HeartWare Battery recallZ-1266-2022 · Medtronic Inc · initiated May 5, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly recharging.
FDA root cause category: Process control
What the company is doing
In the United States, on 5/5/2022, a written Medical Device Retrieval letter dated May 2022 was mailed via 2-day UPS delivery to consignees who have received potentially affected product according to Medtronic records. Outside of the US (OUS), beginning 5/5/2022, an Urgent Medical Device Retrieval letter was being delivered via regionally approved methods to consignees who have received potentially affected product according to Medtronic records. The letter notified customers of the issue description and were provided patient management recommendations. This included reminding their patients to always keep two sources of power connected to their controller and have fully charged spare batteries available at all times. If a [Power Disconnect] alarm occurs while a battery is physically connected, that battery should be taken out of service. The letter references instructions from the patient manual. The consignee actions included immediately identifying and quarantining all affected batteries listed in the attachment; immediately notify patients with affected batteries that Medtronic is conducting this retrieval and have them remove these batteries from service; informed the consignee that the Medtronic Field Representative can assist them in the return of the affected product; complete and return the enclosed Customer Confirmation Form and follow the instructions to initiate an exchange; and share the letter with all those who need to be aware within their organization or to any organization where the affected batteries have been transferred.
Recall details
- Recall number
- Z-1266-2022
- Status
- Open, Classified
- Initiated by the company
- May 5, 2022
- Recalling firm
- Medtronic Inc
- Quantity
- 429 batteries
- Product code
- DSQ
- FDA submissions
- P100047
- Distribution
- Worldwide Distribution. US nationwide, Austria, Belgium, Finland, France, Germany, Hungary, Italy, Netherlands, New Zealand, North Macedonia, Norway, Poland, Spain, Switzerland, and the United Kingdom.
Product description
HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
GTIN 00888707008778. Serial numbers BAT935012, BAT935059, BAT935008, BAT935009, BAT935018, BAT935020, BAT935022, BAT935023, BAT935047, BAT934999, BAT935003, BAT935010, BAT935329, BAT934928, BAT934934, BAT935160, BAT935078, BAT934997, BAT935027, BAT935048, BAT934953, BAT934933, BAT934948, BAT934853, BAT934854, BAT934858, BAT934895, BAT934899, BAT934901, BAT934903, BAT934904, BAT934905, BAT934927, BAT934932, BAT934935, BAT934940, BAT934979, BAT934981, BAT935109, BAT935112, BAT935116, BAT935117, BAT935120, BAT935122, BAT935124, BAT935127, BAT935132, BAT935133, BAT935134, BAT935136, BAT934846, BAT934848, BAT934849, BAT934850, BAT935137, BAT935142, BAT935143, BAT935145, BAT935146, BAT935148, BAT935150, BAT935151, BAT935152, BAT935154, BAT935155, BAT935156, BAT935157, BAT935158, BAT935161, BAT935162, BAT935163, BAT935164, BAT935260, BAT935301, BAT935302, BAT935303, BAT935201, BAT935202, BAT935203, BAT935172, BAT935173, BAT935174, BAT935175, BAT935176, BAT935177, BAT935178, BAT935179, BAT935180, BAT935181, BAT935182, BAT935183, BAT935184, BAT935185, BAT935186, BAT935187, BAT934910, BAT934912, BAT934914, BAT934915, BAT934916, BAT934919, BAT935317, BAT935318, BAT935319, BAT935320, BAT935322, BAT935324, BAT935217, BAT935264, BAT935269, BAT935276, BAT935277, BAT935282, BAT935283, BAT935285, BAT935286, BAT935287, BAT935312, BAT935313, BAT935314, BAT935188, BAT935189, BAT935190, BAT935191, BAT935192, BAT935193, BAT935194, BAT935195, BAT935196, BAT935197, BAT935198, BAT935199, BAT934851, BAT934852, BAT934855, BAT934857, BAT934859, BAT934861, BAT934862, BAT934863, BAT934865, BAT934866, BAT935207, BAT935210, BAT935222, BAT935256, BAT935257, BAT935230, BAT935231, BAT935233, BAT935234, BAT935236, BAT935237, BAT935238, BAT935268, BAT