Palindrome recallZ-1316-2022 · Covidien Llc · initiated Jun 8, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

FDA root cause category: Process control

What the company is doing

Medtronic distributed a recall notification to customers by mail on 06/08/2022. Customers were instructed to immediately quarantine and discontinue use for specific lots of chronic hemodialysis catheters, which are to be returned to Medtronic. If customers have further distributed product, they are asked to forward the information from the recall notification to those recipients. All customers are asked to complete and return the Consignee Product Retrieval Confirmation Form to [email protected]. Additionally, consignees are asked to share the notice within their organization and retain it for their records regarding the retrieval of unused product. If a consignee is aware of any adverse events associated with recalled product, they are to submit a report to FDA via Medwatch. Customers with any questions can contact their Medtronic Representative or call Customer Service at 800-962-9888, option 2. ***Updated 8/16/22*** Medtronic has created a website to help customers identify affected product: https://www.medtronic.com/covidien/en-us/products/c/renal-care-recall-lookup-tool.html#lotLookupTool

Recall details

Recall number
Z-1316-2022
Status
Open, Classified
Initiated by the company
Jun 8, 2022
Posted by the FDA
Jul 7, 2022
Recalling firm
Covidien Llc
Quantity
576,018 units
Product code
MSD
FDA submissions
K111372
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, DC, GU, & PR. The countries of Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Ireland, Israel, Italy, Jamaica, Japan, Kenya, Republic of Korea, Kosovo, Kuwait, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Malta, Mayotte, Mexico, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, United States, Viet Nam, Virgin Islands, and Zimbabwe.

Product description

Palindrome Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.

Code information (models, lots, serial numbers)

Show the full code information (58k characters)

Code Information organized as "Model No., Serial No., GTIN": 8888119360, 1916500132, 20884521012803; 8888119364, 1831000099, 20884521056555; 8888119364, 1832300114, 20884521056555; 8888119364, 2019500276, 20884521056555; 8888119364, 2106900226, 20884521056555; 8888123404, 1912600108, 20884521056586; 8888123404, 1934400172, 20884521056586; 8888123404, 2012100136, 20884521056586; 8888123404, 2017500130, 20884521056586; 8888123404, 2034400075, 20884521056586; 8888123404, 2104600236, 20884521056586; 8888123404, 2125800094, 20884521056586; 8888123404, 2127200415, 20884521056586; 8888123410, 2019500253, 20884521012988; 8888123410, 2104100079, 20884521012988; 8888128450, 2127200431, 20884521013008; 8888128454, 2017800061, 10884521056619; 8888128454, 1822600121, 20884521056616; 8888128454, 2017800061, 20884521056616; 8888128460, 2019500266, 20884521013053; 8888145014, 1715100153, 10884521013148; 8888145014, 1715800117, 10884521013148; 8888145014, 1715900177, 10884521013148; 8888145014, 1716100154, 10884521013148; 8888145014, 1717200138, 10884521013148; 8888145014, 1717400127, 10884521013148; 8888145014, 1717900133, 10884521013148; 8888145014, 1718600101, 10884521013148; 8888145014, 1719300100, 10884521013148; 8888145014, 1719800114, 10884521013148; 8888145014, 1720000100, 10884521013148; 8888145014, 1720700111, 10884521013148; 8888145014, 1721300160, 10884521013148; 8888145014, 1723400096, 10884521013148; 8888145014, 1724900105, 10884521013148; 8888145014, 1726300130, 10884521013148; 8888145014, 1726400083, 10884521013148; 8888145014, 1727100074, 10884521013148; 8888145014, 1728600063, 10884521013148; 8888145014, 1729700112, 10884521013148; 8888145014, 1731900212, 10884521013148; 8888145014, 1733300157, 10884521013148; 8888145014, 1733300163, 10884521013148; 8888145014, 1734700107, 10884521013148; 8888145014, 1800300139, 10884521013148; 8888145014, 1801000129, 10884521013148; 8888145014, 1801000130, 10884521013148; 8888145014, 1801700128, 10884521013148; 8888145014, 1804500149, 10884521013148; 8888145014, 1804500152, 10884521013148; 8888145014, 1804500153, 10884521013148; 8888145014, 1805200094, 10884521013148; 8888145014, 1805900132, 10884521013148; 8888145014, 1807300172, 10884521013148; 8888145014, 1807300173, 10884521013148; 8888145014, 1808600067, 10884521013148; 8888145014, 1809400139, 10884521013148; 8888145014, 1812200179, 10884521013148; 8888145014, 1812700099, 10884521013148; 8888145014, 1819000134, 10884521013148; 8888145014, 1819000143, 10884521013148; 8888145014, 1820700110, 10884521013148; 8888145014, 1820700112, 10884521013148; 8888145014, 1820700114, 10884521013148; 8888145014, 1821300120, 10884521013148; 8888145014, 1821300121, 10884521013148; 8888145014, 1821300122, 10884521013148; 8888145014, 1823400084, 10884521013148; 8888145014, 1823400087, 10884521013148; 8888145014, 1824700121, 10884521013148; 8888145014, 1827600136, 10884521013148; 8888145014, 1828200125, 10884521013148; 8888145014, 1829500108, 10884521013148; 8888145014, 1830300168, 10884521013148; 8888145014, 1830300169, 10884521013148; 8888145014, 1831600200, 10884521013148; 8888145014, 1832300111, 10884521013148; 8888145014, 1832300112, 10884521013148; 8888145014, 1833100110, 10884521013148; 8888145014, 1835100115, 10884521013148; 8888145014, 1835300233, 10884521013148; 8888145014, 1835300234, 10884521013148; 8888145014, 1835300235, 10884521013148; 8888145014, 1900700140, 10884521013148; 8888145014, 1900700142, 10884521013148; 8888145014, 1900700143, 10884521013148; 8888145014, 1900700144, 10884521013148; 8888145014, 1901400101, 10884521013148; 8888145014, 1902800139, 10884521013148; 8888145014, 1902800141, 10884521013148; 8888145014, 1903500106, 10884521013148; 8888145014, 1903500107, 10884521013148; 8888145014, 1904200095, 10884521013148; 8888145014, 1904900121, 10884521013148; 8888145014, 1904900123, 10884521013148; 8888145014, 1905600057, 10884521013148; 8888145014, 1905600125, 10884521013148; 8888145014, 1905600126, 10884521013148; 8888145014, 1906000129, 10884521013148; 8888145014, 1909400132, 10884521013148; 8888145014, 1910200091, 10884521013148; 8888145014, 1911500109, 10884521013148; 8888145014, 1911500110, 10884521013148; 8888145014, 1911500111, 10884521013148; 8888145014, 1911500112, 10884521013148; 8888145014, 1911500113, 10884521013148; 8888145014, 1912600100, 10884521013148; 8888145014, 1912600101, 10884521013148; 8888145014, 1912600102, 10884521013148; 8888145014, 1914000267, 10884521013148; 8888145014, 1914000270, 10884521013148; 8888145014, 1914000271, 10884521013148; 8888145014, 1914400141, 10884521013148; 8888145014, 1914400142, 10884521013148; 8888145014, 1914800172, 10884521013148; 8888145014, 1914800204, 10884521013148; 8888145014, 1915800096, 10884521013148; 8888145014, 1915800097, 10884521013148; 8888145014, 1915800098, 10884521013148; 8888145014, 1919700101, 10884521013148; 8888145014, 1919800096, 10884521013148; 8888145014, 1919800097, 10884521013148; 8888145014, 1919800098, 10884521013148; 8888145014, 1919800099, 10884521013148; 8888145014, 1919800109, 10884521013148; 8888145014, 1919800111, 10884521013148; 8888145014, 1921100180, 10884521013148; 8888145014, 1921100182, 10884521013148; 8888145014, 1922100141, 10884521013148; 8888145014, 1922500100, 10884521013148; 8888145014, 1922500101, 10884521013148; 8888145014, 1922500102, 10884521013148; 8888145014, 1922500103, 10884521013148; 8888145014, 1922500106, 