867, 111868, 111869, 111870, 111871, 111872, 111873, 111874, 111875, 111876, 111877, 111878, 111879, 111880, 111881, 111882, 111884, 111885, 111886, 111888, 111889, 111890, 111891, 111893, 111894, 111895, 111896, 111897, 111898, 111899, 111900, 111902, 111903, 111904, 111905, 111906, 111907, 111908, 111909, 111910, 111911, 111912, 111913, 111914, 111915, 111916, 111917, 111918, 111920, 111921, 111922, 111923, 111924, 111925, 111926, 111928, 111930, 111931, 111932, 111933, 111934, 111935, 111936, 111937, 111938, 111939, 111940, 111941, 111942, 111943, 111944, 111945, 111946, 111948, 111949, 111951, 111952, 111953, 111954, 111955, 111956, 111957, 111958, 111960, 111961, 111963, 111964, 111965, 111966, 111967, 111968, 111969, 111970, 111971, 111972, 111973, 111975, 111976, 111977, 111978, 111979, 111980, 111981, 111982, 111983, 111984, 111985, 111986, 111987, 111989, 111990, 111992, 111993, 111994, 111995, 111996, 111997, 111999, 112000, 112001, 112002, 112003, 112004, 112005, 112006, 112008, 112009, 112010, 112011, 112012, 112013, 112014, 112015, 112016, 112017, 112018, 112019, 112020, 112021, 112022, 112023, 112024, 112025, 112026, 112027, 112028, 112029, 112030, 112031, 112032, 112033, 112034, 112035, 112037, 112038, 112039, 112040, 112041, 112042, 112043, 112044, 112045, 112046, 112047, 112048, 112049, 112050, 112052, 112053, 112054, 112055, 112056, 112057, 112058, 112059, 112060, 112061, 112062, 112063, 112064, 112065, 112066, 112067, 112068, 112069, 112070, 112071, 112072, 112073, 112074, 112075, 112076, 112077, 112078, 112079, 112080, 112081, 112082, 112083, 112084, 112085, 112086, 112087, 112088, 112089, 112090, 112091, 112092, 112093, 112094, 112095, 112096, 112098, 112099, 112100, 112101, 112102, 112103, 112105, 112106, 112107, 112108, 112109, 112110, 112111, 112112, 112113, 112114, 112115, 112116, 112117, 112118, 112119, 112120, 112121, 112122, 112123, 112124, 112125, 112126, 112127, 112128, 112130, 112131, 112133, 112134, 112135, 112136, 112137, 112139, 112140, 112141, 112142, 112143, 112144, 112145, 112146, 112147, 112148, 112149, 112151, 112152, 112153, 112154, 112155, 112156, 112157, 112158, 112160, 112161, 112162, 112163, 112164, 112166, 112167, 112168, 112169, 112170, 112171, 112172, 112173, 112175, 112176, 112177, 112178, 112179, 112180, 112181, 112182, 112183, 112184, 112185, 112186, 112187, 112188, 112190, 112191, 112192, 112195, 112196, 112197, 112198, 112199, 112200, 112201, 112202, 112203, 112204, 112205, 112206, 112207, 112208, 112209, 112210, 112211, 112212, 112213, 112214, 112215, 112216, 112217, 112218, 112219, 112220, 112221, 112222, 112223, 112224, 112225, 112227, 112228, 112229, 112230, 112231, 112232, 112233, 112234, 112236, 112237, 112238, 112239, 112240, 112241, 112242, 112243, 112244, 112245, 112246, 112247, 112248, 112249, 112250, 112251, 112252, 112253, 112254, 112255, 112256, 112257, 112258, 112260, 112261, 112262, 112263, 112264, 112265, 112266, 112267, 112268, 112269, 112270, 112271, 112272, 112273, 112274, 112275, 112276, 112277, 112278, 112279, 112280, 112281, 112282, 112283, 112285, 112286, 112287, 112288, 112289, 112290, 112291, 112292, 112293, 112294, 112295, 112296, 112298, 112299, 112300, 112301, 112302, 112303, 112304, 112305, 112306, 112307, 112308, 