935270, BAT935271, BAT935272, BAT935273, BAT935275, BAT935279, BAT935280, BAT935288, BAT935290, BAT935291, BAT935292, BAT935284, BAT934844, BAT935229, BAT934917, BAT934918, BAT934920, BAT934924, BAT935232, BAT935235, BAT935247, BAT935261, BAT934867, BAT934868, BAT935265, BAT935266, BAT935204, BAT935205, BAT935206, BAT935212, BAT935213, BAT935216, BAT935250, BAT935251, BAT935252, BAT935253, BAT935254, BAT935267, BAT934869, BAT935200, BAT935262, BAT935218, BAT935219, BAT935227, BAT935228, BAT934871, BAT934873, BAT934874, BAT934875, BAT934876, BAT934877, BAT934878, BAT934880, BAT934881, BAT934882, BAT934883, BAT934884, BAT934886, BAT934909, BAT935138, BAT935140, BAT935293, BAT935294, BAT935296, BAT935259, BAT935297, BAT935298, BAT935299, BAT935300, BAT934894, BAT935326, BAT935331, BAT934982, BAT934987, BAT934988, BAT934989, BAT935032, BAT935038, BAT935039, BAT935040, BAT935042, BAT935043, BAT935051, BAT935068, BAT935074, BAT935085, BAT935089, BAT935093, BAT935094, BAT935095, BAT935098, BAT935099, BAT935100, BAT935101, BAT935102, BAT935103, BAT935104, BAT935105, BAT935106, BAT935107, BAT935108, BAT935141, BAT935144, BAT935147, BAT935149, BAT935159, BAT935165, BAT935166, BAT935167, BAT935168, BAT935169, BAT935170, BAT935171, BAT935258, BAT934887, BAT934889, BAT934891, BAT934893, BAT934870, BAT935239, BAT935240, BAT935241, BAT935242, BAT935243, BAT935244, BAT935245, BAT935246, BAT935248, BAT935249, BAT934906, BAT934908, BAT935033, BAT935036, BAT935044, BAT935017, BAT935045, BAT935001, BAT935026, BAT935114, BAT935118, BAT935121, BAT935123, BAT935263, BAT935274, BAT935278, BAT935281, BAT935315, BAT935084, BAT935086, BAT934896, BAT934897, BAT934898, BAT934949, BAT934950, BAT935323, BAT935325, BAT935327, BAT935328, BAT935330, BAT935332, BAT935005, BAT935006, BAT935025, BAT935062, BAT935064, BAT935065, BAT935208, BAT934922, BAT934926, BAT934929, BAT934930, BAT934942, BAT934943, BAT934944, BAT934945, BAT934955, BAT934959, BAT934961, BAT934964, BAT934967, BAT934968, BAT934969, BAT934970, BAT934976, BAT934977, BAT934978, BAT935002, BAT935014, BAT935049, BAT934973, BAT934974, BAT935087, BAT935091, BAT935097, BAT934962, BAT935041, BAT935046, BAT935050, BAT935053, BAT935057, BAT935058, BAT935061, BAT935063, BAT935067, BAT935071, BAT935077, BAT935080, BAT935081, BAT935088, BAT935090, BAT935092, BAT934952, BAT934954, BAT934958, BAT934921, BAT934923, BAT934925, BAT934931, BAT934941, BAT934951, BAT935015, BAT935016, BAT935019, BAT935021, BAT935096, BAT935111, BAT934998, BAT934956, BAT935024, BAT935028, BAT935034, BAT935037, BAT935069, BAT935070, BAT935215, BAT935321, BAT935135, BAT935029, BAT935030, BAT935031, BAT934946, BAT934983, BAT934984, BAT934986, BAT934992, BAT934971, BAT934972, BAT934975, BAT934980, BAT934995, BAT935000, BAT935007, BAT935013, BAT934947, BAT934996, BAT935011, BAT935052, BAT935054, BAT935056, BAT935082, BAT935072, BAT935073, BAT935075, BAT935076, BAT935004, BAT934990, BAT934991, BAT934993, BAT934994, BAT935079, BAT934985, BAT934963, BAT934965, and BAT934966. Expiration date 30-APR-22.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass2 recalls606 reports
Product lines
Other devices under the same FDA submission
32 devices cleared under P100047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HeartWare® Controller 1403USHeartware Inc. · Ventricular (Assisst) Bypass3 recalls10 deaths1,495 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assisst) Bypass1 recall1,126 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall615 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass7 recalls36 deaths203 reports
- HVAD® Controller 1403USHeartware Inc. · Ventricular (Assist) Bypass548 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall487 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assist) Bypass1 recall281 reports
- HeartWare® Battery 1650DEHeartware Inc. · Ventricular (Assist) Bypass1 recall159 reports
- HVAD® Pump Implant Kit MCS1705PUHeartware Inc. · Ventricular (Assist) Bypass4 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page