Artis Q biplane, Model Number 10848282 recallZ-0976-2022 · Siemens Medical Solutions USA, Inc · initiated Feb 23, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

FDA root cause category: Device Design

What the company is doing

An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 2/23/22 was sent to customers. Which steps have to be taken by the user to avoid the possible risks associated with this problem? As also described in the Operator Manual, we recommend that the system operator checks the water level of the cooling circuit at least every three months and refills water, if necessary: 1. Open the filling gland of the cooling unit. The water surface must be clearly visible above the cooling ribs. 2. Replenish with water (drinking water quality) if cooling liquid is lacking. Please inform the service technician in case of lacking cooling liquid. How will the corrective action be implemented and what is the efficiency of the corrective actions? The tube cooling unit will be equipped with a water level sensor and the software will be updated accordingly. In case the coolant level drops below a certain threshold the following message will be displayed: Tube cooling water level low: Refill as per Operator Manual . Siemens will correct the error with a hardware and software update via Update Instruction AX002/20/S and AX004/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand t

Recall details

Recall number
Z-0976-2022
Status
Open, Classified
Initiated by the company
Feb 23, 2022
Posted by the FDA
Apr 21, 2022
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
269 systems
Product code
OWB
FDA submissions
K181407
Distribution
US Nationwide.

Product description

Artis Q biplane, Model Number 10848282

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI: 4056869009995; Serial Numbers: 134007 121314 134008 121396 121590 121665 121564 122025 121627 121185 121430 121406 121336 122012 121471 121195 121467 121612 122022 121530 121089 121094 134004 121169 121006 122021 122028 121013 121040 121501 121368 121526 121208 121548 121515 121445 121446 121623 121082 121459 121011 122037 121039 121521 121371 122005 121374 121204 121614 121305 121536 121053 121054 121422 121014 121112 121325 121357 121621 121167 121378 121369 121604 121133 134011 121200 121490 121163 121364 121642 121018 121154 121418 121605 121547 121365 121205 122034 121137 121324 121012 121049 121078 121096 121332 121333 121344 121183 134003 121455 121504 121019 121491 121381 121457 121460 121352 121563 121635 121434 121165 121625 121415 121355 121010 121171 121139 121210 121573 121215 121542 121366 121323 121370 121493 121556 121391 121384 121015 121487 122017 121395 121472 121558 121405 121309 121513 121342 121301 121081 121551 121633 121649 121190 121315 121629 121529 121138 121527 121600 121568 121193 121478 121479 121554 121020 121615 121161 121140 121071 121216 121537 121566 121470 121417 121151 121188 121168 122001 121321 134500 122033 121411 121120 121121 121122 121520 121541 121645 121144 121383 121663 121047 122010 121312 121407 121557 121148 121436 121201 121088 121027 121359 121618 121622 122007 121136 121322 121444 121511 121004 122032 121648 121535 122016 121175 121486 121173 121462 121009 121056 121057 121419 121435 121468 121008 121152 121155 121160 121433 121610 121466 121578 121608 121449 122013 121576 121611 121525 121379 121319 121085 121361 121206 121431 121034 121653 121544 121211 121187 121197 121569 121375 121377 121412 121505 121601 121393 121518 121506 121339 121616 121199 121310 121362 121509 121609 121440 121594 121510 121533 121079 121212 122024 122027 121041 121591 121037 121046 121502 121313 122014 Additional Serial Numbers as of 10/4/23: 134502 121340 121461 121051 121179 121016 121022

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

11 devices cleared under K181407 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 89860).