112309, 112310, 112311, 112312, 112313, 112314, 112315, 112316, 112317, 112318, 112319, 112320, 112321, 112322, 112323, 112324, 112325, 112326, 112327, 112328, 112329, 112330, 112332, 112333, 112334, 112335, 112337, 112339, 112340, 112341, 112342, 112343, 112344, 112346, 112347, 112348, 112349, 112352, 112353, 112354, 112355, 112356, 112357, 112358, 112359, 112360, 112361, 112363, 112364, 112365, 112366, 112368, 112370, 112371, 112372, 112373, 112374, 112375, 112376, 112377, 112379, 112380, 112381, 112382, 112383, 112384, 112385, 112386, 112388, 112389, 112390, 112391, 112392, 112393, 112394, 112395, 112396, 112397, 112398, 112399, 112400, 112401, 112402, 112403, 112404, 112405, 112406, 112407, 112408, 112409, 112410, 112411, 112412, 112413, 112414, 112415, 112416, 112417, 112418, 112419, 112420, 112421, 112422, 112423, 112424, 112425, 112426, 112427, 112428, 112429, 112430, 112431, 112432, 112433, 112434, 112435, 112436, 112437, 112438, 112439, 112441, 112442, 112443, 112444, 112445, 112446, 112447, 112448, 112449, 112450, 112451, 112452, 112453, 112454, 112455, 112456, 112457, 112458, 112460, 112461, 112462, 112463, 112464, 112465, 112468, 112470, 112471, 112472, 112474, 112475, 112476, 112477, 112478, 112479, 112481, 112482, 112483, 112484, 112485, 112486, 112487, 112488, 112489, 112490, 112491, 112492, 112493, 112494, 112495, 112496, 112498, 112499, 112501, 112502, 112503, 112504, 112505, 112506, 112507, 112508, 112509, 112510, 112511, 112512, 112513, 112514, 112515, 112516, 112517, 112518, 112519, 112520, 112521, 112522, 112523, 112524, 112525, 112526, 112527, 112528, 112529, 112530, 112531, 112532, 112533, 112534, 112535, 112536, 112537, 112538, 112539, 112540, 112541, 112542, 112543, 112544, 112545, 112546, 112547, 112548, 112549, 112550, 112551, 112552, 112553, 112554, 112555, 112556, 112557, 112559, 112560, 112561, 112562, 112563, 112564, 112565, 112566, 112568, 112569, 112570, 112571, 112572, 112573, 112575, 112576, 112579, 112580, 112581, 112582, 112583, 112584, 112585, 112586, 112587, 112588, 112590, 112591, 112592, 112593, 112594, 112595, 112596, 112597, 112598, 112599, 112600, 112601, 112602, 112603, 112604, 112605, 112606, 112607, 112608, 112609, 112610, 112611, 112612, 112613, 112614, 112615, 112616, 112617, 112618, 112619, 112620, 112621, 112622, 112623, 112624, 112625, 112626, 112627, 112628, 112629, 112630, 112631, 112632, 112633, 112635, 112637, 112638, 112639, 112640, 112641, 112643, 112645, 112646, 112648, 112649, 112651, 112652, 112653, 112655, 112656, 112657, 112658, 112659, 112661, 112662, 112664, 112666, 112669, 112670, 112672, 112673, 112674, 112676, 112678, 112679, 112680, 112681, 112682, 112683, 112684, 112686, 112687, 112688, 112689, 112690, 112692, 112693, 112694, 112695, 112696, 112697, 112698, 112699, 112700, 112702, 112703, 112704, 112705, 112707, 112708, 112709, 112712, 112714, 112717, 112718, 112719, 112720, 112721, 112722, 112723, 112725, 112726, 112727, 112728, 112729, 112731, 112732, 112735, 112736, 112737, 112738, 112739, 112740, 112741, 112743, 112744, 112745, 112747, 112748, 112749, 112750, 112751, 112752, 112753, 112754, 112755, 112756, 112758, 112759, 112760, 112761, 112762, 112763, 112764, 112765, 112766, 112767, 112768, 112769, 112770, 112771, 112772, 