10884521013148; 8888145014, 1922500199, 10884521013148; 8888145014, 1926100305, 10884521013148; 8888145014, 1926800390, 10884521013148; 8888145014, 1926800391, 10884521013148; 8888145014, 1926800392, 10884521013148; 8888145014, 1926800393, 10884521013148; 8888145014, 1927600099, 10884521013148; 8888145014, 1927600100, 10884521013148; 8888145014, 1928200085, 10884521013148; 8888145014, 1928800200, 10884521013148; 8888145014, 1928800201, 10884521013148; 8888145014, 1928800202, 10884521013148; 8888145014, 1928800205, 10884521013148; 8888145014, 1928800207, 10884521013148; 8888145014, 1928800246, 10884521013148; 8888145014, 1928800247, 10884521013148; 8888145014, 1928800248, 10884521013148; 8888145014, 1930200112, 10884521013148; 8888145014, 1930200114, 10884521013148; 8888145014, 1930400108, 10884521013148; 8888145014, 1930900092, 10884521013148; 8888145014, 1930900097, 10884521013148; 8888145014, 1930900099, 10884521013148; 8888145014, 1931600123, 10884521013148; 8888145014, 1931600127, 10884521013148; 8888145014, 1933700122, 10884521013148; 8888145014, 1933700123, 10884521013148; 8888145014, 1933700124, 10884521013148; 8888145014, 1934400139, 10884521013148; 8888145014, 1934400167, 10884521013148; 8888145014, 1934400168, 10884521013148; 8888145014, 1934400169, 10884521013148; 8888145014, 1934400170, 10884521013148; 8888145014, 2002100141, 10884521013148; 8888145014, 2008400124, 10884521013148; 8888145014, 2008500083, 10884521013148; 8888145014, 2008500085, 10884521013148; 8888145014, 2008500086, 10884521013148; 8888145014, 2012100129, 10884521013148; 8888145014, 2012100130, 10884521013148; 8888145014, 2012100131, 10884521013148; 8888145014, 2012100132, 10884521013148; 8888145014, 2012100133, 10884521013148; 8888145014, 2015100056, 10884521013148; 8888145014, 2016400033, 10884521013148; 8888145014, 2016400034, 10884521013148; 8888145014, 2019500255, 10884521013148; 8888145014, 2019500274, 10884521013148; 8888145014, 2019500279, 10884521013148; 8888145014, 2020400091, 10884521013148; 8888145014, 2020400092, 10884521013148; 8888145014, 2020400094, 10884521013148; 8888145014, 2020400095, 10884521013148; 8888145014, 2020400096, 10884521013148; 8888145014, 2020400097, 10884521013148; 8888145014, 2020400098, 10884521013148; 8888145014, 2023300123, 10884521013148; 8888145014, 2023300124, 10884521013148; 8888145014, 2023300148, 10884521013148; 8888145014, 2023300149, 10884521013148; 8888145014, 2023300156, 10884521013148; 8888145014, 2023300157, 10884521013148; 8888145014, 2028200219, 10884521013148; 8888145014, 2028200220, 10884521013148; 8888145014, 2028200221, 10884521013148; 8888145014, 2028200222, 10884521013148; 8888145014, 2028200223, 10884521013148; 8888145014, 2029600112, 10884521013148; 8888145014, 2029600113, 10884521013148; 8888145014, 2029600114, 10884521013148; 8888145014, 2029600115, 10884521013148; 8888145014, 2029600116, 10884521013148; 8888145014, 2031800147, 10884521013148; 8888145014, 2031800148, 10884521013148; 8888145014, 2031800149, 10884521013148; 8888145014, 2031800151, 10884521013148; 8888145014, 2031800152, 10884521013148; 8888145014, 2034400121, 10884521013148; 8888145014, 2034400137, 10884521013148; 8888145014, 2102500096, 10884521013148; 8888145014, 2102600119, 10884521013148; 8888145014, 2104100081, 10884521013148; 8888145014, 2104100082, 10884521013148; 8888145014, 2104100083, 10884521013148; 8888145014, 2104100096, 10884521013148; 8888145014, 2104100097, 10884521013148; 8888145014, 2104100098, 10884521013148; 8888145014, 2104100099, 10884521013148; 8888145014, 2104100100, 10884521013148; 8888145014, 2104600115, 10884521013148; 8888145014, 2104600165, 10884521013148; 8888145014, 2104600169, 10884521013148; 8888145014, 2104600240, 10884521013148; 8888145014, 2106900191, 10884521013148; 8888145014, 2106900192, 10884521013148; 8888145014, 2106900193, 10884521013148; 8888145014, 2106900194, 10884521013148; 8888145014, 2110900115, 10884521013148; 8888145014, 2110900155, 10884521013148; 8888145014, 2110900156, 10884521013148; 8888145014, 2113300442, 10884521013148; 8888145014, 2113300443, 10884521013148; 8888145014, 2113300444, 10884521013148; 8888145014, 2113300445, 10884521013148; 8888145014, 2113300446, 10884521013148; 8888145014, 2113400058, 10884521013148; 8888145014, 2113400064, 10884521013148; 8888145014, 2113400070, 10884521013148; 8888145014, 2113400082, 10884521013148; 8888145014, 2113400099, 10884521013148; 8888145014, 2116500091, 10884521013148; 8888145014, 2116500093, 10884521013148; 8888145014, 2116500125, 10884521013148; 8888145014, 2116500126, 10884521013148; 8888145014, 2116500127, 10884521013148; 8888145014, 2116500128, 10884521013148; 8888145014, 2116500129, 10884521013148; 8888145014, 2116500154, 10884521013148; 8888145014, 2116500155, 10884521013148; 8888145014, 2116600068, 10884521013148; 8888145014, 2116600092, 10884521013148; 8888145014, 2124500100, 10884521013148; 8888145014, 2124500101, 10884521013148; 8888145014, 2124500103, 10884521013148; 8888145014, 2124500104, 10884521013148; 8888145014, 2124500105, 10884521013148; 8888145014, 2124500107, 10884521013148; 8888145014, 2124500114, 10884521013148; 8888145014, 2124500117, 10884521013148; 8888145014, 2124600219, 10884521013148; 8888145014, 2124600221, 10884521013148; 8888145014, 2124600223, 10884521013148; 8888145014, 2124600224, 10884521013148; 8888145014, 2124600225, 10884521013148; 8888145014, 2124600226, 10884521013148; 8888145014, 2124600259, 10884521013148; 8888145014, 2124600260, 10884521013148; 8888145014, 2124600261, 10884521013148; 8888145014, 2135700266, 10884521013148; 8888145014, 2135700301, 10884521013148; 8888145014, 2201000059, 10884521013148; 8888145014, 1714400164, 20884521013145; 8888145014, 1715100153, 20884521013145; 8888145014, 1715300104, 20884521013145; 8888145014, 1715800117, 20884521013145; 8888145014, 1715900177, 20884521013145; 8888145014, 1716100154, 20884521013145; 8888145014, 1717200138, 20884521013145; 8888145014, 1717400127, 20884521013145; 8888145014, 1717900133, 20884521013145; 8888145014, 1718600101, 20884521013145; 8888145014, 1719300100, 20884521013145; 8888145014, 1719800114, 20884521013145; 8888145014, 1720000100, 20884521013145; 8888145014, 1720700111, 20884521013145; 8888145014, 1721300160, 20884521013145; 8888145014, 1722100144, 20884521013145; 8888145014, 1722100145, 20884521013145; 8888145014, 1723400096, 20884521013145; 8888145014, 1724900105, 20884521013145; 8888145014, 1726300130, 20884521013145; 8888145014, 1726400083, 20884521013145; 8888145014, 1727100074, 20884521013145; 8888145014, 1728600063, 20884521013145; 8888145014, 1729100098, 20884521013145; 8888145014, 1729700112, 20884521013145; 8888145014, 1731900212, 20884521013145; 8888145014, 1733300163, 20884521013145; 8888145014, 1734700107, 20884521013145; 8888145014, 1800300136, 20884521013145; 8888145014, 1800300139, 20884521013145; 8888145014, 1801000129, 20884521013145; 8888145014, 1801000130, 20884521013145; 8888145014, 1801700128, 20884521013145; 8888145014, 1804500149, 20884521013145; 8888145014, 1804500152, 20884521013145; 8888145014, 1804500153, 20884521013145; 8888145014, 1805200094, 20884521013145; 8888145014, 1805900132, 20884521013145; 8888145014, 1806600079, 20884521013145; 8888145014, 1806600080, 20884521013145; 8888145014, 1807300172, 20884521013145; 8888145014, 1807300173, 20884521013145; 8888145014, 1809400139, 20884521013145; 8888145014, 1810000168, 20884521013145; 8888145014, 1810000169, 20884521013145; 8888145014, 1811500158, 20884521013145; 8888145014, 1812200179, 