112773, 112774, 112775, 112776, 112778, 112779, 112780, 112781, 112782, 112783, 112784, 112785, 112786, 112787, 112788, 112789, 112790, 112791, 112792, 112793, 112794, 112796, 112797, 112798, 112801, 112802, 112803, 112804, 112806, 112807, 112808, 112809, 112810, 112811, 112812, 112813, 112814, 112815, 112816, 112817, 112818, 112819, 112820, 112822, 112823, 112824, 112825, 112826, 112827, 112828, 112829, 112830, 112831, 112832, 112833, 112834, 112835, 112837, 112838, 112839, 112840, 112841, 112842, 112843, 112844, 112845, 112846, 112847, 112848, 112849, 112851, 112852, 112853, 112854, 112855, 112856, 112857, 112858, 112860, 112861, 112862, 112863, 112864, 112865, 112866, 112868, 112869, 112870, 112871, 112872, 112873, 112874, 112875, 112877, 112878, 112879, 112880, 112881, 112882, 112883, 112884, 112885, 112886, 112887, 112888, 112889, 112890, 112891, 112892, 112894, 112895, 112896, 112897, 112898, 112899, 112900, 112902, 112903, 112904, 112905, 112906, 112907, 112908, 112909, 112910, 112911, 112912, 112913, 112914, 112915, 112916, 112917, 112918, 112919, 112920, 112921, 112922, 112923, 112924, 112925, 112926, 112927, 112928, 112929, 112930, 112931, 112932, 112934, 112935, 112936, 112937, 112938, 112939, 112940, 112941, 112942, 112943, 112944, 112945, 112946, 112947, 112948, 112949, 112950, 112951, 112952, 112954, 112955, 112957, 112959, 112960, 112961, 112962, 112963, 112964, 112966, 112967, 112968, 112969, 112970, 112971, 112972, 112974, 112975, 112976, 112977, 112978, 112979, 112980, 112981, 112983, 112984, 112985, 112986, 112987, 112988, 112989, 112990, 112991, 112992, 112993, 112995, 112996, 112997, 112998, 113000, 113001, 113004, 113005, 113006, 113007, 113008, 113010, 113011, 113013, 113014, 113015, 113016, 113017, 113018, 113019, 113020, 113021, 113022, 113023, 113024, 113026, 113027, 113028, 113029, 113030, 113031, 113032, 113033, 113034, 113035, 113036, 113037, 113038, 113039, 113040, 113044, 113045, 113046, 113047, 113048, 113049, 113050, 113051, 113052, 113053, 113054, 113055, 113056, 113060, 113062, 113063, 113064, 113065, 113066, 113067, 113068, 113069, 113071, 113072, 113073, 113074, 113075, 113076, 113077, 113078, 113080, 113082, 113083, 113087, 113088, 113089, 113090, 113091, 113092, 113093, 113094, 113095, 113096, 113097, 113098, 113099, 113100, 113103, 113104, 113105, 113106, 113107, 113108, 113110, 113111, 113112, 113114, 113115, 113116, 113118, 113119, 113120, 113121, 113122, 113123, 113124, 113125, 113127, 113128, 113129, 113130, 113131, 113132, 113133, 113134, 113135, 113136, 113138, 113139, 113140, 113141, 113142, 113145, 113146, 113147, 113148, 113149, 113150, 113151, 113152, 113153, 113154, 113155, 113156, 113157, 113158, 113160, 113161, 113162, 113163, 113164, 113165, 113167, 113168, 113169, 113170, 113171, 113172, 113173, 113174, 113176, 113177, 113179, 113180, 113181, 113183, 113184, 113185, 113186, 113187, 113188, 113189, 113190, 113191, 113193, 113194, 113195, 113196, 113197, 113198, 113200, 113201, 113202, 113204, 113205, 113206, 113207, 113208, 113209, 113210, 113211, 113213, 113214, 113215, 113216, 113217, 113218, 113219, 113220, 113221, 113222, 113223, 113224, 113225, 113226, 113227, 113228, 113229, 113230, 113231, 113232, 113233, 113234, 113235, 113236, 113238, 113239, 113240, 113241, 113242, 113244, 113246, 113247, 113248, 113249, 113250, 113251, 113253, 113255, 113257, 113258, 113259, 113260, 113261, 113262, 113263, 113264, 113265, 113266, 113267, 113268, 113269, 113270, 113271, 113272, 113273, 113274, 113275, 113276, 113277, 113278, 113279, 113280, 113281, 113283, 113284, 113285, 113287, 113288, 113289, 113290, 113291, 113292, 113293, 113298, 113299, 113300, 113301, 113303, 113304, 113305, 113306, 113307, 113308, 113309, 113310, 113311, 113312, 113313, 113314, 113315, 113316, 113317, 113318, 113319, 113323, 113324, 113325, 113326, 113327, 113328, 113329, 113330, 113331, 113332, 113333, 113336, 113337, 113338, 113339, 113340, 113341, 113342, 113343, 113344, 113345, 113346, 113347, 113348, 113349, 113350, 113351, 113352, 113353, 113354, 113355, 113357, 113358, 113359, 113360, 113361, 113362, 113363, 113364, 113366, 113367, 113368, 113369, 113370, 113371, 113372, 113373, 113375, 113376, 113378, 113379, 113380, 113381, 113382, 113383, 113384, 113385, 113386, 113387, 113388, 113389, 113390, 113391, 113393, 113394, 113395, 113396, 113397, 113398, 113399, 113400, 113402, 113403, 113404, 113405, 113406, 113407, 113408, 113410, 113411, 113412, 113413, 113414, 113415, 113418, 113419, 113420, 113422, 113423, 113424, 113425, 113426, 113427, 113428, 113429, 113430, 113431, 113433, 113434, 113435, 113436, 113437, 113439, 113440, 113441, 113442, 113443, 113445, 113446, 113447, 113448, 113450, 113451, 113452, 113453, 113454, 113455, 113457, 113458, 113459, 113461, 113462, 113463, 113464, 113465, 113466, 113467, 113468, 113470, 113471, 113472, 113473, 113474, 113475, 113476, 113477, 113478, 113479, 113480, 113481, 113482, 113483, 113484, 113486, 113487, 113488, 113489, 113490, 113491, 113492, 113493, 113494, 113495, 113496, 113497, 113498, 113499, 113500, 113501, 113502, 113503, 113505, 113506, 113507, 113508, 113510, 113512, 113514, 113515, 113516, 113517, 113518, 113519, 113520, 113521, 113522, 113523, 113524, 113525, 113527, 113528, 113529, 113530, 113531, 113532, 113533, 113534, 113535, 113536, 113537, 113538, 113539, 113540, 113541, 113542, 113543, 113544, 113545, 113547, 113548, 113549, 113551, 113552, 113554, 113555, 113556, 113557, 113558, 113559, 113560, 113561, 113562, 113563, 113564, 113565, 113566, 113567, 113568, 113569, 113570, 113571, 113572, 113574, 113575, 113576, 113577, 113578, 113579, 113580, 113581, 113582, 113583, 113584, 113585, 113586, 113587, 113588, 113589, 113590, 113591, 113592, 113593, 113594, 113595, 113596, 113597, 113598, 113599, 113600, 113601, 113602, 113603, 113604, 113605, 113606, 113607, 113608, 113609, 113610, 113611, 113612, 113613, 113615, 113616, 113617, 113618, 113620, 113621, 113623, 113624, 113625, 113626, 113627, 113628, 113629, 113630, 113631, 113632, 113633, 113635, 113636, 113637, 113639, 113640, 113641, 113643, 113644, 113645, 113646, 113648, 113649, 113650, 113651, 113652, 113653, 113654, 113655, 113656, 113657, 113658, 113659, 113660, 113662, 113663, 113665, 113666, 113667, 113668, 113669, 113671, 113672, 113673, 113674, 113675, 113676, 113677, 113678, 113679, 113680, 113681, 113682, 113684, 113685, 113686, 113687, 113688, 113689, 113691, 113692, 113693, 113694, 113695, 113696, 113697, 113699, 113700, 113702, 113703, 