20884521013145; 8888145014, 1812700099, 20884521013145; 8888145014, 1819000134, 20884521013145; 8888145014, 1819000143, 20884521013145; 8888145014, 1820700110, 20884521013145; 8888145014, 1820700112, 20884521013145; 8888145014, 1820700114, 20884521013145; 8888145014, 1821300120, 20884521013145; 8888145014, 1821300121, 20884521013145; 8888145014, 1821300122, 20884521013145; 8888145014, 1822600145, 20884521013145; 8888145014, 1823400083, 20884521013145; 8888145014, 1823400084, 20884521013145; 8888145014, 1823400086, 20884521013145; 8888145014, 1823400087, 20884521013145; 8888145014, 1824700120, 20884521013145; 8888145014, 1824700121, 20884521013145; 8888145014, 1827600136, 20884521013145; 8888145014, 1827600137, 20884521013145; 8888145014, 1828200125, 20884521013145; 8888145014, 1828200126, 20884521013145; 8888145014, 1829500107, 20884521013145; 8888145014, 1829500108, 20884521013145; 8888145014, 1830300168, 20884521013145; 8888145014, 1830300169, 20884521013145; 8888145014, 1831600199, 20884521013145; 8888145014, 1831600200, 20884521013145; 8888145014, 1832300111, 20884521013145; 8888145014, 1832300112, 20884521013145; 8888145014, 1833100110, 20884521013145; 8888145014, 1833800143, 20884521013145; 8888145014, 1835100115, 20884521013145; 8888145014, 1835300230, 20884521013145; 8888145014, 1835300231, 20884521013145; 8888145014, 1835300232, 20884521013145; 8888145014, 1835300233, 20884521013145; 8888145014, 1835300234, 20884521013145; 8888145014, 1835300235, 20884521013145; 8888145014, 1900700140, 20884521013145; 8888145014, 1900700141, 20884521013145; 8888145014, 1900700142, 20884521013145; 8888145014, 1900700143, 20884521013145; 8888145014, 1900700144, 20884521013145; 8888145014, 1901400100, 20884521013145; 8888145014, 1901400101, 20884521013145; 8888145014, 1902800139, 20884521013145; 8888145014, 1902800140, 20884521013145; 8888145014, 1902800141, 20884521013145; 8888145014, 1903500106, 20884521013145; 8888145014, 1903500107, 20884521013145; 8888145014, 1904200095, 20884521013145; 8888145014, 1904200096, 20884521013145; 8888145014, 1904900121, 20884521013145; 8888145014, 1904900122, 20884521013145; 8888145014, 1904900123, 20884521013145; 8888145014, 1904900124, 20884521013145; 8888145014, 1905600057, 20884521013145; 8888145014, 1905600122, 20884521013145; 8888145014, 1905600124, 20884521013145; 8888145014, 1905600125, 20884521013145; 8888145014, 1905600126, 20884521013145; 8888145014, 1905600127, 20884521013145; 8888145014, 1906000128, 20884521013145; 8888145014, 1906000129, 20884521013145; 8888145014, 1909400132, 20884521013145; 8888145014, 1910200088, 20884521013145; 8888145014, 1910200089, 20884521013145; 8888145014, 1910200090, 20884521013145; 8888145014, 1910200091, 20884521013145; 8888145014, 1911500109, 20884521013145; 8888145014, 1911500110, 20884521013145; 8888145014, 1911500111, 20884521013145; 8888145014, 1911500112, 20884521013145; 8888145014, 1911500113, 20884521013145; 8888145014, 1912600099, 20884521013145; 8888145014, 1912600100, 20884521013145; 8888145014, 1912600101, 20884521013145; 8888145014, 1912600102, 20884521013145; 8888145014, 1913000093, 20884521013145; 8888145014, 1914000267, 20884521013145; 8888145014, 1914000268, 20884521013145; 8888145014, 1914000269, 20884521013145; 8888145014, 1914000270, 20884521013145; 8888145014, 1914000271, 20884521013145; 8888145014, 1914400140, 20884521013145; 8888145014, 1914400141, 20884521013145; 8888145014, 1914400142, 20884521013145; 8888145014, 1914800172, 20884521013145; 8888145014, 1914800204, 20884521013145; 8888145014, 1919800096, 20884521013145; 8888145014, 1919800097, 20884521013145; 8888145014, 1919800098, 20884521013145; 8888145014, 1919800099, 20884521013145; 8888145014, 1919800109, 20884521013145; 8888145014, 1919800111, 20884521013145; 8888145014, 1921100180, 20884521013145; 8888145014, 1921100182, 20884521013145; 8888145014, 1922100141, 20884521013145; 8888145014, 1922500102, 20884521013145; 8888145014, 1922500106, 20884521013145; 8888145014, 1922500107, 20884521013145; 8888145014, 1922500108, 20884521013145; 8888145014, 1924000188, 20884521013145; 8888145014, 1926100305, 20884521013145; 8888145014, 1926700064, 20884521013145; 8888145014, 1926800390, 20884521013145; 8888145014, 1926800391, 20884521013145; 8888145014, 1926800392, 20884521013145; 8888145014, 1926800393, 20884521013145; 8888145014, 1927600099, 20884521013145; 8888145014, 1927600100, 20884521013145; 8888145014, 1927600101, 20884521013145; 8888145014, 1927600102, 20884521013145; 8888145014, 1927600103, 20884521013145; 8888145014, 1928200085, 20884521013145; 8888145014, 1928200087, 20884521013145; 8888145014, 1928800199, 20884521013145; 8888145014, 1928800200, 20884521013145; 8888145014, 1928800201, 20884521013145; 8888145014, 1928800202, 20884521013145; 8888145014, 1928800205, 20884521013145; 8888145014, 1928800207, 20884521013145; 8888145014, 1928800246, 20884521013145; 8888145014, 1928800247, 20884521013145; 8888145014, 1928800248, 20884521013145; 8888145014, 1930200112, 20884521013145; 8888145014, 1930200113, 20884521013145; 8888145014, 1930200114, 20884521013145; 8888145014, 1930400108, 20884521013145; 8888145014, 1930900092, 20884521013145; 8888145014, 1930900094, 20884521013145; 8888145014, 1930900097, 20884521013145; 8888145014, 1930900098, 20884521013145; 8888145014, 1930900099, 20884521013145; 8888145014, 1931600123, 20884521013145; 8888145014, 1931600124, 20884521013145; 8888145014, 1931600125, 20884521013145; 8888145014, 1931600127, 20884521013145; 8888145014, 1933700122, 20884521013145; 8888145014, 1933700123, 20884521013145; 8888145014, 1933700124, 20884521013145; 8888145014, 1934400139, 20884521013145; 8888145014, 1934400167, 20884521013145; 8888145014, 1934400168, 20884521013145; 8888145014, 1934400169, 20884521013145; 8888145014, 1934400170, 20884521013145; 8888145014, 2002100141, 20884521013145; 8888145014, 2008400124, 20884521013145; 8888145014, 2008500083, 20884521013145; 8888145014, 2008500085, 20884521013145; 8888145014, 2008500086, 20884521013145; 8888145014, 2012100129, 20884521013145; 8888145014, 2012100130, 20884521013145; 8888145014, 2012100131, 20884521013145; 8888145014, 2012100132, 20884521013145; 8888145014, 2012100133, 20884521013145; 8888145014, 2015100056, 20884521013145; 8888145014, 2016400033, 20884521013145; 8888145014, 2016400034, 20884521013145; 8888145014, 2017100092, 20884521013145; 8888145014, 2019500255, 20884521013145; 8888145014, 2019500274, 20884521013145; 8888145014, 2019500279, 20884521013145; 8888145014, 2020400091, 20884521013145; 8888145014, 2020400092, 20884521013145; 8888145014, 2020400093, 20884521013145; 8888145014, 2020400094, 20884521013145; 8888145014, 2020400095, 20884521013145; 8888145014, 2020400096, 20884521013145; 8888145014, 2020400097, 20884521013145; 8888145014, 2020400098, 20884521013145; 8888145014, 2023300123, 20884521013145; 8888145014, 2023300124, 20884521013145; 8888145014, 2023300148, 20884521013145; 8888145014, 2023300149, 20884521013145; 8888145014, 2023300156, 20884521013145; 8888145014, 2023300157, 20884521013145; 8888145014, 2028200219, 20884521013145; 8888145014, 2028200220, 20884521013145; 8888145014, 2028200221, 20884521013145; 8888145014, 2028200222, 20884521013145; 8888145014, 2028200223, 20884521013145; 8888145014, 2029600112, 20884521013145; 8888145014, 2029600113, 20884521013145; 8888145014, 2029600114, 20884521013145; 8888145014, 2029600115, 20884521013145; 8888145014, 2029600116, 20884521013145; 8888145014, 2031800147, 20884521013145; 8888145014, 2031800148, 20884521013145; 8888145014, 2031800149, 20884521013145; 8888145014, 2031800151, 20884521013145; 8888145014, 2031800152, 20884521013145; 8888145014, 2034400121, 20884521013145; 8888145014, 2034400133, 20884521013145; 8888145014, 2034400137, 20884521013145; 8888145014, 2102500096, 20884521013145; 8888145014, 2102600119, 20884521013145; 8888145014, 2104100080, 20884521013145; 8888145014, 2104100081, 20884521013145; 8888145014, 2104100082, 20884521013145; 8888145014, 2104100083, 20884521013145; 8888145014, 2104100084, 20884521013145; 8888145014, 2104100096, 20884521013145; 8888145014, 2104100097, 20884521013145; 8888145014, 2104100098, 20884521013145; 8888145014, 2104100099, 20884521013145; 8888145014, 2104100100, 20884521013145; 8888145014, 2104100101, 20884521013145; 8888145014, 2104600115, 20884521013145; 8888145014, 2104600165, 20884521013145; 8888145014, 2104600169, 20884521013145; 8888145014, 2104600207, 20884521013145; 8888145014, 2104600235, 20884521013145; 8888145014, 2104600240, 20884521013145; 8888145014, 2104900108, 20884521013145; 8888145014, 2106900191, 20884521013145; 8888145014, 2106900192, 20884521013145; 8888145014, 2106900193, 20884521013145; 8888145014, 2106900195, 20884521013145; 8888145014, 2106900196, 20884521013145; 8888145014, 2110900115, 20884521013145; 8888145014, 2110900155, 20884521013145; 8888145014, 2110900156, 20884521013145; 8888145014, 2110900157, 20884521013145; 8888145014, 2113300442, 20884521013145; 8888145014, 2113300443, 20884521013145; 8888145014, 2113300444, 20884521013145; 8888145014, 2113300445, 20884521013145; 8888145014, 2113300446, 20884521013145; 8888145014, 2113400064, 20884521013145; 8888145014, 2113400070, 20884521013145; 8888145014, 2113400099, 20884521013145; 8888145014, 2116500091, 20884521013145; 8888145014, 2116500093, 20884521013145; 8888145014, 2116500125, 20884521013145; 8888145014, 2116500126, 20884521013145; 8888145014, 2116500127, 20884521013145; 8888145014, 2116500128, 20884521013145; 8888145014, 2116500129, 20884521013145; 8888145014, 2116500154, 20884521013145; 8888145014, 2116500155, 20884521013145; 8888145014, 2116500158, 20884521013145; 8888145014, 2116600068, 20884521013145; 8888145014, 2116600092, 20884521013145; 8888145014, 2124500100, 20884521013145; 8888145014, 2124500101, 20884521013145; 8888145014, 2124500103, 20884521013145; 8888145014, 2124500104, 20884521013145; 8888145014, 2124500105, 20884521013145; 8888145014, 2124500107, 20884521013145; 8888145014, 2124500114, 20884521013145; 8888145014, 2124500116, 20884521013145; 8888145014, 2124500117, 20884521013145; 8888145014, 2124600218, 20884521013145; 8888145014, 2124600219, 20884521013145; 8888145014, 2124600220, 20884521013145; 8888145014, 2124600221, 20884521013145; 8888145014, 2124600223, 20884521013145; 8888145014, 2124600224, 20884521013145; 8888145014, 2124600225, 20884521013145; 8888145014, 2124600226, 20884521013145; 8888145014, 2124600260, 20884521013145; 8888145014, 2124600261, 20884521013145; 8888145014, 2135700301, 20884521013145; 8888145014, 2201000059, 20884521013145; 8888145015, 1715100154, 10884521013162; 8888145015, 1715800118, 10884521013162; 8888145015, 1716300113, 10884521013162; 8888145015, 1716500176, 10884521013162; 8888145015, 1717600073, 10884521013162; 8888145015, 1717900134, 10884521013162; 8888145015, 1718600102, 10884521013162; 8888145015, 1719300101, 10884521013162; 8888145015, 1720700112, 10884521013162; 8888145015, 1721300142, 10884521013162; 8888145015, 1722100146, 10884521013162; 8888145015, 1728600082, 10884521013162; 8888145015, 1729100099, 10884521013162; 8888145015, 1729700113, 10884521013162; 8888145015, 1730500135, 10884521013162; 8888145015, 1731200153, 10884521013162; 8888145015, 1731200181, 10884521013162; 8888145015, 1732600133, 10884521013162; 8888145015, 1732600135, 10884521013162; 8888145015, 1733300157, 10884521013162; 8888145015, 1733700167, 10884521013162; 8888145015, 1733700189, 10884521013162; 8888145015, 1734700111, 10884521013162; 8888145015, 1801700127, 10884521013162; 8888145015, 1801700130, 10884521013162; 8888145015, 1801700132, 10884521013162; 8888145015, 1802400117, 10884521013162; 8888145015, 1803100116, 10884521013162; 8888145015, 1803100120, 10884521013162; 8888145015, 1803800132, 10884521013162; 8888145015, 1803800136, 10884521013162; 8888145015, 1804500150, 10884521013162; 8888145015, 1805900135, 10884521013162; 8888145015, 1807300170, 10884521013162; 8888145015, 1807300178, 10884521013162; 8888145015, 1808600075, 10884521013162; 8888145015, 1810400144, 10884521013162; 8888145015, 1811500159, 10884521013162; 8888145015, 1812200180, 10884521013162; 8888145015, 1812900135, 10884521013162; 8888145015, 1812900136, 10884521013162; 8888145015, 1812900137, 10884521013162; 8888145015, 1813600095, 10884521013162; 8888145015, 1814400061, 10884521013162; 8888145015, 1814400062, 10884521013162; 8888145015, 1815800080, 10884521013162; 8888145015, 1819000135, 10884521013162; 8888145015, 1819300096, 10884521013162; 8888145015, 1822600126, 10884521013162; 8888145015, 1822600128, 10884521013162; 8888145015, 1822600129, 10884521013162; 8888145015, 1822600130, 10884521013162; 8888145015, 1823900108, 10884521013162; 8888145015, 1823900109, 10884521013162; 8888145015, 1823900110, 10884521013162; 8888145015, 1824700122, 10884521013162; 8888145015, 1824700123, 10884521013162; 8888145015, 1828200104, 10884521013162; 8888145015, 1828900143, 10884521013162; 8888145015, 1829200087, 10884521013162; 8888145015, 1829500120, 10884521013162; 8888145015, 1831000105, 10884521013162; 8888145015, 1831600204, 10884521013162; 8888145015, 1832300113, 10884521013162; 8888145015, 1833100111, 10884521013162; 8888145015, 1833800147, 10884521013162; 8888145015, 1900700135, 10884521013162; 8888145015, 1900700136, 10884521013162; 8888145015, 1901400119, 10884521013162; 8888145015, 1901400122, 10884521013162; 8888145015, 1901400123, 10884521013162; 8888145015, 1901400124, 10884521013162; 8888145015, 1902100130, 10884521013162; 8888145015, 1902100131, 10884521013162; 8888145015, 1902100132, 10884521013162; 8888145015, 1902100141, 10884521013162; 8888145015, 1902100142, 10884521013162; 8888145015, 1902800120, 10884521013162; 8888145015, 1904200099, 10884521013162; 8888145015, 1904200100, 10884521013162; 8888145015, 1904200101, 10884521013162; 8888145015, 1904200102, 10884521013162; 8888145015, 1906000140, 10884521013162; 8888145015, 1906000141, 10884521013162; 8888145015, 1906000142, 10884521013162; 8888145015, 1906700186, 10884521013162; 8888145015, 1907700095, 10884521013162; 8888145015, 1911500114, 10884521013162; 8888145015, 1911500115, 10884521013162; 8888145015, 1911500116, 10884521013162; 8888145015, 1912600103, 10884521013162; 8888145015, 1912600104, 10884521013162; 8888145015, 1912600105, 10884521013162; 8888145015, 1915800089, 10884521013162; 8888145015, 1915800090, 10884521013162; 8888145015, 1915800092, 10884521013162; 8888145015, 1917600088, 10884521013162; 8888145015, 1917700081, 10884521013162; 8888145015, 1917700082, 10884521013162; 8888145015, 1917700083, 10884521013162; 8888145015, 1917700084, 10884521013162; 8888145015, 1917700085, 10884521013162; 8888145015, 1919700104, 10884521013162; 8888145015, 1919700105, 10884521013162; 8888145015, 1919700106, 10884521013162; 8888145015, 1919700107, 10884521013162; 8888145015, 1919700108, 10884521013162; 8888145015, 1919800095, 10884521013162; 8888145015, 1922500099, 10884521013162; 8888145015, 1922500220, 10884521013162; 8888145015, 1922500221, 10884521013162; 8888145015, 1922500222, 