113704, 113705, 113707, 113708, 113709, 113710, 113711, 113713, 113714, 113715, 113716, 113717, 113718, 113720, 113721, 113722, 113723, 113724, 113725, 113726, 113727, 113728, 113729, 113730, 113731, 113732, 113733, 300712, 300713, 300714, 300716, 300717, 300718, 300719, 300720, 300721, 300722, 300723, 300724, 300725, 300726, 300727, 300728, 300730, 300731, 300732, 300733, 300734, 300735, 300736, 300737, 300738, 300739, 300740, 300741, 300742, 300743, 300744, 300745, 300746, 300747, 300748, 300749, 300750, 300751, 300752, 300753, 300754, 300755, 300756, 300757, 300758, 300759, 300760, 300762, 300763, 300764, 300765, 300766, 300767, 300768, 300770, 300771, 300772, 300773, 300774, 300775, 300776, 300778, 300779, 300780, 300781, 300782, 300783, 300784, 300785, 300786, 300787, 300788, 300789, 300790, 300791, 300792, 300793, 300794, 300795, 300796, 300797, 300798, 300799, 300800, 300801, 300802, 300803, 300805, 300806, 300807, 300808, 300809, 300810, 300812, 300813, 300814, 300815, 300817, 300818, 300819, 300820, 300821, 300822, 300823, 300824, 300825, 300826, 300828, 300829, 300830, 300831, 300832, 300833, 300834, 300835, 300836, 300842, 300844, 300845, 300846, 300847, 300848, 300849, 300850, 300851, 300852, 300853, 300854, 300855, 300857, 300858, 300859, 300861, 300863, 300864, 300865, 300866, 300867, 300868, 300869, 300870, 300871, 300872, 300873, 300874, 300875, 300876, 300877, 300878, 300879, 300880, 300881, 300882, 300883, 300884, 300885, 300886, 300887, 300888, 300889, 300890, 300891, 300892, 300894, 300896, 300897, 300898, 300899, 300900, 300901, 300902, 300903, 300904, 300905, 300906, 300907, 300908, 300909, 300910, 300911, 300912, 300913, 300914, 300915, 300916, 300917, 300918, 300919, 300920, 300921, 300922, 300923, 300925, 300926, 300927, 300929, 300930, 300931, 300932, 300933, 300934, 300936, 300937, 300938, 300939, 300940, 300941, 300942, 300943, 300944, 300945, 300946, 300947, 300948, 300949, 300950, 300951, 300953, 300954, 300956, 300957, 300958, 300959, 300960, 300961, 300962, 300963, 300964, 300965, 300966, 300967, 300968, 300969, 300970, 300971, 300972, 300973, 300974, 300975, 300976, 300977, 300978, 300979, 300981, 300982, 300983, 300984, 300985, 300986, 300987, 300988, 300989, 300990, 300992, 300993, 300994, 300995, 300996, 300997, 300998, 300999, 301000, 301001, 301002, 301003, 301004, 301005, 301006, 301007, 301008, 301009, 301010, 301011, 301012, 301013, 301014, 301015, 301016, 301018, 301019, 301020, 301021, 301022, 301023, 301024, 301025, 301026, 301027, 301028, 301029, 301030, 301031, 301032, 301033, 301034, 301035, 301037, 301038, 301039, 301040, 301041, 301042, 301043, 301044, 301045, 301046, 301049, 301050, 301051, 301052, 301053, 301054, 301055, 301056, 301057, 301058, 301060, 301061, 301062, 301063, 301064, 301065, 301066, 301067, 301068, 301069, 301070, 301071, 301073, 301074, 301075, 301076, 301077, 301078, 301079, 301080, 301081, 301082, 301083, 301084, 301086, 301087, 301088, 301089, 301090, 301091, 301092, 301093, 301094, 301095, 301096, 301097, 301099, 301100, 301101, 301102, 301103, 301104, 301105, 301106, 301107, 301108, 301110, 301111, 301112, 301113, 301114, 301115, 301116, 301117, 301118, 301119, 301120, 301122, 301123, 301124, 301125, 301126, 301127, 301128, 301129, 301130, 301131, 301133, 301134, 301135, 301136, 301137, 301138, 301139, 301140, 301141, 301142, 301143, 301144, 301145, 301146, 301147, 301148, 301149, 301150, 301151, 301153, 301154, 301156, 301157, 301158, 301159, 301161, 301162, 301163, 301165, 301166, 301167, 301169, 301170, 301171, 301172, 301173, 301174, 301175, 301176, 301177, 301178, 301179, 301180, 301181, 301182, 301183, 301184, 301186, 301187, 301188, 301189, 301190, 301191, 301192, 301193, 301194, 301195, 301196, 301197, 301198, 301199, 301200, 301202, 301203, 301204, 301205, 301206, 301207, 301208, 301209, 301210, 301211, 301212, 301213, 301214, 301215, 301216, 301217, 301218, 301219, 301220, 301221, 301222, 301223, 301224, 301225, 301226, 301228, 301229, 301230, 301231, 301232, 301234, 301235, 301236, 301237, 301238, 301239, 301240, 301241, 301242, 301243, 301244, 301245, 301246, 301247, 301248, 301249, 301250, 301251, 301252, 301253, 301279, 301304, 301308, 301309, 301310, 301311, 301312, 301313, 301314, 301315, 301316, 301317, 301318, 301319, 301320, 301321, 301322, 301323, 301324, 301325, 301326, 301327, 301328, 301329, 301330, 301331, 301332, 301333, 301334, 301336, 301337, 301338, 301339, 301340, 301341, 301342, 301343, 301344, 301346, 301347, 301348, 301350, 301351, 301352, 301355, 301356, 301357, 301372, 301375, 301379, 301380, 301381, 301384, 301386, 301389, 301391, 301392, 301396, 301397, 301399, 301400, 301401, 301402, 301403, 301404, 301405, 301406, 301407, 301408, 301409, 301410, 301412, 301413, 301415, 301416, 301417, 301418, 301419, 301420, 301421, 301422, 301423, 301424, 301425, 301426, 301427, 301428, 301429, 301430, 301431, 301432, 301433, 301434, 301435, 301436, 301437, 301438, 301441, 301442, 301443, 301444, 301445, 301446, 301447, 301448, 301449, 301450, 301452, 301453, 301454, 301455, 301456, 301457, 301458, 301459, 301460, 301461, 301462, 301469, 301470, 301492
Other devices under the same FDA submission
9 devices cleared under P030005 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VALITUDE™ X4 CRT-P U128Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls26 deaths2,073 reports
- VISIONIST™ X4 CRT-P U228Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall4 deaths819 reports
- INVIVE™ V173Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death670 reports
- VALITUDE™ CRT-P U125Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls2 deaths558 reports
- VISIONIST™ CRT-P U225Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall2 deaths253 reports
- INTUA™ V273Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)264 reports
- VISIONIST™ CRT-P U226Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls18 reports
- INTUA™ V272Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)23 reports
- INVIVE™ V172Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 88107).
- Z-2017-2021 · Open, ClassifiedVITALIO DR EL (pacemaker), Model Numbers: a) J274 b) J277 c) K274 (added 7/15/2021) d) K277 (added 7/15/2021) e) K284 (added 7/15/2021) f) K287 (added 7/15/2021)
- Z-2018-2021 · Open, ClassifiedINVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W173
- Z-2019-2021 · Open, ClassifiedINTUA CRT-P, Model Numbers: a) V272 b) V273 c) W273
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page