10884521013162; 8888145015, 1922500223, 10884521013162; 8888145015, 1922500224, 10884521013162; 8888145015, 1922500225, 10884521013162; 8888145015, 1922500226, 10884521013162; 8888145015, 1926800394, 10884521013162; 8888145015, 1926800395, 10884521013162; 8888145015, 1926800396, 10884521013162; 8888145015, 1926800397, 10884521013162; 8888145015, 1926800398, 10884521013162; 8888145015, 1928800242, 10884521013162; 8888145015, 1929500114, 10884521013162; 8888145015, 1930200103, 10884521013162; 8888145015, 1930200104, 10884521013162; 8888145015, 1930200105, 10884521013162; 8888145015, 1930400106, 10884521013162; 8888145015, 1931500079, 10884521013162; 8888145015, 1931600121, 10884521013162; 8888145015, 1931600122, 10884521013162; 8888145015, 1932300198, 10884521013162; 8888145015, 1932300200, 10884521013162; 8888145015, 1932300201, 10884521013162; 8888145015, 1933800172, 10884521013162; 8888145015, 2000700088, 10884521013162; 8888145015, 2003700039, 10884521013162; 8888145015, 2003700045, 10884521013162; 8888145015, 2007000053, 10884521013162; 8888145015, 2007000054, 10884521013162; 8888145015, 2007000055, 10884521013162; 8888145015, 2007000056, 10884521013162; 8888145015, 2007000057, 10884521013162; 8888145015, 2012100125, 10884521013162; 8888145015, 2012100126, 10884521013162; 8888145015, 2012100127, 10884521013162; 8888145015, 2015100053, 10884521013162; 8888145015, 2015100054, 10884521013162; 8888145015, 2015100055, 10884521013162; 8888145015, 2016500072, 10884521013162; 8888145015, 2017100091, 10884521013162; 8888145015, 2018100025, 10884521013162; 8888145015, 2018100026, 10884521013162; 8888145015, 2018200029, 10884521013162; 8888145015, 2019100005, 10884521013162; 8888145015, 2019500273, 10884521013162; 8888145015, 2023300109, 10884521013162; 8888145015, 2023300146, 10884521013162; 8888145015, 2023300155, 10884521013162; 8888145015, 2027200210, 10884521013162; 8888145015, 2027200216, 10884521013162; 8888145015, 2028200190, 10884521013162; 8888145015, 2028200191, 10884521013162; 8888145015, 2028200192, 10884521013162; 8888145015, 2028200193, 10884521013162; 8888145015, 2028200199, 10884521013162; 8888145015, 2028200213, 10884521013162; 8888145015, 2031800127, 10884521013162; 8888145015, 2031800128, 10884521013162; 8888145015, 2034400085, 10884521013162; 8888145015, 2034400094, 10884521013162; 8888145015, 2034400099, 10884521013162; 8888145015, 2034400103, 10884521013162; 8888145015, 2034400110, 10884521013162; 8888145015, 2034400120, 10884521013162; 8888145015, 2034400132, 10884521013162; 8888145015, 2034400136, 10884521013162; 8888145015, 2104600168, 10884521013162; 8888145015, 2104600206, 10884521013162; 8888145015, 2104600210, 10884521013162; 8888145015, 2106900103, 10884521013162; 8888145015, 2106900182, 10884521013162; 8888145015, 2106900197, 10884521013162; 8888145015, 2106900229, 10884521013162; 8888145015, 2110900148, 10884521013162; 8888145015, 2110900154, 10884521013162; 8888145015, 2113400069, 10884521013162; 8888145015, 2113400080, 10884521013162; 8888145015, 2116600125, 10884521013162; 8888145015, 2116600162, 10884521013162; 8888145015, 2124500091, 10884521013162; 8888145015, 2124500097, 10884521013162; 8888145015, 2124500113, 10884521013162; 8888145015, 2124500115, 10884521013162; 8888145015, 2124500123, 10884521013162; 8888145015, 2124500124, 10884521013162; 8888145015, 2124600238, 10884521013162; 8888145015, 2124600246, 10884521013162; 8888145015, 2124600250, 10884521013162; 8888145015, 2124600251, 10884521013162; 8888145015, 2124600254, 10884521013162; 8888145015, 2124600257, 10884521013162; 8888145015, 2124600263, 10884521013162; 8888145015, 2124600264, 10884521013162; 8888145015, 2127200481, 10884521013162; 8888145015, 2200300056, 10884521013162; 8888145015, 2200300059, 10884521013162; 8888145015, 2201000060, 10884521013162; 8888145015, 2201000061, 10884521013162; 8888145015, 2201000063, 10884521013162; 8888145015, 1715100154, 20884521013169; 8888145015, 1715800118, 20884521013169; 8888145015, 1716300113, 20884521013169; 8888145015, 1716500176, 20884521013169; 8888145015, 1717600073, 20884521013169; 8888145015, 1717900134, 20884521013169; 8888145015, 1718600102, 20884521013169; 8888145015, 1719300101, 20884521013169; 8888145015, 1720100134, 20884521013169; 8888145015, 1720400112, 20884521013169; 8888145015, 1720700112, 20884521013169; 8888145015, 1721300142, 20884521013169; 8888145015, 1729700113, 20884521013169; 8888145015, 1730500135, 20884521013169; 8888145015, 1731200153, 20884521013169; 8888145015, 1731200181, 20884521013169; 8888145015, 1732600133, 20884521013169; 8888145015, 1732600135, 20884521013169; 8888145015, 1733300157, 20884521013169; 8888145015, 1733700167, 20884521013169; 8888145015, 1733700189, 20884521013169; 8888145015, 1734700111, 20884521013169; 8888145015, 1800300142, 20884521013169; 8888145015, 1801700127, 20884521013169; 8888145015, 1801700130, 20884521013169; 8888145015, 1801700132, 20884521013169; 8888145015, 1802400117, 20884521013169; 8888145015, 1803100116, 20884521013169; 8888145015, 1803100120, 20884521013169; 8888145015, 1803800132, 20884521013169; 8888145015, 1803800136, 20884521013169; 8888145015, 1804500150, 20884521013169; 8888145015, 1804600099, 20884521013169; 8888145015, 1805900135, 20884521013169; 8888145015, 1807300170, 20884521013169; 8888145015, 1807300178, 20884521013169; 8888145015, 1808600075, 20884521013169; 8888145015, 1812200180, 20884521013169; 8888145015, 1812900135, 20884521013169; 8888145015, 1812900136, 20884521013169; 8888145015, 1812900137, 20884521013169; 8888145015, 1813600095, 20884521013169; 8888145015, 1814400061, 20884521013169; 8888145015, 1814400062, 20884521013169; 8888145015, 1815800080, 20884521013169; 8888145015, 1819000135, 20884521013169; 8888145015, 1822600126, 20884521013169; 8888145015, 1822600128, 20884521013169; 8888145015, 1822600130, 20884521013169; 8888145015, 1823900108, 20884521013169; 8888145015, 1823900109, 20884521013169; 8888145015, 1823900110, 20884521013169; 8888145015, 1824700122, 20884521013169; 8888145015, 1824700123, 20884521013169; 8888145015, 1828200104, 20884521013169; 8888145015, 1828900143, 20884521013169; 8888145015, 1829200087, 20884521013169; 8888145015, 1829500120, 20884521013169; 8888145015, 1831000105, 20884521013169; 8888145015, 1831600204, 20884521013169; 8888145015, 1832300113, 20884521013169; 8888145015, 1833800147, 20884521013169; 8888145015, 1900700135, 20884521013169; 8888145015, 1900700136, 20884521013169; 8888145015, 1901400119, 20884521013169; 8888145015, 1901400122, 20884521013169; 8888145015, 1901400123, 20884521013169; 8888145015, 1901400124, 20884521013169; 8888145015, 1902100130, 20884521013169; 8888145015, 1902100131, 20884521013169; 8888145015, 1902100132, 20884521013169; 8888145015, 1902100141, 20884521013169; 8888145015, 1902100142, 20884521013169; 8888145015, 1902800120, 20884521013169; 8888145015, 1904200099, 20884521013169; 8888145015, 1904200100, 20884521013169; 8888145015, 1904200101, 20884521013169; 8888145015, 1904200102, 20884521013169; 8888145015, 1906000140, 20884521013169; 8888145015, 1906000141, 20884521013169; 8888145015, 1906000142, 20884521013169; 8888145015, 1906700186, 20884521013169; 8888145015, 1906700187, 20884521013169; 8888145015, 1907700095, 20884521013169; 8888145015, 1911500114, 20884521013169; 8888145015, 1911500115, 20884521013169; 8888145015, 1911500116, 20884521013169; 8888145015, 1912600103, 20884521013169; 8888145015, 1912600104, 20884521013169; 8888145015, 1915800089, 20884521013169; 8888145015, 1915800090, 20884521013169; 8888145015, 1915800091, 20884521013169; 8888145015, 1915800092, 20884521013169; 8888145015, 1917600088, 20884521013169; 8888145015, 1917700081, 20884521013169; 8888145015, 1917700082, 20884521013169; 8888145015, 1917700083, 20884521013169; 8888145015, 1917700084, 20884521013169; 8888145015, 1917700085, 20884521013169; 8888145015, 1919700104, 20884521013169; 8888145015, 1919700105, 20884521013169; 8888145015, 1919700106, 20884521013169; 8888145015, 1919700107, 20884521013169; 8888145015, 1919700108, 20884521013169; 8888145015, 1922500220, 20884521013169; 8888145015, 1922500222, 20884521013169; 8888145015, 1922500223, 20884521013169; 8888145015, 1922500224, 20884521013169; 8888145015, 1922500225, 20884521013169; 8888145015, 1926800394, 20884521013169; 8888145015, 1926800395, 20884521013169; 8888145015, 1926800396, 20884521013169; 8888145015, 1926800397, 20884521013169; 8888145015, 1926800398, 20884521013169; 8888145015, 1928800242, 20884521013169; 8888145015, 1929500114, 20884521013169; 8888145015, 1930200103, 20884521013169; 8888145015, 1930200104, 20884521013169; 8888145015, 1930200105, 20884521013169; 8888145015, 1930400106, 20884521013169; 8888145015, 1931500079, 20884521013169; 8888145015, 1931600121, 20884521013169; 8888145015, 1931600122, 20884521013169; 8888145015, 1932300200, 20884521013169; 8888145015, 1932300201, 20884521013169; 8888145015, 1933800172, 20884521013169; 8888145015, 2000700088, 20884521013169; 8888145015, 2003700039, 20884521013169; 8888145015, 2003700045, 20884521013169; 8888145015, 2007000053, 20884521013169; 8888145015, 2007000054, 20884521013169; 8888145015, 2007000055, 20884521013169; 8888145015, 2007000056, 20884521013169; 8888145015, 2007000057, 20884521013169; 8888145015, 2012100125, 20884521013169; 8888145015, 2012100126, 20884521013169; 8888145015, 2012100127, 20884521013169; 8888145015, 2015100053, 20884521013169; 8888145015, 2015100054, 20884521013169; 8888145015, 2015100055, 20884521013169; 8888145015, 2016500072, 20884521013169; 8888145015, 2017100091, 20884521013169; 8888145015, 2018100025, 20884521013169; 8888145015, 2018100026, 20884521013169; 8888145015, 2018200029, 20884521013169; 8888145015, 2019100005, 20884521013169; 8888145015, 2019500273, 20884521013169; 8888145015, 2020400099, 20884521013169; 8888145015, 2023300109, 20884521013169; 8888145015, 2023300146, 20884521013169; 8888145015, 2023300155, 20884521013169; 8888145015, 2027200210, 20884521013169; 8888145015, 2027200216, 20884521013169; 8888145015, 2028200190, 20884521013169; 8888145015, 2028200191, 20884521013169; 8888145015, 2028200192, 20884521013169; 8888145015, 2028200193, 20884521013169; 8888145015, 2028200199, 20884521013169; 8888145015, 2028200213, 20884521013169; 8888145015, 2031800127, 20884521013169; 8888145015, 2031800128, 20884521013169; 8888145015, 2034400085, 20884521013169; 8888145015, 2034400103, 20884521013169; 8888145015, 2034400132, 20884521013169; 8888145015, 2034400136, 20884521013169; 8888145015, 2104600168, 20884521013169; 8888145015, 2104600206, 20884521013169; 8888145015, 2104600210, 20884521013169; 8888145015, 2106900103, 20884521013169; 8888145015, 2106900197, 20884521013169; 8888145015, 2106900229, 20884521013169; 8888145015, 2110900148, 20884521013169; 8888145015, 2110900154, 20884521013169; 8888145015, 2113400069, 20884521013169; 8888145015, 2116600125, 20884521013169; 8888145015, 2116600162, 20884521013169; 8888145015, 2124500091, 20884521013169; 8888145015, 2124500097, 20884521013169; 8888145015, 2124500115, 20884521013169; 8888145015, 2124500123, 20884521013169; 8888145015, 2124500124, 20884521013169; 8888145015, 2124600238, 20884521013169; 8888145015, 2124600246, 20884521013169; 8888145015, 2124600250, 20884521013169; 8888145015, 2124600251, 20884521013169; 8888145015, 2124600252, 20884521013169; 8888145015, 2124600254, 20884521013169; 8888145015, 2124600257, 20884521013169; 8888145015, 2124600263, 20884521013169; 8888145015, 2124600264, 20884521013169; 8888145015, 2127200481, 20884521013169; 8888145015, 2200300059, 20884521013169; 8888145015, 2201000061, 20884521013169; 8888145015, 2201000063, 20884521013169; 8888145016, 1715800119, 10884521013179; 8888145016, 1720700113, 10884521013179; 8888145016, 1723000087, 10884521013179; 8888145016, 1725000111, 10884521013179; 8888145016, 1725600112, 10884521013179; 8888145016, 1730500130, 10884521013179; 8888145016, 1731200169, 10884521013179; 8888145016, 1731900153, 10884521013179; 8888145016, 1732600128, 10884521013179; 8888145016, 1733300162, 10884521013179; 8888145016, 1733700186, 10884521013179; 8888145016, 1800300130, 10884521013179; 8888145016, 1800300138, 10884521013179; 8888145016, 1800300144, 10884521013179; 8888145016, 1801000132, 10884521013179; 8888145016, 1801000134, 10884521013179; 8888145016, 1802400123, 10884521013179; 8888145016, 1803100124, 10884521013179; 8888145016, 1803800138, 10884521013179; 8888145016, 1804500145, 10884521013179; 8888145016, 1804800056, 10884521013179; 8888145016, 1805200095, 10884521013179; 8888145016, 1806600081, 10884521013179; 8888145016, 1807300174, 10884521013179; 8888145016, 1811700503, 10884521013179; 8888145016, 1825400103, 10884521013179; 8888145016, 1826200148, 10884521013179; 8888145016, 1826800149, 10884521013179; 8888145016, 1835400081, 10884521013179; 8888145016, 1903500115, 10884521013179; 8888145016, 1903500116, 10884521013179; 8888145016, 1906000143, 10884521013179; 8888145016, 1907800095, 10884521013179; 8888145016, 1908800155, 10884521013179; 8888145016, 1911500091, 10884521013179; 8888145016, 1914000291, 10884521013179; 8888145016, 1914000292, 10884521013179; 8888145016, 1914400120, 10884521013179; 8888145016, 1914400121, 10884521013179; 8888145016, 1915800094, 10884521013179; 8888145016, 1915800095, 10884521013179; 8888145016, 1919800105, 10884521013179; 8888145016, 1922100146, 10884521013179; 8888145016, 1922500104, 10884521013179; 8888145016, 1922500105, 10884521013179; 8888145016, 1922500242, 10884521013179; 8888145016, 1922500243, 10884521013179; 8888145016, 1934400184, 10884521013179; 8888145016, 2000700090, 10884521013179; 8888145016, 2003700041, 10884521013179; 8888145016, 2007900092, 10884521013179; 8888145016, 2012100134, 10884521013179; 8888145016, 2017100094, 10884521013179; 8888145016, 2019500275, 10884521013179; 8888145016, 2023300113, 10884521013179; 8888145016, 2027300038, 10884521013179; 8888145016, 2028200218, 10884521013179; 8888145016, 2031800129, 10884521013179; 8888145016, 2034400122, 10884521013179; 8888145016, 2106900169, 10884521013179; 8888145016, 2113300433, 10884521013179; 8888145016, 2113400083, 10884521013179; 8888145016, 2127200498, 10884521013179; 8888145016, 1715800119, 20884521013176; 8888145016, 1720700113, 20884521013176; 8888145016, 1723000087, 20884521013176; 8888145016, 1725000111, 20884521013176; 8888145016, 1725600112, 20884521013176; 8888145016, 1729100103, 20884521013176; 8888145016, 1730500130, 20884521013176; 8888145016, 1731200169, 20884521013176; 8888145016, 1731900153, 20884521013176; 8888145016, 1732600128, 20884521013176; 8888145016, 1733300162, 20884521013176; 8888145016, 1733700186, 20884521013176; 8888145016, 1800300130, 20884521013176; 8888145016, 1800300138, 20884521013176; 8888145016, 1800300144, 20884521013176; 8888145016, 1801000132, 20884521013176; 8888145016, 1801000134, 20884521013176; 8888145016, 1802400123, 20884521013176; 8888145016, 1803100124, 20884521013176; 8888145016, 1803800138, 20884521013176; 8888145016, 1804500145, 20884521013176; 8888145016, 1804800056, 20884521013176; 8888145016, 1805200095, 20884521013176; 8888145016, 1806600081, 20884521013176; 8888145016, 1807300174, 20884521013176; 8888145016, 1808600068, 20884521013176; 8888145016, 1825400103, 20884521013176; 8888145016, 1826200148, 20884521013176; 8888145016, 1826800149, 20884521013176; 8888145016, 1903500115, 20884521013176; 8888145016, 1903500116, 20884521013176; 8888145016, 1906000143, 20884521013176; 8888145016, 1907800095, 20884521013176; 8888145016, 1908800155, 20884521013176; 8888145016, 1911500091, 20884521013176; 8888145016, 1914000291, 20884521013176; 8888145016, 1914000292, 20884521013176; 8888145016, 1914400120, 20884521013176; 8888145016, 1914400121, 20884521013176; 8888145016, 1919800105, 20884521013176; 8888145016, 1922500104, 20884521013176; 8888145016, 1922500105, 20884521013176; 8888145016, 1922500242, 20884521013176; 8888145016, 1922500243, 20884521013176; 8888145016, 1934400184, 20884521013176; 8888145016, 2000700090, 20884521013176; 8888145016, 2003700041, 20884521013176; 8888145016, 2007900092, 20884521013176; 8888145016, 2012100134, 20884521013176; 8888145016, 2017100094, 20884521013176; 8888145016, 2019500275, 20884521013176; 8888145016, 2023300113, 20884521013176; 8888145016, 2027300038, 20884521013176; 8888145016, 2028200218, 20884521013176; 8888145016, 2031800129, 20884521013176; 8888145016, 2034400122, 20884521013176; 8888145016, 2106900169, 20884521013176; 8888145016, 2113300433, 20884521013176; 8888145016, 2113400083, 20884521013176; 8888145016, 2127200497, 20884521013176; 8888145016, 2127200498, 20884521013176; 8888145016, 2201000068, 20884521013176; 8888145017, 1716500177, 10884521013193; 8888145017, 1723000089, 10884521013193; 8888145017, 1724200115, 10884521013193; 8888145017, 1726300121, 10884521013193; 8888145017, 1731200168, 10884521013193; 8888145017, 1731200173, 10884521013193; 8888145017, 1734700106, 10884521013193; 8888145017, 1801000130, 10884521013193; 8888145017, 1802400118, 10884521013193; 8888145017, 1803100121, 10884521013193; 8888145017, 1803800139, 10884521013193; 8888145017, 1804500151, 10884521013193; 8888145017, 1808000126, 10884521013193; 8888145017, 1811500163, 10884521013193; 8888145017, 1812200183, 10884521013193; 8888145017, 1822000072, 10884521013193; 8888145017, 1823900111, 10884521013193; 8888145017, 1825400104, 10884521013193; 8888145017, 1826800150, 10884521013193; 8888145017, 1834500199, 10884521013193; 8888145017, 1834500200, 10884521013193; 8888145017, 1902800144, 10884521013193; 8888145017, 1907900140, 10884521013193; 8888145017, 1912600109, 10884521013193; 8888145017, 1913000187, 10884521013193; 8888145017, 1915800082, 10884521013193; 8888145017, 1915800083, 10884521013193; 8888145017, 1919700102, 10884521013193; 8888145017, 1919700103, 10884521013193; 8888145017, 1922500201, 10884521013193; 8888145017, 1922500202, 10884521013193; 8888145017, 1922500204, 10884521013193; 8888145017, 1930200111, 10884521013193; 8888145017, 1930900100, 10884521013193; 8888145017, 1932300205, 10884521013193; 8888145017, 2000700092, 10884521013193; 8888145017, 2008400130, 10884521013193; 8888145017, 2023300114, 10884521013193; 8888145017, 2026000068, 10884521013193; 8888145017, 2027200260, 10884521013193; 8888145017, 2034300097, 10884521013193; 8888145017, 2034300099, 10884521013193; 8888145017, 2034400111, 10884521013193; 8888145017, 1716500177, 20884521013190; 8888145017, 1723000089, 20884521013190; 8888145017, 1724200115, 20884521013190; 8888145017, 1726300121, 20884521013190; 8888145017, 1731200168, 20884521013190; 8888145017, 1731200173, 20884521013190; 8888145017, 1734700106, 20884521013190; 8888145017, 1802400118, 20884521013190; 8888145017, 1803100121, 20884521013190; 8888145017, 1803800139, 20884521013190; 8888145017, 1804500151, 20884521013190; 8888145017, 1808000126, 20884521013190; 8888145017, 1811500163, 20884521013190; 8888145017, 1812200183, 20884521013190; 8888145017, 1822000072, 20884521013190; 8888145017, 1823900111, 20884521013190; 8888145017, 1825400104, 20884521013190; 8888145017, 1826800150, 20884521013190; 8888145017, 1834500199, 20884521013190; 8888145017, 1834500200, 20884521013190; 8888145017, 1902800144, 20884521013190; 8888145017, 1907900140, 20884521013190; 8888145017, 1912600109, 20884521013190; 8888145017, 1913000187, 20884521013190; 8888145017, 1919700102, 20884521013190; 8888145017, 1919700103, 20884521013190; 8888145017, 1922500201, 20884521013190; 8888145017, 1922500202, 20884521013190; 8888145017, 1922500204, 20884521013190; 8888145017, 1930200111, 20884521013190; 8888145017, 1930900100, 20884521013190; 8888145017, 1932300205, 20884521013190; 8888145017, 2000700092, 20884521013190; 8888145017, 2008400130, 20884521013190; 8888145017, 2023300114, 20884521013190; 8888145017, 2026000068, 20884521013190; 8888145017, 2027200260, 20884521013190; 8888145017, 2034300097, 20884521013190; 8888145017, 2034300099, 20884521013190; 8888145017, 2034400111, 20884521013190; 8888145018, 1722100138, 10884521013216; 8888145018, 1723400097, 10884521013216; 8888145018, 1725600109, 10884521013216; 8888145018, 1725600116, 10884521013216; 8888145018, 1733300155, 10884521013216; 8888145018, 1733300164, 10884521013216; 8888145018, 1800300129, 10884521013216; 8888145018, 1800300141, 10884521013216; 8888145018, 1802400121, 10884521013216; 8888145018, 1803100119, 10884521013216; 8888145018, 1803800131, 10884521013216; 8888145018, 1803800137, 10884521013216; 8888145018, 1803800140, 10884521013216; 8888145018, 1804800057, 10884521013216; 8888145018, 1804900082, 10884521013216; 8888145018, 1805200108, 10884521013216; 8888145018, 1808000125, 10884521013216; 8888145018, 1809400134, 10884521013216; 8888145018, 1810400155, 10884521013216; 8888145018, 1907800096, 10884521013216; 8888145018, 1912600110, 10884521013216; 8888145018, 1917600089, 10884521013216; 8888145018, 1922100143, 10884521013216; 8888145018, 1922100145, 10884521013216; 8888145018, 1922500228, 10884521013216; 8888145018, 2000700093, 10884521013216; 8888145018, 2002100142, 10884521013216; 8888145018, 2009200086, 10884521013216; 8888145018, 2017000127, 10884521013216; 8888145018, 2027200179, 10884521013216; 8888145018, 2028000086, 10884521013216; 8888145018, 2028200210, 10884521013216; 8888145018, 2034400104, 10884521013216; 8888145018, 2102600144, 10884521013216; 8888145018, 2106900216, 10884521013216; 8888145018, 2113400071, 10884521013216; 8888145018, 1723400097, 20884521013213; 8888145018, 1725600109, 20884521013213; 8888145018, 1725600116, 20884521013213; 8888145018, 1733300155, 20884521013213; 8888145018, 1733300164, 20884521013213; 8888145018, 1800300129, 20884521013213; 8888145018, 1800300141, 20884521013213; 8888145018, 1802400121, 20884521013213; 8888145018, 1803100119, 20884521013213; 8888145018, 1803800131, 20884521013213; 8888145018, 1803800137, 20884521013213; 8888145018, 1803800140, 20884521013213; 8888145018, 1804800057, 20884521013213; 8888145018, 1804900082, 20884521013213; 8888145018, 1805200108, 20884521013213; 8888145018, 1808000125, 20884521013213; 8888145018, 1809400134, 20884521013213; 8888145018, 1810400155, 20884521013213; 8888145018, 1907800096, 20884521013213; 8888145018, 1912600110, 20884521013213; 8888145018, 1917600089, 20884521013213; 8888145018, 1922500228, 20884521013213; 8888145018, 2000700093, 20884521013213; 8888145018, 2002100142, 20884521013213; 8888145018, 2009200086, 20884521013213; 8888145018, 2017000127, 20884521013213; 8888145018, 2027200179, 20884521013213; 8888145018, 2028000086, 20884521013213; 8888145018, 2028200210, 20884521013213; 8888145018, 2034400104, 20884521013213; 8888145018, 2102600144, 20884521013213; 8888145018, 2106900216, 20884521013213; 8888145018, 2113400071, 20884521013213; 8888145039, 1727100077, 10884521056657; 8888145039, 1727700137, 10884521056657; 8888145039, 1733300150, 10884521056657; 8888145039, 1800300134, 10884521056657; 8888145039, 1804500147, 10884521056657; 8888145039, 1815800081, 10884521056657; 8888145039, 1816400081, 10884521056657; 8888145039, 1816400082, 10884521056657; 8888145039, 1822000074, 10884521056657; 8888145039, 1823200094, 10884521056657; 8888145039, 1823400099, 10884521056657; 8888145039, 1826200169, 10884521056657; 8888145039, 1826800151, 10884521056657; 8888145039, 2002100135, 10884521056657; 8888145039, 2023300158, 10884521056657; 8888145039, 2026000062, 10884521056657; 8888145039, 2027200266, 10884521056657; 8888145039, 2110900106, 10884521056657; 8888145039, 2113400065, 10884521056657; 8888145039, 2116600098, 10884521056657; 8888145039, 2124500111, 10884521056657; 8888145039, 1726300124, 20884521056654; 8888145039, 1727100077, 20884521056654; 8888145039, 1727700137, 20884521056654; 8888145039, 1733300150, 20884521056654; 8888145039, 1800300134, 20884521056654; 8888145039, 1803100123, 20884521056654; 8888145039, 1804500147, 20884521056654; 8888145039, 1807300171, 20884521056654; 8888145039, 1815800081, 20884521056654; 8888145039, 1816400081, 20884521056654; 8888145039, 1816400082, 20884521056654; 8888145039, 1822000074, 20884521056654; 8888145039, 1823200094, 20884521056654; 8888145039, 1823400099, 20884521056654; 8888145039, 1826200169, 20884521056654; 8888145039, 1826800151, 20884521056654; 8888145039, 2002100135, 20884521056654; 8888145039, 2023300158, 20884521056654; 8888145039, 2026000062, 20884521056654; 8888145039, 2027200266, 20884521056654; 8888145039, 2110900106, 20884521056654; 8888145039, 2113400065, 20884521056654; 8888145039, 2116600098, 20884521056654; 8888145039, 2124500111, 20884521056654; 8888145040, 1725600115, 10884521056664; 8888145040, 1727100078, 10884521056664; 8888145040, 1734700108, 10884521056664; 8888145040, 1803800130, 10884521056664; 8888145040, 1805200085, 10884521056664; 8888145040, 1808600080, 10884521056664; 8888145040, 1810400151, 10884521056664; 8888145040, 1812200186, 10884521056664; 8888145040, 1814300201, 10884521056664; 8888145040, 1834500211, 10884521056664; 8888145040, 1901400125, 10884521056664; 8888145040, 1903500122, 10884521056664; 8888145040, 1908400321, 10884521056664; 8888145040, 1913000177, 10884521056664; 8888145040, 1914000272, 10884521056664; 8888145040, 1914000273, 10884521056664; 8888145040, 2000700094, 10884521056664; 8888145040, 2003700044, 10884521056664; 8888145040, 2009200088, 10884521056664; 8888145040, 2023300150, 10884521056664; 8888145040, 2026000070, 10884521056664; 8888145040, 2027200238, 10884521056664; 8888145040, 2031800140, 10884521056664; 8888145040, 2034400100, 10884521056664; 8888145040, 2034400141, 10884521056664; 8888145040, 2104600241, 10884521056664; 8888145040, 2116600127, 10884521056664; 8888145040, 2119400277, 10884521056664; 8888145040, 2127200511, 10884521056664; 8888145040, 1725600115, 20884521056661; 8888145040, 1727100078, 20884521056661; 8888145040, 1734700108, 20884521056661; 8888145040, 1803800130, 20884521056661; 8888145040, 1805200085, 20884521056661; 8888145040, 1808600080, 20884521056661; 8888145040, 1810400151, 20884521056661; 8888145040, 1812200186, 20884521056661; 8888145040, 1814300201, 20884521056661; 8888145040, 1834500211, 20884521056661; 8888145040, 1901400125, 20884521056661; 8888145040, 1903500122, 20884521056661; 8888145040, 1908400321, 20884521056661; 8888145040, 1913000177, 20884521056661; 8888145040, 1914000272, 20884521056661; 8888145040, 1914000273, 20884521056661; 8888145040, 2000700094, 20884521056661; 8888145040, 2003700044, 20884521056661; 8888145040, 2009200088, 20884521056661; 8888145040, 2023300150, 20884521056661; 8888145040, 2026000070, 20884521056661; 8888145040, 2027200238, 20884521056661; 8888145040, 2031800140, 20884521056661; 8888145040, 2034400100, 20884521056661; 8888145040, 2034400141, 20884521056661; 8888145040, 2104600241, 20884521056661; 8888145040, 2116600127, 20884521056661; 8888145040, 2119400277, 20884521056661; 8888145040, 2127200511, 20884521056661; 8888145041, 1731900209, 10884521056671; 8888145041, 1814300202, 10884521056671; 8888145041, 1829500123, 10884521056671; 8888145041, 1913000178, 10884521056671; 8888145041, 1918300149, 10884521056671; 8888145041, 2009300122, 10884521056671; 8888145041, 2034400142, 10884521056671; 8888145041, 2124500099, 10884521056671; 8888145041, 1731900209, 20884521056678; 8888145041, 1814300202, 20884521056678; 8888145041, 1829500123, 20884521056678; 8888145041, 1913000178, 20884521056678; 8888145041, 1918300149, 20884521056678; 8888145041, 2009300122, 20884521056678; 8888145041, 2034400142, 20884521056678; 8888145041, 2124500099, 20884521056678; 8888145042, 1729700115, 10884521056688; 8888145042, 1808600079, 10884521056688; 8888145042, 1812200185, 10884521056688; 8888145042, 1814300203, 10884521056688; 8888145042, 2009300123, 10884521056688; 8888145042, 2026000057, 10884521056688; 8888145042, 2110900107, 10884521056688; 8888145042, 2113400072, 10884521056688; 8888145042, 1729700115, 20884521056685; 8888145042, 1808600079, 20884521056685; 8888145042, 1812200185, 20884521056685; 8888145042, 1814300203, 20884521056685; 8888145042, 2009300123, 20884521056685; 8888145042, 2026000057, 20884521056685; 8888145042, 2110900107, 20884521056685; 8888145042, 2113400072, 20884521056685; 8888145058, 1809400137, 10884521013339; 8888145058, 1917600087, 10884521013339; 8888145058, 2102600145, 10884521013339; 8888145058, 2110900113, 10884521013339; 8888145058, 2113400085, 10884521013339; 8888145058, 2116500159, 10884521013339; 8888145058, 1715300100, 20884521013336; 8888145058, 1717900076, 20884521013336; 8888145058, 1723100081, 20884521013336; 8888145058, 1731900197, 20884521013336; 8888145058, 1809400137, 20884521013336; 8888145058, 1904200105, 20884521013336; 8888145058, 1910200106, 20884521013336; 8888145058, 1916800083, 20884521013336; 8888145058, 1917600086, 20884521013336; 8888145058, 1917600087, 20884521013336; 8888145058, 1919800106, 20884521013336; 8888145058, 2102600145, 20884521013336; 8888145058, 2104100102, 20884521013336; 8888145058, 2110900113, 20884521013336; 8888145058, 2113400085, 20884521013336; 8888145058, 2116500159, 20884521013336; 8888145058, 2119400279, 20884521013336; 8888145059, 1825400105, 10884521013346; 8888145059, 1830300165, 10884521013346; 8888145059, 2000700102, 10884521013346; 8888145059, 1718100122, 20884521013343; 8888145059, 1722100149, 20884521013343; 8888145059, 1728600065, 20884521013343; 8888145059, 1729100104, 20884521013343; 8888145059, 1805200086, 20884521013343; 8888145059, 1815000192, 20884521013343; 8888145059, 1822000076, 20884521013343; 8888145059, 1825400105, 20884521013343; 8888145059, 1826200143, 20884521013343; 8888145059, 1830300165, 20884521013343; 8888145059, 1902100112, 20884521013343; 8888145059, 2000700102, 20884521013343; 8888145059, 2001400115, 20884521013343; 8888145059, 2106900114, 20884521013343; 8888145059, 2110900050, 20884521013343; 8888145060, 2002100134, 10884521013360; 8888145060, 2002100134, 20884521013367; 8888145061, 2012100135, 20884521013381

Devices named in this recall

20 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

23 devices cleared under K111372 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 90311).